The Kneehab XP is intended to provide transcutaneous Neuromuscular Electrical Stimulation (NMES) and Transcutaneous Electrical Nerve Stimulation (TENS).
Device Story
Kneehab XP is a portable, battery-operated combination device providing NMES and TENS. It utilizes 'multipath' technology via a conductive garment fastened around the thigh and above the kneecap to deliver focused NMES muscle contractions. The system includes a rechargeable control unit, garment, custom adhesive electrodes, and charger. The device features nine pre-set programs (six NMES, three TENS) with 20-minute durations. It is operated by the patient or clinician to manage pain or facilitate muscle rehabilitation. The control unit is powered by a 3.6V NiMH battery. The healthcare provider selects the appropriate program based on the patient's clinical needs (e.g., post-surgical strengthening or chronic pain management). The device output is delivered through the garment and electrodes, providing electrical stimulation to the peripheral nervous system to induce muscle contraction or modulate pain signals. This aids in muscle re-education, atrophy prevention, and symptomatic pain relief, potentially improving patient mobility and recovery outcomes.
Clinical Evidence
Bench testing only. The device was evaluated for electrical, mechanical, and thermal safety in accordance with IEC 60601-1 and IEC 60601-2-10 standards. Biocompatibility of electrodes was confirmed via reference to previous 510(k) clearances (K000947). No clinical trial data was required or provided for this submission.
Indicated for NMES (Programs 1-6): maintain/increase range of motion, prevent/retard disuse atrophy, re-educate muscles, early post-surgical quadriceps strengthening, improve post-surgical knee stability, relax muscle spasms, increase local blood circulation. Indicated for TENS (Programs 7-9): symptomatic relief/management of chronic intractable pain, adjunctive treatment for acute post-surgical/post-traumatic pain, relief of pain associated with arthritis, adjunctive therapy for pain/symptoms of knee osteoarthritis.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{1}------------------------------------------------
K110350
Image /page/0/Picture/1 description: The image shows the logo for "bio-medical research ltd.". The logo consists of the letters "bmr" in a stylized font, with the words "bio-medical research ltd." written in a smaller font below the letters. The letters "bmr" are connected to each other.
# Bio-Medical Research Ltd.
AUG - 3 2011 Parkmore Business Park West, Galway, Ireland Tel: +353 (0)91 774300 - Fax: +353 (0)91 774301
This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92.
# 1. Contact Details
| Name: | Anne-Marie Keenan |
|------------|-------------------------------------------------------------------------------|
| Title: | Quality/Regulatory Engineer |
| Address: | Bio-Medical Research Ltd.,<br>Parkmore Business Park, West<br>Galway, Ireland |
| Telephone: | +353 91 774300 |
| Fax: | +353 91 774302 |
| E-Mail: | akeenan@bmr.ie |
| Prepared: | July 25, 2011. |
#### 2. Device Name
| Trade Name of Device: | Kneehab XP, Type 412/421 |
|-------------------------|-----------------------------------------------------------------------------------------|
| Common Name: | Transcutaneous Electrical Nerve Stimulator<br>Powered Muscle Stimulator |
| Regulation Number: | 21 CFR 882.5890<br>21 CFR 890.5850 |
| Regulation Description: | Transcutaneous electrical nerve stimulator for pain relief<br>Powered muscle stimulator |
| Product Code: | IPF, GZJ, NYN |
| Device Class: | 2 |
# 3. Identification of Equivalent Legally Marketed Device
| 510(k) Number: | K083105 |
|----------------|-----------------------------------------|
| Manufacturer: | Bio-Medical Research Ltd. |
| Trade Name: | Kneehab XP Conductive Garment, Type 411 |
Directors: T. Kirwan (Chairman), Dr. P.A. Smith (Managing), Dr. B. McDonnell, C. Minogue, P. McGrath
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a blurry, high contrast image of what appears to be a collection of small, dark spots against a white background. The spots are irregularly shaped and scattered across the image. Due to the blurriness, it is difficult to determine the exact nature or origin of the spots.
