TENS mode: - Symptomatic relief of chronic (long term) intractable pain - Symptomatic relief of post-tramautic acute pain and post surgical pain Russian mode: - Temporary relaxation of muscle spasms in back area - Prevention or retardation of disuse atrophy of the lumbar paraspinal muscles
Device Story
StimaWELL 120MTRS is a 12-channel electro-therapy system for muscle stimulation, pain relief, and back warming. System comprises a stimulation mat, controller, and remote control. Inputs include user-selected treatment sequences via touchscreen or chip cards. Device transforms electrical energy into medium-frequency alternating current (MFAC) or modulated low-frequency alternating current to stimulate nerves and muscles; resistive heating elements provide warming. Output is delivered through the mat electrodes to the patient's back. Used in clinical or home settings; operated by patients or clinicians. Remote control allows patients to pause treatment and adjust intensity. Output affects clinical decision-making by providing non-pharmacological pain management and muscle therapy. Benefits include pain relief, muscle spasm reduction, and atrophy prevention.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through electrical safety, EMC, biocompatibility (ISO 10993), risk management (ISO 14971), software (IEC 62304), and usability (IEC 62366) testing.
Technological Characteristics
Materials: ABS plastic. Energy: AC powered (100-240V, 50/60Hz). Stimulation: 12-channel mat, MFAC and modulated LFAC. Heating: Resistive heating in mat. Connectivity: Touchscreen controller, chip card interface, remote control. Software: Microprocessor-controlled. Safety: Double isolation, patient leakage <0.063 mA (single fault).
Indications for Use
Indicated for patients requiring symptomatic relief of chronic intractable pain, post-traumatic acute pain, or post-surgical pain (TENS mode), and for patients requiring temporary relaxation of back muscle spasms or prevention/retardation of lumbar paraspinal muscle disuse atrophy (Russian mode).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
August 13, 2020
Pierenkemper GmbH % Andre Kindsvater Senior Consultant RA & QA Emergo Global Consulting 2500 Bee Cave Rd, Building 1, Suite 300 Austin, Texas 78746
Re: K181645
Trade/Device Name: StimaWELL 120MTRS Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ, IPF Dated: May 10, 2020 Received: May 15, 2020
Dear Andre Kindsvater:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Amber Ballard, PhD Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181645
Device Name StimaWELL®120MTRS
### Indications for Use (Describe)
TENS mode:
- Symptomatic relief of chronic (long term) intractable pain
- Symptomatic relief of post-tramautic acute pain and post surgical pain
Russian mode:
- Temporary relaxation of muscle spasms in back area
- Prevention or retardation of disuse atrophy of the lumbar paraspinal muscles
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## StimaWELL 120MTRS
# K181645
#### 1. Submission Sponsor
Pierenkemper GmbH Am Geiersberg 6 35630 Ehringshausen Germany
Phone: (+49) 6443 8333-574 Fax: (+49) 6444 8333-580 Contact: Thorsten Reichel Title: Technical Director
#### 2. Submission Correspondent
Emergo Global Consulting 2500 Bee Cave Rd, Suite 300 Austin, TX 78746 Cell Phone: +41 78 843 7077 Office Phone: (512) 327 9997
Fax: (512) 327.9998 Contact: André Kindsvater, Senior Consultant, RA/QA Email: LST.AUS.ProjectManagement@ul.com.
#### 3. Date Prepared
August 6th, 2020
#### 4. Device Identification
| Trade/Proprietary Name: | StimaWELL 120MTRS |
|-------------------------|-----------------------------------------------------------------------------------------|
| Common/Usual Name: | Nerve and Muscle stimulator for pain relief |
| Classification Name: | Transcutaneous electrical nerve stimulator for pain relief<br>Powered muscle stimulator |
| Regulation Number | 882.5890<br>890.5850 |
| Product Codes | GZJ<br>IPF |
| Regulation Name | Stimulator, Nerve, Transcutaneous, For Pain Relief<br>Stimulator, Muscle,Powered |
{4}------------------------------------------------
#### 5. Legally Marketed Predicate Devices
a) Electro-therapy (primary) K131917, Neurodyn II Ibramed Equipamentos Medico, Sao Paulo, Brasil
#### 6. Device Description
The StimaWELL 120MTRS has been designed for the use of medium frequency electro-therapy for muscle stimulation and pain relief therapy as well as for the warming and wellness application in the back region.
