K043041 · Aesculap, Inc. · HCH · Dec 9, 2004 · Neurology
Device Facts
Record ID
K043041
Device Name
AESCULAP T-BAR ANEURYSM CLIPS (TITANIUM & PHYNOX)
Applicant
Aesculap, Inc.
Product Code
HCH · Neurology
Decision Date
Dec 9, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aesculap T-Bar Aneurysm Clips are intended for occlusion of cerebral aneurysms in either a temporary or permanent manner.
Device Story
Aesculap T-Bar Aneurysm Clips are surgical implants used for temporary or permanent occlusion of cerebral aneurysms. Devices are fenestrated clips with jaws at 90 and 45-degree angles. Used in neurosurgical procedures; applied by surgeons using previously cleared Aesculap aneurysm clip appliers. Clips provide mechanical vessel occlusion to prevent aneurysm rupture or facilitate surgical access. No electronic or software components.
Clinical Evidence
Bench testing only. Biomechanical testing performed per FDA guidance demonstrated substantial equivalence to currently marketed predicate devices.
Technological Characteristics
Materials: Phynox (cobalt alloy) per ISO 5832/7 or Titanium Alloy per ISO 5832/3. Design: Fenestrated clips with 90 and 45-degree jaw angles. Mechanical device; no energy source, software, or connectivity.
Indications for Use
Indicated for occlusion of cerebral aneurysms in patients requiring temporary or permanent vessel occlusion during neurosurgical procedures.
Regulatory Classification
Identification
An aneurysm clip is a device used to occlude an intracranial aneurysm (a balloonlike sac formed on a blood vessel) to prevent it from bleeding or bursting.
Submission Summary (Full Text)
{0}------------------------------------------------
Page 1 of 1
# C. 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (in Accordance with SMDA of 1990)
DEC - 9 2004
#### Aesculap T-Bar Aneurysm Clips October 2004
| COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
|-----------------|--------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Matthew M. Hull<br>800-258-1946 x 5072 (phone)<br>610-791-6882 (fax) |
| TRADE NAME: | Aesculap T-Bar Aneurysm Clips |
| COMMON NAME: | Aneurysm Clips |
| DEVICE CLASS: | Class II |
| PRODUCT CODE: | 84 HCH |
| CLASSIFICATION: | 882.5200 - Clip, Aneurysm |
| REVIEW PANEL: | Neurology |
## INDICATIONS FOR USE
The Aesculap T-Bar Aneurysm Clips are intended for occlusion of cerebral aneurysms in either a temporary or permanent manner.
## DEVICE DESCRIPTION
The Aesculap T-Bar Aneurysm Clips are designed for temporary or permanent occlusion of vessels during neurosurgical procedures. They are made from either Phynox (oobalt alloy) per ISO 5832/7 or Titanium Alloy per ISO 5832/3. The clips are a fennstrated design with the jaws at 90 and 45 degree angles. They are applied with previously cleared Aesculap aneurysm clip appliers.
## PERFORMANCE DATA
All required testing per "Draft Guidance for the Preparation of Premarket Notifications (510(k)s)" was completed for Aesculap T-Bar Aneurysm Clips. Biomechanical testing results demonstrate the Aesculap T-Bar Aneurysm Clips are substantially equivalent to our current aneurysm clips currently on the market.
## SUBSTANTIAL EQUIVALENCE
Aesculap®, Inc. believes that the Aesculap T-Bar Aneurysm Clips are substantially equivalent to our currently marketed Aneurysm Clips.
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 9 2004
Mr. Matthew M. Hull, RAC Regulatory Affairs Manager Aesculap, Inc. 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K043041
Trade/Device Name: Aesculap T-Bar Aneurysm Clips Regulation Number: 21 CFR 882.5200 Regulation Name: Aneurysm Clip Regulatory Class: II Product Code: HCH Dated: November 1, 2004 Received: November 9, 2004
Dear Mr. Hull:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
#### Page 2 - Mr. Matthew M. Hull, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Muriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT B.
K043041 510(k) Number:
Device Name: Aesculap T-Bar Aneurysm Clips
Indication for Use:
The Aesculap T-Bar Aneurysm Clips are intended for occlusion of cerebral aneurysms in either a temporary or permanent manner.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
Page 1 of 1
**510(k) Number** K043041
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.