← Product Code [GZJ](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZJ) · K202186

# TrueRelief Device (K202186)

_Truerelief · GZJ · Mar 23, 2021 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZJ/K202186

## Device Facts

- **Applicant:** Truerelief
- **Product Code:** [GZJ](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZJ.md)
- **Decision Date:** Mar 23, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The TrueRelief Device is intended for use under the supervision of a Healthcare Professional. TrueRelief is intended for temporary symptomatic relief of chronic intractable pain, and as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

## Device Story

TrueRelief is an AC-powered transcutaneous electrical nerve stimulator (TENS) used in clinical settings (physician offices, physical therapy clinics, hospitals, nursing homes, chiropractic clinics) under healthcare professional supervision. The device comprises a pulse generator, a primary probe, and a secondary probe. It delivers electromagnetic energy transcutaneously to painful tissue using a proprietary waveform with small voltage fluctuations. The healthcare provider operates the device to apply electrical pulses to the patient's tissue. By stimulating nerves, the device aims to provide temporary symptomatic relief of chronic intractable pain and manage acute post-surgical or post-traumatic pain, potentially reducing patient discomfort.

## Clinical Evidence

Bench testing only. Performance testing included electrical safety (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-6), electromagnetic compatibility (IEC 60601-1-2), and usability/human factors evaluation. No clinical data was provided.

## Technological Characteristics

AC-powered (115 VAC, 60Hz) TENS device. Patient contact material: 316L stainless steel. Output: single channel, transformer-isolated, regulated current. Waveform: rectangular. Pulse rate: 1-400 pps (low), 4,000-20,000 pps (high). Pulse duration: 0.24-0.74 ms (low), 15-25 µs (high). Max current: 8.9 mA. Max charge per pulse: 7 µC. Electrical safety: Type BF. Non-sterile.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- TrueRelief Device (originally NewLife Sciences TMR device) ([K070474](/device/K070474.md))
- DJO Chattanooga Revolution Wireless ([K153696](/device/K153696.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

TrueRelief % Kristen Allen Founder and Principal Consultant AllenBridge Consulting 2221 Oleander Drive Wilmington, North Carolina 28403

March 23, 2021

Re: K202186

Trade/Device Name: TrueRelief Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: March 16, 2021 Received: March 17, 2021

Dear Kristen Allen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name TrueRelief Device

Indications for Use (Describe)

The TrueRelief Device is intended for use under the supervision of a Healthcare Professional.

TrueRelief is intended for temporary symptomatic relief of chronic intractable pain, and as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for TrueRelief. Above the name is a graphic of a sound wave in shades of orange, purple, and blue. The name TrueRelief is written in a blue font, with the word True in a darker shade of blue than the word Relief.

## 510(k) Summary

## TrueRelief Device

| Submitter:                                 | True Relief<br>100 Wilshire Boulevard, Suite 700<br>Santa Monica, CA 90401<br>310.918.9064 (P)                            |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                            | Kristen Allen<br>Principal Consultant, Regulatory, Quality and<br>Compliance<br>910-612-4153 (P)<br>kallen@truerelief.com |
| Date Prepared:                             | March 23, 2021                                                                                                            |
| Trade Name:                                | TrueRelief Device                                                                                                         |
| Common Name:                               | Transcutaneous Electrical Nerve Stimulator (TENS) for pain<br>relief                                                      |
| Device Product Code<br>and Classification: | Regulation Number: 21 CFR 882.5890<br>GZJ, Class II, Transcutaneous Electrical Nerve Stimulator<br>(TENS) for pain relief |
| Primary Predicate:                         | TrueRelief Device (originally NewLife Sciences TMR device)<br>K070474                                                     |
| Additional Predicates:                     | DJO Chattanooga Revolution Wireless, K153696                                                                              |

#### Device Description:

The TrueRelief device is an AC power operated device that consists of three main components and associated power cables: the pulse generator, a primary probe and a secondary probe. The signal generator applies electromagnetic energy transcutaneously through the primary and secondary probes into painful tissue, similar to Transcutaneous Electrical Nerve Stimulation (TENS) device. The TrueRelief device uses a proprietary waveform with small voltage fluctuations to achieve pain relief.

