The sponsor used retrospective clinical data from patients undergoing activity-based therapy to demonstrate the safety and effectiveness of the RT300 device when operating at higher pulse widths (500-3000 usec).
Multi-arm clinical therapy evaluation; Multi-arm study with patients serving as their own control; Follow-up/Duration: Total of 434 (cycling), 920 (LE NMES), 172 (Trunk NMES), 65 (UE NMES), 324 (Pediatric) sessions
Patients undergoing activity-based therapy (adults and pediatric patients ages 2-13); Sample Size: Total of 15 (cycling), 33 (LE NMES), 13 (Trunk NMES), 14 (UE NMES), 8 (Pediatric)
The RT300 is intended for general rehabilitation for: 1. Relaxation of muscle spasms 2. Prevention or retardation of disuse atrophy 3. Increasing local blood circulation 4. Maintaining or increasing range of motion 5. Muscle re-education The RT300 pediatric version is intended for population ages 4 to 12 years. The RT300 is intended for use with a surface electrical stimulation garment for population ages 12 and above. The RT300 is for prescription use only.
Device Story
RT300 FES Cycle Ergometer is a rehabilitation system for patients with impaired extremity movement. It combines a motorized leg cycle ergometer, optional arm crank, and FES controller. Input data includes ergometer angle information and pulse oximetry (SpO2/pulse). The system uses Sage software on a Windows tablet to coordinate electrical stimulation to muscle groups based on cycle angle, facilitating smooth active or passive cycling. Used in clinical settings by patients under professional supervision. Output includes therapy session logs and real-time performance data. Healthcare providers use this data to monitor progress and adjust therapy settings. Benefits include improved muscle function, circulation, and range of motion. The device supports remote database storage for therapy settings and logs.
Clinical Evidence
Multi-arm clinical study (n=15-33 per arm) evaluated high pulse width (500-3000µsec) stimulation. Primary endpoints: safety (adverse reactions, strain, fracture) and effectiveness (NRS, Berg Balance, MVC torque, Overground Standing, SATCo). Results: No serious adverse reactions reported. Significant improvements observed: NRS scores increased 8-88%, Berg scores 47-50%, MVC torque increased up to 3796%, and standing/sitting times improved significantly. Study confirms safety and effectiveness profile comparable to predicate.
Technological Characteristics
System includes motorized cycle ergometer, tablet PC controller, and 6-channel FES stimulators (up to 12 channels total). Stimulation: 0-140mA, pulse width up to 3,000µsec. Connectivity: Wireless stimulation channels, pulse oximeter interface, remote database interface. Software: Windows-based Sage software. Compliance: IEC 60601-1, IEC 60601-2-10, IEC 60601-1-11, IEC 60601-1-2.
Indications for Use
Indicated for general rehabilitation (spasm relaxation, disuse atrophy prevention, circulation, range of motion, muscle re-education) in patients with impaired upper/lower extremity movement. Pediatric version for ages 4-12; garment-based stimulation for ages 12+.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with three heads, suggesting a focus on people and services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 11, 2017
Restorative Therapies, Inc. Andrew Barriskill CEO 1434 Fleet St. Baltimore, Maryland 21231
Re: K162470
Trade/Device Name: RT300 FES Cycle Ergometer Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI Dated: June 7, 2017 Received: June 8, 2017
Dear Mr. Barriskill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162470
Device Name RT300 FES Cycle Ergometer
Indications for Use (Describe)
The RT300 is intended for general rehabilitation for:
- 1. Relaxation of muscle spasms
2. Prevention or retardation of disuse atrophy
3. Increasing local blood circulation
4. Maintaining or increasing range of motion
5. Muscle re-education
The RT300 pediatric version is intended for population ages 4 to 12 years.
The RT300 is intended for use with a surface electrical stimulation garment for population ages 12 and above. The RT300 is for prescription use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary
## RT300 510(k) Summary
#### (1) Submitter's name, address, telephone number, a contact person, and the date the summary was prepared:
Restorative Therapies Inc 1434 Fleet St Baltimore, MD 21231
Contact Person: Andrew Barriskill Phone: 800 609-9166
Prepared on July 16, 2016.
