K153613 · Acera Surgical, Inc. · GXQ · Mar 16, 2016 · Neurology
Device Facts
Record ID
K153613
Device Name
Cerafix Dura Substitute
Applicant
Acera Surgical, Inc.
Product Code
GXQ · Neurology
Decision Date
Mar 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5910
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cerafix Dura Substitute is indicated as a dura substitute for the repair of dura mater. This device is indicated for defects of 1.9 in2 (12.5 cm2) or less in area. For example, 1.2 in x 1.6 in (3 cm x 4 cm) would be an acceptable defect size.
Device Story
Cerafix Dura Substitute is a resorbable, porous polymer matrix implant used for dural repair. Surgeon cuts device to size and places over dural defect using tensionless sutures; requires hydration prior to placement. Device acts as a scaffold for tissue repair. Used in clinical settings by surgeons. Provides physical barrier for dura mater; prevents cerebrospinal fluid leakage. Benefits patient by facilitating dural healing and closure.
Clinical Evidence
No human clinical data. Evidence consists of bench testing (thickness, mass, tensile strength, suture pull-out, burst strength, shrink temperature, fiber diameter, pore size) and side-by-side animal implantation studies comparing subject device to predicate. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, irritation, hemolysis, genotoxicity, pyrogenicity, systemic toxicity, and 90/180-day chronic toxicity).
Technological Characteristics
Porous PGLA/PDO polymer matrix; resorbable. Supplied sterile (SAL 10^-6). Non-pyrogenic. Dimensions vary; requires hydration. Mechanical properties verified via ASTM standards. Biocompatibility per ISO 10993.
Indications for Use
Indicated for repair of dura mater defects in patients requiring dural substitute, specifically for defects 12.5 cm2 or less.
Regulatory Classification
Identification
A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the top half of the circle. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W 066-G609 Silver Spring, MD 20993-0002
March 24, 2016
Acera Surgical. Inc. % Linda Braddon, PhD Consultant Secure BioMed Evaluations 7828 Hickory Flat Highway, Suite 120 Woodstock, Georgia 30188
Re: K153613
Trade/Device Name: Cerafix Dura Substitute Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXQ Dated: December 17, 2015 Received: December 17, 2015
Dear Dr. Braddon:
This letter corrects our substantially equivalent letter of March 16, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153613
Device Name Cerafix Dura Substitute
Indications for Use (Describe)
The Cerafix Dura Substitute is indicated as a dura substitute for the repair of dura mater. This device is indicated for defects of 1.9 in2 (12.5 cm2) or less in area. For example, 1.2 in x 1.6 in (3 cm x 4 cm) would be an acceptable defect size.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><div> </div><div>☑ Prescription Use (Part 21 CFR 801 Subpart D)</div></div> | <div style="display:flex; align-items:center;"><div> </div><div>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div></div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Image /page/3/Picture/1 description: The image shows the logo for Acera Surgical. The logo consists of a green triangle on the left, with the word "ACERA" in green letters to the right of the triangle. Below the word "ACERA" is the word "SURGICAL" in smaller, green letters. The logo is simple and modern, and the use of green suggests a focus on health and wellness.
In accordance with 21 CFR 807.87 (h) and 21 CRF 807.92, the 510(k) summary for the Acera Surgical Cerafix® Dura Substitute is provided below.
| Date Summary Prepared | March 16, 2016 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by | Acera Surgical, Inc.<br>10880 Baur Blvd<br>St. Louis, MO 63132<br>Phone 844-879-2237 |
| 510(k) Contact | Secure BioMed Evaluations<br>Linda Braddon, Ph.D.<br>7828 Hickory Flat Highway<br>Suite 120<br>Woodstock, GA 30188<br>770-837-2681 (direct)<br>855-MED-DEV1 (office)<br>LGB@SecureBME.com |
| Trade Name | Cerafix® Dura Substitute |
| Common Name | Dura substitute |
| Code —Classification | GXQ 21 CFR 882.5910 : Class II |
| Primary Predicate Device | K991413 Ethisorb™ Dura Patch |
| Reference Device | K092388 DuraGen Plus™ Dural Regeneration Matrix |
#### Device Description
Cerafix® Dura Substitute is a resorbable implant for repair of dural defects and is to be used with tensionless sutures. Cerafix® Dura Substitute is a soft, white, pliable, nonfriable, porous polymer matrix. Cerafix® Dura Substitute is available in a variety of sizes and is supplied sterile and nonpyrogenic in a single-use nested pouch configuration, which is enclosed within a protective chipboard envelope.
