K221376 · Boston Scientific Neuromodulation · GZB · Jun 10, 2022 · Neurology
Device Facts
Record ID
K221376
Device Name
Nalu Neurostimulation System
Applicant
Boston Scientific Neuromodulation
Product Code
GZB · Neurology
Decision Date
Jun 10, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Spinal Cord Stimulation (SCS) This system is indicated as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or limbs, including unilateral pain. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device. Peripheral Nerve Stimulation (PNS) This system is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. The system is not intended to treat pain in the craniofacial region. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
Device Story
Implantable neurostimulation system for chronic, intractable pain; delivers pulsed electrical current to spinal cord or peripheral nerves to inhibit pain signal transmission. Components: Implantable Pulse Generator (IPG), leads with 4 or 8 electrodes, Therapy Disc (battery/RF electronics), and remote programmer. Used in hospital/home settings by physicians (orthopedic, neurosurgeon, anesthesiologist) and patients. Clinician uses programmer to set stimulation parameters; patient uses remote to control therapy. System uses RF wireless transmission for power and control. Updated software allows optional model-based current steering to optimize stimulation delivery. Patient remains in feedback loop to determine therapy effectiveness.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench and non-clinical performance testing, including software/firmware validation.
Technological Characteristics
Active implantable neurostimulator; RF wireless power transmission (40.68 MHz). Leads with 4 or 8 cylindrical electrodes. Waveform: charge-balanced, delayed, asymmetrical biphasic; decaying exponential pulse shape. Connectivity: wireless communication between Therapy Disc, IPG, and remote programmer. Software: Moderate level of concern; supports manual and model-based current steering. Standards: ISO 14708-1/3, IEC 60601-1/1-2/1-11, EN ISO 14971.
Indications for Use
Indicated for adults with severe intractable chronic pain of peripheral nerve origin or chronic, intractable pain of the trunk/limbs (SCS). Used as sole therapy or adjunct in multidisciplinary approach. Not for craniofacial pain. Trial devices limited to 30 days.
Regulatory Classification
Identification
An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services-USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
June 10, 2022
Nalu Medical, Inc. Chelsea Gutierrez VP Regulatory and Quality 2320 Faraday Avenue, Suite 100 Carlsbad, California 92008
Re: K221376
Trade/Device Name: Nalu Neurostimulation System Regulation Number: 21 CFR 882.5880 Regulation Name: Implanted Spinal Cord Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZB, GZF Dated: May 12, 2022 Received: May 12, 2022
Dear Chelsea Gutierrez:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-
combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221376
Device Name Nalu Neurostimulation System
## Indications for Use (Describe)
Spinal Cord Stimulation (SCS)
This system is indicated as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or limbs, including unilateral pain. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
### Peripheral Nerve Stimulation (PNS)
This system is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. The system is not intended to treat pain in the craniofacial region.
The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span>☑</span> <span>Research that is Part 21 CFR 58 Subject to Bioresearch Monitoring Program</span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span>☐</span> <span>One Time Contracted (to Part 21 CFR 58) Study</span> </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1. Submission Sponsor
Nalu Medical, Incorporated 2320 Faraday Ave., Suite 100 Carlsbad, CA 92008 Phone: (714) 401-0608 Fax: (760) 448-2377 Contact: Chelsea Gutierrez Vice President of Regulatory Affairs and Quality Assurance
#### 2. Date Prepared: June 10, 2022
#### 3. Device Names and Classification
| Primary Product Code | |
|---------------------------|-------------------------------------------------------------------|
| Proprietary Name | Nalu Neurostimulation System |
| Common Names | Stimulator, Spinal-Cord, Implanted (Pain Relief) |
| Class | II |
| Classification Regulation | 21 CFR 882.5880; Implanted spinal cord stimulator for pain relief |
| Product Code | GZB |
| Review Panel | Division of Neurology |
| Secondary Product Code | |
|---------------------------|------------------------------------------------------------------------|
| Proprietary Name | Nalu Neurostimulation System |
| Common Names | Stimulator, Peripheral Nerve, Implanted (Pain Relief) |
| Class | II |
| Classification Regulation | 21 CFR 882.5870; Implanted peripheral nerve stimulator for pain relief |
| Product Code | GZF |
| Review Panel | Division of Neurology |
#### 4. Predicate Device(s)
Nalu Neurostimulation System, K203547
For software differences between the Nalu Neuromodulation System and the predicate(s) (K203547), substantial equivalence to the predicate device is demonstrated.
