The Stryker Universal Neuro III System is intended for reconstruction and/or rigid fixation of non loadbearing bony areas subsequent to cramictomy, cramiectomy and cranial fractures in adults and adolescents (age 12 and higher).
Device Story
Stryker Universal Neuro III System AXS screws are bone fasteners used for cranial reconstruction and fixation. This submission introduces two new sterile packaging configurations: sterile procedure packs and sterile screw cartridges. The device is used in clinical settings by surgeons to secure bone plates or cover burr holes following neurosurgical procedures. The screws are implanted to provide rigid fixation of non-loadbearing cranial bone. The device is provided sterile via gamma radiation. The healthcare provider uses the screws to stabilize bone segments, facilitating healing and structural integrity of the cranium. The primary benefit is the provision of pre-sterilized components to streamline surgical workflows and maintain sterility in the operating room.
Clinical Evidence
No clinical data was required or provided. Substantial equivalence was supported by non-clinical bench testing, including sterile barrier integrity, transport safety, and process validation for sterilization and cleaning.
Technological Characteristics
The device consists of bone plates and screws for cranial fixation. Materials and manufacturing processes are identical to the predicate. Sterilization is achieved via gamma radiation (SAL 1 x 10^-6). Packaging utilizes sterile barrier systems: flexible bags for procedure packs and sealed blisters for screw cartridges. No software or electronic components are involved.
Indications for Use
Indicated for reconstruction, stabilization, and rigid fixation of non-loadbearing bony areas following craniotomy, craniectomy, and cranial fractures in adults and adolescents aged 12 and older.
Regulatory Classification
Identification
A cranioplasty plate fastener is a screw, wire, or other article made of tantalum, vitallium, or stainless steel used to secure a plate to the patient's skull to repair a skull defect.
Predicate Devices
Stryker Universal Neuro III System AXS Screw (K151387)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2017
Stryker Craniomaxillofacial Jonathan Schell Staff Regulatory Affairs Specialist 750 Trade Centre Way, Suite 200 Portage. Michigan 49002
Re: K171152
Trade/Device Name: Stryker Universal Neuro III System AXS Screw Regulation Number: 21 CFR 882.5360 Regulation Name: Cranioplasty Plate Fastener Regulatory Class: Class II Product Code: HBW Dated: April 19, 2017 Received: April 20, 2017
Dear Mr. Schell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Hoffmann -S
for
Carlos L. Peña. PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171152
Device Name
Stryker Universal Neuro III System AXS Screw
Indications for Use (Describe)
The Stryker Universal Neuro III System is intended for reconstruction and/or rigid fixation of non loadbearing bony areas subsequent to cramictomy, cramiectomy and cranial fractures in adults and adolescents (age 12 and higher).
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase except for the first letter, which is capitalized. A registered trademark symbol is located to the upper right of the word. The word is in black and the background is white.
## Section 5. 510(k) Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
- SUBMITTER I.
| 510(k) Owner: | Stryker Leibinger GmbH& Co. KG<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Germany |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter/ Contact<br>Person: | Jonathan Schell<br>Staff Regulatory Affairs Specialist<br>Stryker Craniomaxillofacial<br>750 Trade Centre Way<br>Portage, MI 49002<br>Phone: 269-389-5596<br>Fax: 877-648-7114 |
| Date prepared: | May 12, 2017 |
| II.<br>DEVICE | |
| Trade Name: | Stryker Universal Neuro III System AXS screw |
| Common or Usual<br>name: | Cranioplasty plate fastener |
| Classification<br>name: | Cranioplasty Plate Fastener 21 CFR §882.5360 |
| Regulatory Class: | Class II |
| Product Code: | HBW |
#### PREDICATE DEVICE III.
Predicate: Stryker Universal Neuro III System AXS Screw – K151387 Reference Devices: K120352 Stryker QuikFlap Sterile Procedure Pack, and K131775 Universal Neuro 3 System
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The Stryker Universal Neuro III System (UN III) consists of an assortment of different bone plates and screws. The predicate device, the UN III AXS screw, is a part of the overall UN III System, and was cleared in K151387. Here, this special 510(k) is submitted to show the ability to market the UN III AXS screw sterile in two different sterile packaging configurations which are:
- 1. UN III AXS Screws packed in sterile procedure packs, and
- 2. UN III AXS Screws packed in a sterile screw cartridge.