#### Indications and Intended use:

The TrueRelief Device is intended for use under the supervision of a Healthcare Professional.

TrueRelief is intended for temporary symptomatic relief of chronic intractable pain, and as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for TrueRelief. Above the name is a graphic of four diamonds in a row, each a different color. The name TrueRelief is written in two different colors, with True in blue and Relief in purple.

## Comparison of Technological Characteristics With The Predicate Device:

The subject and predicate devices have similar technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. The following characteristics are similar between the subject device and predicate devices:

- . Indications for use
- . Principle of operation
- Technological characteristics .
- . Performance testing

### Comparison with Predicate Devices

| Item                                                                     | Primary Predicate Device<br>K070474 TMR                                                                                                                                                                                                   | Reference Device<br>K153696                                                                                                                                                                                                                                                                                                                                         | Subject Device<br>TrueRelief                                                                                                                                                                                                                                                                                               |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for<br>Use                                 | TrueRelief is intended for<br>temporary symptomatic relief<br>of chronic intractable pain, and<br>as an adjunctive treatment in<br>the management of post-<br>surgical and post-traumatic<br>acute pain.                                  | The Chattanooga<br>Revolution Wireless is a<br>clinical electrotherapy<br>device intended for use<br>under the supervision of<br>a Healthcare<br>Professional.<br>As a TENS device,<br>indications are for the<br>following conditions:<br>- Symptomatic relief and<br>management of chronic,<br>intractable pain - Post-<br>surgical and post-trauma<br>acute pain | The TrueRelief Device is<br>intended for use under<br>the supervision of a<br>Healthcare Professional.<br><br>TrueRelief is intended for<br>temporary symptomatic<br>relief of chronic<br>intractable pain, and as<br>an adjunctive treatment in<br>the management of post-<br>surgical and post-<br>traumatic acute pain. |
| Prescription/<br>Professional Use<br>Only or OTC                         | Prescription                                                                                                                                                                                                                              | Prescription/<br>Professional Use Only                                                                                                                                                                                                                                                                                                                              | Prescription/<br>Professional Use Only                                                                                                                                                                                                                                                                                     |
| Where used                                                               | Physician office, physical<br>therapy clinic, hospital nursing<br>home, post acute care,<br>chiropractic clinic                                                                                                                           | Physician office,<br>physical therapy clinic,<br>hospital nursing home,<br>post acute care,<br>chiropractic clinic                                                                                                                                                                                                                                                  | Physician office, physical<br>therapy clinic, hospital<br>nursing home, post acute<br>care, chiropractic clinic                                                                                                                                                                                                            |
| Target population                                                        | Adult population                                                                                                                                                                                                                          | Adult population                                                                                                                                                                                                                                                                                                                                                    | Adult population                                                                                                                                                                                                                                                                                                           |
| Battery or mains<br>powered                                              | Mains                                                                                                                                                                                                                                     | Mains and battery<br>operated                                                                                                                                                                                                                                                                                                                                       | Mains                                                                                                                                                                                                                                                                                                                      |
| Principle of<br>Operation                                                | External nerve stimulator that<br>generates an electrical pulse<br>that is delivered to the tissue<br>through the primary and<br>secondary probes                                                                                         | External nerve stimulator<br>that generates an<br>electrical pulse that is<br>delivered to the tissue<br>through the primary and<br>secondary probes                                                                                                                                                                                                                | External nerve stimulator<br>that generates an<br>electrical pulse that is<br>delivered to the tissue<br>through the primary and<br>secondary probes                                                                                                                                                                       |
| Sterility                                                                | Non-Sterile                                                                                                                                                                                                                               | Non-sterile                                                                                                                                                                                                                                                                                                                                                         | Non-sterile                                                                                                                                                                                                                                                                                                                |
| Materials                                                                | Patient contact material: 316L<br>stainless steel                                                                                                                                                                                         | Patient contact material:<br>stainless steel                                                                                                                                                                                                                                                                                                                        | Patient contact material:<br>316L stainless steel                                                                                                                                                                                                                                                                          |
| Connection of<br>device to<br>electrodes                                 | The device uses two<br>electrodes(probes) using a<br>single channel. At one end of<br>each probe is a 316 stainless                                                                                                                       | Stimulation Module is<br>directly connected to the<br>custom male SNAP<br>assembled in the                                                                                                                                                                                                                                                                          | The device uses two<br>electrodes(probes) using<br>a single channel. At one<br>end of each probe is a                                                                                                                                                                                                                      |