#### 5. Device Description
The Nalu Neurostimulation System has been cleared by the FDA for spinal cord stimulation (SCS; K203547) and peripheral nerve stimulation (PNS; K203547) to provide therapeutic relief for chronic, intractable pain of the trunk and/or limbs including unilateral, bilateral nerve pain. The Nalu Neurostimulation therapy utilizes pulsed
{4}------------------------------------------------
electrical current to create an energy field that acts on nerves in the spinal cord or peripheral nerve to inhibit the transmission of pain signals to the brain. The Nalu System is implanted only following a successful trial period using the Nalu Neurostimulation trial system.
The Nalu Neurostimulation System consists of five (5) components. The implantable pulse generator (IPG) provides electrical stimulation pulses that are transmitted through the leads, to the desired location, either on the spinal cord or peripheral nerve site. The leads are implantable and designed to deliver electrical pulses to the nerves via an array of four (4) or eight (8) cylindrical electrodes at the distal end. The Trial Therapy Disc or the Therapy Disc houses the battery and electronics for RF power and controls the IPG for therapy delivery via the remote programmer. Implantation of the Nalu IPG and lead components for Spinal Cord Stimulation (SCS) or Peripheral Nerve Stimulation (PNS) is performed via standard surgical tools and techniques, as described in (K203547).
#### 6. Indications for Use
## Spinal Cord Stimulation (SCS)
This system is indicated as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or limbs, including unilateral or bilateral pain.
The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long-term) device.
## Peripheral Nerve Stimulator (PNS)
This system is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. The system is not intended to treat pain in the craniofacial region.
The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long-term) device.
{5}------------------------------------------------
#### 7. Comparison with the Predicate Device
| Table 1: Nalu Neuromodulation System for Spinal Cord Stimulation (SCS) | | | |
|------------------------------------------------------------------------|--|--|--|
|------------------------------------------------------------------------|--|--|--|
| Device | Nalu<br>Neuromodulation System (Subject Device, K221376) | Nalu<br>Neuromodulation System (Predicate Device, K203547) | Analysis of Technological Differences |
|---------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K221376 | K203547 | N/A |
| Product Code and Class | GZB and GZF, Class II | GZB and GZF, Class II | Same |
| Regulation Number | 21 CFR 882.5880 (GZB)<br>21 CFR 882.5870 (GZF) | 21 CFR 882.5880 (GZB)<br>21 CFR 882.5870 (GZF) | Same |
| Classification Name | Implanted spinal cord stimulator for pain relief. (GZB)<br>Implanted peripheral nerve stimulator for pain relief. (GZF) | Implanted spinal cord stimulator for pain relief. (GZB)<br>Implanted peripheral nerve stimulator for pain relief. (GZF) | Same |
| Trade Name | Nalu<br>Neuromodulation System | Nalu<br>Neurostimulation System | Same |
| Manufacturer | Nalu Medical, Inc. | Nalu Medical, Inc. | Same |
| Intended Use | The Nalu<br>Neuromodulation system is intended for the stimulation of the spinal cord for treatment of chronic, intractable pain (GZB).<br>Stimulation of peripheral nerves for chronic, intractable pain (GZF). | The Nalu<br>Neuromodulation system is intended for the stimulation of the spinal cord for treatment of chronic, intractable pain (GZB).<br>Stimulation of peripheral nerves for chronic, intractable pain (GZF). | Same |
| Indications for Use | For Spinal Cord Stimulation-<br>This system is indicated as the | For Spinal Cord Stimulation-<br>This system is indicated as the | Same |
| Device | Nalu<br>Neuromodulation<br>System (Subject<br>Device, K221376) | Nalu<br>Neuromodulation<br>System (Predicate<br>Device, K203547) | Analysis of<br>Technological<br>Differences |