#### (2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name:
Proprietary name: RT300 FES Cycle Ergometer Common name: External Functional Neuromuscular Stimulator Classification name:External Functional Neuromuscular Stimulator (21 CFR 882.5810) Regulatory Class: II Product Code: GZI
#### (3) Identification of the legally marketed device to which the submitter claims equivalence:
RESTORATIVE THERAPIES, INC. product: "RT300", K090750 RESTORATIVE THERAPIES, INC. product: "Xcite Clinical Station", K160614
#### (4) A description of the device that is the subject of the premarket notification submission.
The RT300 is a Functional Electrical Stimulation (FES) cycle ergometer which is composed of:
1. A motorized leg cycle ergometer (RTI part number SA100047 for adults and SA100044 for children)
2. An optional motorized arm crank (RTI part number PP102663)
3. An FES controller with up to two 6 channel stimulators (RTI part number SA216172)
Image /page/3/Picture/15 description: The image shows the logo for RTI International. The logo consists of the letters "RTI" in a sans-serif font, stacked vertically. To the left of the letters is a stylized graphic of curved lines that converge to a point. The logo is in a light blue color.
| Wide pulse width 510(k) submission | | | | Project | R&D | |
|------------------------------------|---|----|----|---------|-----|----------|
| Page: | 1 | of | 10 | Version | 2 | EN211366 |
{5}------------------------------------------------
4. Up to 5 additional wireless single channel stimulators (RTI part number SA216128)
5. A leg and optional arm stimulation cable which connects the stimulator to cutaneous electrodes
6. Cutaneous electrodes (up to 24 electrodes for up to 12 stimulation channels)
7. An interface to a remote database for the storage and retrieval of therapy settings and the storage of therapy session logs
8. An interface to a pulse oximeter for the display and recording of pulse and SpO2 levels and provision of alarming based on the data
9. A garment incorporating electrodes for lower extremity cycling in population ages 12 and above (RTI part number FA105486)
This system allows a person with impaired upper or lower extremity movement to undertake cycle erqometry both actively (utilizing FES evoked upper or lower extremity muscle contractions) and passively (utilizing power developed by the ergometer's motor). The system combines a Windows based tablet, cycle ergometer, and stimulators in a single unit. The system is controlled by the Sage software that runs on the tablet. The Sage software delivers stimulation to the designated muscle groups in a coordinated fashion based on angle information from the ergometer. In doing so the system facilitates a smooth cycling motion throughout the therapy session.
#### (5) Indications for Use:
The RT300 is intended for general rehabilitation for:
- 1. Relaxation of muscle spasms
- 2. Prevention or retardation of disuse atrophy
- 3. Increasing local blood circulation
- 4. Maintaining or increasing range of motion
- 5. Muscle re-education
The RT300 pediatric version is intended for population ages 4 to 12 years. The RT300 is intended for use with a surface electrical stimulation garment for population ages 12 and above.
The RT300 is for prescription use only.
The Indications for Use described in the numbered items above are not identical to the predicate device, however, the differences do not alter the intended
| | Wide pulse width 510(k) submission | | | | | Project | R&D |
|-----|------------------------------------|---|----|----|---------|---------|----------|
| RTI | Page: | 2 | of | 10 | Version | 2 | EN211366 |
{6}------------------------------------------------
therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for general rehabilitation.
## (6) Technological Characteristics
The function of the RT300 is the same as the predicate devices however there are certain technological similarities and differences as described below:
| Technology | RT300 | RT300 predicate:<br>K090750 | Xcite predicate<br>K160614 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Ergometer | Cycle ergometer | Cycle ergometer | No ergometer |
| Power<br>source<br>(energy<br>used) | Mains power and<br>rechargeable battery for<br>RT50 stimulators | Mains power and<br>rechargeable battery for<br>RT50 stimulators | Mains power and<br>rechargeable<br>battery for RT60<br>and RT50<br>stimulators |
| Controller | Based on Tablet PC<br>running custom<br>software. | Based on Pocket PC<br>running custom<br>software. | Based on tablet<br>PC running<br>custom software. |
| Stimulator<br>(energy<br>delivered) | Built in AC mains<br>powered 0-140mA 6<br>channel charge<br>balanced stimulator.<br>Optionally, the RT300<br>can accommodate two<br>6-channel stimulators for<br>a total of 12 channels. | Built in AC mains<br>powered 0-140mA 6<br>channel charge<br>balanced stimulator. | Up to 2 DC<br>powered 0-140mA<br>6 channel charge<br>balanced<br>stimulators<br>("RT60"s). |
| Stimulation<br>pulse width | Up to 3,000usec | Up to 500usec | Up to 500usec |
| Additional<br>stimulation<br>channels | Up to 5 additional<br>wireless battery<br>powered stimulation<br>channels delivering 0-<br>140mA charge balanced<br>stimulation. | Up to 5 additional<br>wireless battery<br>powered stimulation<br>channels delivering 0-<br>140mA charge<br>balanced stimulation. | Up to 4 additional<br>wireless battery<br>powered<br>stimulation<br>channels<br>delivering 0-<br>140mA charge<br>balanced<br>stimulation. |
| Stand<br>alone<br>stimulation<br>mode | Wireless battery<br>powered stimulation<br>channels may be used<br>in standalone mode<br>without the cycle<br>ergometer. | Wireless battery<br>powered stimulation<br>channels may be used<br>in standalone mode<br>without the cycle<br>ergometer. | Always used in<br>standalone<br>stimulation mode. |
| Stimwear | Stimwear garment | Stimwear garment | No stimwear |
Image /page/5/Picture/6 description: The image shows the logo for RTI International. The logo consists of the letters "RTI" in a teal sans-serif font. To the left of the letters are two curved teal lines that form a vertical oval shape. The logo is simple and modern.