| 510(k) Number: | K082011 |
|----------------|---------------------------------------|
| Manufacturer: | Bio-Medical Research Ltd. |
| Trade Name: | MediStim XP, Type 281 |
| 510(k) Number: | K082011 |
| Manufacturer: | Bio-Medical Research Ltd. |
| Trade Name: | MediTens XP, Type 458 |
| 510(k) Number: | K061516 |
| Manufacturer: | Compex Technologies, Inc |
| Trade Name: | Staodyn Max Preset |
| 510(k) Number: | K021100 |
| Manufacturer: | Empi |
| Trade Name: | 300 PV Complete Electrotherapy System |
| 510(k) Number: | K971437 |
| Manufacturer: | BioniCare Medical Technologies |
| Trade Name: | Bionicare® Stimulator System |
#### 4. Description of Device
The Kneehab XP, Type 412/421 is a portable, battery operated, combination device which can provide both neuromuscular electronic stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS). The device incorporates multipath®, a patented technology developed by neurotech® which enables the Kneehab XP Conductive Garment to deliver highly focused and accurate NMES muscle contractions. This device also provides a method of pain management and relief through the use of TENS technology.
The Kneehab XP pack consists of a rechargeable control unit, a left or right universally sized garment, a pack of custom adhesive electrodes, a battery charger and instructions for use. The garment is fastened around the thigh and above the kneecap and contains a connector socket into which the control unit is plugged. Power is derived from a 3.6V NiMH rechargeable battery pack that is pre-installed in the unit. A battery charger is included with
Page 2 of 10
{3}------------------------------------------------
the device and the device cannot be used while being charged. The adhesive electrodes have an estimated usage capability of 20 sessions when used under the recommended conditions of use.
All internal connections of the unit are over molded to prevent moisture ingress. The user has no access to the wiring or connectors within the garment and is unable to alter the current path. There are nine treatment programs in total (six NMES and three TENS) with a duration of 20 minutes each. Program details are included in the instructions for use. For purposes of hygiene, the garment may be cleaned and instructions for device care are included in the user manual.
#### 5. Statement of Intended Use and Indications for Use
Kneehab XP, Type 412/421 delivers stimulation based on the principles of NMES and TENS. NMES may be defined as the application of electrical stimulation of the peripheral nervous system to contract a muscle, either through the direct activation of the motor neurons in the mixed peripheral nerve, or indirectly through reflex recruitment. TENS can be defined as a pain therapy based on the application of electrical stimuli to the skin via stimulation of the nerve fibers.
In NMES mode (Programs 1-6), the Kneehab XP, Types 412/421. is indicated for use as follows:
- Maintain or increase the range of motion. 1.
- Prevention or retardation of disuse atrophy 2.
- Re-educate muscles 3.
- 4. Early post-surgical quadriceps strengthening and improved post surgical knee stability secondary to quadriceps strengthening
- Relax muscle spasms 5.
- Increase blood circulation 6.
Programs 1-6 use multipath technology.
{4}------------------------------------------------
| Program<br>Number | Duration<br>(Minutes) | Frequency/<br>Rate (Hz) | Pulse<br>Width<br>(µsec) | Ramp<br>Up<br>Time<br>(seconds) | Contraction<br>Time<br>(seconds) | Ramp<br>Down<br>Time<br>(seconds) | Relaxation<br>Time<br>(seconds) | Additional<br>Function | Indication<br>No. |
|-------------------|-----------------------|-------------------------|--------------------------|---------------------------------|----------------------------------|-----------------------------------|---------------------------------|------------------------|-------------------|
| P1 | 20 | 50 | 300-400 | 1 | 5 | 0.5 | 10 | Trigger | 4 |
| P2 | 20 | 50 | 300-400 | 1 | 10 | 0.5 | 10 | Trigger | 1, 2, 3, 5, 6 |
| P3 | 20 | 50 | 300-400 | 1 | 10 | 0.5 | 20 | Trigger | 1, 2, 3, 5, 6 |
| P4 | 20 | 50 | 300-400 | 1 | 10 | 0.5 | 30 | Trigger | 1, 2, 3, 5, 6 |
| P5 | 20 | 35 | 300-400 | 1 | 5 | 0.5 | 5 | Trigger | 1, 2, 3, 5, 6 |
| P6 | 20 | 70 | 300-400 | 1 | 10 | 0.5 | 50 | Trigger | 1, 2, 3, 5, 6 |
Neuromuscular Electrical Stimulation (NMES) Programs on Kneehab XP:
In TENS Mode (Programs 7 - 10), the Kneehab XP, Type 412/421 is indicated for use as follows:
- Provide symptomatic relief and management of chronic, intractable pain 7.
