INTEGRA FLOW REGULATING VALVE SYSTEMS OSV II MODELS AND INTEGRA FLOW REGULATING VALVE LOW FLOW MODELS
Applicant
Integra LifeSciences Corporation
Product Code
JXG · Neurology
Decision Date
Jan 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Models for ventricular application: The Integra Flow Regulating Valve Systems are implantable systems used in the treatment of patients with hydocephalus, to shunt CSF from the ventricles to the peritoneal cavity or other appropriate drainage site such as the heart's right atrium. Models for lumbar application: The Lumbar Integra Flow Regulating Valve Systems are implantable systems used in the treatment of patients with communicating hydrocephalus to shunt CSF from the lumbar subarachnoid region to the peritoneal cavity.
Device Story
Implantable hydrocephalus shunt system; regulates CSF drainage from ventricles or lumbar subarachnoid space to peritoneal cavity or heart's right atrium. Device utilizes variable resistance valve mechanism to maintain constant drainage rate (standard: ~20ml/hr; low flow: ~10ml/hr) within physiological intracranial pressure ranges; includes safety pressure relief mode to prevent intracranial hypertension. System components include proximal catheter, valve unit, drainage catheter, and optional antechamber/burr hole cap. Used by neurosurgeons in clinical/surgical settings. Provides controlled CSF diversion to manage hydrocephalus symptoms; benefits patient by preventing excessive intracranial pressure accumulation.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing including load testing (up to 15N), water testing (37°C for 28 days), and biocompatibility testing (sensitization, genotoxicity, pyrogenicity) per ISO 10993 and USP standards. MR conditional testing performed per ASTM F2182-09, F2119-07, F2052-06e1, and F2213-06.
Technological Characteristics
Implantable variable resistance valve system. Materials include medical-grade components with a modified epoxy glue. Features two flow ranges (standard 18-30ml/h; low 8-17ml/h). MR Conditional per ASTM F2182-09, F2119-07, F2052-06e1, F2213-06. Sterilized via Ethylene Oxide; single-use. Mechanical flow regulation based on pressure-sensitive resistance.
Indications for Use
Indicated for patients with hydrocephalus (ventricular application) or communicating hydrocephalus (lumbar application) requiring CSF shunting to the peritoneal cavity or heart's right atrium.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
OSV II Lumbar and Integra Flow Regulating Valve Low Flow (Lumbar) (K092395)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6, 2015
Integra Lifesciences Corporation Janet Kay Director, Regulatory Affairs 22 Terry Avenue Burlington, Massachusetts 01803
Re: K140722
> Trade/Device Name: OSV II Valve Systems, Integra Flow Regulating Valve Low Flow Valve Systems Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt And Components Regulatory Class: Class II Product Code: JXG Dated: December 22, 2014 Received: December 24, 2014
Dear Ms. Kay,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Carlos L. Pena =5/△
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140722
Device Name OSV II Valve Systems Integra Flow Regulating Valve Low Flow Valve Systems
Indications for Use (Describe)
Models for ventricular application:
The Integra Flow Regulating Valve Systems are implantable systems used in the treatment of patients with hydocephalus, to shunt CSF from the ventricles to the peritoneal cavity or other appropriate drainage site such as the heart's right atrium.
Models for lumbar application:
The Lumbar Integra Flow Regulating Valve Systems are implantable systems used in the treatment of patients with communicating hydrocephalus to shunt CSF from the lumbar subarachnoid region to the peritoneal cavity.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
| 807.92(a)(1) - Submitter information | | |
|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Name | Integra LifeSciences Corporation | |
| Address | 22 Terry Avenue | |
| | Burlington, MA 01830 | |
| Phone Number | (781) 565-1347 | |
| Fax Number | (781) 238-0645 | |
| Establishment | | |
| Registration Number | 1222895 | |
| Name of Contact Person | Janet Kay | |
| Date Prepared | January 5, 2015 | |
| 807.92(a)(2) – Name of device | | |
| Trade or Propriety Names | OSV II® Valve Systems | |
| | Integra Flow Regulating Valve Low Flow Valve Systems | |
| | | |
| Common or Usual Name | Hydrocephalus Shunt System and Components | |
| Classification Name | Central Nervous System Fluid Shunt and Components | |
| Classification Panel | Neurology | |
| Regulation | Class II, under 21 CFR 882.5550 | |
| Product Code(s) | JXG | |
| 807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed | | |
| Equivalence is claimed to current Integra Flow Regulating Valves cleared under the | | |
| following Premarket Notifications: | | |
| K042192 - Integra Flow Regulating Valve Low Flow (Standard) | | |
| K081773 – OSV II and OSV II Low Pro | | |
| K093968- Integra Flow Regulating Valve Low Flow (Mini) | | |