| Wide pulse width 510(k) submission | | | | Project | R&D | |
|------------------------------------|---|----|----|---------|-----|----------|
| Page: | 3 | of | 10 | Version | 2 | EN211366 |
{7}------------------------------------------------
| Technology | RT300 | RT300 predicate:<br>K090750 | Xcite predicate<br>K160614 |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| garment | incorporating electrodes<br>available for lower<br>extremity cycling, ages<br>12 and above. | incorporating electrodes<br>available for lower<br>extremity cycling, ages<br>12 and above. | garment. |
| Muscles<br>available<br>for<br>stimulation | Quadriceps, hamstrings,<br>gluteals, gastroc,<br>anterior tibialis,<br>shoulder, biceps,<br>triceps, anterior,<br>posterior and middle<br>deltoid, wrist, grasp,<br>abdominals, erector<br>spinae. | Quadriceps,<br>hamstrings, gluteals,<br>gastroc, anterior tibialis,<br>shoulder, biceps,<br>triceps, anterior,<br>posterior and middle<br>deltoid, wrist, grasp,<br>abdominals, erector<br>spinae. | Quadriceps,<br>hamstrings,<br>gluteals, gastroc,<br>anterior tibialis,<br>shoulder, biceps,<br>triceps, anterior,<br>posterior and<br>middle deltoid,<br>wrist extensors<br>and flexors, grasp,<br>abdominals,<br>erector spinae. |
| Flywheel | Uses leg / arm crank<br>motor to create flywheel<br>effect with reduced<br>weight and space. | Uses leg / arm crank<br>motor to create flywheel<br>effect with reduced<br>weight and space. | N/A |
| Seating | Allows user to remain in<br>their own seating, e.g<br>wheelchair eliminating<br>the need for transfer. | Allows user to remain in<br>their own seating, e.g<br>wheelchair eliminating<br>the need for transfer. | N/A |
| Passive<br>cycling | Utilizes motor to provide<br>assistance during<br>passive cycling. | Utilizes motor to<br>provide assistance<br>during passive cycling. | N/A |
| Database<br>interface | Utilizes database<br>interface for storage and<br>retrieval of patient<br>therapy settings and<br>storage of session logs. | Utilizes database<br>interface for storage<br>and retrieval of patient<br>therapy settings and<br>storage of session logs. | Utilizes database<br>interface for<br>storage and<br>retrieval of patient<br>therapy settings<br>and storage of<br>session logs. |
| Motorized<br>arm crank | Allows active / passive<br>arm cycling with FES | Allows active / passive<br>arm cycling with FES | N/A |
| Pulse<br>oximeter<br>interface | Utilize pulse and SpO2<br>data for display,<br>recording and alarming | Utilize pulse and SpO2<br>data for display,<br>recording and alarming | Utilize pulse and<br>SpO2 data for<br>display, recording<br>and alarming |
| Bilateral or<br>Unilateral<br>stimulation | Uses bilateral or<br>unilateral stimulation. | Uses bilateral or<br>unilateral stimulation. | Uses bilateral or<br>unilateral<br>stimulation. |
| Image: RTI logo | Wide pulse width 510(k) submission | | | Project | R&D | | |
|-----------------|------------------------------------|---|-------|---------|-----|--|----------|
| | Page: | 4 | of 10 | Version | 2 | | EN211366 |
{8}------------------------------------------------
| Technology | RT300 | RT300 predicate:<br>K090750 | Xcite predicate<br>K160614 |
