← Product Code [GZI](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZI) · K141812

# MYGAIT STIMULATION SYSTEM (K141812)

_Otto Bock Healthcare Products GmbH · GZI · Apr 2, 2015 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZI/K141812

## Device Facts

- **Applicant:** Otto Bock Healthcare Products GmbH
- **Product Code:** [GZI](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZI.md)
- **Decision Date:** Apr 2, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5810
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The stimulation system is intended to provide ankle dorsiflexion or extension in individuals with foot drop and thigh muscle weakness following an upper motor neuron injury or disease. During gait, the stimulation system sends electric stimuli to muscles in the affected leg, initiating dorsiflexion of the foot and knee extension or flexion and may thus improve the individual's gait. The stimulation system may also prevent or retard atrophy caused by inactivity, facilitate muscle reeducation, maintain or improve the range of motion in the joints and promote local blood circulation.

## Device Story

MyGait Stimulation System is a battery-operated, two-channel external functional neuromuscular stimulator for gait improvement. Input: wireless heel switch (worn in sock) registers heel lift/ground contact to detect swing/stance phases. Processing: stimulator (worn in cuff) receives wireless signals and delivers electrical stimulation to common peroneal nerve (dorsiflexion) and/or thigh muscles (knee flexion/extension). Output: controlled muscle contraction. Used in everyday environments by patients; settings adjusted by clinicians via MyGait Adjustment Tool and MyGait Soft PC software. Benefits: improved gait, muscle reeducation, atrophy prevention, increased local blood circulation, maintained joint range of motion.

## Clinical Evidence

Bench testing only. Performance testing included electrical safety (IEC 60601-1, 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), environmental testing (vibration, shock, temperature, humidity), and biocompatibility (ISO 10993-1, 5, 10).

## Technological Characteristics

Two-channel neuromuscular stimulator; ABS housing; 3.7V 1100mAh Li-ion battery; wireless heel switch; biphasic symmetrical/asymmetrical waveforms; constant current output; microprocessor-controlled; connectivity via wireless remote and PC software; IP-rated enclosure; medical class II power adapter for charging.

## Regulatory Identification

An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.

## Predicate Devices

- Walkaide ([K123972](/device/K123972.md))
- Empi 300pv ([K021100](/device/K021100.md))
- NESS L300 Plus System ([K103343](/device/K103343.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are arranged in a way that suggests unity and forward movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 2, 2015

Otto Bock Healthcare Products GmbH Rene Urtz Regulatory Affairs Specialist Brehmstrabe 16 1110 Vienna AUSTRIA

Re: K141812

Trade/Device Name: MyGait Stimulation System Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI, IPF Dated: February 23, 2015 Received: February 27, 2015

Dear Mr. Urtz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

## Felipe Aquel -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K141812

Device Name MyGait Stimulation System

#### Indications for Use (Describe)

The stimulation system is intended to provide ankle dorsiflexion or extension in individuals with foot drop and thigh muscle weakness following an upper motor neuron injury or disease.

During gait, the stimulation system sends electric stimuli to muscles in the affected leg, initiating dorsiflexion of the foot and knee extension or flexion and may thus improve the individual's gait. The stimulation system may also prevent or retard atrophy caused by inactivity, facilitate muscle reeducation, maintain or improve the range of motion in the joints and promote local blood circulation.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "ottobock." in a sans-serif font. The word is written in a dark blue color. The letters are closely spaced together, and the period at the end is small.

# 510(K) SUMMARY

## 1 Submitter Information

| Manufacturer:               | Otto Bock Healthcare Products GmbH                                   |
|-----------------------------|----------------------------------------------------------------------|
| Establishment Registration: | 9615892                                                              |
| Adress:                     | Brehmstraße 16<br>1110 Vienna<br>Austria                             |
| Contact Person:             | Rene Urtz<br>Regulatory Affairs Specialist<br>rene.urtz@ottobock.com |
| Telephone:                  | +43 1 523 37 86 707                                                  |
| Fax Number:                 | +43 1 523 22 64                                                      |
| Date Prepared:              | March 31, 2015                                                       |

