This device is an electrical nerve stimulator intended for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is intended for use for (1) Symptomatic relief and management of chronic (long term) intractable pain and (2) adjunctive treatment in the management of post surgical and post traumatic pain problems. This device is an electrically powered muscle stimulator intended for use for medical purposes to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is intended for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation. (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy.
Device Story
Comfy Series (EV-804, EV-805, EV-806) are TENS/EMS stimulators; deliver electrical current to nerves/muscles via skin electrodes; parameters (pulse rate, width, timer, contraction/relaxation/ramp times) controlled via push buttons; LCD interface displays settings; EV-804 (TENS), EV-805 (EMS), EV-806 (Combination); used by patients for pain management or muscle therapy; output intensity adjustable; device provides therapeutic benefit through nerve stimulation or muscle contraction.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Powered muscle/nerve stimulator; electrical current output; push-button control interface; liquid crystal display (LCD); lead wire electrode connection; battery-operated; Class II device.
Indications for Use
Indicated for patients requiring pain relief (chronic intractable, post-surgical, post-traumatic) via TENS, or muscle stimulation (spasm relaxation, circulation increase, venous thrombosis prevention, re-education, range of motion maintenance, disuse atrophy prevention) via EMS.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
K07/951/
SEP - 5 2007
# 510(k) Summary As Required by 21 section 807.92 ( c )
- Everyway Medical Instruments Co., Ltd 1-Submitter Name: 3 FL., No.5, LANE 155, Sec. 3, Peishen Rd 2-Address: Shen Keng Hsiang. 222 Taipei Hsien. Taiwan (ROC) (886) 2 2662 0038
- 3-Phone: (886) 2 2664 5566
- 4-Fax:
Mr Robert Tu (General Manager) 5-Contact Person:
- 6-Date summary prepared: July 10th, 2007
- 7 Official Correspondent: Mansour Consulting LLC
- 8- Address: 845 Aronson Lake Court. Roswell, GA 30075 USA
- 9- Phone: 678-908-8180
- 10- Fax: 678-623-3765
- 11- Contact Person: Jay Mansour, President
#### 12-Device Trade or Proprietary Name:
- (a) Comfy TENS (Model EV-804)
- (b) Comfy EMS (Model EV-805)
- (c) Comfy STIM (Model EV-806)
#### 13-Device Common or usual name:
- (a) TENS UNIT
- (b) EMS UNIT
- (c) TENS/EMS COMBINATION UNIT
#### 14-Device Classification Name:
- (a) Stimulator, nerve, transcutaneous, for pain relief
- (b) Stimulator, muscle, powered
- (c) Combination of (i) Stimulator, nerve, transcutaneous, for pain relief AND (ii) Stimulator, muscle, powered
- 15-Substantial Equivalency is claimed against the following device:
- · Digital TENS, model EV-803 from Everyway Medical Instruments Co., Ltd. 510k# K020749
- · Digital EMS, model EV-807 from Everyway Medical Instruments Co., Ltd. 510k# K020750
#### 16-Description of the Device:
The Comfy Series, which includes models EV-804, EV-805, EV-806, are Transcutaneous Electrical Nerve Stimulator for pain relief and/or Electrical Muscle Stimulator. The stimulator sends gentle electrical current to underlying nerves and muscle group via electrodes applied on the skin. The parameters of units are controlled by the press buttons. Its intensity level is adjustable according to the needs of patients.
The three models have the same housing. The process to set the parameter and attach lead wires to the unit are also the same. Yet, they have different liguid crystal display and parameters for patients to create their own settings.
The EV-804 is a TENS device with 5 modes and adjustable pulse rate, pulse width and timer. The EV-805 is an EMS device with 3 modes and adjustable pulse rate, pulse width, contraction time, relaxation time, ramp time and timer. The EV-806 is
{1}------------------------------------------------
a combination unit with both TENS and EMS functions. The function can be selected by press buttons. The range of settings are identical to those of EV-804 and EV-805. The difference on the three units can be identified by the liquid crystal display.
#### 17-Intended use of the device: (refer to FDA form attached)
#### The intended Use of Comfy TENS, Model EV-804 TENS is as follows:
This device is an electrical nerve stimulator intended for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is intended for use for (1) Symptomatic relief and management of chronic (long term) intractable pain and (2) adjunctive treatment in the management of post surgical and post traumatic pain problems
#### The Intended Use of Comfy EMS, Model EV-805 EMS is as follows:
This device is an electrically powered muscle stimulator intended for use for medical purposes to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is intended for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation. (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy
### The Intended Use of Comfy Stim, Model EV-806 TENS/EMS Combination Unit are as follows:
This combination device is intended for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is intended for use for (1) Symptomatic relief and management of chronic (long term) intractable pain and (2) adjunctive treatment in the management of post surgical and post traumatic pain problems
It is also intended for use for medical purposes to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is intended for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation, (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy
## 18-Safety and Effectiveness of the device:
This series is safe and effective as the predicate devices cited above. The design modifications that EV-803 and EV-807 underwent did not affect safety and effectiveness of the new series EV-804, EV-805 and EV-806 as detailed within this submission.
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Everyway Medical Instruments Co., Ltd. C/O Jay Mansour President Mansour Consulting LLC 845 Aronson Lake Court Roswell, Georgia 30075
SEP - 5 2007
Re: K071951
Trade Name: Comfy Series, Models EV-804, EV-805, and EV-806 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF and GZJ Dated: August 10, 2007 Received: August 15, 2007
Dear Mr. Mansour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
{3}------------------------------------------------
Page 2 -- Mr. Jay Mansour
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Neil R. Oyler
J Mark N. Melkerson, M.S. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K071951
Device Name: COMFY SERIES EV-804, EV-805 & EV-806
Indications For Use:
APPLICABLE FOR EV-804 (TENS) AND EV-806 (TENS/EMS COMBINATION UNIT):
This device is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for (1) Symptomatic relief and management of chronic (long term) intractable pain and (2) adjunctive treatment in the management of post surgical and post traumatic pain problems
APPLICABLE FOR EV-805 (EMS) AND EV-806 (TENS/EMS COMBINATION UNIT):
This device is an electrically powered muscle stimulator indicated for use for medical purposes to repeatedly contract muscles by passing electrical currents through electrodes contacting the affected body area. In particular, this device is indicated for use for (1) Relaxing muscle spasms, (2) Increasing local blood circulation, (3) Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, (4) Muscle re-education, (5) Maintaining or increasing range of motion, and (6) Preventing or retarding disuse atrophy
V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil RP Ogden for mxm
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 1 of 1
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.