|-------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications | 1.Relaxation of muscle<br>spasms<br>2.Prevention or<br>retardation of disuse<br>atrophy<br>3.Increasing local blood<br>circulation; and<br>4.Maintaining or<br>increasing range of<br>motion<br>5. Muscle re-education<br><br>The RT300 pediatric<br>version is intended for<br>population ages 4 to 12<br>years.<br><br>The RT300 is intended<br>for use with a surface<br>electrical stimulation<br>garment for population<br>ages 12 and above.<br><br>The RT300 is for<br>prescription use only. | 1.Relaxation of muscle<br>spasms<br>2.Prevention or<br>retardation of disuse<br>atrophy<br>3.Increasing local blood<br>circulation; and<br>4.Maintaining or<br>increasing range of<br>motion<br><br>The RT300 pediatric<br>version is intended for<br>population ages 4 to 12<br>years.<br><br>The RT300 is intended<br>for use with a surface<br>electrical stimulation<br>garment for population<br>ages 12 and above.<br><br>The RT300 is for<br>prescription use only. | 1.Relaxation of<br>muscle spasms<br>2.Prevention or<br>retardation of<br>disuse atrophy<br>3.Increasing local<br>blood circulation<br>4.Maintaining or<br>increasing range<br>of motion<br>5.Muscle re-education |
Table 1 Device technology comparison
## (b) Performance data
Non clinical testing to determine equivalence has been primarily composed of the following tests:
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the RT300, consisting of the tablet, leg ergometer, arm ergometer, 2 RT60s and stimulation cable. The system complies with IEC 60601-1, IEC 60601-2-10, and IEC 60601-1-11 standards for safety and IEC 60601-1-2 for EMC.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices: The software for this device was considered "moderate" level
| Wide pulse width 510(k) submission | | | | | Project | R&D | |
|------------------------------------|-------|---|----|----|---------|-----|----------|
| RTI | Page: | 5 | of | 10 | Version | 2 | EN211366 |
{9}------------------------------------------------
of concern, since a failure of latent flaw in the software could directly result in minor injury to the patient or operator.
| Test or procedure | Description |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Review of user documentation for<br>predicate device | Ensure that equivalent functionality is<br>specified and implemented in the new<br>device. |
| Review of 510(k) submission for<br>predicate device | Confirm technical specifications for<br>completion of predicate details in<br>comparison tables. |
| Output characteristic measurement of<br>new device | Confirm technical specifications for<br>completion of new device details in<br>comparison tables. |
| Conduct of system testing | Conduct system testing to verify<br>performance to specification. |
#### Clinical Testing
Clinical testing consisted of a multi-arm study with patients serving as their own control. The protocol's objective was to evaluate the safety and effectiveness of external functional neuromuscular stimulators operating at high pulse widths (500-3000 usec) stimulation. Subjects used multi-muscle, high frequency, and pulse width with the RT300 stimulator to generate neuromuscular activity. Subjects were evaluated for 4 interventions; Cycling with Lower Extremity NMES (NeuroMuscular Electrical Stimulation), Lower Extremity NMES, Trunk NMES, and Upper Extremity NMES. Additionally pediatric subjects were assessed
#### Cycling with Lower Extremity NMES
#### Intervention:
Fifteen patients (15) undertook activity based therapy using NMES during cvcling on an RT300 cycle.
#### Primary safety endpoint:
The primary safety endpoint consisted of treatment-related adverse reactions, normal erythema, blancheable erythema, muscle strain, joint strain, or fracture.
#### Effectiveness
The primary endpoint for the Cycling intervention focused on the risk assessment of cycling with high pulse width (500-3000 usec, with 8 patients receiving the maximum of 3000us) stimulation. Fifteen (15) patients completed 434 cycling sessions - 3 had normal erythema and none had an adverse reaction, blancheable erythema, muscle strain, joint strain nor fracture. Therefore this intervention met its primary endpoint with no adverse reactions or other injuries.