The device consists of a 12 channel (electrodes) Stimulation Mat, a Controller and a Remote Control Unit. Various accessories (see 11.3.2 below) support the StimaWELL. The warming is provided through controlled resistive heating of the stimulation mat.
Interaction with the device is via a touchscreen on the controller. Selection and set-up of various treatment sequences is via the controller touchscreen or can be recalled from chip cards. The patient can interact with the device by using the remote control unit. With that he can pause the treatment at any time and also adjust the overall intensity.
#### 7. Indication for Use Statement
TENS mode: - Symptomatic relief of chronic (long term) intractable pain - Symptomatic relief of post-traumatic acute pain and post surgical pain
- Temporary relaxation of muscle spasms in back area Russian mode: - Prevention or retardation of disuse atrophy of the lumbar paraspinal muscles
#### 8. Substantial Equivalence Discussion
The following tables compares the Muscle StimaWELL 120MTRS to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence.
{5}------------------------------------------------
| Manufacturer | Pierenkemper GmbH | IBRAMED EQUIPAMENTOS MEDICOS |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | StimaWELL 120MTRS | Neurodyn II |
| 510(k) Number | K181645 | K131917 |
| Product Code | GZJ<br>IPF | GZJ<br>IPF<br>GZI |
| Regulation Number | 882.5890<br>890.5850 | 882.5890<br>890.5850<br>882.5810 |
| Regulation Name | Transcutaneous electrical nerve<br>stimulator for pain relief | Transcutaneous electrical nerve<br>stimulator for pain relief |
| | Powered muscle stimulator | Powered muscle stimulator |
| | | External functional neuromuscular<br>stimulator |
| | Indications for Use | N/A<br>The StimaWELL is not used as a FES<br>device |
| TENS mode:<br>- Symptomatic relief of chronic (long<br>term) intractable pain<br>- Symptomatic relief of post-traumatic<br>acute pain and post surgical pain | | As a TENS device:<br>- Symptomatic relief of chronic (long<br>term) intractable pain<br>- Symptomatic relief of post-traumatic<br>acute pain and post surgical pain |
| Russian mode:<br>- Temporary relaxation of muscle<br>spasms in back area<br>- Prevention or retardation of disuse<br>atrophy of the lumbar paraspinal<br>muscles | | As a Burst Modulated Alternating<br>Current<br>(Russian) device:<br>- Temporary relaxation of muscle<br>spasms<br>- Prevention or retardation of disuse<br>atrophy in post-injury type conditions<br>- Increase local blood circulation<br>- Muscle re-education<br>- Maintaining or increasing range of<br>motion |
| Technological<br>characteristic | Medium-frequency<br>alternating current (MFAC);<br>modulated low-frequency alternating<br>current | Medium-frequency<br>alternating current (MFAC);<br>Low-frequency alternating current<br>(LFAC) |
Table 5-1 – Comparison of Characteristics
{6}------------------------------------------------
| Electro-therapy<br>Mechanism of Action | Transcutaneous muscle stimulation | Transcutaneous muscle stimulation |
|------------------------------------------------------|----------------------------------------------------------------------------------------|-----------------------------------|
| Operating Modes:<br>TENS<br>Russian<br>FES | Yes<br>Yes<br>No | Yes<br>Yes<br>Yes |
| Setting of Mat Size | yes | N/A |
| Treatment Timer | 20-30 minutes | Not publicly available |
| Automatic Shutoff | No | No |
| Device Material | ABS Plastic and LCD display | ABS Plastic and LCD display |
| Number of Channels | 12 | 4 |
| Technology Overview<br>Software Micro-<br>processor | Yes | Yes |
| Method of line<br>current isolation | Double Isolation | Double Isolation |
| Patient leakage<br>control-normal<br>condition | < 0.041 mA | 0.0347mA |
| Patient leakage<br>control-single fault<br>condition | < 0.063 mA | 0.0162mA |
