K162718 · Shenzhen Xft Medical Limited · GZI · Nov 30, 2017 · Neurology
Device Facts
Record ID
K162718
Device Name
Foot Drop System (Model XFT-2001D)
Applicant
Shenzhen Xft Medical Limited
Product Code
GZI · Neurology
Decision Date
Nov 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
XFT-2001D Foot Drop System is intended to address the lack of ankle dorsiflexion in patients who have sustained damage to upper motor neurons or pathways to the spinal cord. During the swing phase of walking, the XFT-2001D electrically stimulates the appropriate muscles that cause ankle dorsiflexion and may thus improve the individual's gait. Medical benefits of Functional Electrical Stimulation (FES) may include prevention/retardation of disuse atrophy, increased local blood flow, muscle re-education, and maintained or increased joint range of motion.
Device Story
Battery-operated external functional electrical stimulator (FES) for foot drop; attaches to leg below knee near fibula head. Uses tilt sensor to detect gait cycle; stimulates common peroneal nerve to trigger tibialis anterior muscle contraction during swing phase; improves gait. Comprises patient kit (stimulator, cuff) and clinician kit (management software). Used in clinic or home; operated by patient or clinician. Provides muscle re-education, increased blood flow, and range of motion maintenance. Healthcare provider uses output to monitor gait improvement; patient benefits from restored dorsiflexion during walking.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes electrical safety and EMC testing per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10. Biocompatibility testing for cuff performed per ISO 10993-5 and ISO 10993-10. Software verification and validation conducted per FDA guidance for moderate level of concern.
Indicated for adult patients with lack of ankle dorsiflexion due to upper motor neuron or spinal cord pathway damage. Contraindicated for patients with demand-type cardiac pacemakers or defibrillators, history of seizure disorders, or presence of malignant tumors, thrombosis, or electrode placement over the carotid sinus, throat, or mouth.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 30, 2017
Shenzhen XFT Medical Limited Mrs. Jiang Xiaoying Manager RM 203, BLD1, 14 Jinhui Road, New District Shenzhen, Guangdong CN
Re: K162718
Trade/Device Name: Foot Drop System, Model 2001-D Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI Dated: October 27, 2017 Received: October 31, 2017
Dear Mrs. Xiaoying:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162718
Device Name Foot Drop System (Model XFT-2001D)
#### Indications for Use (Describe)
XFT-2001D Foot Drop System is intended to address the lack of ankle dorsiflexion in patients who have sustained damage to upper motor neurons or pathways to the spinal cord. During the swing phase of walking, the XFT-2001D electrically stimulates the appropriate muscles that cause ankle dorsiflexion and may thus improve the individual's gait. Medical benefits of Functional Stimulation (FES) may include prevention/retardation of disuse attophy, increased local blood flow, muscle re-education, and maintained or increased joint range of motion.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 100%;"> <input checked="true" type="checkbox"/> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 100%;"> <input type="checkbox"/> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | November 30, 2017 |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Shenzhen XFT Medical Limited<br>Room203, Building 1, Biomedicine Innovations Industrial Park,<br>#14 Jinhui Road, Pingshan New District, Shenzhen, China |
| Contact Person: | Mrs. Jiang Xiaoying |
Co xfthr3@xft.cn Tel: +86-(0) 755-29888818
Device: Trade Name: Foot Drop System
> Model: XFT-2001D series
External Neuromuscular Functional Stimulator Classification Names:
Product Code: GZI Regulation Number: 21 CFR 882.5810 Device Class: Class II Submission Purpose: New Device Predicate Device(s): K123972
{5}------------------------------------------------
- Nerve and Muscle Stimulator (Foot Drop System) is a battery-operated, Device
Description: electrical stimulator that can be used for functional electrical stimulation (FES).
