The Transcutaneous Electrical Nerve Stimulator (model: TENS WMPS6-1) is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, the TENS WMPS6-1 is indicated for use for symptomatic relief of chronic intractable pain and adjunctive treatment in the management of post-surgical and post traumatic pain.
Device Story
TENS WMPS6-1 is an electrical nerve stimulator for pain relief; applies electrical current to skin via electrodes. Device features 18 programs (13 standard, 5 editable); powered by 6 1.5V batteries or AC 100-240V. Six outlet sockets connect to electrodes via lead wires. Operated by clinician or patient (prescription use); used in clinical or home settings. Device transforms electrical input into biphase rectangular pulse outputs to stimulate nerves. Healthcare providers use device to manage chronic, post-surgical, or post-traumatic pain. Benefits include non-pharmacological pain management. Device includes microprocessor control for stimulation parameters.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing only, including electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance requirements for nerve/muscle stimulators (IEC 60601-2-10).
Technological Characteristics
ABS housing; microprocessor-controlled; 6 output channels; biphase rectangular pulse waveform; 1-100Hz frequency; 50-250μs pulse width; 50V/100mA max output @500Ω. Powered by 6x1.5V batteries or AC adapter. Complies with ANSI AAMI ES60601-1, IEC 60601-1-2, and IEC 60601-2-10.
Indications for Use
Indicated for patients requiring pain relief, specifically for symptomatic relief of chronic intractable pain and as adjunctive treatment for post-surgical and post-traumatic pain.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 24, 2023
Wuxi Jiajian Medical Instrument Co., Ltd Wu Zhifang General manager No.35 Baiqiao Rd., Ehu Town, Xishan District, WuXi, JiangSu 214116 China
Re: K222879
Trade/Device Name: Transcutaneous Electrical Nerve Stimulator (model: TENS WMPS6-1) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: December 7, 2022 Received: December 8, 2022
Dear Wu Zhifang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lauren E. Woodard -S
for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222879
Device Name
Transcutaneous Electrical Nerve Stimulator (model: TENS WMPS6-1)
Indications for Use (Describe)
The Transcutaneous Electrical Nerve Stimulator (model: TENS WMPS6-1) is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, the TENS WMPS6-1 is indicated for use for symptomatic relief of chronic intractable pain and adjunctive treatment in the management of post-surgical and post traumatic pain.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|------------------------------------------------|
| | ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
| | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary [As required by 21 CFR 807.92]
| 1. Submission Information | |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number: | K222879 |
| Date: | January 23, 2023 |
| Type of 510(k) Submission: | Traditional 510(k) |
| Submitter/Manufacturer: | Wuxi Jiajian Medical Instrument Co., Ltd. |
| | No. 35 Baiqiao Rd., Ehu Town, Xishan District, Wuxi, Jiangsu, China<br>214116 |
| | Contact:Wu Zhifang |
| | E-mail: doris.d@ceve.org.cn |
| | Tel: 86 510-88745788 |
| 2. Device Description | |
| Proprietary Name: | Transcutaneous Electrical Nerve Stimulator (model: TENS WMPS6- 1) |
| Common Name: | TENS (Transcutaneous Electrical Nerve Stimulator) |
| Classification Name: | Transcutaneous electrical nerve stimulator for pain relief |
| Product Code: | GZJ |
| Device Class: | 2 |
| Regulation Number: | 21 CFR 882.5890 |
| Review Panel: | Neurology |
| Indications for use: | The Transcutaneous Electrical Nerve Stimulator (model: TENS<br>WMPS6-1) is an electrical nerve stimulator indicated for use for pain relief<br>by applying an electrical current to electrodes on a patient's skin to treat<br>pain. In particular, the TENS WMPS6-1 is indicated for use for<br>symptomatic relief of chronic intractable pain and adjunctive treatment in<br>the management of post-surgical and post traumatic pain. |
| Device Description: | Transcutaneous Electrical Nerve Stimulator (model: TENS WMPS6-1) is<br>an electrical nerve stimulator indicated for use for pain relief by applying<br>an electrical current to electrodes on a patient's skin to treat pain.<br>The device has 18 programs (13 standard programs and 5 editable<br>programs), powered by 6 pieces of 1.5V batteries or AC 100-240V,<br>comprising electronic stimulatory module and accessories of lead wires,<br>electrodes.<br>Six outlet sockets are used to connect skin electrodes by lead wires. The<br>accessories of electrodes is 510(k) cleared device (K192568), Size:<br>50*50mm. |
