M AND R REUSABLE ELECTRODES (MULTIPLE)

K963125 · M & R Mfg., Inc. · GZJ · Feb 6, 1997 · Neurology

Device Facts

Record IDK963125
Device NameM AND R REUSABLE ELECTRODES (MULTIPLE)
ApplicantM & R Mfg., Inc.
Product CodeGZJ · Neurology
Decision DateFeb 6, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

M&R Reusable Electrodes are intended for single-patient/multiple-application use in TENS ("traditional" TENS, microcurrent TENS, etc.) or NMES ("traditional" NMES, interferential, high-volt pulsed galvanic, functional electrical stimulation, etc.).

Device Story

M&R Reusable Electrodes are flexible, laminated cutaneous electrodes designed to interface between standard TENS/NMES stimulation devices and patient skin. They deliver electrical stimulation for pain management or neuromuscular therapy. The device consists of various backing materials (polyester, polyethylene foam, tricot), conductive elements (conductive rubber, conductive film, Ag/AgCl coating), and biocompatible conductive gels. They are available in multiple sizes and shapes with either direct pin lead wire connections or integral pre-wired leads. Used in clinical or home settings, the electrodes are applied by the patient or clinician to the skin surface. They are intended for single-patient, multiple-application use. By providing a stable electrical interface, they facilitate the delivery of therapeutic currents from the stimulator to the target tissue, potentially reducing pain or improving muscle function.

Clinical Evidence

No clinical data. Substantial equivalence is based on design, material composition, and intended use comparisons to the predicate device.

Technological Characteristics

Laminated, flexible cutaneous electrodes. Materials: polyester, polyethylene foam, tricot, conductive rubber, conductive film, Ag/AgCl coating, biocompatible adhesive, and conductive gel. Connectivity: standard .080 in. female pin connectors or integral lead wires. Compatible with standard TENS/NMES devices. Multi-model form factor (37 variations).

Indications for Use

Indicated for patients requiring TENS or NMES therapy for pain management or neuromuscular stimulation. Compatible with standard TENS/NMES devices. Contraindicated for patients with known sensitivity to electrode materials or adhesives.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Submission Summary (Full Text)