| Item                                                                     | Primary Predicate Device<br>K070474 TMR                                                                                                                                                                                                   | Reference Device<br>K153696                                                                                                                                                                                                                                                                                                                                         | Subject Device<br>TrueRelief                                                                                                                                                                                                                                                                                               |
|                                                                          | steel tip that touches the<br>patient and the other end of<br>the stainless steel tip screws<br>into the probe. On the other<br>end of the probe is a cable<br>which connects each probe to<br>the device using a connector<br>with pins. | electrode. User Interface<br>(LCD and buttons) is<br>physically separated<br>(Remote Control) and<br>communicates wirelessly<br>with up to four (4)<br>stimulation modules.<br>Stimulation safety<br>remains fully managed<br>by Stimulation Module<br>electronic circuit itself.                                                                                   | 316 stainless steel tip that<br>touches the patient and<br>the other end of the<br>stainless steel tip screws<br>into the probe. On the other<br>end of the probe is<br>a cable which connects<br>each probe to the device<br>using a connector with<br>pins.                                                              |
| Power Supply                                                             | 115 VAC, 60Hz input<br>12 volt DC output                                                                                                                                                                                                  | Rechargeable battery                                                                                                                                                                                                                                                                                                                                                | 115 VAC, 60Hz input<br>12 volt DC output                                                                                                                                                                                                                                                                                   |
| Electrical Type                                                          | Type BF                                                                                                                                                                                                                                   | NA<br>Battery operated device                                                                                                                                                                                                                                                                                                                                       | Type BF                                                                                                                                                                                                                                                                                                                    |
| Patient Leakage<br>Current - Normal<br>Condition (µA)                    | <100uA patient leakage                                                                                                                                                                                                                    | NA<br>Battery operateddevice                                                                                                                                                                                                                                                                                                                                        | <100uA patient leakage                                                                                                                                                                                                                                                                                                     |
| Patient Leakage<br>Current - Single<br>Fault Condition<br>(μΑ)           | <300uA line leakage                                                                                                                                                                                                                       | NA<br>Battery operateddevice                                                                                                                                                                                                                                                                                                                                        | <300uA line leakage                                                                                                                                                                                                                                                                                                        |
| Number of<br>Output Channels                                             | 1                                                                                                                                                                                                                                         | 0, 2, or 4                                                                                                                                                                                                                                                                                                                                                          | 1                                                                                                                                                                                                                                                                                                                          |
| Method of<br>Channel Isolation                                           | Transformer isolated.                                                                                                                                                                                                                     | Each channel is the<br>middle of a H Bridge.<br>Except when it is<br>activated, each<br>channel is always in high<br>impedance state                                                                                                                                                                                                                                | Transformer isolated.                                                                                                                                                                                                                                                                                                      |
| Regulated<br>Current or<br>Requlated<br>Voltage (output<br>signals only) | Regulated current on only<br>channel                                                                                                                                                                                                      | Regulated current on all<br>channels                                                                                                                                                                                                                                                                                                                                | Regulated current on<br>only channel                                                                                                                                                                                                                                                                                       |
| Maximum<br>Current (RMS)<br>Density<br>(mA/cm2)                          | 14mA/cm2<br>Average of 2.1mA/cm2                                                                                                                                                                                                          | 1.34mA/cm2                                                                                                                                                                                                                                                                                                                                                          | 14mA/cm2<br>Average of 2.1mA/cm2                                                                                                                                                                                                                                                                                           |
