← Product Code [JXG](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/JXG) · K221840

# Hakim Programmable Valves, Hakim Precision Fixed Pressure Valves (K221840)

_Integra Lifesciences Production Corporation · JXG · Oct 14, 2022 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/JXG/K221840

## Device Facts

- **Applicant:** Integra Lifesciences Production Corporation
- **Product Code:** [JXG](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/JXG.md)
- **Decision Date:** Oct 14, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5550
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Codman Hakim Precision Fixed Pressure Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus. The Codman Hakim Programmable Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.

## Device Story

Implantable, sterile, single-use pressure-regulating valves; manage hydrocephalus by diverting excess cerebrospinal fluid (CSF) from brain ventricles to another body cavity for absorption. Hakim Precision Valves offer fixed opening pressures (5 ranges); Hakim Programmable Valves allow non-invasive adjustment (18 settings). Devices maintain constant intraventricular pressure. Used by neurosurgeons in clinical settings. Output is mechanical regulation of CSF flow; clinical decision-making relies on valve pressure settings to manage intracranial pressure. Benefits include relief of intracranial pressure and management of hydrocephalus symptoms.

## Clinical Evidence

No clinical studies performed. Substantial equivalence supported by bench testing and biocompatibility assessments. Bench testing included functional testing (ISO 7197, ASTM F647), shelf-life (ASTM F1980), MRI spatial gradient (ASTM F2052), and X-ray (ASTM F640). Biocompatibility testing followed ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, implantation, genotoxicity, hemolysis) and USP standards. All tests passed.

## Technological Characteristics

Implantable pressure-regulating valves. Materials: Polysulfone (PSU). Standards: ISO 7197, ASTM F647, ASTM F1980, ASTM F2052, ASTM F640, ISO 10993 series. Energy: Passive mechanical pressure regulation. Connectivity: None. Sterilization: Not specified (equivalency assessment performed).

## Regulatory Identification

A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

## Predicate Devices

- Hakim Precision Valves and Hakim Programmable Valves ([K172022](/device/K172022.md))

## Reference Devices

- Codman Certas Plus Programmable Valve ([K143111](/device/K143111.md), [K182265](/device/K182265.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.

October 14, 2022

Integra LifeSciences Production Corporation Amanda Erwin Manager, Regulatory Affairs 11 Cabot Boulevard Mansfield, Massachusetts 02048

Re: K221840

Trade/Device Name: Hakim Programmable Valves, Hakim Precision Fixed Pressure Valves Regulation Number: 21 CFR 882.5550 Regulation Name: Central nervous system fluid shunt and components Regulatory Class: Class II Product Code: JXG Dated: July 18, 2022 Received: July 18, 2022

Dear Amanda Erwin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221840

Device Name Hakim Precision Fixed Pressure Valves Hakim Programmable Valves

#### Indications for Use (Describe)

The Codman Hakim Precision Fixed Pressure Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.

The Codman Hakim Programmable Valves Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.

Type of Use (Select one or both, as applicable)

| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Hakim® Precision Fixed Pressure Valves and Hakim® Programmable Valves

#### 510(k) Summary

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

| 807.92(a) (1) Submitter Information |                                                                                                                                                                                              |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                | Integra LifeSciences Production Corporation                                                                                                                                                  |
| Address                             | 11 Cabot Boulevard<br>Mansfield, MA 02048                                                                                                                                                    |
| Telephone number                    | (609) 627-9053                                                                                                                                                                               |
| Primary Contact                     | Amanda Erwin                                                                                                                                                                                 |
| Date Summary Prepared               | June 23, 2022                                                                                                                                                                                |
| 807.92(a) (2) Name of Device        |                                                                                                                                                                                              |
| Trade or Proprietary Name           | Hakim Precision Valves<br>Hakim Programmable Valves                                                                                                                                          |
| Common Name                         | Central Nervous System Fluid Shunt and Components                                                                                                                                            |
| Classification Name                 | Central Nervous System Fluid Shunt and Components<br>(21 CFR 882.5550)                                                                                                                       |
| Device Class                        | II                                                                                                                                                                                           |
| Product Code                        | JXG                                                                                                                                                                                          |
| 807.92(a) (3) Predicate Information |                                                                                                                                                                                              |
| Predicate Device                    | Hakim Precision Valves and Hakim Programmable Valves:<br>K172022<br>The following reference device is used in this submission:<br>Codman Certas Plus Programmable Valve: K143111,<br>K182265 |
| 807.92(a) (4) Device Description    |                                                                                                                                                                                              |

The Codman Hakim® Precision Fixed Pressure and Programmable Valves are implantable, sterile, single use devices that provide constant intraventricular pressure and drainage of cerebrospinal fluid (CSF) for the management of hydrocephalus. Hydrocephalus is a condition caused by excessive accumulation of CSF in the ventricles of the brain due to a disturbance of

{4}------------------------------------------------

Hakim® Precision Fixed Pressure Valves and Hakim® Programmable Valves

CSF secretion, flow, or absorption, which causes a rise in intracranial pressure (ICP). To relieve ICP, CSF can be diverted through a shunting device, such as a Hakim Precision Valve or Hakim Programmable Valve, to another body cavity where it is a subsequently absorbed. Both the Codman Hakim Precision Fixed Pressure and Programmable Valves are pressure regulating valves which maintain intraventricular pressure at a constant level. The Hakim Precision valves are fixed pressure valves and are available in 5 different opening pressure ranges. The Codman Hakim Programmable Valves, not having fixed pressures, permit non-invasive adjustment of the valve opening pressure. The Codman Hakim Programmable Valves can be adjusted to 18 different opening pressure settings.