{1} FEB - 6 1997 K963125 M & R MANUFACTURING Telephone 614-694-6511 11987 Upper Fredericktown Amity Rd. Fredericktown, OH 43019 M & R Manufacturing, Inc. M&R Reusable Electrodes 510(k) Summary of Safety and Effectiveness The following information is submitted in accordance with the requirements of 21 CFR §807.92: Contact Person: Russell A. Latimer Date Prepared: 30 July 1996 Proprietary Name: M&R Reusable Electrodes Common/Usual Name: Reusable electrodes for Transcutaneous Electrical Nerve Stimulator (TENS) and Neuromuscular Electrical Stimulation (NMES) Classification Name: Electrodes, cutaneous (84GXY); Class II in accordance with 21 CFR § 882.1320 Device Description: M&R Reusable Electrodes are laminated, flexible structures composed of materials commonly used in this application (e.g., various cloths, tapes, etc. with biocompatible adhesive, various electrical conductors, and various biocompatible conductive gels). The electrodes are available in models of various sizes with either direct pin lead wire connection or via an integral electrode "prewire" lead wire, and are compatible with all standard, marketed TENS and NMES stimulation devices. Intended Use: M&R Reusable Electrodes are intended for single-patient/multiple-application use in TENS ("traditional" TENS, microcurrent TENS, etc.) or NMES ("traditional" NMES, interferential, high-volt pulsed galvanic, functional electrical stimulation, etc.). Predicate Device: Medi-Source, Inc. GeoHesive Reusable Electrodes K903402. A chart comparing characteristics of M&R Reusable Electrodes to the predicate device, demonstrating substantial equivalence, is attached. Submitted by: Russell A. Latimer Secretary & Treasurer M & R Manufacturing, Inc. 11967 Upper Fredericktown Amity Road Fredericktown, Ohio 43019 614/694-6511 101 {2} Device Comparison Chart | Device Characteristic | Medi-Source GeoHesive Reusable Electrodes - K903402 | M&R Reusable Electrodes | | --- | --- | --- | | Number of Models | 3 models | 37 models | | Electrode Sizes (in.)/Shapes * M&R models 33000 & 35000 are 2-electrode assemblies, each being of the size shown (*). | 1. 2.000 x 2.000 clover leaf 2. 2.000 x 3.000 dog bone 3. 2.000 x 6.000 dog bone | 1. 1.250 x 1.250 clover leaf 2. 1.875 x 1.875 clover leaf 3. 2.875 x 1.875 dog bone 4. 1.875 x 1.875 square 5. 2.000 x 1.250 rectangle 6. 1.750 x 1.625 rectangle 7. 3.875 x 1.875 rectangle 8. 3.875 x 2.500 rectangle * 9. 6.000 x 3.875 butterfly 10. 1.875 dia. round 11. 2.750 dia. round | | Electrode Construction | Vinyl w/ adhesive Lead wire assembly w/ conductive tab Conductive film w/ Ag/AgCl coating Biocompatible gel | 1. Polyester w/ adhesive Conductive rubber strip Biocompatible gel 2. Polyester/polyester film laminate w/ adhesive Conductive rubber strip Biocompatible gel 3. Polyethylene foam w/ adhesive Conductive rubber strip Biocompatible gel 4. Tricot/polyester laminate Conductive rubber strip Biocompatible gel | {3} Device Comparison Chart (cont.) | Device Characteristic | Medi-Source GeoHesive Reusable Electrodes - K903402 | M&R Reusable Electrodes | | --- | --- | --- | | Electrode Construction (cont.) | | 5. Polyester film w/adhesive Conductive rubber strip Biocompatible gel 6. Polyester w/ adhesive Conductive film Lead wire assembly Biocompatible gel 7. Polyester w/ adhesive Conductive film w/ Ag/AgCl coating Lead wire assembly Biocompatible gel 8. Polyethylene foam w/ adhesive Conductive film Lead wire assembly Biocompatible gel 9. Polyethylene foam w/ adhesive Conductive film w/ Ag/AgCl coating Lead wire assembly Biocompatible gel 10. Tricot/polyester laminate w/ adhesive Conductive film Lead wire assembly Biocompatible gel | {4} Device Comparison Chart (cont.) | Device Characteristic | Medi-Source GeoHesive Reusable Electrodes - K903402 | M&R Reusable Electrodes | | --- | --- | --- | | Electrode Construction (cont.) | | 11.Tricot/polyester laminate w/ adhesive Conductive film w/ Ag/AgCl coating Lead wire assembly Biocompatible gel 12.Polyester/polyester film laminate w/ adhesive Conductive film Lead wire assembly Biocompatible gel 13.Polyester/polyester film laminate w/ adhesive Conductive film w/ Ag/AgCl coating Lead wire assembly Biocompatible gel 14.Polyester film w/ adhesive Conductive film w/ Ag/AgCl coating Lead wire assembly Biocompatible gel | | Stimulator Compatibility | Common, marketed TENS & NMES devices | Common, marketed TENS & NMES devices | {5} Device Comparison Chart (cont.) | Device Characteristic | Medi-Source GeoHesive Reusable Electrodes - K903402 | M&R Reusable Electrodes | | --- | --- | --- | | Lead Wire Connector Types | Standard .080 in. female connection (built-in electrode lead with connector) | 1. Standard .080 in. female connection (built-in electrode lead with connector) 2. Standard .080 in. female connection (hole in conductive rubber strip) | | Labeled as reusable or single-use disposable? | Reusable/for single-patient, multiple-application use | Reusable/for single-patient, multiple-application use | | Indications For Use | TENS and NMES | TENS and NMES | | Stated potential adverse reactions | Skin irritation | Skin irritation |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...