| K092395 - OSV II Lumbar and Integra Flow Regulating Valve Low Flow (Lumbar) | | |
| 807.92(a)(4) - Device description | | |
| The Integra Flow Regulating Hydrocephalus Valve Systems are implantable devices for | | |
| controlled cerebrospinal fluid (CSF) drainage from the ventricles to the peritoneal cavity | | |
| or other appropriate drainage site such as the heart's right atrium (ventricular application) | | |
| or from the lumbar subarachnoid region to the peritoneal cavity (lumbar application). | | |
| Unlike conventional valves, these are variable resistance valves which maintain a | | |
| | | |
| 807.92(a)(5) – Intended use of the device | | |
| Indications for Use | | |
| | Models for ventricular application:<br>The Integra Flow Regulating Valve Systems are implantable<br>systems used in the treatment of patients with<br>hydrocephalus, to shunt CSF from the ventricles to the<br>peritoneal cavity or other appropriate drainage site such as<br>the heart's right atrium. | |
| | Models for lumbar application:<br>The Lumbar Integra Flow Regulating Valve Systems are<br>implantable systems used in the treatment of patients with<br>communicating hydrocephalus to shunt CSF from the<br>lumbar subarachnoid region to the peritoneal cavity. | |
drainage rate constant within the physiological range (for the specified populations and disorders) of intracranial pressure. The mechanism incorporates a safety pressure relief mode to prevent accidental intracranial hypertension.
The various models include two flow regulating ranges, two antechamber sizes, and
{4}------------------------------------------------
various accessories allowing either a ventricular or a lumbar application. The two flow regulating ranges are:
Standard flow regulating: maintains a drainage rate close to the normal CSF secretion rate, around 20ml/hr (18-30ml/h).
Low flow regulating: maintains a lower drainage rate than OSV II Valve Systems, around 10ml/hr (8-17ml/h).
The systems are available in different configurations. A system generally consists of a ventricular or lumbar (proximal) catheter, a valve unit, and a drainage (distal) catheter. Some configurations contain an antechamber (standard or low profile) or a burr hole cap.
Various accessories are available for the implantation procedure or to be implanted in addition to the valves.
All products are sold sterilized, for single use only
807.92(a)(5) - Intended use of the device
807.92(a)(6) Summary of the technological characteristics of the device compared to the predicate
The modified devices have the same design, materials (except a change of epoxy glue), indications, contraindications, packaging, manufacturing and sterilization processes as the predicate devices. There is no change to the products except for the epoxy glue.
## 807.92(b)(1-2) – Nonclinical tests submitted
The following tests were performed and described below. Samples were environmentally conditioned, aged, and 2x sterilized by Ethylene Oxide.
Load Testing Load testing was conducted up to 15N (1.5kg).
Water Test
{5}------------------------------------------------
Water testing was conducted at 37℃ for 28 days.
Biocompatibility testing
The biocompatibility evaluation for the modification to the Integra Flow Regulating Valve Systems (Integra Flow Regulating Valve Low Flow (Standard), OSV II and OSV II Low Pro and Integra Flow Regulating Valve Low Flow (Mini)) and the Lumbar Integra Flow Regulating Valve Systems (OSV II Lumbar and Integra Flow Regulating Valve Low Flow (Lumbar)) devices were conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995 and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
| Sensitization | ISO 10993-10:2010 |
|----------------------------------------------------------|-------------------|
| Genotoxicity | |
| Bacterial Reverse Mutation Study | ISO 10993-3:2003 |
| In Vitro Chromosomal Aberration Study in Mammalian Cells | ISO 10993-3:2003 |
| Mouse Peripheral Blood Micronucleus Test | ISO 10993-3:2003 |
| Pyrogen | |
| Rabbit Pyrogen Study | USP35, NF30<151> |
The flow regulating valve systems are considered implantable devices. tissue and/or bone contacting and are permanent implants (>30 days). The biocompatibility testing and design verification testing have verified that the change of glue has no impact on the product specifications, performance and reliability.
Additional testing was conducted and concluded to the MR Conditional status of the valves. The battery of testing included the following tests: Radio Frequency Induced Heating Test ASTM F2182-09 MR Image Artifacts ASTM F2119-07 Magnetically Induced Displacement Force test ASTM F2052-06e1 Magnetically Induced Torque Test ASTM F2213-06
### 807.92(b)(3) - Conclusions drawn from non-clinical data
The predicate devices were cleared based on the results of non-clinical data. The modified devices and predicate device performance data were compared to support the safety of the modified devices and demonstrate that the Integra Flow Regulating Hydrocephalus Valve Systems should perform as intended in the specified use conditions.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.