Rebuilder Nerve Guidance Conduit is indicated for the reconstruction of a peripheral nerve discontinuity up to 20 mm in patients who have sustained a complete division of a nerve.
Device Story
Permeable, bioresorbable poly(lactide-co-caprolactone)-based tube; serves as interface between nerve and surrounding tissue; provides protective environment for peripheral nerve repair; guides axonal growth across nerve gaps. Used in surgical settings for nerve repair. Provided sterile, non-pyrogenic, single-use; various sizes (1.5mm to 10mm inner diameter). Surgeon implants conduit at site of nerve injury; facilitates nerve regeneration and functional recovery. Benefits patient by providing structural guidance for axonal regrowth across discontinuities.
Clinical Evidence
No human clinical data. Evidence based on non-clinical bench testing and in vivo animal study. Bench testing: mechanical strength, suture retention, compression, chemical analysis, degradation, and biocompatibility (ISO 10993-1:2018). Animal study: rat sciatic nerve transection model (n=not specified) compared subject device to Neurolac Nerve Guide. Evaluated at 1, 3, and 6 months. Endpoints: macroscopic tissue adhesion, clinical pathology, animal wellness, Sciatic Function Index (walking tracks), and histological assessment (inflammation, nerve fiber density, axonal myelination). Results demonstrated substantial equivalence in facilitating nerve repair and functional recovery.
Technological Characteristics
Bioresorbable tube composed of poly(lactide-co-caprolactone), poly(lactic-co-glycolic acid), and polylactic acid-b-polyethylene glycol. Permeable design; inner diameters 1.5mm–10mm; length 3cm. Sterilized via Ethylene Oxide to SAL 10^-6. Single-use. Biocompatibility per ISO 10993-1:2018.
Indications for Use
Indicated for reconstruction of peripheral nerve discontinuity up to 20 mm in patients with complete nerve division.
Regulatory Classification
Identification
A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).
{0}------------------------------------------------
January 29, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. To the right of the square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
CelestRay Biotech Company, LLC Shanshan Xu CTO and Partner P.O. Box 341754 Bethesda, Maryland 20827-1754
Re: K230794
Trade/Device Name: Rebuilder Nerve Guidance Conduit Regulation Number: 21 CFR 882.5275 Regulation Name: Nerve Cuff Regulatory Class: Class II Product Code: JXI Dated: December 23, 2023 Received: December 29, 2023
Dear Shanshan Xu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Adam D. Digitally signed by Adam D. Pierce -S Digitally signed by Pierce -S 13:55:20 -05'00'
Adam Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological
{2}------------------------------------------------
and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230794
Device Name Rebuilder Nerve Guidance Conduit
Indications for Use (Describe)
Rebuilder Nerve Guidance Conduit is indicated for the reconstruction of a peripheral nerve discontinuity up to 20 mm in patients who have sustained a complete division of a nerve.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary K230794
### Applicant information
| Applicant Name: | CelestRay Biotech company, LLC |
|--------------------|----------------------------------------|
| Applicant Address: | PO BOX 341754, BETHESDA, MD 20827-1754 |
| Phone: | 202-808-5668 |
| Fax: | -- |
| Contact Person: | Shanshan Xu<br>CTO and Partner |
| Date Prepared: | January 25, 2024 |
#### Name of Device
| Device Common Name: | Nerve Guide |
|-----------------------------|----------------------------------|
| Device Trade Name: | Rebuilder Nerve Guidance Conduit |
| Device Classification Name: | Nerve Cuff |
| | 882.5275 |
| | Class II |
| | JXI |
#### Legally Marketed Devices to Which Substantial Equivalence is Claimed
| Primary Predicate Device | Neurolac Nerve guide, K112267<br>Polyganics BV. |
|-----------------------------|-------------------------------------------------------------------|
| Additional Predicate Device | Neurolac Nerve guide, K050573<br>Polyganics BV. |
| Reference Device | NeuroGen Nerve Guide, K011168<br>Integra LifeSciences Corporation |
#### Description of the device
The Rebuilder Nerve Guidance Conduit is designed to be a permeable and bioresorbable poly(lactide-co-caprolactone) based tube. It is designed to serve as an interface between the nerve and the surrounding tissue, providing a protective environment for peripheral nerve repair after injury. The Rebuilder Nerve Guidance Conduit offers guidance and protection for axonal growth across a nerve gap.
{5}------------------------------------------------
The Rebuilder Nerve Guidance Conduit is provided sterile, non-pyrogenic, for single use only, in various sizes (with product inner diameter ranging from 1.5mm to 10 mm). The Rebuilder Nerve Guidance Conduit is provided in double peel package.
#### Indications for Use
Rebuilder Nerve Guidance Conduit is indicated for the reconstruction of a peripheral nerve discontinuity up to 20 mm in patients who have sustained a complete division of a nerve.