## 2 Device Information

| Trade Name:                  | MyGait® Stimulation System                           |
|------------------------------|------------------------------------------------------|
| Common or Usual Name:        | Stimulator, Neuromuscular, External                  |
| Classification Name:         | External functional neuromuscular stimulator         |
|                              | (21 CFR 882.5810)                                    |
| Classification Product Code: | GZI (stimulator, neuromuscular, external functional) |
| Subsequent Product Code:     | IPF (stimulator, muscle, powered)                    |

{4}------------------------------------------------

# 3 Identification of Legally Marketed Predicate Devices

| Name:         | Walkaide                      |
|---------------|-------------------------------|
| Manufacturer: | Innovative Neurotronics, Inc. |
| 510k Number:  | K123972                       |
| Date Cleared: | April 4th, 2013               |

| Name:         | Empi 300pv      |
|---------------|-----------------|
| Manufacturer: | Empi, Inc       |
| 510k Number:  | K021100         |
| Date Cleared: | June 18th, 2002 |

| Name:         | NESS L300 Plus System*      |
|---------------|-----------------------------|
| Manufacturer: | Bioness Neuromodulation Ltd |
| 510k Number:  | K103343                     |
| Date Cleared: | April 29th, 2011            |

* Primary predicate

{5}------------------------------------------------

## 4 Indications for Use

The stimulation system is intended to provide ankle dorsiflexion or extension in individuals with foot drop and thigh muscle weakness following an upper motor neuron injury or disease. During gait, the stimulation system sends electric stimuli to muscles in the affected leq, initiating dorsiflexion of the foot and knee extension or flexion and may thus improve the individual's gait. The stimulation system may also prevent or retard atrophy caused by inactivity, facilitate muscle reeducation, maintain or improve the range of motion in the joints and promote local blood circulation.

## 5 Device Description

The MyGait® Stimulation System is a battery-operated two-channel stimulator and was developed exclusively for everyday use during walking in an everyday environment.

The patient wears a wireless heel switch in a special sock. If the user is not able to put sufficient weight on the affected leg, he or she can wear the heel switch on the other foot. The heel switch registers heel lift and ground contact and thus the swing and stance phase. It transmits the information to the stimulator wirelessly.

The stimulator is located in a cuff that holds the electrodes. It is easy to put on, even with only one hand. The correct placement of the electrodes is achieved by the cuff. The stimulator delivers the electrical stimulation to the common peroneal nerve. The nerve stimulates for controlled dorsiflexion of the foot during the swing phase.

The second channel can be used independently of the first channel to provide stimulation to the knee flexors or knee extensors in patients with thigh muscle weakness.

Using the wireless remote control, the patient can control and adjust the stimulator settings. When the patient sits for an extended period, the sleep mode helps save energy.

The adjustment tool enables qualified personnel to adjust the stimulation parameters to the patient's individual needs. Stimulation parameters are used to affect movements in the foot. Examples of stimulation parameters are current, pulse width and shape, frequency, stimulation timing,

The "MyGait Soft" PC software is used to manage and analyze gait analysis data and stimulation parameters. The data is loaded from the stimulator to the adjustment tool. The stimulation parameters can be restored to the stimulator from the PC.