- Provide an adjunctive treatment in the management of acute, post-surgical or post-8. traumatic pain
- Provide symptomatic relief and management of intractable pain and relief of pain 9. associated with arthritis
- Provide an adjunctive therapy in reducing the level of pain and symptoms 10. associated with osteoarthritis of the knee
- 11.
Transcutaneous Electrical Nerve Stimulation (TENS) Programs on Kneehab XP:
| Program<br>Number | Duration<br>(Minutes) | Frequency/<br>Rate (Hz) | Pulse<br>Width<br>(µsec) | Ramp<br>Up<br>Time<br>(seconds) | Contraction<br>Time<br>(seconds) | Ramp<br>Down<br>Time<br>(seconds) | Relaxation<br>Time<br>(seconds) | Additional<br>Function | Indication<br>No. |
|-------------------|-----------------------|-------------------------|--------------------------|---------------------------------|----------------------------------|-----------------------------------|---------------------------------|------------------------|-------------------|
| P7 | 30 | 99 | 300 | N/A | Continuous | N/A | N/A | No Trigger | 7-10 |
| P8 | 30 | 4 | 300 | N/A | Continuous | N/A | N/A | No Trigger | 7, 8, 9 |
| P9 | 30 | 125 | 175 | N/A | Continuous | N/A | N/A | No Trigger | 7, 8, 9 |
# 6. Summary of Technological Characteristics
There are no new technological characteristics that could affect safety or effectiveness of the Kneehab XP, Type 412/421 device. A summary of the technological characteristics of the new device in comparison to the predicate device has been included below:
{5}------------------------------------------------
| General Comparison<br>Name of Device: | Proposed Device<br>Knechab XP | Predicate Device<br>Knechab XP<br>Conductive Garment | Predicate Device<br>MediTens XP | Predicate Device<br>MediStim XP | Predicate Device<br>Staodyn Max | Predicate Device<br>300 PV Complete<br>Electrotherapy System | Predicate Device<br>Blonicare® Stimulator<br>System |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number: | K110350 | K083105 | K082011 | K082011 | K061516 | K021100 | K971437 |
| Intended Use | The Kneehab XP is<br>intended to provide<br>transcutaneous<br>Neuromuscular Electrical<br>Stimulation (NMES) and<br>Transcutaneous<br>Electrical Nerve<br>Stimulation (TENS). | The Knechab XP is<br>intended to provide<br>transcutaneous<br>Neuromuscular Electrical<br>Stimulation (NMES) and<br>Transcutaneous<br>Electrical Nerve<br>Stimulation (TENS) | The MediTens XP is<br>intended to provide<br>Transcutaneous<br>Electrical Nerve<br>Stimulation (TENS). | The Knechab XP is<br>intended to provide<br>transcutaneous<br>Neuromuscular Electrical<br>Stimulation (NMES) and<br>Transcutaneous<br>Electrical Nerve<br>Stimulation (TENS) | The Knechah XP is<br>intended to provide<br>transcutaneous<br>Neuromuscular Electrical<br>Stimulation (NMES) and<br>Transcutaneous<br>Electrical Nerve<br>Stimulation (TENS) | The 300 PV is a<br>multifunction<br>electrotherapy device<br>intended to provide<br>Neuromuscular Electrical<br>Stimulation (NMES),<br>Transcutaneous<br>Electrical Nerve<br>Stimulation (TENS).<br>Interferential Current<br>Stimulation (IFS) and<br>Functional Electrical<br>Stimulation (FES). | The Bionicare is<br>battery operated TENS<br>stimulator producing<br>pulses at 100 Hz<br>Electrodes are applied to<br>the knee and thigh. |
| Prescriptive Use | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| Indications for Use | In NMES mode<br>(Programs 1-6) the<br>Knechab XP is intended<br>to:<br>Maintain or increase<br>range of motion,<br>Prevention or retardation<br>of disuse atrophy,<br>Re-educate muscles,<br>Early post-surgical<br>quadriceps strengthening<br>and improved post<br>surgical knee stability<br>secondary to quadriceps<br>strengthening. Relax<br>muscle spasms &<br>Increase local blood<br>circulation<br><br>In TENS Mode<br>(Programs 7 - 9) the<br>Knechab XP is intended<br>to:<br>Provide symptomatic<br>relief and management of<br>chronic, intractable pain,<br>Provide an adjunctive<br>treatment in the<br>management of acute,<br>post-surgical or post-<br>traumatic pain, Provide | Muscle re-education of<br>the quadriceps.