Axium Detachable Coil; Axium Prime Detachable Coil
Applicant
Micro Therapeutics Lnc. D/B/A Ev3 Neurovascular
Product Code
HCG · Neurology
Decision Date
Nov 7, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Axium™ Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. Axium™ Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Prime Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. The Axium™ Prime Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Prime Detachable Coil is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Prime Detachable Coils are also intended for arterial and venous embolizations in the peripheral vasculature.
Device Story
Neurovascular embolization system; consists of platinum/tungsten alloy coil attached to composite stainless steel hypotube delivery pusher; includes radiopaque positioning marker and fluorosafe markers on proximal pusher end. Used in endovascular procedures; operated by interventional neuroradiologists/physicians. Coil delivered via microcatheter to target site; detached mechanically using hand-held Instant Detacher (I.D.) unit. Provides permanent occlusion of aneurysms/vascular abnormalities; benefits patient by preventing rupture or reducing blood flow to malformations. Fluorosafe markers improve visualization during procedure.
Clinical Evidence
No clinical or animal data provided. Substantial equivalence supported by non-clinical bench testing, including visual inspection, marker dimensional inspection, corrosion resistance (ISO 10555-1), and simulated-use clinical evaluation of marker visibility.
Technological Characteristics
Platinum/tungsten alloy coil; polypropylene tension-safe filament; stainless steel hypotube pusher with PTFE jacket/liner. Fluorosafe markers added to proximal pusher. MR Conditional. Sterilized via Ethylene Oxide (EO). Single-use.
Indications for Use
Indicated for endovascular embolization of intracranial aneurysms, arteriovenous malformations, and arteriovenous fistulae. Axium™ Prime models also indicated for arterial and venous embolizations in peripheral vasculature.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
Predicate Devices
Axium™ Detachable Coil; Axium™ Prime Detachable Coil (K203432)
Submission Summary (Full Text)
{0}------------------------------------------------
November 7, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Lekshmi Pillai, MS Senior Regulatory Affairs Specialist 5290 California Avenue Irvine, California 92617
# Re: K233420
Trade/Device Name: Axium Detachable Coil; Axium Prime Detachable Coil Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: October 9, 2023 Received: October 10, 2023
# Dear Lekshmi Pillai:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233420
Device Name Axium™ Detachable Coil; Axium™ Prime Detachable Coil
# Indications for Use (Describe)
Axium™ Detachable Coil:
Axium™ Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. Axium™ Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
Axium™ Prime Detachable Coil:
(Models: APB-X-Y-3D-ES, APB-X-Y-3D-SS, APB-X-Y-HX-ES, APB-X-Y-HX-SS)
The Axium™ Prime Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. The Axium™ Prime Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
Axium™ Prime Detachable Coil:
(Models: FC-X-Y-3D)
The Axium™ Prime Detachable Col is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Prime Detachable Coils are also intended for arterial and venous embolizations in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
| | <span> <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--|---------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K233420 510(k) Summary
| 510(k) Owner: | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular<br>9775 Toledo Way<br>Irvine, CA 92618, USA<br>Establishment Registration: 2029214 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Lekshmi Pillai<br>Senior Regulatory Affairs Specialist<br>Telephone: (917) 291-3810<br>Email : lekshmiajaykumar.pillai@medtronic.com |
| Date Summary<br>Prepared: | 06 November 2023 |
|---------------------------|-------------------------------------------------------------------|
| Trade Name of Device: | Axium™ Detachable Coil<br>Axium™ Prime Detachable Coil |
| Common Name of<br>Device: | Neurovascular Embolization Device, Vascular Embolization Device |
| Review Panel: | Neurology, Cardiovascular |
| Product Code: | HCG, KRD |
| Regulation Number: | 21 CFR 882.5950; 21 CFR 870.3300 |
| Regulation Name: | Neurovascular Embolization Device, Vascular Embolization Device |
| Device Classification | Class II |
| Predicate Device: | K203432<br>Axium™ Detachable Coil<br>Axium™ Prime Detachable Coil |
## Device Description:
The Axium™ Detachable Coil and Axium™ Prime Detachable Coil consist of a platinum embolization coil attached to a composite implant delivery pusher with a radiopaque positioning marker and a hand-held Instant Detacher (I.D.) which when activated detaches the coil from the delivery pusher tip. The I.D. is sold separately.