#### Lower Extremity NMES
Intervention:
R
| | Wide pulse width 510(k) submission | | Project | R&D | | |
|-------|------------------------------------|----|---------|---------|---|----------|
| Page: | 6 | of | 10 | Version | 2 | EN211366 |
{10}------------------------------------------------
Thirty-three (33) patients undertook activity based therapy using NMES during the following motor tasks: lying supine, sit to stand, standing.
#### Primary effectiveness endpoint:
Clinically significant improvements in functional outcome measures. These measures included evaluation using the Adult Neuromuscular Recovery Scale (NRS), Berg Balance Scale, Biodex Dynamometer Torque test for Force Production from Maximum Voluntary Contractions (MVC), Overground Standing Test.
#### Primary safety endpoint:
The primary safety endpoint consisted of treatment-related adverse reactions, normal erythema, blancheable erythema, muscle strain, joint strain, or fracture.
#### Effectiveness
The primary endpoint for the Lower Extremity intervention focused on the changes to the outcome measures after intervention with high pulse width (500-3000 usec, with 10 patients receiving the maximum of 3000us). Twenty-three (23) patients increased their NRS by 16% (average pre assessment score of 5.43 to average post of 6.45), their lower extremity NRS by 11% (3.27 to 3.71); 15 patients increased their Berg scores by 47% (average pre assessment score of 6.8 to average post of 10); MVC torque increased for 8 lower extremity muscle groups (Right and Left knee extensors, Right and Left plantarflexors and Right and Left dorsiflexors) in an average of 9 patients by a minimum of 152% (average torque in Nm at pre assessment of 3.5 to a post assessment of 7.13) increase in right plantarflexor torque output to a maximum 3796% (average torque in Nm at pre assessment of 0.1 to a post of 21.87) for left plantarflexor output, and 8 patients increased their Overground Standing Test times by 7108% (average pre time standing of 0 to average post time of 300 seconds). Therefore this intervention met its primary endpoint of an increase in all outcome measures.
#### Safetv
After 920 sessions twenty-seven (27) patients experienced no adverse reactions, 8 patients experienced normal erythema. 1 patient experienced blancheable erythema and none experienced muscle strain, joint strain nor fracture.
#### Trunk NMES
Intervention:
Thirteen (13) patients undertook activity based therapy using NMES during the following motor tasks: sit, sit up and reverse sit up.
#### Primary effectiveness endpoint:
Clinically significant improvements in functional outcome measures. These measures included evaluation using the Adult Neuromuscular Recovery Scale
Image /page/9/Picture/14 description: The image shows the logo for RTI International. The logo consists of the letters "RTI" in a teal color, with a stylized teal line to the left of the letters. The line is curved, resembling a stylized "K" shape.
| | Wide pulse width 510(k) submission | | Project | R&D |
|-------|------------------------------------|---------|---------|----------|
| Page: | 7 of 10 | Version | 2 | EN211366 |
{11}------------------------------------------------
(NRS), Berg Balance Scale, Modified Functional Reach, Independent Sitting, and Trunk Scale.
#### Primary safety endpoint:
The primary safety endpoint consisted of treatment-related adverse reactions, normal erythema, blancheable erythema, muscle strain, joint strain, or fracture.
#### Effectiveness
The primary endpoint for the Lower Extremity intervention focused on the changes to the outcome measures after intervention with high pulse width (500-3000 usec with 2 patients receiving the maximum of 3000us) stimulation. Eleven (11) patients increased their NRS by 18% (average pre assessment score of 5.26 to average post of 5.81), 5 patients increased their Berg scores by 50% (average pre assessment score of 9 to average post of 14); Modified reach scale increased 9 patients by 196% (average reach in inches at pre assessment of 6.45 to a post assessment of 18): Independent Sitting times increased in 3 patients 26583% (pre assessment time of 0 seconds to a post of 600 seconds); Trunk scale increased in 2 patients by 180% (pre assessment score of 0 to a post of 4). Therefore this intervention met its primary endpoint of an increase in all outcome measures.
#### Safety
After 172 sessions ten (10) patients experienced no adverse reactions, 3 patients experienced normal erythema. 1 patient experienced blancheable erythema and none experienced muscle strain, joint strain nor fracture.
#### Upper Extremity NMES
#### Intervention:
Fourteen (14) patients undertook activity based therapy using NMES during different motor tasks including reaching, grasping, pressing overhead, manipulating materials, and opening a door.