| Maximum Power<br>Density<br>[mW/cm²]                                     | 79 mW/cm2 @5000<br>Average of 11mW/cm2<br>@5000                                                                                                                                                                                           | 14.4 [mW/cm²] @500Ω                                                                                                                                                                                                                                                                                                                                                 | 79 mW/cm2 @5000<br>Average of 11mW/cm2<br>@5000                                                                                                                                                                                                                                                                            |
| Output Voltage                                                           | Range of normal use: 50-60 V<br>Peak pulse amplitude: 200 V                                                                                                                                                                               | 60 V @ 500 Ω<br>180V @ 2 kΩ<br>180 V @ 10 kΩ                                                                                                                                                                                                                                                                                                                        | Range of normal use: 50-<br>60 V<br>Peak pulse amplitude:<br>200 V                                                                                                                                                                                                                                                         |
| Pulse Rate                                                               | 1 to 400 (+10%) Pulse /<br>Second (Low)<br>4,000 (-5%) to 20,000 (+10%)<br>Pulse/Second (High)                                                                                                                                            | 35 - 80 Hz                                                                                                                                                                                                                                                                                                                                                          | 1 to 400 (+10%) Pulse /<br>Second (Low)<br>4,000 (-5%) to 20,000<br>(+10%) Pulse/Second<br>(High)                                                                                                                                                                                                                          |
| Item                                                                     | Primary Predicate Device<br>K070474 TMR                                                                                                                                                                                                   | Reference Device<br>K153696                                                                                                                                                                                                                                                                                                                                         | Subject Device<br>TrueRelief                                                                                                                                                                                                                                                                                               |
| Pulse Duration                                                           | 0.24-0.74 millisecond (Low)<br>15-25 microsecond (High)                                                                                                                                                                                   | 300 to 400 [μs]<br>(microseconds)                                                                                                                                                                                                                                                                                                                                   | 0.24-0.74 millisecond<br>(Low)<br>15-25 microsecond (High)                                                                                                                                                                                                                                                                 |
| Output Current<br>(maximum)                                              | 8.9 milliamps (at body<br>resistance > 10.11K ohms)                                                                                                                                                                                       | 120 mA @ 500 Ω<br>90 mA @ 2 kΩ<br>18 mA @ 10 kΩ                                                                                                                                                                                                                                                                                                                     | 8.9 milliamps (at body<br>resistance > 10.11K<br>ohms)                                                                                                                                                                                                                                                                     |
| Maximum Charge<br>Per Pulse                                              | 7 μC                                                                                                                                                                                                                                      | 36 [μC] @ 500Ω                                                                                                                                                                                                                                                                                                                                                      | 7 μC                                                                                                                                                                                                                                                                                                                       |
| Wave Form<br>Shape                                                       | Rectangular                                                                                                                                                                                                                               | Rectangular                                                                                                                                                                                                                                                                                                                                                         | Rectangular                                                                                                                                                                                                                                                                                                                |
| Maximum<br>Amplitude                                                     | No load - 200 V Peak<br>With 50K ohm load - 175 V<br>Peak                                                                                                                                                                                 | 60 V @ 500 Ω<br>180V @ 2 kΩ<br>180 V @ 10 kΩ                                                                                                                                                                                                                                                                                                                        | No load - 200 V Peak<br>With 50K ohm load - 175<br>V Peak                                                                                                                                                                                                                                                                  |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for TrueRelief. The logo consists of the word "TrueRelief" in a sans-serif font, with "True" in a dark blue color and "Relief" in a lighter blue color. Above the word is a graphic of four diamond shapes in a row, with the colors of the diamonds being orange, pink, purple, and blue.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for TrueRelief. Above the name is a graphic of a sound wave in different shades of purple and orange. The name TrueRelief is written in a combination of blue and purple.

## Summary of Performance Testing:

Electrical Safety: The TrueRelief Device was tested and certified to comply with recognized standards for electrical safety (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-6).

Electromagnetic Compatibility: The TrueRelief Device was tested and certified to comply with recognized standards for electromagnetic compatibility (IEC 60601-1-2).

Usability/Human Factors: Usability/Human Factors were evaluated, which demonstrated that the established requirements for usability were met, and the device's design is appropriate for the intended users and use environment. The result of this evaluation substantiates the acceptability of the use-related risks identified during the risk assessment activities.

## Conclusion:

Based on the performance testing and the similarities of the indications for use and the technological characteristics, it can be concluded that the TrueRelief Device is as safe and effective as, and substantially equivalent to, the predicate device(s).

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZJ/K202186](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZJ/K202186)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