## 3. Predicate Device Identification
| Predicate 510(k) Number: | K220578 |
|---------------------------|--------------------------------------------|
| Marketing clearance date: | May 25, 2021 |
| Product name: | Transcutaneous Electrical Nerve Stimulator |
| Manufacturer: | Bozhou Rongjian Medical Appliance Co.,Ltd. |
# 4. Substantially Equivalent Comparison Conclusion
{4}------------------------------------------------
| Parameters | New Device | Predicate Device | Comparison | |
|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|-------------------|
| 510(k) Number | K222879 | K220578 | -- | |
| Device Name | Transcutaneous Electrical<br>Nerve Stimulator (model: TENS<br>WMPS6-1) | Transcutaneous Electrical<br>Nerve Stimulator (RJTENS-1) | -- | |
| Manufacturer | Wuxi Jiajian Medical<br>Instrument Co., Ltd | Bozhou Rongjian Medical<br>Appliance Co.,Ltd. | -- | |
| Intended use | Transcutaneous Electrical<br>Nerve Stimulator(model: TENS<br>WMPS6-1) is an electrical<br>nerve stimulator indicated for<br>use for pain relief by applying<br>an electrical current to<br>electrodes on a patient's skin to<br>treat pain. In particular, this<br>device is indicated for use for<br>symptomatic relief of chronic<br>intractable pain and adjunctive<br>treatment in the management of<br>post surgical and post traumatic<br>pain. | Transcutaneous Electrical<br>Nerve Stimulator is an electrical<br>nerve stimulator indicated for<br>use for pain relief by applying<br>an electrical current to<br>electrodes on a patient's skin to<br>treat pain. In particular, this<br>device is indicated for use for<br>symptomatic relief of chronic<br>intractable pain and adjunctive<br>treatment in the management of<br>post surgical and post traumatic<br>pain. | Same | |
| Type of use | Prescription use | Prescription use | Same | |
| Power Source(s) | DC9V(6×1.5V disposable<br>alkaline battery R14/UM2 type)<br>Power adapter:<br>Model:GTM96060-0612-3.0<br>Input:100-240V~,50/60Hz;<br>Output:DC9V,0.5A | 1.5Vx4 AAA alkaline battery | Similar<br>Note 1 | |
| - Method of Line<br>Current Isolation | N/A | N/A | Same | |
| - Patient Leakage<br>Current | -- | -- | -- | |
| - Normal<br>Condition (μΑ)<br>- Single Fault<br>Condition (μA) | 2μA<br>NA | 2μA<br>NA | Same | |
| Average DC current<br>through electrodes<br>when device is on but<br>no pulses are being<br>applied (μΑ) | <0.01μA | <0.01μA | Same | |
| Number of program | 18 | 16 | Similar<br>Note 2 | |
| Number of Output<br>channels: | 6 (No more than two channels<br>can be used on the same | 2 | Similar<br>Note 2 | |
| | - Synchronous<br>or Alternating? | person.) | Synchronous | Same |
| | - Method of<br>Channel Isolation | By Transformer | By Transformer | Same |
| | Regulated Current or<br>Regulated Voltage? | Current control | Current control | Same |
| | Software/Firmware/<br>Microprocessor<br>Control? | Yes | Yes | Same |
| | Automatic Overload<br>Trip? | No | No | Same |
| | Automatic No-Load<br>Trip? | No | No | Same |
| | Automatic Shut Off? | Yes | Yes | Same |
| | User Override<br>Control? | Yes | Yes | Same |
| | On/Off<br>Status? | Yes | Yes | Same |
| | Indicat<br>or Low<br>Battery? | Yes | Yes | Same |
| | Display Voltage/Cu<br>rrent<br>Level? | Yes | Yes | Same |
| | Timer Range<br>(minutes) | 1~99 min | 10~90 min | Similar<br>Note 2 |
| | Compliance with<br>Voluntary Standards? | ANSI AAMI ES60601-1, IEC<br>60601-1-2, IEC 60601-2-10 | ANSI AAMI ES60601-1, IEC<br>60601-1-2, IEC 60601-2-10 | Same |
| | Compliance with 21<br>CFR 898? | Yes | Yes | Same |
| | Weight (grams) | Approx.650g without battery | Approx.96g without battery | Similar<br>Note 3 |
| | Dimensions (mm) [W<br>x H x D] | 238*175*72mm | 140*64*28 mm | Similar<br>Note 3 |
| | Housing Materials &<br>Construction | ABS | ABS | Same |
| | Waveform | Biphase | Monophasic | Similar<br>Note 4 |
| | Shape | Rectangular pulse | Rectangular pulse | Same |
| | Maximum Output<br>Voltage (volts) | 50V±20% @500Ω | 30V±10% @500Ω | |
| | Maximum Output<br>Current (specify<br>units) | 100mA±20% @500Ω | 60mA±10% @500Ω | Similar<br>Note 5 |
| | Pulse width (usec) | 50-250μs±20% | 75-300μs±10% | |
| Pulse Period (msec) | 10-1000ms | 8.33-1000ms | | |