#### Primary effectiveness endpoint:
Clinically significant improvements in functional outcome measures. These measures included evaluation using the Adult Neuromuscular Recovery Scale (NRS), Modified Functional Reach, Bilateral Upper Extremity Basic Strength, and Modified Ashworth Upper Extremity scale.
#### Primary safety endpoint:
The primary safety endpoint consisted of treatment-related adverse reactions, normal erythema, blancheable erythema, muscle strain, joint strain, or fracture.
#### Effectiveness
The primary endpoint for the Upper Extremity intervention focused on the outcome measures after intervention with high pulse width (500-3000 usec with 1
Image /page/10/Picture/16 description: The image shows the logo for RTI International. The logo consists of the letters "RTI" in a sans-serif font, stacked vertically. To the left of the letters is a curved line that resembles a parenthesis. The color of the logo is a light blue.
| Wide pulse width 510(k) submission | | | | Project | R&D | |
|------------------------------------|---|----|----|---------|-----|----------|
| Page: | 8 | of | 10 | Version | 2 | EN211366 |
{12}------------------------------------------------
patient receiving the maximum of 3000us) stimulation. Eleven (11) patients increased their NRS by 8% (average pre assessment score of 3.91 to average post of 4.08), Modified reach scale increased in 4 patients by 122% (average reach in inches at pre assessment of 1.5 to a post assessment of 5.8); Bilateral Upper Extremity Basic Strength increased in 3 patients by 8% (pre assessment scale 2 to 2.1 post assessment) and Modified Ashworth Upper Extremity scale increased in 3 patients by 3%. Therefore this intervention met its primary endpoint of an increase in all outcome measures.
#### Safety
After 65 sessions seven (7) patients experienced no adverse reactions, 8 patients experienced normal erythema, and none experienced blancheable erythema, muscle strain, joint strain nor fracture.
#### Pediatric NMES
#### Intervention:
8 (eight) patients between the ages of 2 and 13 yrs old undertook activity based therapy for the upper, trunk and lower extremities using NMES during the different motor tasks (eg. reaching, grasping, standing).
#### Primary effectiveness endpoint:
Clinically significant improvements in functional outcome measures. These measures included evaluation using the Pediatric Neuromuscular Recovery Scale (NRS) and Segmental Assessment of Trunk Control (SATCo),
#### Primary safety endpoint:
The primary safety endpoint consisted of treatment-related adverse reactions, normal erythema, blancheable erythema, muscle strain, joint strain, or fracture.
## Effectiveness
The primary endpoint for the Lower Extremity intervention focused on the outcome measures after intervention with high pulse width (500-1000 usec with all 8 patients receiving the maximum of 1000us) stimulation. Four (4) patients increased their NRS by 88% (average pre assessment score of 2.38 to average post of 4.13) and 3 patients increased their SATCo by 35% (pre assessment score of 8 to a post of 10.7)
## Safetv
After 324 sessions 8 patients experienced no adverse reactions, 6 patients experienced normal erythema, and none experienced blancheable erythema, muscle strain, joint strain nor fracture.
## Summary
Image /page/11/Picture/15 description: The image shows the logo for RTI International. The logo consists of the letters "RTI" in a teal sans-serif font. To the left of the letters are two curved teal lines that form a stylized abstract shape. The logo is simple and modern in design.
| Wide pulse width 510(k) submission | | | Project | R&D | |
|------------------------------------|---|-------|---------|-----|----------|
| Page: | 9 | of 10 | Version | 2 | EN211366 |
{13}------------------------------------------------
Based on the clinical performance as documented in the clinical testing, the RT300 was found to have a safety and effectiveness profile that is similar to the predicate device
#### RTI concludes that:
The RT300 has the same intended use as the RT300 predicate device.
The RT300 with available wider stimulation pulses utilizes the same stimulators and control system as the predicate device including controller, stimulators, motor controller and drive assembly. The different technological characteristics do not raise new questions of safety and effectiveness.
In conclusion, RTI's clinical testing has demonstrated that the RT300 is as safe and effective as the predicate devices.
Image /page/12/Picture/7 description: The image shows the logo for RTI International. The logo features the letters "RTI" in a sans-serif font, with the "R" and "I" in a darker shade of blue than the "T". To the left of the letters, there are two curved lines that resemble a stylized "X". The overall design is simple and modern.
| Wide pulse width 510(k) submission | | | | | Project | R&D |
|------------------------------------|----|----|----|---------|---------|----------|
| Page: | 10 | of | 10 | Version | 2 | EN211366 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.