## Indications for Use Statement:
| Axium™ Detachable Coil and Axium Prime Detachable Coil Indications for Use | | |
|----------------------------------------------------------------------------|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model Numbers | Trade Name | Indications for Use |
| QC-X-Y-HELIX | | |
| QC-X-Y-3D | Axium™<br>Detachable<br>Coil | Axium™ Detachable Coils are intended for the endovascular embolization<br>of intracranial aneurysms. Axium™ Detachable Coils are also intended for<br>the embolization of other neuro vascular abnormalities such as<br>arteriovenous malformations and arteriovenous fistulae. |
| NC-X-Y-HELIX | | |
| PC-X-Y-HELIX | | |
| PC-X-Y-3D | | |
{4}------------------------------------------------
| Axium™ Detachable Coil and Axium Prime Detachable Coil Indications for Use | | |
|----------------------------------------------------------------------------|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model Numbers | Trade Name | Indications for Use |
| APB-X-Y-3D-ES | Axium™<br>Prime<br>Detachable<br>Coil | The Axium™ Prime Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. The Axium™ Prime Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. |
| APB-X-Y-3D-SS | | |
| APB-X-Y-HX-ES | | |
| APB-X-Y-HX-SS | | |
| FC-X-Y-3D | Axium™<br>Prime<br>Detachable<br>Coil | The Axium™ Prime Detachable Coil is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Prime Detachable Coils are also intended for arterial and venous embolizations in the peripheral vasculature. |
The model numbers are formatted to summarize sizes available, where X is the coil loop outer diameter in mm and Y is the implant length in cm.
## Proposed Change:
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular requests clearance for the Axium™ Detachable Coil and Axium™ Prime Detachable Coil with addition of fluorosafe markers on the proximal end of the coil delivery pusher.
#### Device Comparison:
| Design Feature | Predicate Devices (K203432):<br>Axium™ Detachable Coil;<br>Axium™ Prime Detachable Coil | Subject Devices:<br>Axium™ Detachable<br>Coil;<br>Axium™ Prime<br>Detachable Coil |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Axium™ Detachable Coil:<br>Axium™ Detachable Coils are intended for the endovascular<br>embolization of intracranial aneurysms. Axium™ Detachable Coils<br>are also intended for the embolization of other neuro vascular<br>abnormalities such as arteriovenous malformations and<br>arteriovenous fistulae.<br><br>Axium™ Prime Detachable Coil (APB-X-Y-3D-ES, APB-X-Y-3D-SS,<br>APB-X-Y-HX-ES, APB-X-Y-HX-SS):<br>The Axium™ Prime Detachable Coils are intended for the<br>endovascular embolization of intracranial aneurysms. The Axium™<br>Prime Detachable Coils are also intended for the embolization of<br>other neuro vascular abnormalities such as arteriovenous<br>malformations and arteriovenous fistulae.<br><br>Axium™ Prime Detachable Coil (FC-X-Y-3D):<br>The Axium™ Prime Detachable Coil is intended for the endovascular<br>embolization of intracranial aneurysms and other neurovascular<br>abnormalities, such as arteriovenous malformations and<br>arteriovenous fistulae. The Axium™ Prime Detachable Coils are also | Same |
| Design Feature | Predicate Devices (K203432):<br>Axium™ Detachable Coil;<br>Axium™ Prime Detachable Coil | Subject Devices:<br>Axium™ Detachable<br>Coil;<br>Axium™ Prime<br>Detachable Coil |
| | intended for arterial and venous embolizations in the peripheral<br>vasculature. | |
| Visual Markers | None | Fluorosafe<br>markers |
| Dimensions | | |