## 807.92(a) (5) Indications for Use

The Codman Hakim Precision Fixed Pressure Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.

The Codman Hakim Programmable Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.

807.92(a) (6) Technological Characteristics Compared to Predicate

The proposed Hakim Precision Valves and Hakim Programmable Valves have the same intended use, sterility, design principles and fundamental operation as the predicate valves. The proposed minor change in technological characteristics for the valves consists of a material change and minor dimensional and tolerance changes for various components. The changes do not raise any new questions of safety and/or effectiveness.

| Component Affected                                                   | Proposed Modification                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Rationale                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Various Internal<br>Components                                       | • Replacement of<br>Polyethersulfone<br>(PES) with<br>Polysulfone (PSU) in<br>Hakim Precision<br>Fixed Pressure<br>Valves and Hakim<br>Programmable Valves                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Polyethersulfone will no<br>longer be available for<br>Integra to manufacture the<br>impacted internal<br>components with. Nylon will<br>no longer be available to<br>manufacture the impacted<br>Hakim Programmable                                                                                                                                                                                                                                                                                               |
| Labeling                                                             | for various internal<br>components. Minor<br>dimensional and<br>tolerance changes to<br>internal components<br>will also be made to<br>the internal<br>components to<br>accommodate the<br>material change.<br>Replacement of<br>Nylon with<br>Polysulfone (PSU) in<br>Hakim programmable<br>Valve internal<br>component. Minor<br>dimensional changes<br>will be made to<br>accommodate the<br>material change.   Updates made to<br>reflect the proposed<br>material change, as<br>well as administrative<br>updates and<br>corrections. Inclusion of Patient<br>Implant Cards and<br>Patient Information<br>Leaflet The labelling for<br>Hakim Programmable<br>Valves has also been<br>updated to comply<br>with Regulation (EU)<br>2017/745. | Valves Needle Guard<br>internal component with.<br>The minor dimensional and<br>tolerance changes to the<br>internal components have<br>been made to accommodate<br>the material change or<br>standardize component<br>design across Integra's<br>hydrocephalus valve<br>portfolio.<br>Update labelling with<br>current information. Patient Implant Cards<br>and Patient<br>Information Leaflets<br>are included as the<br>devices are MR<br>conditional implants. To comply with<br>Regulation (EU)<br>2017/745. |
| 807.92(b) 1-2: Summary of Nonclinical and Clinical Testing Performed |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{5}------------------------------------------------

Hakim® Precision Fixed Pressure Valves and Hakim® Programmable Valves

{6}------------------------------------------------

Hakim® Precision Fixed Pressure Valves and Hakim® Programmable Valves

The following performance testing has been conducted in support of the substantial equivalence determination. The testing utilized well-established methods, including those from FDA consensus standards. All testing was performed on production equivalent devices.

| Performance Bench Test Results              |            |
|---------------------------------------------|------------|
| Test                                        | Conclusion |
| Product Shelf Life Testing per ASTM F1980   | Pass       |
| Functional Testing per ISO 7197             | Pass       |
| Functional Testing per ASTM F647            | Pass       |
| MRI Spatial Gradient Testing per ASTM F2052 | Pass       |
| X-Ray Testing per ASTM F640                 | Pass       |

| Biocompatibility Testing Results                                                                                                          |            |
|-------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Test                                                                                                                                      | Conclusion |
| MTT and MTS Cytotoxicity Studies per ISO<br>10993-5                                                                                       | Pass       |
| Guinea Pig Maximization Sensitization Studies per<br>ISO 10993-10                                                                         | Pass       |
| Intracutaneous Irritation Studies in Rabbits per ISO<br>10993-10                                                                          | Pass       |
| Acute Systemic Toxicity Study in Mice per ISO<br>10993-11                                                                                 | Pass       |
| Rabbit Pyrogen Studies per United States<br>Pharmacopeia (USP 42 – NF 37)                                                                 | Pass       |
| Subcutaneous Implantation Studies in Rabbits, 1<br>Week and 4 Weeks per ISO 10993-6                                                       | Pass       |
| Systemic Toxicity and Local Effects Study in<br>Rabbits Following Subcutaneous Implantation, 13<br>Weeks per ISO 10993-6 and ISO 10993-11 | Pass       |
| Bacterial Reverse Mutation Studies per ISO 10993-<br>3 and ISO/TR 10993-33                                                                | Pass       |
| In Vitro Mouse Lymphoma Studies per ISO 10993-<br>3 and ISO/TR 10993-33                                                                   | Pass       |
| Hemolysis on Extract Studies per ISO 10993-4 and<br>ASTM F756                                                                             | Pass       |

There are no changes in sterility method as a result of the proposed material change and minor dimensional and tolerance changes; a sterilization equivalency assessment was performed comparing the proposed device to the predicate device and deemed acceptable.

{7}------------------------------------------------

Hakim® Precision Fixed Pressure Valves and Hakim® Programmable Valves

No clinical studies were required as appropriate verification of the subject device was achieved based on the comparison to the predicate device and from the results of testing.

#### 807.92(b) (3) Conclusion

Based upon the intended use, design, comparison to the predicate device, and testing performed, Integra LifeSciences believes that the proposed modifications to the Hakim Precision Valves and Hakim Programmable Valves do not raise any new questions of safety and effectiveness, and is therefore, substantially equivalent to the predicate Hakim Precision Valves and Hakim Programmable Valves.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/JXG/K221840](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/JXG/K221840)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