The specific devices included are:
| Description | Packaging Configuration Type |
|-----------------------------------------|------------------------------------|
| 2-Hole Plate Set, SD Screw | UN III AXS Sterile Procedure Pack |
| 2-Hole Plate Set, ST Screw | UN III AXS Sterile Procedure Pack |
| 2-Hole Plate Set, Low Profile with Tab | UN III AXS Sterile Procedure Pack |
| 2-Hole Plate / Burr Hole Cover 14mm Set | UN III AXS Sterile Procedure Pack |
| 2-Hole Plate / Burr Hole Cover 20mm Set | UN III AXS Sterile Procedure Pack |
| Burr Hole Cover 14mm Set | UN III AXS Sterile Procedure Pack |
| UN III AXS Screw, ST, 1.5x4mm | UN III AXS Sterile Screw Cartridge |
| UN III AXS Screw, ST, 1.5x5mm | UN III AXS Sterile Screw Cartridge |
| UN III AXS Screw, ST, 1.5x6mm | UN III AXS Sterile Screw Cartridge |
| UN III AXS Screw, SD, 1.5x3mm | UN III AXS Sterile Screw Cartridge |
| UN III AXS Screw, SD, 1.5x4mm | UN III AXS Sterile Screw Cartridge |
| UN III AXS Screw, SD, 1.5x5mm | UN III AXS Sterile Screw Cartridge |
| UN III AXS Screw, Emergency, 1.7x4mm | UN III AXS Sterile Screw Cartridge |
#### V. INDICATIONS FOR USE
The Stryker Universal Neuro III System is intended for reconstruction, stabilization and/or rigid fixation of non load-bearing bony areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 and higher).
Offering the AXS screw sterile does not alter the Indications for Use statement for the proposed device. The Indications for Use is identical to the predicate device from the Stryker Universal Neuro III System (K151387).
{5}------------------------------------------------
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device is compared to its predicate device for substantial equivalence of technological characteristics based on the following criteria:
- A. Principle of Operation
- B. Technological Characteristics
## A. Principle of Operation
The basic operational principle of the UN III AXS screw is unchanged from the predicate and remains the same.
#### B. Technological Characteristics
The technological characteristics remain the same as the predicate and are unchanged.
#### VII. PERFORMANCE DATA
In Table 5.1 below, the relevant topics, as given in "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile Guidance for Industry and Food and Drug Administration Staff (2016)," are referenced.
#### Table 5-1: Reference of testing
| Category A Sterilization Methods –<br>Established Method | Gamma Radiation |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Sterilization Chamber: | Terminally sterilized in final packaging. |
| SAL | SAL 1 x 10-6 for all products |
| Pyrogenicity | Acceptance Criteria met |
| Packaging Description | Sterile Barrier Systems:<br>Procedure Packs: Flexible bag (sealed)<br>Screws: Blister (sealed) |
#### Process Validations
Overall the subject device is unchanged and has the same implant material and manufacturing process, and the same duration or location of contact to the patient as the predicate device. The only difference between subject device and predicate device lies in the sterile packaging process required with the two sterile barrier systems referenced above.
{6}------------------------------------------------
To evaluate the influence of the changes compared to the predicate device, and based on the Risk Analysis performed on the modification of providing the UN III AXS screw sterile, the following validations were performed and referenced in support of substantial equivalence:
- . Process Cleaning Validation (Bioburden level, Process residuals, Endotoxin Level)
- . Sealing Validation of seals closing the sterile barrier systems
- Process Sterilization Validation Gamma Radiation with VDmax25 Cycle .
A Biocompatibility Assessment was performed to support substantial equivalence respecting the additional manufacturing process steps due to the sterile packaging compared to the predicate device. Substantial equivalence was proven.
## Performance Bench Testing
The following performance bench tests were completed and referenced.
- . Transport safety of product and sterile barrier system including weight and compression test
- Shelf life and integrity of sterile barrier system .
- . Handling test of overall packaging functionality
The subject device met all pre-defined acceptance criteria. Overall, the results of the performance bench tests support the substantial equivalence of the subject device to the predicate device.
## Animal Testing
Animal testing was not required as a basis for substantial equivalence.
## Clinical Testing
Clinical testing was not required as a basis for substantial equivalence.
## VII. CONCLUSIONS
The results of the non-clinical data demonstrate that the modified UN III AXS sterile screws perform as intended in the specified use conditions. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.