<br>Maintaining or increase<br>range of motion of the<br>knee joint. Prevention or<br>retardation of disuse<br>atrophy in the<br>quadriceps, Early post-<br>surgical quadriceps<br>strengthening and<br>improved post-surgical<br>knee stability secondary<br>to quadriceps<br>strengthening &<br>Increasing local blood<br>circulation. | The symptomatic relief<br>and management of<br>chronic intractable pain.<br>It is also an adjunctive<br>treatment in the<br>management of post-<br>surgical and post-<br>traumatic pain. The<br>device has no curative<br>value and should only be<br>used in conjunction with<br>medical supervision. | Neuromuscular Electrical<br>Stimulation (NMES) for<br>relaxation of muscle<br>spasms. prevention or<br>retardation of disuse<br>atrophy, increasing local<br>blood circulation, muscle<br>re-education, immediate<br>post-surgical stimulation<br>of calf muscles to<br>prevent venous<br>thrombosis and<br>maintaining or increasing<br>range of motion.<br><br>Transcutaneous<br>Electrical Nerve<br>Stimulation (TENS) for<br>an adjunctive treatment<br>in the management of<br>post-surgical and post-<br>traumatic acute pain<br>problems. | The Staodyn® Max<br>Preset Transcutaneous<br>Electrical Nerve<br>Stimulator Device is used<br>for the symptomatic<br>relief and management of<br>chronic intractable pain<br>and relief of pain<br>associated with arthritis.<br>It is also used as an<br>adjunctive treatment in<br>the management of post-<br>surgical and post-<br>traumatic pain | (TENS/NMES<br>Indications for use)<br>As a NMES device, the<br>300 PV is indicated for<br>the following conditions:<br>Re-educating muscles,<br>Relaxation of muscle<br>spasm. Increasing local<br>blood circulation,<br>Retarding or preventing<br>disuse atrophy.<br>Maintaining or<br>increasing range of<br>motion & Prevention of<br>venous thrombosis of the<br>calf muscles immediately<br>after surgery<br><br>As a TENS device, the<br>300 PV is indicated for<br>the following conditions:<br>Symptomatic relief and<br>management of chronic,<br>intractable pain and<br>Adjunctive treatment for<br>post-surgical and post-<br>trauma acute pain | Provide an adjunctive<br>therapy in reducing the<br>level of pain and<br>symptoms associated<br>with osteoarthritis of the<br>knee and for overall<br>improvement of the knee |
ン
:
.
-.
.
:
Page 5 of 10
{6}------------------------------------------------
| General Comparison<br>Name of Device: | Proposed Device<br>Kneehab XP | Predicate Device<br>Kneehab XP<br>Conductive Garment | Predicate Device<br>MediTens XP | Predicate Device<br>MediStim XP | Predicate Device<br>Staodyn Max | Predicate Device<br>300 PV Complete<br>Electrotherapy System | Predicate Device<br>Bionicare® Stimulator<br>System | |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-------------------------|
| 510k Number: | K110350 | K083105 | K082011 | K082011 | K061516 | K021100 | K971437 | |
| | symptomatic relief and<br>management of<br>intractable pain. Relief of<br>pain associated with<br>arthritis.<br>Program 7 provides an<br>adjunctive therapy in<br>reducing the level of pain<br>and symptoms associated | | | | | | | |
| | with osteoarthritis of the<br>knee. | | | | | | | |
| Energy Used or<br>Delivered: | 3.6V<br>Rechargeable<br>Pack | NiMH<br>Battery | Same as proposed device | 9V Battery (type 6LR61) | 9V Battery (type 6LR61) | 3 x AAA Batteries | 2 x AA | 9V Battery (type 6LR61) |
| Unit: | Constructed from<br>injection moulded<br>thermosetting plastic<br>(ABS-PA-757) | Same as proposed device | Same as proposed device | Same as proposed device | Not available | Not available | 9V Battery (type 6LR61) | |