| Duration of<br>stimulation | 30 minutes | Not publicly available |
| Mode of operation | Continuous | Continuous |
| Energy Output | Yes | Yes |
| Sterile | No | No |
| Single-Use | No | No |
| AC Powered | 100 to 240V<br>50/60Hz | 100 to 240V<br>50/60Hz |
| Operating<br>Environment | 50°F - 104 °F<br>(10° C - 40° C)<br>30%-90% relative humidity<br>700 hPa to 1060 hPa | 41°F to 113°F<br>(5°C- 45°C) |
| Storage and<br>Transport<br>Environment | 14°F - 131 °F<br>(-10 °C to 55 °C)<br>30%-90% relative humidity<br>500 hPa to 1060 hPa | 59°F to 104°F<br>(15°C- 40°C) |
| Complies with ISO<br>10993-1 | Yes | Not publicly available |
| Electrical Safety<br>Testing Passed | Yes | Yes |
{7}------------------------------------------------
## 9. Comparison of Output Specifications (per Section 3)
### Table 5-2 TENS Mode
| Trade Name > | StimaWELL 120MTRS | Neurodyn II |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|-----------------------------------------------|
| a) Waveform<br>(e.g., pulsed monophasic, biphasic) | biphasic | Not publicly available |
| b) Shape<br>(e.g., rectangular, spike, rectified<br>sinusoidal) | rectangular | Not publicly available |
| c) Maximum Output Voltage<br>(specify units) | | Not publicly available |
| @ 500 Ω, | 58.7 Vpp<br>(+/- 10 %) | Not publicly available |
| @ 2 kΩ | 57.6 Vpp<br>(+/- 10 %) | Not publicly available |
| @ 10 kΩ | 68.3 Vpp<br>(+/- 10 %) | Not publicly available |
| d) Maximum Output Current<br>(specify units) | | 0 to 120mA peak to peak |
| @ 500 Ω, | 58.7 mA<br>(+/- 10 %) | 124mA<br>Tolerance not publicly available |
| @ 2 kΩ | 14.4 mA<br>(+/- 10 %) | 110mA<br>Tolerance not publicly available |
| @ 10 KΩ | 3.42 mA<br>(+/- 10 %) | 39.2mA<br>Tolerance not publicly available |
| e) Pulse Width (specify units) | 500 ms to 5 ms | Not publicly available |
| f) Frequency (Hz) | 1 Hz to 100 Hz | 0.5 Hz to 250Hz |
| g) For interferential modes only:<br>Beat Frequency (Hz) | N / A | Not publicly available |
| h) For multiphasic waveforms only:<br>Phase<br>Symmetrical phases? (yes/no),<br>Phase Duration (include units)<br>(state range, if applicable) (both phases,<br>if asymmetrical) | N / A (Biphasic is<br>noted above) | Not publicly available |
| i) Net Charge (μC per pulse) @ 500 Ω<br>(If zero, state method of achieving zero<br>net charge.) | N / A | Not publicly available |
| j) Maximum Phase Charge, (μC) @ 500 Ω | N / A | Not publicly available |
| Trade Name > | StimaWELL 120MTRS | Neurodyn II |
| k) Maximum Current Density. (mA/cm²)<br>@ 500 Ω | 3.87 mA/cm² | Not publicly available |
| l) Maximum Power Density, (W/cm²) @<br>500 Ω<br>(using smallest electrode conductive<br>surface area) 15.18 cm² | 0.11 W/cm² | 0.038 W/cm² |
| m) Burst Mode (i.e., pulse trains) | N/A | Not publicly available |
| a. Pulses per burst | | |
| b. Bursts per second | | |
| c. Burst duration (seconds) | | |
| d. Duty Cycle<br>[Line (b) x Line (c)] | | |
| n) ON Time (seconds) | N/A | Not publicly available |
| o) OFF Time (seconds) | N/A | Not publicly available |
| p) Additional features (if applicable) | N/A | N/A |
| Trade Name > | StimaWELL 120MTRS | Neurodyn II |
| a) Waveform<br>(e.g., pulsed monophasic, biphasic) | Biphasic | Not publicly available |
| b) Shape<br>(e.g., rectangular, spike, rectified<br>sinusoidal) | Rectangular | Not publicly available |
| c) Maximum Output Voltage<br>(specify units) | | Not publicly available |
| @ 500 Ω, | 62.7 Vpp<br>(+/- 10 %) | Not publicly available |
| @ 2 kΩ | 67.1 Vpp<br>(+/- 10 %) | Not publicly available |
| @ 10 kΩ | 68.7 Vpp<br>(+/- 10 %) | Not publicly available |
| d) Maximum Output Current<br>(specify units) | | 0 to 120mA peak to peak |
| @ 500 Ω, | 62.7 mAp<br>(+/- 10 %) | 124mA<br>Tolerance not publicly<br>available |
| @ 2 kΩ | 16.8 mAp<br>(+/- 10 %) | 110mA<br>Tolerance not publicly<br>available |
