K233006 · Myndtec, Inc. · GZI · Jun 25, 2024 · Neurology
Device Facts
Record ID
K233006
Device Name
MyndMove, MyndMove 2.0
Applicant
Myndtec, Inc.
Product Code
GZI · Neurology
Decision Date
Jun 25, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MyndMove and MyndMove 2.0 are electrical stimulation devices indicated for the following uses: Functional electrical stimulation (FES) Improvement of arm and hand function and active range of motion in patients with hemplegia due to stroke or upper limb paralysis due to C3-T1 spinal cord injury. NeuroMuscular Electrical Stimulation (NMES) for general rehabilitation for: - maintenance and/or increase of arm and hand range of motion, - prevention and/or retardation of disuse atrophy, - increase in local blood circulation, - reduction in muscle spasm, and - re-education of muscles. MyndMove and MyndMove 2.0 therapies can only be administered by Occupational or Physical Therapy professionals that have completed MyndMove training by MyndTec on the use of the MyndMove 2.0 Systems.
Device Story
Neuromodulation device delivering electrical pulses to stimulate muscle contractions; enhances motor recovery post-stroke or spinal cord injury. Comprises stimulator, electrodes, cables, hand/foot switches, and integrated software. Used in clinical settings (and home for MyndMove 2.0) by trained therapists. Therapist selects pre-programmed protocols via touch screen UI; customizes intensity. Device connects via IEEE 802.11 wireless to cloud-based backend for patient/protocol management. Patient performs functional tasks while therapist triggers stimulation bursts; provides efferent/afferent input to promote neuroplasticity. Benefits include improved motor function and range of motion.
Clinical Evidence
No clinical data; bench testing only. Device conforms to previously cleared standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11, IEC 62304). Biocompatibility of patient-contacting accessories confirmed per ISO 10993-1.
Indicated for patients with hemiplegia due to stroke or upper limb paralysis due to C3-T1 spinal cord injury. Used for FES to improve arm/hand function and active range of motion, and NMES for general rehabilitation (maintenance/increase of range of motion, prevention/retardation of disuse atrophy, increased local blood circulation, reduction of muscle spasm, muscle re-education). Administered by trained Occupational or Physical Therapy professionals.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 25, 2024
MyndTec Inc. Yesmil Pena Quality Assurance Manager 1900 Minnesota Court Mississauga, Ontario L5N 3C9 Canada
Re: K233006
Trade/Device Name: MyndMove, MyndMove 2.0 Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI, IPF Dated: May 31, 2024 Received: June 10, 2024
Dear Yesmil Pena:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Amber T. Ballard -S
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices
{3}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233006
Device Name MyndMove and MyndMove 2.0
#### Indications for Use (Describe)
MyndMove and MyndMove 2.0 are electrical stimulation devices indicated for the following uses:
Functional electrical stimulation (FES)
Improvement of arm and hand function and active range of motion in patients with hemplegia due to stroke or upper limb paralysis due to C3-T1 spinal cord injury.
NeuroMuscular Electrical Stimulation (NMES) for general rehabilitation for:
- · maintenance and/or increase of arm and hand range of motion,
- · prevention and/or retardation of disuse atrophy,
- · increase in local blood circulation,
- · reduction in muscle spasm, and
- · re-education of muscles.
MyndMove and MyndMove 2.0 therapies can only be administered by Occupational or Physical Therapy professionals that have completed MyndMove training by MyndTec on the use of the MyndMove 2.0 Systems.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary - K233006
## Submitter / Owner Information
MyndTec Inc. 1900 Minnesota Court, Suite 122 Mississauga, Ontario Canada, L5N 3C9
#### Contact Person
Yesmil Pena Telephone: 905-363-0564 ext. 225 Toll Free: 1-888-363-0581 1-877-796-4624 Fax: yesmil.pena@myndtec.com
#### Date Prepared:
June 21, 2024
#### Device Identification
| A. Proprietary Name | MyndMove and MyndMove 2.0 |
|--------------------------|------------------------------------------------------------------------------------------------------------------|
| B. Device Classification | Class II |
| C. Panel | Neurology |
| D. Device Product Code | GZI and IPF |
| E. Regulation | 21 CFR 882.5810 (External functional neuromuscular<br>stimulator)<br>21 CFR 890.5850 (Powered muscle stimulator) |
#### Identification of Predicate Devices
| Device | Applicant | 510(k) No. | Date Cleared |
|--------------|--------------|------------|-----------------|
| MyndMove | MyndTec Inc. | K170564 | August 30, 2017 |
| MyndMove 2.0 | MyndTec Inc. | K212149 | March 03, 2022 |
## Device Description
The subject devices of K233006 include MyndMove 2.0. There have been no changes to the technological characteristics of MyndMove or MyndMove 2.0 since the clearances of K170564 and K212149.