| Garment: | Contains an EEPROM.<br>Outer Fabric: 100%<br>Nylon, Inner Fabric: 70%<br>Polychloroprene & 30%<br>Polyurethane,<br>Binding: 82% Nylon &<br>18% Elastane.<br>Fastenings: 100% Nylon | Same as proposed device | Not Applicable (N/A) | N/A | N/A | N/A | Garment wraps around<br>knee and thigh, contains<br>conductive surfaces onto<br>which conductive gel is<br>applied. | |
| Electrode | Electrode A: 194 cm2,<br>Electrode B: 74 cm2,<br>Electrode C: 83 cm2 and<br>Electrode D: 66 cm2 | Same as proposed device | Industry standard<br>electrodes<br>5 x 5cm &<br>7 cm round<br>5 cm x 9cm etc. | Industry standard<br>electrodes<br>5 x 5cm &<br>7 cm round<br>5 cm x 9 cm etc. | Industry standard<br>electrodes<br>5 x 5cm &<br>7 cm round<br>5 cm x 9 cm etc. | Industry standard<br>electrodes<br>5 x 5cm &<br>7 cm round<br>5 cm x 9 cm etc. | 2 of 12 x 9 cm approx | |
| Lead Wires: | Over-moulded SATA<br>connector, splitting to 5<br>leads / studs. | Same as proposed device | Set of two (dark blue and<br>light blue), each lead<br>wire has 2 wires and<br>terminates with a 2mm<br>moulded pin for<br>connection to electrodes.<br>Constructed of PVC<br>insulated, containing 7-<br>strand tinsel copper with<br>interwoven Kevlar<br>reinforcing fibers. | Set of two (dark blue and<br>light blue), each lead<br>wire has 2 wires and<br>terminates with a 2mm<br>moulded pin for<br>connection to electrodes.<br>Constructed of PVC<br>insulated, containing 7-<br>strand tinsel copper with<br>interwoven Kevlar<br>reinforcing fibers. | Set of two, each lead<br>wire has 2 wires and<br>terminates with a 2mm<br>moulded pin for<br>connection to electrodes. | Set of two, each lead<br>wire has 2 wires and<br>terminates with a 2mm<br>moulded pin for<br>connection to electrodes. | Set of two, each lead<br>wire has 2 wires and<br>terminates with a 2mm<br>moulded pin for<br>connection to electrodes. | |
| Charger: | PC/ABS, complies with<br>IEC 60950 and UI, 1950 | Same as proposed device | N/A | N/A | N/A | YES, charges batteries<br>external to unit | N/A | |
| Standards Met | IEC 60601-1 (1998) & | Same as proposed device | Same as proposed device | Same as proposed device | Not available | Not Available | Not Available | |
・
.
Page 6 of 10
.
{7}------------------------------------------------
| General Comparison<br>Name of Device: | Proposed Device<br>Kneehab XP | Predicate Device<br>Kneehab XP<br>Conductive Garment | Predicate Device<br>MediTens XP | Predicate Device<br>MediStim XP | Predicate Device<br>Staodyn Max | Predicate Device<br>300 PV Complete<br>Electrotherapy System | Predicate Device<br>Bionicare® Stimulator<br>System |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|--------------------------------------------|--------------------------------------------|---------------------------------|--------------------------------------------------------------|-----------------------------------------------------|
| 510k Number: | K110350<br>1991, A2: 1995.<br>IEC 60601-2-10 (1987)<br>&<br>A1: 2001.<br>IEC 60601-1-2 (2001),<br>ISO 14971:2007,<br>ISO 10993-1:2009<br>ISO 10993-5:2009<br>ISO 10993-10:2002 &<br>A1:2006<br>21 CFR 898<br>21 CFR 801 | K083105 | K082011 | K082011 | K061516 | K021100 | K971437 |
| Biocompatibility | Electrodes - K000947 | Same as proposed device | Electrodes - K970426,<br>K874469 & K965194 | Electrodes - K970426,<br>K874469 & K965194 | Not available | Not available | Not Available |
| Compatibility with the<br>environment & other<br>devices | Complies to IEC 60601-<br>1-2 Medical electrical<br>equipment - Part 1-2:<br>General requirements for<br>safety - Collateral<br>standard:<br>Electromagnetic<br>compatibility<br>Requirements and tests | Same as proposed device | Same as proposed device | Same as proposed device | Not available | Not Available | Not Available |
| Sterility | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