| @ 10 kΩ | 3.44 mAp<br>(+/- 10 %) | 39.2mA<br>Tolerance not publicly<br>available |
| e) Pulse Width (specify units) | 500 ms to 5 ms | Not publicly available |
| f) Frequency (Hz) | 1 Hz to 100 Hz | 0.5 Hz to 250 Hz |
| g) For interferential modes only:<br>Beat Frequency (Hz) | N / A | Not publicly available |
| h) For multiphasic waveforms only: | N / A<br>(Biphasic is noted<br>above) | Not publicly available |
| Phase<br>Symmetrical phases? (yes/no),<br>Phase Duration (include units)<br>(state range, if applicable) (both phases,<br>if asymmetrical) | | |
| i) Net Charge (μC per pulse) @ 500 Ω<br>(If zero, state method of achieving zero<br>net charge.) | N / A | Not publicly available |
| j) Maximum Phase Charge, (μC) @ 500 Ω | N / A | Not publicly available |
| k) Maximum Current Density, (mA/cm²)<br>@ 500 Ω | 4.13 mA/cm² | Not publicly available |
| l) Maximum Power Density, (W/cm²) @<br>500 Ω<br>(using smallest electrode conductive<br>surface area) | 0.13 W/cm² | 0.038 W/cm² |
| m) Burst Mode (i.e., pulse trains) | N/A | Not publicly available |
| a. Pulses per burst | | |
| b. Bursts per second | | |
| c. Burst duration (seconds) | | |
| d. Duty Cycle<br>[Line (b) x Line (c)] | | |
| n) ON Time (seconds) | N/A | Not publicly available |
| o) OFF Time (seconds) | N/A | Not publicly available |
| p) Additional features (if applicable) | | |
{8}------------------------------------------------
{9}------------------------------------------------
### Table 5-3 Russian Mode
{10}------------------------------------------------
#### Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of the StimaWell 120MTR and in showing substantial equivalence to the predicate device that are subject to this 510(k) submission, Pierenkemper completed a number of non-clinical performance tests against applicable standards.
| Test | Pass / fail criteria | Results | |
|------|----------------------------------|--------------------------------------------------------------------------|--------|
| 1 | Electrical safety | Compliance to IEC 60601-1:2005/AMD1:2012 | Passed |
| 2 | Electromagnetic<br>compatibility | Compliance to IEC 60601-1-2:2014 | Passed |
| 3 | Nerve and muscle<br>stimulators | Compliance to IEC 60601-2-10:2012 | Passed |
| 4 | Biocompatibility | Compliance to ISO 10993-1, ISO 10993-5:2009, and<br>ISO 10993-10:2002 | Passed |
| 5 | Risk Management | Compliance to ISO 14971:2007 | Passed |
| 6 | Software | Compliance to IEC 62304:2006 | Passed |
| 7 | Usability | Compliance to IEC 62366:2007 (Ed. 1) + A1: 2014<br>and EN 60601-1-6:2013 | Passed |
Table 5-2 – Performance Standards Testing Summary
The StimaWELL 120MTRS passed all the testing in accordance with internal requirements, national standards, and international standards shown above, to support substantial equivalence of the subject device.
To demonstrate that the StimaWELL 120MTRS meets all design specifications and performance requirements, nonclinical bench testing was performed in accordance with the internal development
{11}------------------------------------------------
process in compliance with the recommendations of the FDA Guidance Document for Powered Muscle Stimulator 510(k)s [1999].
Also, internal verification and validation testing confirms that product specifications are met. The testing results support that the requirements for performance and electrical safety were met for the acceptance of the device. The StimaWELL 120MTRS passed all testing and supports the claims of substantial equivalence to the predicate device.
#### 10.Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device; or the device has the same intended use and different technological characteristics but can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device.
The StimaWELL 120MTRS as designed and manufactured, is determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.