K231575 · Artaflex, Inc. · LIH · Feb 8, 2024 · Neurology
Device Facts
Record ID
K231575
Device Name
Vecttor VT-300
Applicant
Artaflex, Inc.
Product Code
LIH · Neurology
Decision Date
Feb 8, 2024
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Electrical stimulation delivered by the device is intended for the symptomatic relief of chronic intractable pain and/or management of post-traumatic or post-surgical pain.
Device Story
Vecttor VT-300 is a portable, battery-operated interferential current therapy device. It uses eight electrodes to deliver four channels of electrical stimulation to the patient's body. The device incorporates temperature sensors that monitor finger surface temperature to adjust therapy levels based on specified temperature increases. Operation is controlled via the Vecttor Mobile App on Android or iOS devices, which connects to the VT-300 via Bluetooth. The device is intended for use in home or clinical settings. Healthcare providers or patients use the app to manage treatments and data, which are stored in a cloud-based application. The device aims to provide pain relief through interferential current stimulation, with therapy parameters modulated by real-time temperature feedback.
Clinical Evidence
No clinical data.
Technological Characteristics
Portable device with PCBA and lithium-ion battery in polycarbonate housing. Dimensions: 2.1" x 3.3" x 0.75". Delivers biphasic, pure sinusoid interferential current. Connectivity via Bluetooth to mobile app. Complies with ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, and IEEE ANSI C63.27.
Indications for Use
Indicated for symptomatic relief of chronic intractable pain and management of post-traumatic or post-surgical pain in patients requiring electrical stimulation therapy.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 8, 2024
Artaflex Inc. Gerry Iuliano Owner 181 Whitehall Drive Markham, Ontario L3T 9T1 Canada
Re: K231575
Trade/Device Name: Vecttor VT-300 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: LIH Dated: February 8, 2024 Received: February 8, 2024
Dear Gerry Iuliano:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Robert Kang -S
for Pamela Scott, MS Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231575
Device Name Vecttor VT-300
Indications for Use (Describe)
Electrical stimulation delivered by the device is intended for the symptomatic relief of chronic intractable pain and or management of post-traumatic or post-surgical pain.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
# VECTTOR VT-300
# Submitter's name, address, telephone number, a contact person:
Gerry Iuliano, Owner
Artaflex Inc. - 181 Whitehall Dr., Markham Ontario L3T 9T1 Canada.
TEL: 905.470.0109 | FAX: 905.470.0621
# Date the summary was prepared;
February 6, 2024
#### Name of the device
Trade Name: VECTTOR VT-300 Common Name: Interferential Current Therapy Device Classification Name: Interferential Current Therapy Product Code: LIH
#### Identification of the legally marketed device to which the submitter claims equivalence.
# K121662 VECTTOR VT-200
# Description of the device
VT-300 device is portable and battery operated. VT-300 Device is comprised of a PCBA and a lithium-ion battery enclosed in a polycarbonate housing. The device's dimensions are 2.1 inches wide, 3.3 inches long, and 0.75 inch in height.
Vecttor VT-300 provides Interferential Current Therapy. Eight electrodes attached to a pair of lead wires are intended to be positioned on the patient's body to deliver four channels of current from VT-300 dual output jacks. Temperature sensors continually monitor finger surface temperatures with the intention of identifying and then applying an Interferential Current Therapy level that results in specified temperature increases.
The user controls the VT-300 device using Vecttor Mobile App which is a mobile application downloadable on Bluetooth enabled mobile devices using Android or iOS operating systems from the Google Play or Apple App store. Vecttor Mobile App provides the necessary user interface to establish connection to the VT-300 device via Bluetooth, and operate VT-300 device to provide interferential current therapy.
{5}------------------------------------------------
#### A statement of the intended use of the device
Electrical stimulation delivered by the device is intended for the symptomatic relief of chronic intractable pain and/or management of post-traumatic or post-surgical pain.
#### Comparison of the technological characteristics with predicate device
Vecttor VT-300 represents a modified version of predicate device Vecttor VT-200 (K121662) implementing design changes requested by applicant of K121662 for VT-200 predicate device, Dr. Donald Rhodes:
- Modernized redesign including reduced size, wiring, and relocation of connection ports .
- . Change power supply to on-board battery
- Change thermistor temperature sensor to digital temperature sensor .
- Change SD card with cloud-based application for managing patient treatments and data .