| Electrical &<br>Mechanical Safety | Complies to IEC 60601-1<br>Medical electrical<br>equipment - Part 1:<br>General requirements for<br>safety & IEC 60601-2-10<br>Medical electrical<br>equipment - Part 2-10:<br>Particular requirements<br>for the safety of nerve<br>and muscle stimulators | Same as proposed device | Same as proposed device | Same as proposed device | Not available | Not available | Not available |
| Chemical Safety | MSDS Sheet (Electrode<br>Gel) | Same as proposed device | Same as proposed device | Same as proposed device | Not available | Not available | Not available |
| Thermal Safety | Complies to IEC 60601-1<br>& IEC 60601-2-10 | Same as proposed device | Same as proposed device | Same as proposed device | Not available | Not available | Not available |
| Radiation Safety | N/A | N/A | N/A | N/A | Not available | Not available | Not available |
:
・・
:
:
Page 7 of 10
:
{8}------------------------------------------------
| UNIT Comparison | Proposed Device | Predicate Device | Predicate Device | Predicate Device | Predicate Device | Predicate Device | Predicate Device |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------|----------------------------------|-------------------------|-------------------------|--------------------------------|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Name of Device: | Kneehab XP | Kneehab XP<br>Conductive Garment | MediTens XP | MediStim XP | Staodyn Max | 300 PV Complete<br>Electrotherapy System | |
| 510k Number: | K110350 | K083105 | K082011 | K082011 | K061516 | K021100 | |
| Manufacturer | China Turnkey Solutions<br>Logistics<br>(Shenzhen)<br>Co., Futian Free Trade<br>Zone. CHINA 518038 | Same as proposed device | Same as proposed device | Same as proposed device | Compex Technologies | Empi<br>599 Cardigan Road<br>St. Paul,<br>Minnesota 55126-4099 | BioniCare Medical<br>Technologies, Inc.,<br>47 R Loveton Circle<br>Sparks, MD 21152 |
| -Method of<br>line<br>Isolation | No line connection | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | No line connection |
| Patient Leakage<br>Current | N/A | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | |
| No. of Output Modes | 1 | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Several. eg Hi Volt, FES. | |
| Waveform/Shape | Pulsed, Symmetrical,<br>Biphasic. Rectangular<br>with interphase interval | Same as proposed device | Same as proposed device | Same as proposed device | Symmetrical Biphasic<br>Square | Asymmetric and<br>Symmetric square wave<br>options, Hi Volt pulse<br>option: exponential<br>spikes. | Monophasic spike<br>shaped pulse |
| No. of Output Channels | 2 | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | 2 |
| Synch/Alternating? | Synchronous<br>(Multiplexed) | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Synchronous and<br>alternating | Synchronous |
| -Method of channel<br>isolation | Transistor | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Transformer | unknown |
| Regulated Current or<br>Regulated Voltage | Regulated Current | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Claimed to be regulated<br>voltage |
| Software/Firmware/<br>Microprocessor<br>Control? | Yes | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device |
| Automatic Overload<br>Trip? | Yes, current limited,<br>indefinite short circuit<br>allowed | Same as proposed device | Same as proposed device | Same as proposed device | Unknown | Unknown | unknown |
| Automatic No-Load<br>Trip? | Yes | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | unknown |
| Automatic Shut Off? | Yes | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Same as proposed device | Unknown |
| Patient<br>Control? | Override Yes, pause button | Same as proposed device | Same as proposed device | Same as proposed device | Not available | Yes, Pause, FES modes | Yes stop button ends<br>treatment |
.
ﺮ ﺔ
·
.
Page 8 of 10
.