The MyndMove and MyndMove 2.0 Systems are neuromodulation devices that deliver short electrical pulses to stimulate muscle contractions and enhance motor recovery following Stroke or Spinal Cord Injury. The MyndMove and MyndMove 2.0 systems comprise the Stimulator device, stimulation electrodes and cables, hand and foot switches, and integrated software.
The Functional Electrical Stimulators in the MyndMove 2.0 are eight-channel devices
{6}------------------------------------------------
with a touch screen user interface (UI). The Stimulator software allows the User to choose from a number of pre-programmed stimulation protocols, to customize the stimulation intensities for each patient, and to document the treatments provided.
Each MyndMove/MyndMove 2.0 Stimulator is connected, via an IEEE 802.11 a/b/g/n wireless radio adapter, to a cloud-based backend data management system. The primary function of the backend is to manage the various elements of the MyndMove 2.0 ecosystem including but not limited to patient unique IDs and prescriptions, therapist profiles and stimulation protocols.
MyndMove and MyndMove 2.0 Therapies can only be administered by trained MyndMove Therapy professionals (TMMT) with assigned Therapist Log-In IDs. TMMTs are Occupational or Physical Therapy professionals that have completed MyndMove training by MyndTec on the use of the MyndMove/MyndMove 2.0 system.
MyndMove and MyndMove 2.0 use surface electrical stimulation to produce muscle contractions in different combinations to achieve a wide range of reaching and grasping functions. To achieve these movements, the single use electrodes are placed in various combinations on the surface of muscles in the arm and hand.
MyndMove and MyndMove 2.0 are based on advanced functional electrical stimulation (FES) principles designed to promote neuroplasticity by influencing the efferent and afferent pathways. Proper sequencing of the muscle contractions as per the MyndMove or MyndMove 2.0 protocols achieve a wide range of functions. During the delivery of MyndMove 2.0 therapy, the patient executes functional tasks with assistance from the MyndMove/MyndMove 2,0 device. The patient attempts to perform the movement and is then assisted to complete the movement when the therapist activates the MyndMove 2.0 stimulation protocol, which generates bursts of short electrical pulses, using surface electrodes, to produce muscle contraction. In this manner, the patient gets both efferent and afferent input as they are actively attempting a movement. Non-damaged pathways can be activated and non-damaged areas of the central nervous system can be trained to substitute for injured areas.
## Indications for Use
MyndMove and MyndMove 2.0 are electrical stimulation devices indicated for the following uses:
Functional electrical stimulation (FES)
Improvement of arm and hand function and active range of motion in patients with hemipleqia due to stroke or upper limb paralysis due to C3-T1 spinal cord injury.
NeuroMuscular Electrical Stimulation (NMES) for general rehabilitation for:
- . maintenance and/or increase of arm and hand range of motion,
- . prevention and/or retardation of disuse atrophy,
- . increase in local blood circulation,
- . reduction in muscle spasm, and
- . re-education of muscles.
MyndMove and MyndMove 2.0 therapies can only be administered by Occupational or Physical Therapy professionals that have completed MyndMove training by MyndTec on the use of the MyndMove and MyndMove 2.0 Systems.
{7}------------------------------------------------
# Type of Use
Prescription Use (Rx)
## Summary of Non-Clinical Testing
The subject devices conform to all applicable standards listed below, which were leveraged from the previously cleared predicate devices (K170564 and K212149). No additional non-clinical performance testing was conducted because there was no change to the hardware or software since clearances of MyndMove (K170564) and MyndMove 2.0 (K212149).
| FDA Recognition No. | Standard Title |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 17-16 | IEC 60601-2-10 Edition 2.1 2016-04. Medical electrical equipment - Part 2-10:<br>Particular requirements for the basic safety and essential performance of nerve<br>and muscle stimulators. |
| 19-4 | ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and<br>A2:2010/(R)2012 (Consolidated Text): Medical electrical equipment - Part 1:<br>General requirements for basic safety and essential performance (IEC 60601-<br>1:2005, MOD). |
| 19-8 | IEC 60601-1-2 Edition 4.0 2014-02: Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and essential performance - Collateral<br>Standard: Electromagnetic disturbances - Requirements and tests. |
| 19-14 | IEC 60601-1-11 Edition 2.0 2015-01: Medical electrical equipment - Part 1-11:<br>General requirements for basic safety and essential performance - Collateral<br>Standard: Requirements for medical electrical equipment and medical electrical<br>systems used in the home healthcare environment. |
| 13-79 | IEC 62304:2006 Medical device software - Software life cycle processes. |
The electrode cables included in the system comply with the U.S. Federal Performance Standard as set forth in 21 CFR 898 for electrode lead wires and Patient Cables. The patient contacting accessories included in the system (the electrodes and security tape) comply with the requirements for biocompatibility (ISO 10993-1).