- Change device display screen and tactile buttons interface to mobile device screen and . software application
VT-300 has the same intended use and similar technological characteristics and principles of operation as K121662 for VT-200 predicate device. Any differences in technological characteristics potentially raising questions of safety or effectiveness have been addressed through Bench Performance Testing and compliance with FDA recognized consensus standards.
{6}------------------------------------------------
| OUTPUT SPECIFICATIONS | | | |
|-----------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| # | Characteristics | Subject Device | Predicate Device<br>(K121662) |
| 0 | 510(k) Number | K231575 | K121662 |
| 1 | Waveform (e.g., pulsed<br>monophasic, biphasic) | Biphasic | Biphasic |
| 2 | Shape (e.g., rectangular,<br>spike, rectified sinusoidal) | Pure Sinusoid | Pure Sinusoid |
| 3 | Maximum Output Voltage<br>(volts p/p) | 4.36 @ 500 Ω<br>6.72 @ 2 kΩ<br>8.08 @10 kΩ | 4.36 @ 500 Ω<br>6.72 @ 2 kΩ<br>8.08 @10 kΩ |
| 4 | Maximum Output Current (mA<br>RMS) | 3.08 @ 500 Ω<br>1.19 @ 2 kΩ<br>0.29 @10 kΩ | 3.08 @ 500 Ω<br>1.19 @ 2 kΩ<br>0.29 @10 kΩ |
| 5 | Pulse Width (µsec) | Continuous Sine Wave | Continuous Sine Wave |
| 6 | Frequency (Hz) [or Rate (pps)] | 1500Hz – 1580Hz<br>2500Hz – 2580Hz | 1500Hz – 1580Hz<br>2500Hz – 2580Hz |
| 7 | For interferential modes only: -<br>Beat Frequency (Hz) | 1000Hz | 1000Hz |
| 8 | For multiphasic waveforms<br>only: - Symmetrical phases? | Yes | Yes |
| 9 | - Phase Duration (include<br>units) (state range, if<br>applicable) (both phases, if<br>asymmetrical) | N/A | N/A |
| 10 | Net Charge (mC per pulse) | 0 @ 500 Ω<br>Outputs have no DC<br>output by means of DC<br>blocking capacitor and<br>isolation transformers at<br>outputs | 0 @ 500 Ω<br>Outputs have no DC<br>output by means of DC<br>blocking capacitor and<br>isolation transformers at<br>outputs |
| 11 | Maximum Phase Charge, (µC)<br>@500 Ω | 0.93 uC | 0.93 uC |
| 12 | Maximum Current Density,<br>(mA/cm², r.m.s.) @500 Ω | 0.39 | 0.39 |
| 13 | Maximum Power Density,<br>(W/cm²) using smallest<br>electrode conductive surface<br>area @500 Ω | 0.6 | 0.6 |
| 14 | Burst Mode | N/A | N/A |
| 15 | ON Time (seconds) | 5 | 5 |
| 16 | OFF Time (seconds) | 1 | 1 |
| 17 | Average DC current through<br>electrodes when device is on<br>but no pulses are applied | Zero | Zero |
| | Mode or Program Name | 1 | 1 |
| | Duration of primary<br>(depolarizing) phase (msec) | N/A | N/A |
| | Maximum Average Current (<br>average absolute value), mA<br>@500 Ω | 0<br>(No DC contents<br>present) | 0<br>(No DC contents<br>present) |
{7}------------------------------------------------
#### A brief discussion of the nonclinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence
VT-300 system and software verification & validation has been conducted to ensure design output meets the design input requirements. In addition, VT-300 complies with the following FDA recognized consensus standards:
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012 Medical electrical equipment Part 1: ● General requirements for basic safety and essential performance
- . IEC 60601-1-2 Edition 4.0 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- EC 60601-1-11 Edition 2.0 2015-01 Medical electrical equipment Part 1-11: General . requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-10 Edition 2.1 2016-04 Medical electrical equipment Part 2-10: Particular . requirements for the basic safety and essential performance of nerve and muscle stimulators
- . IEEE ANSI C63.27-2017 American National Standard for Evaluation of Wireless Coexistence
{8}------------------------------------------------
#### A brief discussion of the clinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence.
No clinical tests are being submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence
#### Conclusions
VT-300 has the same intended use as its predicate device. Any differences in technological characteristics potentially raising questions of safety or effectiveness have been addressed through Bench Performance Testing and compliance with FDA recognized consensus standards. In conclusion, VT-300 is substantially equivalent to VT-200.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.