K020180 · Woodside Biomedical, Inc. · GZJ · Mar 21, 2002 · Neurology
Device Facts
Record ID
K020180
Device Name
RELIEFBAND DEVICE, MODELS RB-DL, RB-EL, RB-RL
Applicant
Woodside Biomedical, Inc.
Product Code
GZJ · Neurology
Decision Date
Mar 21, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ReliefBand® device is indicated for use in the relief of nausea and vomiting due to motion sickness, and for the relief of mild to moderate nausea and vomiting associated with pregnancy.
Device Story
ReliefBand® models RB-DL, RB-EL, and RB-RL are non-invasive transcutaneous nerve stimulators worn on the ventral wrist. Device provides electrical stimulation to nerves located between the palmaris longus and flexor carpi radialis tendons. User interface includes a pushbutton and five blinking LEDs to select and display one of five intensity levels; a sixth LED indicates low battery. Powered by two 3V lithium batteries (disposable in RB-DL; replaceable in RB-EL/RB-RL). Device is intended for patient self-use to manage nausea and vomiting symptoms. Stimulation intensity is controlled by the user via the pushbutton, which adjusts peak pulse amplitude. Benefit is the non-pharmacological relief of nausea and vomiting.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological comparison to previously cleared predicate devices.
Technological Characteristics
Transcutaneous nerve stimulator; 21 CFR 882.5890; Product Code GZJ. Powered by two 3V lithium batteries. Wrist-worn form factor. User interface: pushbutton for intensity selection, 5-LED intensity display, 1-LED low battery indicator. Stimulation principle: electrical impulse to wrist nerves. Battery life: 150 hours at setting 3.
Indications for Use
Indicated for OTC relief of nausea and vomiting due to motion sickness and mild to moderate nausea and vomiting associated with pregnancy.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
ReliefBand® Device Models RB-2, RB-6, and RB-R (K982967)
ReliefBand® Device Models WB-2L, WB-6L, and WB-RL (K994387)
Submission Summary (Full Text)
{0}------------------------------------------------
KOZO190
### 510(k) SUMMARY MAR 2 1 2002 Woodside Biomedical Inc. ReliefBand® Device Models RB-DL, RB-EL, and RB-RL
## SUBMITTER INFORMATION
| A. | Company Name: | Woodside Biomedical, Inc. |
|----|--------------------------------|-----------------------------------------------------------------------------------|
| B. | Company Address: | 1915 Aston Avenue<br>Carlsbad, CA 92008 |
| C. | Company Phone:<br>Company Fax: | (760) 804-6900<br>(760) 804-6925 |
| D. | Contact Person: | Tom Grey<br>Vice-President of Product<br>Development<br>Woodside Biomedical, Inc. |
| E. | Date Summary Prepared: | January 16, 2002 |
## DEVICE IDENTIFICATION
| A. | Classification Name: | Transcutaneous Nerve Stimulator |
|----|-------------------------|---------------------------------|
| B. | Trade/Proprietary Name: | ReliefBand® Device |
| C. | Classification: | Class II (21 CFR 882.5890) |
| D. | Product Code: | GZJ |
# IDENTIFICATION OF PREDICATE DEVICE
The Woodside Biomedical, Inc. ReliefBand® device Models RB-DL, RB-EL, and RB-RL are of comparable type and are substantially equivalent to the following predicate devices:
296
p // 2
1-16-02
{1}------------------------------------------------
| Predicate Device | Manufacturer | 510(k) No. | Date Cleared |
|-----------------------------------------------------------------|---------------------------|------------|-------------------|
| ReliefBand® Device<br>Models RB-2, RB-6,<br>and RB-R (OTC) | Woodside Biomedical, Inc. | K982967 | February 23, 1999 |
| ReliefBand® Device<br>Models WB-2L,<br>WB-6L, and WB-RL<br>(Rx) | Woodside Biomedical, Inc. | K994387 | March 16, 2000 |
### DEVICE DESCRIPTION
The ReliefBand® Device Models RB-DL, RB-EL, and RB-RL are non-invasive devices which are indicated for over the counter use in the relief of nausea and vomiting (NV) due to motion sickness, and for the relief of mild to moderate nausea and vomiting associated with pregnancy. The devices are contained within a wristband, and provide relief through electrical stimulation of the nerves in the patient's wrist.