| Device Size Range | Axium™ Bare Helix:<br>1.5-20 mm - Loop OD<br>1-50 cm - Length<br>Axium™ Bare 3D:<br>2-25 mm - Loop OD<br>2-50 cm - Length<br>Axium™ Nylon 3D:<br>2-4 mm – Loop OD<br>1-10 cm – Length<br>Axium™ PGLA Helix:<br>2-10 mm – Loop OD<br>1-30 cm – Length<br>Axium™ PGLA 3D:<br>2-18 mm – Loop OD<br>2-40 cm – Length<br>Axium™ Prime 3D:<br>1-6 mm – Loop OD<br>2-20 cm – Length<br>Axium™ Prime Helix:<br>1-6 mm – Loop OD<br>1-20 cm – Length<br>Axium™ Prime 3D:<br>3-25 mm – Loop OD<br>6-50 cm – Length | Same |
| Coil Shape | Helical and 3D | Same |
| | Compatible Accessories | |
| Catheter<br>Compatibility | Axium™ Detachable Coils (Bare) should be delivered through<br>microcatheters with minimum ID of 0.0165".<br>Axium™ Detachable Coils (Nylon and PGLA) should be delivered<br>through microcatheters with minimum ID of 0.0165"-0.020". | Same<br>Same |
| | Axium™ Prime Detachable Coils should be delivered through<br>microcatheters with minimum ID of 0.0165"-0.017". | Axium™ Prime<br>Detachable Coils<br>should be delivered<br>through |
| Design Feature | Predicate Devices (K203432):<br>Axium™ Detachable Coil;<br>Axium™ Prime Detachable Coil | Subject Devices:<br>Axium™ Detachable Coil;<br>Axium™ Prime<br>Detachable Coil<br>microcatheters with<br>minimum ID of<br>0.0165". |
| Guide Catheter<br>Compatibility | 6-8F | Same |
| Method of Coil<br>Detachment | Instant Detacher - standalone hand-held mechanical unit that, when connected to the proximal end of the pusher, pulls the release element inside of the pusher, resulting in release of the implant from the distal end of the delivery pusher. | Same |
| Material of Construction | | |
| Main Coil | Platinum/Tungsten alloy | Same |
| Tension Safe/<br>Stretch Resistant<br>Filament | Polypropylene | Same |
| Delivery System | Pusher: Composite hypotube stainless steel with PTFE outer jacket and liner and PET & Pt/w alignment marker | Same |
| Introducer Sheath | Polypropylene and HDPE | Same |
| Sterilization | | |
| Method of Supply | Sterile and single use | Same |
| Sterilization<br>Method | Ethylene Oxide (EO) | Same |
| Packaging | | |
| Pouch Material | For PC-X-Y-HELIX and PC-X-Y-3D:<br>PLK235/Uncoated 2FS Tyvek<br><br>For all other models:<br>PET/Tyvek | Same<br><br>For all other models:<br>Nylon/Tyvek |
| Pouch Dimensions | For PC-X-Y-HELIX and PC-X-Y-3D:<br>17" X 9.8" (pre-sterilization)<br><br>For all other models:<br>10 1/8" X 10.838" | Same<br><br>For all other models:<br>11.98" X 10.94" |
| Carton | Cardboard, Solid Bleach Sulfate | Same |
| MRI Card | None | 3 1/2" x 4" (+/-<br>1/8"), 1/1 Black -<br>Black and white,<br>2 sided |
| Stability | | |
| Shelf Life | 36 months | Same |
| MRI Compatibility | Magnetic Resonance Imaging<br>MR Conditional | Same |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
#### Biocompatibility:
The fluorosafe marker (subject to this submission) added onto the proximal end of the delivery pusher is considered external communicating device having indirect blood contact via the fluid path, as it is only in contact with the delivery catheter and does not have direct exposure to circulating blood. The distal end of the delivery pusher is categorized as external communicating device with limited exposure duration (د 24 hrs.) with circulating blood. The embolization coil is considered an implant with long-term (> 30 days) direct contact with circulating blood. There are no material or contact duration changes to the components of the device that directly contact circulating blood in the current submission. Therefore, prior hemocompatibility (complement activation, direct hemolysis, and thrombogenicity) tests demonstrate the biocompatibility of the unmodified regions and components of the subject device that contact blood directly.
To assess the addition of fluorosafe markers on the proximal end of the delivery pusher, the following biocompatibility tests were performed on the final, finished delivery pusher subassemblies incorporating the fluorosafe marker.
