K080219 · Ness-Neuromuscular Electrical Stimulation Systems · GZI · Feb 28, 2008 · Neurology
Device Facts
Record ID
K080219
Device Name
MODIFICATION TO:NESS L300
Applicant
Ness-Neuromuscular Electrical Stimulation Systems
Product Code
GZI · Neurology
Decision Date
Feb 28, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NESS L300 is intended to provide ankle dorsiflexion in individuals with drop foot following an upper motor neuron injury or disease. During the swing phase of gait, the NESS L300 electrically stimulates muscles in the affected leg to provide dorsiflexion of the foot; thus, it may improve the individual's gait. The NESS L300 may also facilitate muscle re-education, prevent/retard disused atrophy, maintain or increase joint range of motion and increase local blood flow.
Device Story
Neuroprosthesis for drop foot; consists of RF foot sensor, waist-mounted control unit, and below-knee orthosis with electrodes/RF stimulator. Sensor detects 'heel off' and 'heel contact' gait events; transmits data to stimulator. Stimulator triggers foot dorsiflexion during swing phase; ensures proper foot clearance and placement. Used by patients; clinician programs device via PDA interface. Benefits include improved gait, muscle re-education, atrophy prevention, and increased blood flow.
Clinical Evidence
No clinical data provided; substantial equivalence supported by performance data regarding hardware and software modifications.
Technological Characteristics
External functional neuromuscular stimulator. Components: lower leg orthosis with integrated RF stimulator/electrodes, waist-mounted control unit, RF foot sensor, and clinician PDA. Connectivity via RF communication. Powered by rechargeable batteries (orthosis/control unit) and non-rechargeable coin battery (sensor).
Indications for Use
Indicated for individuals with drop foot secondary to upper motor neuron injury or disease. Used to provide ankle dorsiflexion, improve gait, facilitate muscle re-education, prevent/retard disused atrophy, maintain/increase joint range of motion, and increase local blood flow.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{0}------------------------------------------------
K080219
ﻥ ﮐﮯ ﺍﯾﮏ ﺭﮨﺎﺋﺸﯽ ﮐﮯ ﺍﻧﺘﻈﺎﻣﯿﺎﺕ ﮐﮯ ﺷﮩﺮ
## 510(k) SUMMARY
# NESS, Ltd.'s NESS L300
# 510(k) Summary:
NESS L300
# Company Name:
NESS-Neuromuscular Electrical Stimulation Systems, Ltd.
# Contact Person:
Amit Dar R&D and Clinical Manager
| Telephone: | +972-9-748-5738 |
|------------|-----------------|
| Fax: | +972-9-748-5740 |
| E-mail: | amit@ness.co.il |
# Authorized US Agent:
Bioness, Inc. 25103 Rye Canyon Loop Valencia, CA 91355
#### Date prepared:
January 29, 2008
#### Trade Name:
NESS L300
#### Classification name:
External functional neuromuscular stimulator
# Class: II
Panel identification: Neurological devices
#### Product code:
GZI and IPF
#### Regulation number:
882.5810 External functional neuromuscular stimulator 890.5850 Powered Muscle Stimulator
{1}------------------------------------------------
## Predicate Devices:
Neuromuscular Electrical Stimulation Systems, Ltd.'s NESS L300 (K053468)
# Purpose of the Special 510(k) notice.
The NESS L300 is a modification to the NESS L300 (K053468).
### Device description:
This neuroprosthesis device consists of a RF foot sensor (Intelli-sense gait sensor), RF control unit (waist mounted / neck strap / in pocket), and a below the knee orthosis containing electrodes and a RF controlled stimulation unit (RF stim unit).
The Intelli-gait sensor detects "heel off" and "heel contact" events during gait and transmits signals to the RF stim unit. Accordingly, the RF stim unit initiates and pauses the stimulation, activating the foot dorsiflexors to ensure proper foot clearance during the swing phase and proper foot placement during the stance phase.
The NESS L300 comprises of 4 main parts:
- 1. A lower leg orthosis with integrated RF stim unit and electrodes, RF communication, and rechargeable battery,
- 2. A waist mounted (or in-pocket) Control Unit (CU), including a PDA (clinician programmer) interface, RF communication, and AAA rechargeable battery.
- 3. Intelli-sense gait sensor with RF communication and non-rechargeable coin battery.
- 4. Clinician Programmer Handheld computer (PDA)
#### Indications for Use:
The NESS L300 is intended to provide ankle dorsiflexion in individuals with drop foot following an upper motor neuron injury or disease. During the swing phase of gait, the NESS L300 electrically stimulates muscles in the affected leg to provide dorsiflexion of the foot; thus, it may improve the individual's gait. The NESS L300 may also facilitate muscle re-education, prevent/retard disused atrophy, maintain or increase joint range of motion and increase local blood flow.
#### Substantial Equivalence:
The modified NESS L300 has the same intended use and similar indications, principles of operation, and technological characteristics as the original, cleared NESS L300. The minor differences in the embedded software, the Intelli-Gait (PDA) software, the hardware, and the accessories do not raise any new questions of safety or effectiveness. Performance data demonstrates that the modified NESS L300 is as safe and effective as the predicate device. Thus, the NESS L300 is substantially equivalent to its predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
FEB 28 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Neuromuscular Electrical Stimulation Systems % Dr. Evan Rosenfeld Bioness Incorporated 25103 Rye Canyon Loop Valencia, CA 91355
Re: K080219 Trade/Device Name: Modification to NESS L300 Regulation Number: 21 CFR 882.5810 Regulation Name: External functional neuromuscular stimulator Regulatory Class: Class II Product Code: GZI Dated: January 29, 2008 Received: January 29, 2008
Dear Dr. Rosenfeld:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Dr. Evan Rosenfeld
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ K080219
Device Name: NESS L300
Indications for Use:
The NESS L300 is intended to provide ankle dorsiflexion in individuals with drop foot following an upper motor neuron injury or disease. During the swing phase of gait, the NESS L300 electrically stimulates muscles in the affected leg to provide dorsiflexion of the foot; thus, it may improve the individual's gait. The NESS L300 may also facilitate muscle re-education, prevent/retard disused atrophy, maintain or increase joint range of motion and increase local blood flow.
Prescription Use X (Per 21 C.F.R. 801.109)
AND/OR
Over-The-Counter Use (Per 21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark M. Millerson
Division of General, Restorative,
and i
K08021
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.