{6}------------------------------------------------

## 6 Substantial Equivalence Discussion

### 6.1 Indications for Use

|                                         | New Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number                             | K141812                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | K123972                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K021100                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K103343                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device<br>Name, Model                   | MyGait                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | WalkAide                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Empi 300pv                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | NESS L300 Plus System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Manufacturer                            | Otto Bock Healthcare<br>Products GmbH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Innovative Neurotronics, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Empi, Inc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Bioness Neuromodulation<br>Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications<br>for Use                  | The stimulation system is<br>intended to provide ankle<br>dorsiflexion and knee flexion or<br>extension in individuals with foot<br>drop and thigh muscle<br>weakness following an upper<br>motor neuron injury or disease.<br>During gait, the stimulation<br>system sends electric stimuli to<br>muscles in the affected leg,<br>initiating dorsiflexion of the foot<br>and knee extension or flexion<br>and may thus improve the<br>individual's gait. The stimulation<br>system may also prevent or<br>retard atrophy caused by<br>inactivity, facilitate muscle<br>reeducation, maintain or<br>improve the range of motion in<br>the joints and promote local<br>blood circulation. | Summary Statement<br>The Innovative Neurotronics<br>WalkAide External Functional<br>Neuromuscular Stimulator<br>(WalkAide System) is intended<br>to address the lack of ankle<br>dorsiflexion in patients who have<br>sustained damage to upper<br>motor neurons or pathways to<br>the spinal cord. During the<br>swing phase of gait, the<br>WalkAide System electrically<br>stimulates the appropriate<br>muscles that cause ankle<br>dorsiflexion and may thus<br>improve the patient's gait.<br>Medical benefits of Functional<br>Electrical Stimulation (FES) may<br>include prevention, retardation<br>of disuse atrophy, increased<br>local blood flow, muscle re-<br>education, and maintained or<br>increased joint range of motion. | Summary Statement<br>The 300 PV is a multifunction<br>electrotherapy device with<br>various treatment modes that<br>allow for neuromuscular<br>electrical stimulation (NMES),<br>transcutaneous electrical nerve<br>stimulation (TENS),<br>interferential current stimulation<br>(IFS) and functional electrical<br>stimulation (FES).<br>As a NMES device, the 300 PV<br>is indicated for the following<br>conditions:<br>Re-educating muscles<br>Relaxation of muscle spasm<br>Increasing local blood circulation<br>Retarding or preventing disuse<br>atrophy<br>Maintaining or increasing range<br>of motion. Prevention of venous<br>thrombosis of the calf muscles<br>immediately after surgery.<br>As a TENS device, the 300 PV<br>is indicated for the following<br>conditions:<br>Symptomatic relief and<br>management of chronic,<br>intractable pain.<br>Adjunctive treatment for post-<br>surgical and post-trauma acute<br>pain. As an IFS device, the 300<br>PV is indicated for the following<br>conditions:<br>Symptomatic relief of acute pain<br>Symptomatic relief and<br>management of chronic pain.<br>As a FES device, the 300 PV is<br>indicated for the following<br>condition:<br>Stimulation of the muscles in<br>the leg and ankle of partially<br>paralyzed<br>patients to provide flexion of the<br>foot and thus improve the<br>patient's gait. | Summary Statement<br>The NESS L300 Plus System is<br>intended to provide ankle<br>dorsiflexion and knee flexion or<br>extension in individuals with foot<br>drop and thigh muscle<br>weakness, following an upper<br>motor neuron injury or disease.<br>During gait, the L300 Plus<br>System electrically stimulates<br>muscles in the affected leg to<br>provide dorsiflexion of the foot<br>and knee flexion or extension,<br>thus it may improve the<br>individual's gait. The L300 Plus<br>System may also:<br>* Facilitate Muscle re-education<br>* Prevent/retard disuse atrophy<br>* Maintain or increase joint<br>range of motion<br>* Increase local blood flow |
|                                         | New Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| 510k Number                             | K141812                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | K123972                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K021100                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K103343                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Name, Model                      | MyGait                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | WalkAide                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Empi 300pv                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | NESS L300 Plus System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Manufacturer                            | Otto Bock Healthcare<br>Products GmbH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Innovative Neurotronics,<br>Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Empi, Inc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Bioness Neuromodulation<br>Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Power Sources                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Battery                                 | built in                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | replaceable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | replaceable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | built in                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| # of batteries                          | 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| type                                    | Lilon                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Alkaline                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Alkaline                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Lilon                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| size                                    | 3.7V, 1100mAh                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | AA, LR6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | AA, LR6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 3.7 V, 750mAh                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Method of Line Current<br>Isolation     | Medical Class II Power<br>Adapter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | n/a                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | n/a                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Patient Leakage<br>Current              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| normal condition                        | <1 μA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Single Fault condition                  | <1μA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Number of Output<br>Modes               | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 5                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Number of Output<br>Channels            | 2…

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZI/K141812](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GZI/K141812)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