{9}------------------------------------------------
| UNIT Comparison | Proposed Device | Predicate Device | Predicate Device | Predicate Device | Predicate Device | Predicate Device | Predicate Device |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------|------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Name of Device: | Kneehab XP | Kneehab XP<br>Conductive Garment | MediTens XP | MediStim XP | Staodyn Max | 300 PV Complete<br>Electrotherapy System | |
| 510k Number: | K110350 | K083105 | K082011 | K082011 | K061516 | K021100 | |
| Timer range (mins) | 20 mins - open | Same as proposed device | Open | 30 minutes - open | 120 mins | 5 mins to open | Open |
| Weight (unit) | 116g (inc. batteries) | Same as proposed device | 142g (inc. batteries) | 142g (inc. batteries) | 145g | 226g | 136g |
| Dimensions<br>(W x H x D) | 60x23x115mm | Same as proposed device | 105x68x28mm | 105x68x28mm | 2.5" x 5.25" x 1.0" | 1.26" x 3.3" x 4.5" | 95x64x30mm |
| Frequency/<br>Phase Duration of<br>program with highest<br>output power. | Program 7<br>99Hz<br>$300\mu S$ | Program 1<br>50Hz<br>$400 \mu S$ | Program 3<br>99 Hz<br>$150\mu S$ | Program 3<br>50Hz<br>$400 \mu S$ | Program A<br>100HZ<br>$350\mu S$ | Custom User 2<br>150Hz<br>$400\mu S$ | 100Hz<br>$640\mu S$ |
| Baseline to Peak<br>Current | $80mA @ 500\Omega$<br>$28mA @ 2k\Omega$<br>$3.9mA @ 10k\Omega$ | $80mA @ 500\Omega$<br>$28mA @ 2k\Omega$<br>$3.9mA @ 10k\Omega$ | $75mA @ 500\Omega +/- 10%$<br>$35mA @ 2k\Omega$<br>$7mA @ 10k\Omega$ | $75mA @ 500\Omega +/- 10%$<br>$35mA @ 2k\Omega$<br>$7mA @ 10k\Omega$ | $60mA @ 500\Omega$ | $100mA @ 500\Omega$ | $24mA @ 500\Omega$ |
| Baseline to Peak<br>Output Voltage | $40.0V @ 500\Omega$<br>$55.6V @ 2k\Omega$<br>$39.3V @ 10k\Omega$ | $40.0V at 500 \Omega$<br>$55.6V @ 2k \Omega$<br>$39.3V at 10k \Omega$ | $37V @ 500\Omega$<br>$70V @ 2k\Omega$<br>$70V @ 10k\Omega$ | $37V @ 500\Omega$<br>$70V @ 2k\Omega$<br>$70V @ 10k\Omega$ | $30V @ 500\Omega$ | $50V @ 500\Omega$ | $12V 500\Omega$ |
| Maximum RMS Output<br>Voltage<br>(+/-10%) Vrms | $9.3 V @ 500 \Omega$<br>$17.1V @ 2k\Omega$<br>$14.1V @ 10k\Omega$ | $9.15V@ 500\Omega$<br>$10.74V @ 2k\Omega$<br>$5.65V @ 10k \Omega$ | $6.4 V @ 500\Omega$<br>$12.1V @ 2k\Omega$<br>$7 V @ 10k\Omega$ | $7.5V @ 500\Omega$<br>$14 @ 2k\Omega$<br>$7 V @ 10k\Omega$ | $8.85V @ 500\Omega$ | $10.5V @ 500\Omega$ | 4.3V<br>Estimated assuming a<br>square pulse shape<br>0.64ms |
| Maximum RMS Output<br>Current<br>(+/-10%) Irms | $18.6 mA@ 500 \Omega$<br>$8.6 mA@ 2k\Omega$<br>$1420 \mu A@ 10k\Omega$ | $18.3mA@ 500\Omega$<br>$5.37mA @ 2k\Omega$<br>$565\mu A @ 10k\Omega$ | $12.8mA @ 500\Omega$<br>$6.1mA @ 2k\Omega$<br>$700\mu A @ 10k\Omega$ | $15mA @ 500\Omega$<br>$7 mA @ 2k\Omega$<br>$490\mu A@ 10k\Omega$ | $15.9mA @ 500\Omega$ | $21mA @ 500\Omega$<br>(As<br>stated in device IFU,<br>however is inconsistent<br>with other stated<br>parameters) | 8.6mA |
| Pulse Width | $640 \mu s$ sum of both<br>phases $300\mu s$ +40 $\mu s$<br>interphase interval | $840\mu s$ (both phases max<br>$400\mu s$ with interphase<br>interval of $40\mu S$ ) | $300\mu S$ ( sum of both<br>phases; $150\mu S$ ) (Prog 1) | $800\mu S$ sum of both<br>phases; $400\mu S$ ) (Prog | $700\mu S$ - sum of both<br>phases; $350\mu S$<br>(Prog A) | $1200\mu S$ . sum of both<br>phases and<br>interphase interval | $640 \mu S$ |
| Net Charge<br>( $\mu C$ per pulse) | $0\mu C @ 500\Omega$ | $0\mu C @ 500\Omega$ | $0\mu C @ 500\Omega$ | $0\mu C @ 500\Omega$ | $0\mu C @ 500\Omega$ | $<20\mu A$ | Not available |
| Maximum Phase<br>Charge @500<br>(+/-20%) | $24 \mu C @ 500\Omega$ | $32.8\mu C$ | $10.5\mu C @ 500\Omega$ | $30 \mu C$ | $21 \mu C @ 500\Omega$ | $40 \mu C$ | $21 \mu C @ 500\Omega$ |