| | sole mitigating<br>agent, or as an<br>adjunct to other<br>modes of therapy<br>used in a<br>multidisciplinary<br>approach for<br>chronic, intractable<br>pain of the trunk<br>and/or limbs,<br>including unilateral<br>or bilateral pain.<br>The trial devices are<br>solely used for trial<br>stimulation (no<br>longer than 30<br>days) to determine<br>efficacy before<br>recommendation<br>for a permanent<br>(long- term) device.<br>For Peripheral<br>Nerve Stimulation-<br>This system is<br>indicated for pain<br>management in<br>adults who have<br>severe intractable<br>chronic pain of<br>peripheral nerve<br>origin, as the sole<br>mitigating agent, or<br>as an adjunct to<br>other modes of<br>therapy used in a<br>multidisciplinary<br>approach. The<br>system is not<br>intended to treat<br>pain in the<br>craniofacial region.<br>The trial devices are | sole mitigating<br>agent, or as an<br>adjunct to other<br>modes of therapy<br>used in a<br>multidisciplinary<br>approach for<br>chronic, intractable<br>pain of the trunk<br>and/or limbs,<br>including unilateral<br>or bilateral pain.<br>The trial devices are<br>solely used for trial<br>stimulation (no<br>longer than 30<br>days) to determine<br>efficacy before<br>recommendation<br>for a permanent<br>(long- term) device.<br>For Peripheral<br>Nerve Stimulation-<br>This system is<br>indicated for pain<br>management in<br>adults who have<br>severe intractable<br>chronic pain of<br>peripheral nerve<br>origin, as the sole<br>mitigating agent, or<br>as an adjunct to<br>other modes of<br>therapy used in a<br>multidisciplinary<br>approach. The<br>system is not<br>intended to treat<br>pain in the<br>craniofacial region.<br>The trial devices are | |
| Device | Nalu<br>Neuromodulation<br>System (Subject<br>Device, K221376) | Nalu<br>Neuromodulation<br>System (Predicate<br>Device, K203547) | Analysis of<br>Technological<br>Differences |
| | stimulation (no<br>longer than 30 day)<br>to determine<br>efficacy before<br>recommendation<br>for a permanent<br>(long term) device. | stimulation (no<br>longer than 30 day)<br>to determine<br>efficacy before<br>recommendation<br>for a permanent<br>(long term) device. | |
| Clinical application | Treatment of<br>chronic, intractable<br>pain of the trunk<br>and/or limbs,<br>including unilateral<br>or bilateral pain. | Treatment of<br>chronic, intractable<br>pain of the trunk<br>and/or limbs,<br>including unilateral<br>or bilateral pain. | Same |
| Prescription Use | Yes | Yes | Same |
| Environmental Use | Hospital, Home | Hospital, Home | Same |
| Intended Clinician | Orthopedic,<br>Neurosurgeon,<br>Anesthesiologist | Orthopedic,<br>Neurosurgeon,<br>Anesthesiologist | Same |
| Intended User | Physician,<br>Layperson | Physician,<br>Layperson | Same |
| Implant site, leads | Epidural space (SCS)<br>or peripheral nerve<br>areas (PNS) | Epidural space (SCS)<br>or peripheral nerve<br>areas (PNS) | Same |
| Principle of<br>Operation | Stimulation of the<br>spinal cord to<br>provide therapeutic<br>relief for chronic,<br>intractable pain of<br>the trunk and/or<br>limbs including<br>unilateral or<br>bilateral pain. | Stimulation of the<br>spinal cord to<br>provide therapeutic<br>relief for chronic,<br>intractable pain of<br>the trunk and/or<br>limbs including<br>unilateral or<br>bilateral pain. | Same |
| Mode of Action | RF wireless<br>transmission of<br>energy to deliver<br>stimulation at<br>stimulator<br>electrodes. | RF wireless<br>transmission of<br>energy to deliver<br>stimulation at<br>stimulator<br>electrodes. | Same |
| Software Level of<br>Concern | Moderate | Moderate | Same |
| Clinician<br>Programmer | Software to<br>communicate to | Not publicly<br>available | The differences<br>do not impact the<br>safety and |
| Device | Nalu<br>Neuromodulation<br>System (Subject<br>Device, K221376) | Nalu<br>Neuromodulation<br>System (Predicate<br>Device, K203547) | Analysis of<br>Technological<br>Differences |
| | Trial Therapy or<br>Therapy Disc<br><br>Manual control of<br>current for selected<br>electrodes and<br>optional model<br>based allocation. | | effectiveness of<br>the device. The<br>safety parameters<br>(charge per phase,<br>charge density<br>and current<br>density) remain<br>unchanged. The<br>updated software<br>allows the<br>clinician to<br>optionally use a<br>mode to set the<br>current values on<br>up to 4 electrodes<br>based on a model.<br>The safety limits<br>continue to be<br>applied as before.<br>The new method<br>is an option that<br>may allow<br>programming to<br>go faster when<br>causing the<br>current steering<br>stimulation mode.<br>As with all<br>programming in<br>neurostimulation<br>the patient<br>remains in the<br>feedback loop and<br>determines the<br>effectiveness of<br>the therapy. |
| Patient Remote<br>Control | Software to pair<br>with Trial Therapy<br>or Therapy Disc<br><br>SW update to<br>reflect changes on<br>the Clinician<br>Programmer and | Not publicly<br>available | - |
| Device | Nalu<br>Neuromodulation<br>System (Subject<br>Device, K221376) | Nalu<br>Neuromodulation<br>System (Predicate<br>Device, K203547)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.