| Test Description | Study Method | Results |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity | Axium™ Detachable Coil and Axium™ Prime<br>Detachable Coil were evaluated per ISO 10993-<br>5: 2009, using the MEM elution method. | The Axium™ Detachable Coil and<br>Axium™ Prime Detachable Coil did<br>not induce cytotoxicity. |
| Sensitization | Axium™ Detachable Coil and Axium™ Prime<br>Detachable Coil were evaluated per ISO 10993-<br>10:2021, using a guinea pig maximization test. | The Axium™ Detachable Coil and<br>Axium™ Prime Detachable Coil<br>were not considered a sensitizer. |
| Irritation | Axium™ Detachable Coil and Axium™ Prime<br>Detachable Coil were evaluated per ISO 10993-<br>23:2021, using a rabbit intracutaneous reactivity<br>test. | The Axium™ Detachable Coil and<br>Axium™ Prime Detachable Coil are<br>considered non-irritant. |
| Acute<br>Systemic<br>Toxicity | Axium™ Detachable Coil and Axium™ Prime<br>Detachable Coil were evaluated per ISO 10993-<br>11:2017, using an acute systemic injection test<br>in mice. | The Axium™ Detachable Coil and<br>Axium™ Prime Detachable Coil<br>showed no mortality or evidence of<br>acute systemic toxicity. |
| Indirect<br>(extract)<br>Hemolysis | Axium™ Detachable Coil and Axium™ Prime<br>Detachable Coil were evaluated per ISO 10993-<br>4:2017, using ASTM hemolysis study. | The Axium™ Detachable Coil and<br>Axium™ Prime Detachable Coil are<br>considered non-hemolytic. |
| Material-Mediated<br>Pyrogenicity | Axium™ Detachable Coil and Axium™ Prime<br>Detachable Coil were evaluated per USP 151<br>(2020), using USP rabbit pyrogen test -<br>material-mediated. | The Axium™ Detachable Coil and<br>Axium™ Prime Detachable Coil met<br>the requirements and were found<br>to be non-pyrogenic. |
The Axium™ Detachable Coil and Axium™ Prime Detachable Coil have been evaluated to meet requirements specified in FDA guidance, "Use of International Standard ISO 10993-1, "Biocompatibility Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process", " issued
{8}------------------------------------------------
on September 8, 2023.
#### Performance Data – Bench:
Non-clinical bench testing was conducted to evaluate the performance of Axium™ Detachable Coil and Axium™ Prime Detachable Coil in a clinically representative neurovascular access model. The successful results of the testing demonstrated that the changes do not raise new questions of safety and effectiveness, supporting the substantial equivalence to the predicate devices.
The following non-clinical bench tests were conducted:
| Design Verification | | |
|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Test | Test Method Summary | Results |
| Visual Inspection | Visual inspection verified for the<br>darkness of the etch mark of<br>fluorosafe markers and the 360°<br>etch mark around the delivery<br>pusher. | Axium™ Detachable Coil and<br>Axium™ Prime Detachable Coil<br>met the acceptance criteria for<br>visual inspection. |
| Marker Dimensional (Marker<br>Position and Total Marker<br>Length) Inspection | Marker dimensional inspection<br>verified the total length of the<br>fluorosafe markers and their<br>position on the delivery pusher. | Axium™ Detachable Coil and<br>Axium™ Prime Detachable Coil<br>met the acceptance criteria for<br>marker dimensional inspection. |
| Corrosion Resistance | The test verified the corrosion<br>resistance on the delivery pusher<br>using the method of soaking the<br>devices in saline followed<br>by<br>immersion in boiling water per the<br>method specified in Annex A of ISO<br>10555-1. | Axium™ Detachable Coil and<br>Axium™ Prime Detachable Coil<br>met the acceptance criteria for<br>corrosion resistance. |
| Design Validation | | | | | |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|--|--|--|
| Test | Test Method Summary | Results | | | |
| Fluorosafe Marker Visibility | The clinical users evaluated the<br>devices<br>under simulated<br>use<br>conditions and rated fluorosafe<br>marker visibility. | The devices met the user needs<br>for which it was designed and<br>tested. | | | |
## Performance Data – Animal and Clinical:
The determination of substantial equivalence is based upon non-clinical bench testing as there is no change to the indications for use statement, fundamental scientific technology, principles of operation or materials of construction except for the addition of fluorosafe markers. No clinical data or animal study were deemed necessary to support the subject device submission.
{9}------------------------------------------------
#### Conclusion:
There is no change to the indications for use statement, fundamental scientific technology or principle of operation of the Axium™ Detachable Coils and Axium™ Prime Detachable Coils in comparison to the predicate devices. The addition of fluorosafe markers on the proximal end of the embolization coil delivery pusher does not raise new questions of safety and effectiveness and is supported by non-clinical bench testing using well-established acceptable scientific methods. The information provided in this submission supports a determination of substantial equivalence for Axium™ Detachable Coils and Axium™ Prime Detachable Coils to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.