| Maximum Current<br>Density @500 $\Omega$ | $=18.6 mA/83 cm^2$<br>$=0.22 mA/cm^2$ | $0.22 mA/cm2$ | Using 2" square<br>electrode<br>$= 12.8 mA/25 cm^2$<br>$= 0.5 mA/ cm^2$ | Using 2" square<br>electrode<br>$= 15 mA/25 cm^2$<br>$= 0.6 mA/ cm^2$ | Using 2" square<br>electrode<br>$=15.9 mA/25 cm^2$<br>$= 0.64mA/ cm^2$ | Using 2" square<br>electrode<br>$=21 mA/25 cm^2$<br>$= 0.84mA/ cm^2$ | Using 12 x 9 square<br>electrodes<br>$=8.6 mA/108 cm^2$<br>$=0.08 mA/cm^2$ |
| Maximum Power<br>Density @500 $\Omega$<br>(using smallest<br>electrode conductive<br>surface area) | Program 7<br>$2.1mW/cm^2 @ 500\Omega$ | $2.55 mW/cm2$ | $3.3 mW/cm^2$ | $4.5 mW/cm^2$ | $2.9 mW/cm^2$ | $8.8 mW/cm^2$ | $0.34 mW/cm^2$ |
| Burst Mode<br>(ie pulse trains) | YES, NMES on off cycle | Yes, NMES on off cycle | Yes, burst mode TENS | Yes, NMES on off cycle | Yes, burst mode TENS | Yes. NMES and TENS<br>modes | Yes |
Page 9 of 10
{10}------------------------------------------------
#### 7. Substantial Equivalence
Bio-Medical Research Ltd (BMR) has over 30 years experience in the research, design, manufacture and marketing of medical grade products for both muscle strengthening and pain relief. Bio-Medical Research Ltd. complies with 21 CFR 820 and is registered to I.S. EN ISO 13485:2003, Medical Device Quality Management System for the design. manufacture and distribution of electro-medical devices.
Kneehab XP, Type 412/421 device complies with the following international safety standards:
- ロ IEC 60601-1 (1998) + A1: 1991, A2: 1995, Medical Electrical Equipment Part 1: General Requirements for Safety
- D IEC 60601-1-2 (2001). Medical Electrical Equipment Part 1-2: General Requirements for Safety; Electromagnetic Compatibility Requirements & Tests.
- ロ IEC 60601-2-10 (1987) + Al: 2001 Medical electrical equipment Part 2-10: Particular requirements for the safety of nerve and muscle stimulators
A risk management plan has been carried out to I.S. EN ISO 14971 2007.
{11}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Bio-Medical Research LTD. c/o Anne Marie Keenan Quality/Regulatory Engineer Parkmore Business Park, West Galway Ireland EI
Re: K110350
AUG - 3 2011
Trade/Device Name: Kneehab XP, Type 412/421 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Codes: GZJ, IPF, NYN Dated: June 24, 2011 Received: June 30, 2011
Dear Ms. Keenan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{12}------------------------------------------------
Page 2 - Ms. Anne Marie Keenan
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
# Indications for Use
KI 10350 510(k) Number (if known):
Device Name:
Kneehab XP, Type 412/421
Indications for Use:
The Kneehab XP, Type 412/421, Indications for use are as follows:
In NMES mode (Programs 1-6) the Kneehab XP is intended to:
- Maintain or increase the range of motion. ◆
- . Prevention or retardation of disuse atrophy
- . Re-educate muscles
- Early post-surgical quadriceps strengthening and improved post surgical knee . stability secondary to quadriceps strengthening
- . Relax muscle spasms
- . Increase blood circulation
In TENS Mode (Programs 7 - 9) the Kneehab XP is intended to:
- Provide symptomatic relief and management of chronic, intractable pain .
- . Provide an adjunctive treatment in the management of acute, post-surgical or post-traumatic pain
- . Provide symptomatic relief and management of intractable pain and relief of pain associated with arthritis
- . Provide an adjunctive therapy in reducing the level of pain and symptoms associated with osteoarthritis of the knee
Prescription Use ﺟ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use . (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Stan Olchov
(Division Sien-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K110350
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.