## Basis for Substantial Equivalence
There have been no changes made to the technological characteristics and therapy protocols of MyndMove or MyndMove 2.0 since their clearance in K170564 or K212149. Both MyndMove and MyndMove 2.0 have the same intended use and deliver functional electrical and neuromuscular stimulation using coded therapeutic algorithmic protocols using an eight- channel stimulator with a maximum 20 mA output current for all 9cm x 5cm, 5cm x 5cm, 2.5cm diameter and 1cm x 3cm single use electrodes. MyndMove and MyndMove 2.0 devices of the present submission have identical technical specifications when compared to previously cleared MyndMove and MyndMove 2.0, as illustrated in the table below.
{8}------------------------------------------------
| Substantial Equivalence Discussion for Subject Devices (MyndMove and MyndMove 2.0) and Primary Predicate Device (MyndMove, K170564)<br>and Secondary Predicate Device (MyndMove 2.0, K212149) | | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Devices<br>(MyndMove and MyndMove 2.0) | Primary Predicate Device<br>(MyndMove) | Secondary Predicate Device<br>(MyndMove 2.0) | SE Comparison | |
| 510(k)<br>Number | K233006 | K170564 | K212149 | | |
| Indications<br>for Use | MyndMove and MyndMove 2.0 are<br>electrical stimulation devices indicated<br>for the following uses:<br>Functional electrical stimulation (FES)<br>Improvement of arm and hand function<br>and active range of motion in patients<br>with hemiplegia due to stroke or upper<br>limb paralysis due to C3-T1 spinal cord<br>injury.<br>NeuroMuscular Electrical Stimulation<br>(NMES) for general rehabilitation for:<br>• maintenance and/or increase of arm<br>and hand range of motion,<br>• prevention and/or retardation of<br>disuse atrophy,<br>• increase in local blood circulation,<br>• reduction in muscle spasm, and<br>• re-education of muscles.<br>MyndMove and MyndMove 2.0 therapies<br>can only be administered by<br>Occupational or Physical Therapy<br>professionals that have completed<br>MyndMove training by MyndTec on the<br>use of the MyndMove and MyndMove<br>2.0 Systems. | MyndMove is an electrical stimulation<br>device indicated for the following uses:<br>Functional electrical stimulation (FES)<br>Improvement of arm and hand function<br>and active range of motion in patients<br>with hemiplegia due to stroke or upper<br>limb paralysis due to C3-T1 spinal cord<br>injury.<br>NeuroMuscular Electrical Stimulation<br>(NMES)<br>• maintenance and/or increase of arm<br>and hand range of motion,<br>• prevention and/or retardation of<br>disuse atrophy,<br>• increase in local blood circulation,<br>• reduction in muscle spasm, and<br>• re-education of muscles.<br>MyndMove therapy can only be<br>administered by Occupational or Physical<br>Therapists that have completed<br>MyndMove training by MyndTec on the<br>use of the MyndMove System. | MyndMove is an electrical stimulation<br>device indicated for the following uses:<br>Functional electrical stimulation (FES)<br>Improvement of arm and hand<br>function and active range of motion in<br>patients with hemiplegia due to stroke<br>or upper limb paralysis due to C3-T1<br>spinal cord injury.<br>• NeuroMuscular Electrical<br>Stimulation (NMES)<br>• maintenance and/or increase of arm<br>and hand range of motion,<br>• prevention and/or retardation of<br>disuse atrophy,<br>• increase in local blood circulation,<br>• reduction in muscle spasm, and<br>• re-education of muscles.<br>MyndMove therapy can only be<br>administered by Occupational or<br>Physical Therapy professionals that<br>have completed MyndMove training by<br>MyndTec on the use of the MyndMove<br>System. | Similar | |