| Haracteristic | Subject Device | Predicate | Predicate | Reference |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Device | Device | Device |
| Device Name | Rebuilder Nerve<br>Guidance<br>Conduit | Neurolac<br>Nerve Guide<br>(K112267) | Neurolac<br>Nerve Guide<br>(K050573) | Neurogen<br>Nerve Guide<br>(K01168) |
| Indications<br>for use | Rebuilder Nerve<br>Guidance<br>Conduit is<br>indicated for the<br>reconstruction of<br>a peripheral<br>nerve<br>discontinuity up<br>to 20 mm in<br>patients who<br>have sustained a<br>complete<br>division of a<br>nerve. | The Neurolac<br>nerve guide is<br>indicated for<br>the<br>reconstruction<br>of a peripheral<br>nerve<br>discontinuity<br>up to 20 mm in<br>patients who<br>have sustained<br>complete<br>a<br>division of a<br>nerve. | The Neurolac<br>nerve guide is<br>indicated for<br>the<br>reconstruction<br>of a peripheral<br>nerve<br>discontinuity<br>up to 20 mm in<br>patients who<br>have sustained<br>complete<br>a<br>division of a<br>nerve. | The Collagen<br>Nerve Cuff is<br>intended<br>for<br>use in repair of<br>peripheral<br>nerve<br>discontinuities<br>where<br>gap<br>closure can be<br>achieved<br>by<br>flexion of the<br>extremity. |
| Principle<br>of<br>Operation | To serve as an<br>interface<br>between the<br>nerve and the<br>surrounding<br>tissue, providing<br>a protective<br>environment for | To provide a<br>protective<br>environment<br>for peripheral<br>nerve<br>regeneration<br>after<br>injury<br>and to create a<br>conduit<br>to | To provide a<br>protective<br>environment<br>for peripheral<br>nerve<br>regeneration<br>after after<br>injury<br>and to create a<br>conduit<br>to | To provide a<br>protective<br>environment<br>for peripheral<br>nerve repair<br>after injury, to<br>be an interface<br>between<br>the<br>nerve<br>and |
| | peripheral nerve<br>repair after<br>injury, and to<br>offer guidance<br>and protection<br>for axonal<br>growth across a<br>nerve gap. | guide axonal<br>growth across<br>a nerve gap. | guide axonal<br>growth across<br>a nerve gap. | surrounding<br>tissue and to<br>create<br>a<br>conduit for<br>axonal growth<br>across a nerve<br>gap. |
| Design | A permeable tube<br>with inner<br>diameter ranging<br>from 1.5mm to<br>10 mm, and<br>length of 3 cm | A flexible and<br>transparent tube<br>with internal<br>diameter of 1.5-<br>3 mm, and<br>length of 3 cm. | A flexible and<br>transparent tube<br>with internal<br>diameter of 4-10<br>mm, and length<br>of 3 cm. | A porous tube<br>with inner<br>diameter of 2 -<br>7mm,<br>and<br>length of 2 cm<br>to 4 cm. |
| Material | poly(lactide-co-<br>caprolactone),<br>Poly<br>(lactic-co-<br>glycolic acid) and<br>Polylactic acid-b-<br>Polyethylene<br>glycol | poly(DL-<br>lactide-co-ε-<br>caprolactone) | poly(DL-<br>lactide-co-ε-<br>caprolactone) | Collagen |
| Resorbable<br>Material | Yes | Yes | Yes | Yes |
| Sterilization | Ethylene Oxide<br>to SAL of 10-6 | Ethylene<br>Oxide to SAL<br>of 10-6 | Ethylene<br>Oxide to SAL<br>of 10-6 | Ethylene<br>Oxide to SAL<br>of 10-6 |
| Single<br>Use<br>/Reuse | Single use only | Single<br>use<br>only | Single use<br>only | Single<br>use<br>only |
| Packaging | Tyvek paper-<br>plastic bag, and<br>an aluminum foil<br>bag | Single use,<br>polycarbonate<br>tray, single<br>Tyvek pouch,<br>cardboard box | Single use,<br>polycarbonate<br>tray, single<br>Tyvek pouch,<br>cardboard box | Double blister<br>packages |
| Product<br>Code | JXI | JXI | JXI | JXI |
| CFR Section | 882.5275 | 882.5275 | 882.5275 | 882.5275 |
# Summary/Comparison of Technical Characteristics
{6}------------------------------------------------
## Performance Testing
The Rebuilder Nerve Guidance Conduit and its predicates have been characterized for performance with mechanical strength to demonstrate substantial equivalence. Non-Clinical Tests were conducted and performed on the subject device and predicate device,
{7}------------------------------------------------
including suture retention strength, compression properties, as well as chemical analysis and degradation properties. The test demonstrates the proposed device is substantial equivalent to the legally marketed predicate device.
The biocompatibility testing of the product was performed according to ISO 10993-1:2018 Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process under the category of implant device, tissue/bone contact, permanent duration (>30 days), including Cytotoxicity, Sensitization, Intracutaneous reactivity, Systemic toxicity(acute), Genotoxicity, Implantation (with histology of the surrounding tissue). Hemolysis, Pyrogenicity, Subchronic systemic toxicity and Chronic systemic toxicity. All test results showed that the Rebuilder Nerve Guidance Conduit is biocompatible. The tests helped demonstrate that the device is substantially equivalent to its predicate.
In vivo animal testing was performed using a rat nerve sciatic nerve transection model. The transected nerves were implanted with either the Rebuilder Nerve Guidance Conduit (test article) to the Neurolac Nerve Guide (comparator control), and rats were evaluated at the 1-, 3- and 6-month evaluation time points for healing responses. Outcome measures included macroscopic assessment of the nerve implant sites for tissue adhesions, clinical pathology (hematology and serum chemistry), in-life assessments of animal wellness, functional recovery as assessed using a Sciatic Function Index of walking tracks, and semiquantitative histological assessments for inflammation, nerve fiber density, axonal myelination. The results demonstrate the substantial equivalence of the Rebuilder Nerve Guidance Conduit to the Neurolac Nerve Guide in facilitating nerve repair and functional recovery.
#### Conclusion of Non-clinical test
The results of the mechanical performance, biocompatibility and animal testing demonstrated that the Rebuilder Nerve Guidance Conduit is substantially equivalent to Neurolac Nerve Guide, K112267.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.