#### Indications for Use
The Cerafix® Dura Substitute is indicated as a dura substitute for the repair of dura mater. This device is indicated for defects of 1.9 in? (12.5cm²) or less in area. For example, 1.2 in x 1.6 in (3 cm x 4 cm) would be an acceptable defect size.
{4}------------------------------------------------
#### Technological Characteristics
The materials used in the subject device are equivalent to the predicate device. Additionally, comparative mechanical testing was performed using the commercially available dura substitute reference device. The comparative mechanical testing showed equivalent performance of the subject device to the reference device. Lastly, physical characteristics are comparable to the predicate device, reference device, and that of native human dura.
Based on test results included in this submission, a maximum allowable defect size has been prescribed for the subject device. The subject device has the same technological characteristics as the predicate device and reference device in terms of principles of operation, materials of construction, material performance, and biocompatibility. Additionally, side-by-side animal studies show the subject device is equivalent for the indicated use of a dura substitute for the repair of dura mater. The subject device has the same technological characteristics as the predicate and reference device as follows:
| Characteristic | Cerafix® Dura<br>Substitute<br>(subject device) | Ethisorb™ Dura Patch<br>(predicate device) | DuraGen™ Plus Dural<br>Regeneration Matrix<br>(reference device) | Comparison |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| 510(k) | K153613 | K991413 | K092388 | N/A |
| Principles of<br>Operation | Device can be cut by<br>surgeon and placed on<br>dural defect with<br>tensionless suture<br>application. Suture line<br>should be 2-3 mm<br>from edge of implant.<br>Implant should be<br>large enough to<br>overlap edge of the<br>remaining dura by at<br>least one (1)<br>centimeter. | Device can be cut by<br>surgeon and placed on<br>dural defect with a<br>running or interrupted<br>suture application.<br>Avoid tensioning of<br>sutures. Suture line<br>should be 2 mm from<br>edge of implant. | Device can be cut by<br>surgeon and placed on<br>dural defect in either<br>an onlay or tensionless<br>suture application.<br>Implant should be<br>large enough to<br>overlap edge of the<br>remaining dura by at<br>least one (1)<br>centimeter. | Equivalent |
| Material of<br>Construction | Porous polymer matrix | Porous polymer matrix | Bovine collagen matrix | Equivalent to<br>predicate<br>device |
| Indications for<br>Use | Indicated as a dura<br>substitute for the<br>repair of dura mater.<br>This device is indicated<br>for defects of 1.9 in²<br>(12.5cm²) or less in<br>area. For example, 1.2<br>in x 1.6 in (3 cm x 4<br>cm) would be an<br>acceptable defect size. | Indicated as an<br>absorbable, synthetic<br>implant for bridging<br>defects of the dura<br>mater. | Indicated as a dura<br>substitute for the<br>repair of dura mater. | Equivalent |
| Size | Variety of Sizes | Variety of Sizes | Variety of Sizes | Equivalent |
| Material<br>Composition | Porous PGLA / PDO<br>matrix | Porous PGLA / PDO<br>matrix | Bovine collagen matrix | Equivalent to<br>predicate<br>device |
{5}------------------------------------------------
| Surgical<br>Application<br>Restrictions | Device does not have<br>requirement for<br>specific orientation | On one side the<br>porous structure of<br>the VICRYL fleece<br>allows tissue on-<br>growth while the PDS<br>film coating minimizes<br>leakage of<br>cerebrospinal fluid. | Device does not have<br>requirement for<br>specific orientation | Equivalent to<br>reference<br>device |
|-----------------------------------------|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|--------------------------------------|
| Sterility | Sterile, SAL 10-6 | Sterile, SAL 10-6 | Sterile, SAL 10-6 | Equivalent |
| Packaging | Double sterile pack.<br>Nested pouch<br>configuration within a<br>chipboard envelope. | Foil pouch within a<br>chipboard box | Double sterile pack.<br>Nested thermoformed<br>trays with Tyvek lids<br>within a chipboard<br>box. | Equivalent to<br>reference<br>device |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Non-pyrogenic | Equivalent |
| Resorbable | Yes | Yes | Not Applicable | Equivalent to<br>predicate<br>device |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible | Equivalent |
The following technological differences exist between the subject and predicate devices:
- Subject device is manufactured with non-woven fiber technique versus the predicate device, which is manufactured with a woven technique
- The predicate device has a polymer film dyed with D&C Violet No. 2, while the subject device has neither a film layer nor dyes.