> Nerve and Muscle Stimulator (Foot Drop System) is an external functional electrical stimulator. It is a small device that attaches to the leg just below the knee, near the head of the fibula. During a gait cycle, Nerve and Muscle Stimulator (Foot Drop System) stimulates the common peroneal nerve, which innervates the tibialis anterior and other muscles that cause dorsiflexion of the ankle. The Nerve and Muscle Stimulator (Foot Drop System) consists of the Patient Kit and the Clinician Kit. The Patient Kit comprises of all the components and accessories that the patient will take home and use. The Clinician System includes the Medical Records Management Software.
- Intended Use: XFT-2001D Foot Drop System is intended to address the lack of ankle dorsiflexion in patients who have sustained damage to upper motor neurons or pathways to the spinal cord. During the swing phase of walking, the XFT-2001D electrically stimulates the appropriate muscles that cause ankle dorsiflexion and may thus improve the individual's gait. Medical benefits of Functional Electrical Stimulation (FES) may include prevention/retardation of disuse atrophy, increased local blood flow, muscle re-education, and maintained or increased joint range of motion.
- Technology: Proposed device and predicate device use a small electrical signal to stimulate the nerves in the leg, causing the muscles contract and produce a movement that can help walking. The most common use of FES is as a treatment for foot drop where disruptions in the nerve pathways between the legs and brain which means the front of your foot cannot be lifted to the correct angle when walking.
The Foot Drop System employs the same fundamental scientific technology as its predicate devices.
{6}------------------------------------------------
Basis for Technological characteristics, features, specifications, materials and Substantial intended uses of the new device are substantially equivalent to the Equivalence: quoted predicated device. The parameters such as waveform, frequency, pulse width, output current and voltage, although not exactly identical to this particular predicate, do not raise any new questions of safety or effectiveness. Applicable performance and safety testing was performed to verify technological and functional characteristics including any design modification. The differences that exist between the devices are insignificant in terms of safety and effectiveness.
The proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the subject device is determined as safe and effectiveness.
Table 1 Comparison between XFT-2001D and the predicate device
{7}------------------------------------------------
| Specification | Predicate Device | Proposed device |
|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K number | K123972 | K162718 |
| Device name | WalkAide | XFT-2001D |
| Manufacturer | Innovative Neurotronics, Inc. | Shenzhen XFT Medical Limited |
| Intended Use/ Indications for<br>Use | The Innovative Neurotronics WalkAide<br>System is intended to address the lack of<br>ankle dorsiflexion in patients who have<br>sustained damage to upper motor<br>neurons or pathways to the spinal cord.<br>During the swing phase of gait, the<br>WalkAide electrically stimulates the<br>appropriate muscles that cause ankle<br>dorsiflexion and may thus improve the<br>patient's gait. Medical benefits of<br>Functional Electrical Stimulation (FES)<br>may include prevention/retardation of<br>disuse atrophy, increased local blood<br>flow, muscle re-education, and<br>maintained or increased joint range of<br>motion. | XFT-2001D Foot Drop System is<br>intended to address the lack of ankle<br>dorsiflexion in patients who have<br>sustained damage to upper motor<br>neurons or pathways to the spinal<br>cord. During the swing phase of<br>walking, the XFT-2001D electrically<br>stimulates the appropriate muscles<br>that cause ankle dorsiflexion and may<br>thus improve the individual's gait.<br>Medical benefits of Functional<br>Electrical Stimulation (FES) may include<br>prevention/retardation of disuse<br>atrophy, increased local blood flow,<br>muscle re-education, and maintained<br>or increased joint range of motion. |
