The ComboCare 2000 is intended for the temporary relief of minor muscle and joint pain, promoting relaxation of muscle tissue, symptomatic relief and management of chronic intractable pain, and adjunctive relief of post-surgical or post-traumatic acute pain.
Device Story
ComboCare 2000 combines multiple therapeutic modalities into a single portable device: pulsed infrared light (870 nm), pulsed red light (640 nm), vibratory massage, and electrical stimulation (E-Stim TENS and Micro-current TENS). Device operates on rechargeable NiMH batteries or wall-mount power. Intended for pain management; used by patients or clinicians. Output includes light therapy, vibration, and electrical pulses delivered via pads. Healthcare providers use device to provide adjunctive pain relief; patient benefits from temporary reduction in muscle/joint pain and muscle relaxation. Device integrates modalities previously found in separate predicate systems.
Clinical Evidence
No clinical data. Substantial equivalence determined via bench testing and design validation against voluntary standards: EN 60601-1 (General safety), EN 60601-1-2 (EMC), and EN 60601-2-10 (Nerve/muscle stimulator safety).
Technological Characteristics
Materials: NiMH batteries. Energy: 100-240 VAC or 9 VDC battery. Modalities: Pulsed IR LED (870 nm), Red LED (640 nm), vibratory massage (3 levels), E-Stim/Micro-current TENS. Waveform: 8 Hz asymmetrical biphasic square wave. Pulse width: 200 μs. Connectivity: None. Sterilization: N/A. Standards: EN 60601-1, EN 60601-1-2, EN 60601-2-10, ISO 14971.
Indications for Use
Indicated for patients requiring temporary relief of minor muscle and joint pain, relaxation of muscle tissue, symptomatic relief/management of chronic intractable pain, and adjunctive relief of post-surgical or post-traumatic acute pain.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
510(k) Number: _ _ K081141
Date: _
Page: 1 of 3
# 510(k) Summary
#### Introduction
This summary is intended to comply with requirements of the SMDA and 21CFR§807.92. FDA may make this summary available to the public within 30 days following a finding of substantial equivalence.
## 510(k) Applicant
Xanacare Technologies LLC 9185 East Kenyon Ave., Ste. 270 Denver, CO 80237 USA Tel: 720-554-9262 / Fax: 720-554-9264
## 510(k) Correspondent
Robert N. Clark, President and Senior Consultant Medical Device Regulatory Advisors, Inc. 13605 West 7th Ave., Golden, CO 80401 USA Tel: 303-463-0900 / Fax: 303-558-3833
## Date Prepared
August 19, 2008
## Trade Name of Device
ComboCare 2000
#### Common Name of Device
Lamp, non-heating, for adjunctive use in pain therapy
#### Classification Name
Infrared Lamp
## 510(k) Classification
Class II
#### Device Description and Intended Use
The ComboCare 2000 is intended for the temporary relief of minor muscle and joint pain, promoting relaxation of muscle tissue, symptomatic relief and management of chronic intractable pain, and adjunctive relief of post-surgical or post-traumatic acute pain.
{1}------------------------------------------------
| 510(k) Number: | K081141 |
|----------------|---------|
| Date: | |
| Page: 2 of 3 | |
## Predicate Devices
The ComboCare 2000 does not differ significantly from predicate devices in its technological characteristics, except that it combines modalities using Red LED light, infrared LED light, therapeutic vibratory massage and E-Stim or Micro-current TENS into a single package. Modalities are similarly combined in the TerraQuant device, except that the TerraQuant uses higher power laser IR diodes and does not include a vibrator. The ComboCare 2000 has technological characteristics similar to the following predicate device(s):
- K071445 Escada International, Inc. TerraQuant MQ2000 v.5 with TQ-ITENS .
- K021755 Apex Medical Corp., Apex TS1211 .
- K061650 EMPI, EMPI Select TENS ●
- K041530 Excalibur Light Therapy System .
- . K041565 Lazr Pulsr 4x
- K050668 GRT LITE Model 8-A Light Therapy System .