The devices can be worn on the ventral or palmar (i.e., inside) surface of either or both wrists, approximately 2 fingers breadth above the distal skin crease of the wrist joint, between the tendons of the palmaris longus and flexor carpi radialis muscles.
The ReliefBand® device Model RB-xL family has a user display that incorporates five blinking LEDs which are used to identify the intensity level (5 discrete LEDs, one for each intensity level), so that the patient can easily select the desired stimulation. Selection of the intensity level is performed via a pushbutton located on the user display, which controls the peak pulse amplitude of the electrical impulse and thereby determines the intensity of the stimulation. A sixth blinking LED is used to display the low battery indicator.
The ReliefBand® device Model RB-xL family is powered by two 3V lithium batteries. The batteries are not user replaceable in the disposable device model RB-DL, but are user
297
1-16-02
{2}------------------------------------------------
replaceable in the reusable device models RB-EL and RB-RL. The battery life for all models is specified to be 150 hours when used at setting 3.
### INDICATIONS FOR USE
The ReliefBand® Devices are indicated for over the counter use in the relief of nausea and vomiting (NV) due to motion sickness, and for the relief of mild to moderate nausea and vomiting associated with pregnancy.
## TECHNOLOGICAL CHARACTERISTICS
A comparison of the technological characteristics of the ReliefBand® Device and the predicate device has been performed. The results of this comparison demonstrate that the ReliefBand® Device is equivalent to the marketed predicate device. The differences between the ReliefBand® Device Models RB-DL, and RB-RL and RB-RL and the predicate models are insignificant and do not affect the safety or effectiveness of the device.
#### PERFORMANCE DATA
The performance data indicate that the ReliefBand® Device Models RB-DL, RB-EL, and RB-RL are substantially equivalent to the predicate ReliefBand® Devices distributed under K982967 and K994387.
#### CONCLUSIONS
Woodside Biomedical, Inc. has demonstrated through its evaluation of the ReliefBand® Device Models RB-DL, RB-EL, and RB-RL that the devices are equivalent to the predicate devices with respect to intended use, technological characteristics, and safety and effectiveness.
298
1-16-02
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAR 2 1 2002
Woodside Biomedical,Inc. C/O Carol L. Patterson Patterson Consulting Group 21911 Erie Lane Lake Forest, California 92630
Re: K020180
Trade/Device Name: ReliefBand® Device Models RB-DL, RB-EL and RB-RL Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator Regulatory Class: Class II Product Code: GZJ Dated: January 17, 2002 Received: January 18, 2002
Dear Ms. Patterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
### Page 2 – Ms. Carol L. Patterson
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality byevelle (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and in your your finding of substantial equivalence of your device to a legally premation fouried.com "Theesults in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you desire specific acrise for for in vitro diagnostic devices), please contact the Office of additionally 21 Of IC For at 659. Additionally, for questions on the promotion and advertising of Compliance at (301) 594-6639 Compliance at (301) 594-4639. Also, please note the your do roos, pitted, "Misbranding by reference to premarket notification" (21CFR Part 807.97). regaration on information on your responsibilities under the Act may be obtained from the Oinor gonetal intornational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark A. Millison
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# OTC INDICATIONS FOR USE
510(k) Number:
KO20180 (To Be Assigned By FDA)
Device Name:
ReliefBand® Device Models RB-DL, RB-EL, and RB-RL
Indications For Use:
The ReliefBand® device is indicated for use in the relief of nausea and vomiting due to motion sickness, and for the relief of mild to moderate nausea and vomiting associated with pregnancy.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use **__** OR Over-The-Counter Use \$\boxed{\checkmark}\$
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of General Restorative
and Neurological Devices
Confidential K020180
510(k) Number **__**25 1-16-02