← Product Code [GXQ](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXQ) · K991413

# CODMAN ETHISORB DURA PATCH (K991413)

_Johnson & Johnson Professionals, Inc. · GXQ · Mar 13, 2000 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXQ/K991413

## Device Facts

- **Applicant:** Johnson & Johnson Professionals, Inc.
- **Product Code:** [GXQ](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXQ.md)
- **Decision Date:** Mar 13, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5910
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The CODMAN ETHISORB™ Dura Patch is a synthetic, absorbable implant intended for bridging defects of the dura mater.

## Device Story

CODMAN ETHISORB™ Dura Patch is a synthetic, absorbable implant used by neurosurgeons in the operating room to bridge dural defects. The device consists of a composite fleece fabric of polyglactin 910 and polydioxanone (PDS) yarns, with a PDS film coating on one surface. The porous fleece structure facilitates tissue on-growth, while the PDS film minimizes cerebrospinal fluid (CSF) leakage. The patch is provided sterile in a foil pouch. Unlike non-absorbable predicate patches, this device is hydrolyzed and absorbed by the body over time after the dural defect has healed. It provides a biocompatible, flexible, and impermeable barrier to support dural integrity during the healing process.

## Clinical Evidence

Evidence includes biocompatibility testing (pyrogenicity, cytotoxicity, tissue reaction) and performance testing in target tissue. A prospective, non-randomized clinical trial was conducted to evaluate clinical performance. Results demonstrated the device is safe and effective as a dura substitute and equivalent to currently marketed products.

## Technological Characteristics

Synthetic, absorbable composite fleece fabric of polyglactin 910 and polydioxanone (PDS) yarns; one surface coated with PDS film dyed with D&C Violet No. 2. Flexible, impermeable sheet form factor. Sterile, single-use implant. No electronic or software components.

## Regulatory Identification

A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).

## Predicate Devices

- Dura-Guard™ Dural Repair Patch by Bio-Vascular, Inc. ([K950956](/device/K950956.md))
- Neuro-Patch by Aesculap®, Inc. ([K960470](/device/K960470.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## MAR 1 3 2000

# 510K Summarv CODMAN ETHISORB™ Dura Patch

March 8, 2000

# K991413

#### 1. Submitter

Codman & Shurtleff, Inc. 325 Paramount Drive Ravnham. MA 02767 Deana Boushell, Manager Regulatory Affairs, (508) 828-3107

#### Device Name 2.

Proprietary Name: Common Name: Classification Name: Regulatory Class:

CODMAN ETHISORB™ Dura Patch Dura Substitute Dura Substitute Class II by 21 CFR §882.5910

#### 3. Intended Use

The CODMAN ETHISORB™ Dura Patch is a synthetic, absorbable implant intended for bridging defects of the dura mater.

#### Device Description 4.

The CODMAN ETHISORB™ Dura Patch is composed of a composite fleece fabric of polyglactin 910 and polydioxanone yams, with a polydioxanone film dyed with D&C Violet No. 2 on one surface. The porous structure of the fleece allows tissue on-growth while the PDS film coating minimizes leakage of cerebrospinal fluid (CSF). The device is provided sterile and ready for implantation in a foil pouch. Animal and clinical trials have demonstrated that the device is suitable for its intended use. The implant material is fully biocompatible and absorbable over time

#### 5. Predicate Device Comparison

The CODMAN ETHISORB™ Dura Patch is similar to other dura substitutes with respect to its design as a thin, flexible, impermeable sheet of material which is both biocompatible and sufficiently strong enough to bridge the dural defect until healing occurs. Examples of predicate products with these same characteristics include the Dura-Guard™ Dural Repair Patch by Bio-Vascular, Inc., (K950956) and the Neuro-Patch by Aesculap®, Inc. (K960470). The Dura Patch differs from the predicate devices only in that once it has fulfilled its function and is no longer needed to cover the defect, it is hydrolyzed and absorbed slowly over time.

{1}------------------------------------------------

#### 6. Device Testing Summary

| Testing Performed           | Comments                                                                                                                                                                                                                                                                                                                  |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility<br>Testing | Although all materials used in the device have been shown to<br>be bicompatible implant materials, additionaltests were<br>performed on final Dura Patch product, including pyrogenicity,<br>cytocoxicity, and implant studies of tissue reaction.                                                                        |
| Performance Testing         | Performance testing was conducted in the target tissue to<br>confirm device suitability for its intended use as a dura<br>substitute.                                                                                                                                                                                     |
| Clinical Evaluation         | A prospective non-randomized clinical trial was conducted to<br>demonstrate the clinical performance characteristics of the<br>device. Results of all testing demonstrate that the<br>ETHISORB™ Dura Patch is equivalent to currently marketed<br>products in its performance as a safe and effective dura<br>substitute. |

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three horizontal lines extending from its body, representing feathers or wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## MAR 1 3 2000

Ms. Deana Boushell, RAC Team Leader, Regulatory Affairs Codman & Shurtleff, Inc. 325 Paramount Drive Raynham. Massachusetts 02767

Re: K991413 Trade Name: ETHISORB™ Dura Patch Regulatory Class: II Product Code: GXQ Dated: January 10, 2000 Received: January 11, 2000

Dear Ms. Boushell:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Ms. Deana Boushell, RAC

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Neil R.P. Ogden
James E. Dillard III

ames E. નીન Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications For Use Device Name: CODMAN ETHISORB™ Dura Patch

K99143

The CODMAN ETHISORB™ Dura Patch is an absorbable, synthetic implant for bridging defects of the dura mater.

(Please do not write below this line - Continue on another page if necessary) Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number. K991413

Prescription Use (Per 21 CFR §801.109)

OR

Over-the-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXQ/K991413](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXQ/K991413)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