| Max. pulse frequency<br>(Hz) [or Rate (pps)] | 1-100Hz±10% | 1-120Hz±10% | | |
| Net Charge ( $\mu$ C per<br>pulse) | 0 $\mu$ C @500Ω(Method: Balanced<br>waveform) | 0.65 $\mu$ C @500Ω | | |
| Maximum Phase<br>Charge, ( $\mu$ C) | 25 $\mu$ C @500Ω | 18 $\mu$ C @500Ω | | |
| Maximum Average<br>Current, (mA) | 2.5mA @500Ω | 2.16mA @500Ω | Similar<br>Note 6 | |
| Maximum Current<br>Density, (mA/cm², r.m.s.) | 0.1mA/cm² @500Ω | 0.09mA/cm² @500Ω | | |
| Maximum Average<br>Power Density, (mW/cm²) | 5mW/cm² @500Ω | 2.59mW/cm² @500Ω | | |
| Accessories | Electrodes, cables,battery(users<br>need to purchase their<br>own),adapter(users need to<br>purchase their own) | Electrodes, cables, battery | Similar<br>Note 3 | |
{5}------------------------------------------------
{6}------------------------------------------------
## Comparison in details:
## Note 1:
The power source(s) and battery charge of the proposed device are a little different from the predicate device, but these differences are insignificant in terms of safety or effectiveness. In addition, the proposed device has passed the electrical safety test and ,so these differences don't raise any new safety and effectiveness issues.
## Note 2:
The number of modes and treatment times are only different by design; the core principles of these two devices are similar, although the number of channels is increased to 6 channels, the same person can use up to two channels, which is the same as the number of predicate device channels, and does not cause security and effectiveness problems.
## Note 3:
The weight, dimensions and accessories of the proposed device are a little different from the predicate device, this is a slightly different product design, but these differences are insignificant in terms of safety or effectiveness.
#### Note 4:
The output waveform of the new device is different from that of the predicate device. In addition, the proposed device has passed AAMI / ANSI ES60601-1, IEC 60601-1-2 and IEC 60601-2-10 tests, so these differences don't raise any new safety and effectiveness issues.
## Note 5:
There are some differences on the maximum output voltage and maximum output current between the proposed device and predicate device. Based on the calculation of maximum current density, maximum average power density, these parameters don't exceed the safety limit. All deviation and the worst case have been considered in risk analysis report, and these parameters have passed IEC 60601-2-10 test codes. So these differences will not raise any new safety and effectiveness issues. Although the pulse period and frequency of the proposed device are a little different from the predicate device, but they are all
{7}------------------------------------------------
compliance with IEC 60601-2-10 requirements. So, the minor differences of function specification will not raise any safety or effectiveness issue.
## Note 6:
The net charge and maximum phase charge of the proposed device are different from the predicate device, and both of them comply with IEC 60601-1 and IEC 60601-2-10 requirements. So, the differences will not raise any safety or effectiveness issue.The maximum average current, maximum current density, maximum average power density have some differences between proposed device and predicate device. Both of them meet maximum average current <10mA, maximum current density <2mA/cm² and maximum average power density <0.25W/cm². Therefore, these differences won't raise any new safety and effectiveness issues.Also, all programs have passed AAMI/ANSI ES 60601-2-10 tests. Therefore, these differences don't raise any new safety and effectiveness issues.
#### 5. Non-Clinical Test Conclusion
Bench tests were conducted on Transcutaneous Electrical Nerve Stimulator (model: TENS WMPS6-1) to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The following tests were performed on the proposed device:
• ANSI AAMI ES60601-1: 2005/(R) 2012 And A1: 2012, C1: 2009(R) 2012 And A2: 2010/(R) 2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1: 2005, MOD);
· IEC 60601-2-10 Edition 2.1 2016-04, Medical Electrical Equipment - Part 2-10: Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators;
· IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests:
## 6. Clinical Test
Clinical data was not including in this submission.
#### 7. Conclusions
The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device identified in the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.