#### Summary of Technological Characteristics Compared to Predicate Devices
| | ComboCare 2000 | Predicate Device |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Electrical Power Source | Class II Wall-Mount Battery Charger Input<br>100-240 VAC 50-60 Hz, 0.6A Output: 9 VDC<br>at 2A | Equivalent to other predicate devices rated for<br>international electrical use (i.e. 100-240 VAC<br>50-60 Hz). |
| Battery Type | 4 rechargeable Size AA NiMH batteries | Equivalent to other predicate devices that use<br>rechargeable batteries for portability. |
| Infrared Light | Pulsed infrared light emitting diodes (IR<br>LED). 24 IR LEDs @ 870 nm (standard pad),<br>48 IR LEDs @ 870 nm (large pad). | Equivalent to other predicate devices cleared<br>as non-heating lamps for adjunctive use in<br>pain therapy. |
| Red Light | Pulsed red light emitting diodes (RED LED).<br>32 RED LEDs @ 640 nm (standard pad), 64<br>RED LEDs @ 640 nm (large pad). | Equivalent to other predicate devices cleared<br>as non-heating lamps for adjunctive use in<br>pain therapy |
| Vibratory massage | Off + 3 levels | Equivalent to other predicate therapeutic<br>electrical massagers and vibrators. |
| | E-Stim TENS mode | |
| Maximum Output Current | Maximum peak output current (1 KΩ load) =<br>30 mA / Maximum peak output current (500 Ω<br>load) = 40 mA | K021755: 0-80 mA (500 Ω load)<br>K061650: 0-60 mA, adjustable (500 Ω load) |
| Maximum Output Voltage | Maximum peak output voltage (1 KΩ load) =<br>30 V / Maximum peak output voltage (500 Ω<br>load) = 20 V | K021755: 40 V (500 Ω load)<br>K061650: 0-60V (1K Ω load) |
| Pulse Frequency | 72 pps/Hz nominal | K021755: 2-150 pps/Hz<br>K061650: 2-150 pps/Hz |
| Pulse Width | 200 μs | K021755: 50-300 μs<br>K061650: 0-400 μs |
| | Micro-Current TENS Mode | |
| Maximum Output Current | Maximum peak output current (1 KΩ load) –<br>9.3 mA / Maximum peak output current (500<br>Ω load) = 14.4 mA | K021755: 0-80 mA (500 Ω load)<br>K061650: 0-60 mA, adjustable (500 Ω load |
| Maximum Output Voltage | Maximum peak output voltage (1 KΩ load) =<br>9.3v V / Maximum peak output voltage (500<br>Ω load) = 7.2 V | K021755: 40 V (500 Ω load)<br>K061650: 0-60V (1K Ω load) |
| Waveform | 8 Hz Asymmetrical biphasic square wave | K021755: Asymmetrical biphasic square wave<br>K061650: Asymmetrical biphasic square wave |
{2}------------------------------------------------
| 510(k) Number: | K081141 |
|----------------|---------|
| Date: | |
| Page: 3 of 3 | |
## Clinical & Non-Clinical Testing
Xanacare did not conduct, nor rely upon, clinical tests to determine substantial equivalence. Nonclinical testing was performed in order to validate the design against the company's specified design requirements, and to assure conformance with the following voluntary design standards:
- EN 60601-1 "Medical electrical equipment Part 1: General requirements for safety". .
- EN 60601-1-2 "Medical electrical equipment -- Part 1-2: General requirements for safety Collateral ● Standard"
- EN 60601-2-10 "Medical electrical equipment Part 2: Particular requirements for the safety of nerve . and muscle stimulators"
#### Risk Management
This device has been designed to either completely eliminate or mitigate known health hazards associated with the use of the device. Health hazard risk reduction has been accomplished by rigorous application of a risk management program according to standard ISO 14971.
## Substantial Equivalence
Based on the above Xanacare Technologies LLC believes that the ComboCare 2000 is safe and effective when used as instructed by knowledgeable and trained personnel, and is substantially equivalent to the legally marketed predicate device(s).
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### AUG 2 9 2008
Xanacare Technologies, LLC. % Medical Device Regulatory Advisors, Inc. Mr. Robert N. Clark President and Senior Consultant 13605 West 7th Avenue Golden, Colorado 80401
Re: K081141
Trade/Device Name: ComboCare 2000 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief. Regulatory Class: Class II Product Code: GZJ, NHN, ISA Dated: August 19, 2008 Received: August 21, 2008
Dear Mr. Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Mulkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K081141 ComboCare 2000 Device Name: _ Indications for Use: The ComboCare 2000 is intended for the temporary relief of minor muscle and joint pain, promoting relaxation of muscle tissue, symptomatic relief and management of chronic intractable pain, and adjunctive relief of post-surgical or post-traumatic acute pain. Prescription Use × AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) .............................................................................................................................................................................. waluation (ODE) Concurrence of CDRH, Office of Device E (Division Sign-C Division and Neur 510(k) Number
Page 1 of 1
ComboCare 2000 K081141/S002 Page 8 of 29
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.