| Substantial Equivalence Discussion for Subject Devices (MyndMove and MyndMove 2.0) and Primary Predicate Device (MyndMove, K170564) and Secondary Predicate Device (MyndMove 2.0, K212149) | | | | | |
| | Subject Devices<br>(MyndMove and MyndMove 2.0) | Primary Predicate Device<br>(MyndMove) | Secondary Predicate Device<br>(MyndMove 2.0) | SE Comparison | |
| 510(k)<br>Number | K233006 | K170564 | K212149 | | |
| Anatomical<br>Sites<br>for<br>Stimulation<br>(Upper Limb) | The MyndMove System stimulates the<br>following Muscles:<br>Extensor Digitorum, Extensor Carpi<br>Radialis & Extensor Carpi Ulnaris<br>Thenar Eminence (Opponens Pollicis<br>Brevis, Flexor Pollicis Brevis & Abductor<br>Pollicis Brevis)<br>Flexor Digitorum Superficialis and Flexor<br>Digitorum Profundus<br>Biceps<br>Triceps<br>Posterior Deltoid, Middle Deltoid, and<br>Anterior Deltoid<br>Pectoralis Major<br>1st, 2nd, and 3rd Lumbricals<br>2nd Dorsal Interosseous<br>Serratus Anterior, Lower Trapezius,<br>Upper Trapezius (Scapula) | The MyndMove System stimulates the<br>following Muscles:<br>Extensor Digitorum, Extensor Carpi<br>Radialis & Extensor Carpi Ulnaris<br>Thenar Eminence (Opponens Pollicis<br>Brevis, Flexor Pollicis Brevis &<br>Abductor Pollicis Brevis)<br>Flexor Digitorum Superficialis and<br>Flexor Digitorum Profundus<br>Biceps<br>Triceps<br>Posterior Deltoid, Middle Deltoid, and<br>Anterior Deltoid<br>Pectoralis Major<br>1st, 2nd, and 3rd Lumbricals<br>2nd Dorsal Interosseous | The MyndMove System stimulates the<br>following Muscles:<br>Extensor Digitorum, Extensor Carpi<br>Radialis & Extensor Carpi Ulnaris<br>Thenar Eminence (Opponens Pollicis<br>Brevis, Flexor Pollicis Brevis &<br>Abductor Pollicis Brevis)<br>Flexor Digitorum Superficialis and<br>Flexor Digitorum Profundus<br>Biceps<br>Triceps<br>Posterior Deltoid, Middle Deltoid, and<br>Anterior Deltoid<br>Pectoralis Major<br>1st, 2nd, and 3rd Lumbricals<br>2nd Dorsal Interosseous<br>Serratus Anterior, Lower Trapezius,<br>Upper Trapezius (Scapula). | Identical | |
| Where Used | MyndMove is intended to be used for<br>therapy sessions in a clinical setting only<br>MyndMove 2.0 is intended to be used<br>for therapy sessions in a clinical setting<br>and for home use by a qualified user. | Used for therapy sessions in a clinical<br>setting only. | Used for therapy sessions in a clinical<br>setting and for home use by a qualified<br>user. | The subject<br>MyndMove is<br>identical to the<br>MyndMove<br>cleared in<br>K170564. The<br>subject<br>MyndMove 2.0 is<br>identical to the | |
| Substantial Equivalence Discussion for Subject Devices (MyndMove and MyndMove 2.0) and Primary Predicate Device (MyndMove, K170564)<br>and Secondary Predicate Device (MyndMove 2.0, K212149) | | | | | |
| | Subject Devices<br>(MyndMove and MyndMove 2.0) | Primary Predicate Device<br>(MyndMove) | Secondary Predicate Device<br>(MyndMove 2.0) | SE Comparison | |
| 510(k)<br>Number | K233006 | K170564 | K212149 | | |
| Regulated<br>current or<br>regulated<br>voltage? | Current Regulated | Current Regulated | Current Regulated | Identical | |
| Software/<br>firmware/<br>microprocess<br>or control?<br>(yes/no) | Yes | Yes | Yes | Identical | |
| Automatic<br>overload<br>trip?<br>(yes/no) | Yes | Yes | Yes | Identical | |
| Automatic<br>no- load trip?<br>(yes/no) | Yes | Yes | Yes | Identical | |
| Automatic<br>shut off?<br>(yes/no) | Yes | Yes | Yes | Identical | |
| User override<br>control?<br>(yes/no) | Yes | Yes | Yes | Identical | |
| Indicator<br>display: | Yes | Yes | Yes | Identical | |
| - On/off<br>status? | Yes | Yes | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.