| Contraindications | Do not use on persons with implanted<br>demand type cardiac pacemakers or<br>defibrillators. | Do not use on persons with implanted<br>demand type cardiac pacemakers or<br>defibrillators |
| | Do not place the electrodes in the<br>carotid sinus region (throat). | Do not place the electrodes in the<br>carotid Sinus region (throat). |
| | Laryngeal or pharyngeal spasms placed<br>across the throat or in the mouth. | Laryngeal or pharyngeal spasms may<br>occur when the electrodes are placed<br>across the throat or in the mouth. |
| | Do not place the electrodes over<br>malignant tumors. | Do not place the electrodes over<br>malignant tumors. |
| | Do not place the electrodes over areas in<br>which symptoms of existing thrombosis<br>are present. | Do not place the electrodes over areas<br>in which symptoms of existing<br>thrombosis are present. |
| | Do not use if person has a history of<br>seizure disorder. | Do not use if person has a history of<br>seizure disorder. |
| Patient Population | Adult | Adult |
| Power Source(s) | 1.5 AA Battery (not rechargeable) | DC3.7V, 400mAh, rechargeable lithium<br>battery (Stim Unit & Remote control) |
| Patient Leakage Current | -Normal condition: 0.5mA | -Normal condition: 0.5mA |
| Specification | Predicate Device | Proposed Device |
| Number of Output Modes | 2 modes: gait mode and training<br>mode | 2 modes: gait mode and training<br>mode |
| Number of Output Channels | 1 | 1 |
| Regulated Current or<br>Regulated Voltage | Voltage | Current |
| Software/Firmware/Microproc<br>essor | Microprocessor | Microprocessor |
| Auto Overload Trip? | No | No |
| Auto No-Load Trip? | No | Yes |
| Auto Shut Off? | No | Yes |
| Patient Override Control? | Yes | Yes |
| Indicator Display | | |
| On/Off Status? | Yes | Yes |
| Low Battery?<br>Voltage/Current Level? | Yes | Yes |
| Timer Range (minutes) | 30 minutes | 20 minutes |
| ON Time (seconds) | 1-5 sec in 1 sec steps | 1-5 sec in 0.5 sec steps |
| OFF Time (seconds) | 1-10 sec in 1 sec steps | 2-10 sec in 1 sec steps |
| Net Charge (µC per pulse) | 16.7-33Hz | 16.7-33Hz |
| | 25-300 microseconds | 100-300 μS |
| Frequency (Pulses per second) | Pules width adjustable in following<br>discrete steps:<br>25/50/100/150/200/250/300μS | Pules width adjustable in following<br>discrete<br>steps:100/150/200/250/300uS |
| Pulse width | 25-300 microseconds | 100-300 μS |
| | Pules width adjustable in following<br>discrete steps:<br>25/50/100/150/200/250/300μS | Pules width adjustable in following<br>discrete steps:<br>100/150/200/250/300uS |
| | 41.2µC @ 500 ohm | 27.0μC @ 500 ohm |
| Max Phase Charge(μC) | 28µC @ 1K ohm | 27.0µC @ 1K ohm |
| Max Current Density,<br>(mA/cm²) | 1.8mA/cm² (500 ohm) | 0.7mA/cm² (500 ohm) |
| Maximum Power Density,<br>(W/cm²) | 0.0128W/cm² (500 ohm) | 0.005W/cm² (500 ohm) |
| Specification | Predicate Device | Proposed Device |
| Max Output Voltage | unknown@ 500 ohm | 45V @ 500 ohm |
| | 121V@ 1 K ohm | 90V @ 1 K ohm |
| | unknown @ 2 K ohm | 156V @ 2 K ohm |
| | unknown @ 10 K ohm | 158V@ 10 K ohm |
| Max Output Current | 20.7mA rms@ 500 ohm | 14mA rms@ 500 ohm |
| | unknown @ 2 K ohm | 11mA rms@ 2 K ohm |
| | unknown @ 10 K ohm | 2.2mA rms@10 K ohm |
| Burst Mode | | |
| -Pulses per burst | Burst length and PPS dependent | Burst length and PPS dependent |
| -Bursts per second | Depends on ON and OFF times | Depends on ON and OFF times |
| -Burst duration<br>(seconds)<br>-Duty Cycle [Line (b) x Line (c)] | 1 to 5 sec in 1 sec steps Depends on<br>ON and OFF times | 1 to 5 sec in 1 sec steps Depends on<br>ON and OFF times |
| Stimulation Trigger Source | Tilt Sensor or Foot Sensor | Tilt Sensor |
| Anatomical Sites | Limb | Limb…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.