NeuroGen" Nerve Guide is indicated for repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.
Device Story
NeuroGen™ Nerve Guide is a sterile, porous, collagen-based tubular implant designed for peripheral nerve repair. It acts as a conduit between nerve ends and surrounding tissue, providing a protective environment to facilitate axonal growth across nerve gaps. The device is flexible, allowing for joint and tendon movement while resisting occlusive forces from surrounding tissue. It is intended for surgical use by physicians to bridge nerve discontinuities. By maintaining a stable, protective environment, the device supports natural nerve regeneration, potentially improving functional recovery for patients with peripheral nerve injuries.
Clinical Evidence
Evidence includes a long-term (3.5 years) primate study and rodent animal models. Results demonstrated performance substantially equivalent to nerve graft, direct suture, and silicone tubes. Biocompatibility testing confirmed the device is non-cytotoxic, non-pyrogenic, non-irritating, non-sensitizing, non-toxic, and non-mutagenic. Mechanical and physical testing verified suture retention, compression resistance, and appropriate porosity/permeability for axonal growth.
Technological Characteristics
Tubular, porous collagen conduit; bioresorbable material. Designed for flexibility and resistance to occlusive forces. Meets ISO 10993 biocompatibility standards. Provided sterile in double blister packaging.
Indications for Use
Indicated for repair of peripheral nerve discontinuities in patients where gap closure is achievable via extremity flexion.
Regulatory Classification
Identification
A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).
Predicate Devices
Salumedica™ Nerve Cuff
NeuroTube®
Fastube™ Nerve Cuff
Submission Summary (Full Text)
{0}------------------------------------------------
# NeuroGen™ Nerve Guide
JUN 2 2 2001
## 510(K) SUMMARY
#### Submitter's name and address:
Integra LifeSciences Corporation 105 Morgan Lane Plainsboro, NJ 08536 USA
#### Contact person and telephone number:
Judith E. O'Grady Senior Vice President, Regulatory Affairs, Quality Assurance and Clinical Research (609) 275-0500
#### Date Summary was prepared:
April 16, 2001
#### Name of the device:
| Proprietary Name: | NeuroGen™ |
|----------------------|--------------------------------|
| Common Name: | Nerve Guide |
| Classification Name: | Nerve Cuff, Product Code 84JXI |
#### Substantial Equivalence:
NeuroGen™ Nerve Guide is substantially equivalent in function and intended use to the following products which have been cleared to market under Premarket Notifications 510(k): Salumedica" Nerve Cuff, NeuroTube and Fastube" Nerve Cuff.
## Device Description:
NeuroGen" Nerve Guide is an implant designed for repair of peripheral nerve discontinuities. NeuroGen™ Nerve Guide provides a protective environment for peripheral nerve repair after injury. NeuroGen" Nerve Guide is designed to be an interface between the nerve and surrounding tissue and to create a conduit for axonal growth across a nerve gap.
NeuroGen" Nerve Guide is flexible to accommodate movement of joint and associated tendons while retaining its shape and is resistant to occlusive forces from surrounding tissue. When hydrated, NeuroGen" is an easy to handle, soft, pliable, nonfriable, porous collagen tube. NeuroGen™ Nerve Guides are provided sterile in double blister packages in a variety of sizes.
## Intended Use:
NeuroGen" Nerve Guide is indicated for repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.
{1}------------------------------------------------
## Safety
Biocompatibility studies have demonstrated NeuroGen™ Nerve Guides to be: noncytotoxic, nonpyrogenic, nonirritating, and nonsensitizing. The following studies were conducted:
- a) Cytotoxicity
- b) Irritation / Intracutaneous Reactivity
- Sensitization c)
- d) Acute Systemic Toxicity
- e) Subchronic Toxicity
- Chronic Toxicity f)
- Genotoxicity g)
- h) Implantation
- Hemolysis i)
## Performance Characteristics:
The effectiveness of NeuroGen™ Nerve Guide to repair peripheral nerve discontinuities was studied in a long-term (3.5 years) primate study and in rodent animal models. The studies demonstrated that NeuroGen™ Nerve Guides are substantially equivalent to nerve graft, direct suture and silicone tubes.
The mechanical and physical characteristics of the NeuroGen™ Nerve Guides were evaluated in a series of tests. These tests were conducted to ensure that the NeuroGen™ Nerve Guides possess the mechanical properties (suture retention and mechanical compression) as well as physical properties (porosity and permeability) that determine their suitability for use in the human body. Testing has demonstrated that the nerve guides are able to hold a suture, resist repeated compression from surrounding tissues, have a porous outer surface and tube wall, and allow the passage of molecules of specific size through the tube wall.
### Technological Characteristics Compared to Predicate Devices:
NeuroGen™ Nerve Guide is a tubular device which is equivalent to the predicate devices, Salumedica™ Nerve Cuff. NeuroTube® and Fastube™ Nerve Cuff in its design for repair of peripheral nerve discontinuities. Like the predicate devices, NeuroGen" is provided sterile, for single use only. The NeuroGen™ Nerve Guide is manufactured from a bioresorbable material, as is one of the predicate devices, NeuroTube®. NeuroGen™ Nerve Guide meets ISO 10993 requirements for Biocompatibility testing.
### Conclusion
NeuroGen" Nerve Guide is indicated for the repair of nerve discontinuities where gap closure can be achieved by flexion of the extremity. NeuroGen™ Nerve Guide is flexible to accommodate movement of joint and associated tendons while retaining its shape and is resistant to occlusive forces from surrounding tissue.
Biocompatibility studies have demonstrated NeuroGen" Nerve Guide to be non-cytotoxic, nonsensitizing, non-toxic and non-mutagenic. Extensive, long-term evaluations in primates demonstrates NeuroGen™ Nerve Guide to biocompatible and provides an environment for axonal growth.
Valid scientific evidence through substantial testing of descriptive characteristics, Biocompatibility, mechanical and physical property testing and extensive performance testing in a primate model, provide reasonable assurance that NeuroGen™ Nerve Guide is safe and effective under the proposed conditions of use, and substantially equivalent to its predicate devices.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human profiles or a bird in flight, composed of curved lines.
#### Public Health Service
JUN 2 2 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Judith E. O'Grady, RN, MSN Senior Vice President, Regulatory Affairs, Quality Assurance and Clinical Affairs Integra Life Sciences Corporation 105 Morgan Lane Plainsboro, New Jersey 08536
Re: K011168
Trade/Device Name: NeuroGen Nerve Guide Regulation Number: 882.5275 Regulatory Class: II Product Code: JXI Dated: April 16, 2001 Received: April 17, 2001
Dear Ms. O'Grady:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Ms. Judith E. O'Grady, RN, MSN
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Marlen Millenor
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number:
K011168
Device Name: NeuroGen™ Nerve Guide
## Indications for Use
NeuroGen™ Nerve Guide is indicated for repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.
## PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
(Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-the-Counter Use X (Per 21 CFR 801.109)
Mark N. Milhusan
(Division Sign-Off) (Division of General, Restorative and Neurological Devices
510(k) Number -
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.