- Subject device does not have a requirement for specific orientation
- The predicate device does not specify how much overlap should exist between the edge of the device and the remaining dura. The subject device specifies a minimum distance of one centimeter. (Note: the reference device specifies a minimum distance of one centimeter as well).
- . Although the maximum thickness of the subject device is comparable to the predicate device, the subject device has a lower minimum thickness that is comparable to native dura human dura.
- Subject device needs to be hydrated prior to placement, whereas the predicate device can be used without hydration.
Pre-clinical testing confirmed that despite differences in manufacturing techniques, the Cerafix® Dura Substitute is equivalent in function, indication for use, device classification product code, environment of use, and principles of operation to the predicate device.
{6}------------------------------------------------
### Non-Clinical Testing – Mechanical
The subject device was evaluated in side-by-side bench testing versus the predicate or commercially available reference device. The results showed that the subject device demonstrated equivalent properties in the following areas:
| Test | Test Method Summary | Acceptance Criteria<br>Results |
|--------------------------|---------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Thickness | Comparison of Cerafix® Dura<br>Substitute thickness to other dura<br>substitutes on the market | Equivalent to Predicate<br>or Reference Device<br>PASS |
| Mass per Area | Comparison of Cerafix® Dura<br>Substitute mass per area to other<br>dura substitutes on the market | Equivalent to Predicate<br>or Reference Device<br>PASS |
| Tensile Strength | Comparison of Cerafix® Dura<br>Substitute tensile strength to other<br>dura substitutes on the market | Equivalent to Predicate<br>or Reference Device<br>PASS |
| Suture Pull-Out Strength | Comparison of Cerafix® Dura<br>Substitute suture pull-out strength to<br>other dura substitutes on the market | Equivalent to Predicate<br>or Reference Device<br>PASS |
| Burst Strength | Comparison of Cerafix® Dura<br>Substitute burst strength to other<br>dura substitutes on the market | Equivalent to Predicate<br>or Reference Device; and<br>burst strength greater than anticipated<br>intracranial pressures<br>PASS |
| Shrink Temperature | Evaluation of Cerafix® Dura<br>Substitute stability at various<br>temperatures | Show stability at applicable<br>temperatures<br>PASS |
| Fiber Diameter | Evaluation of Cerafix® Dura<br>Substitute fiber diameter via SEM | Meets Final Device Specification<br>PASS |
| Pore Size | Evaluation of Cerafix® Dura<br>Substitute pore size via SEM | Meets Final Device Specification<br>PASS |
#### Non-Clinical Testing - Biocompatibility
Biocompatibility testing was performed in compliance with ISO 10993. The results are summarized in the following table:
| Biocompatibility Tests | Results |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| ISO Cytotoxicity MEM Elution<br>According to ISO 10993-5 Biological evaluation of medical<br>devices: Part 5 Tests for In vitro Cytotoxicity | Cell culture treated with test sample<br>exhibited no reactivity. Therefore, non-<br>cytotoxic. |
| Guinea Pig Maximization - Sensitization<br>According to ISO 10993-10 Biological evaluation of medical<br>devices: Part 10 Tests for irritation and delayed hypersensitivity | Albino guinea pigs treated with test sample<br>did not elicit a sensitization response.<br>Therefore, non-irritant. |
| Intracutaneous Irritation Reactivity<br>According to ISO 10993-10 Biological evaluation of medical<br>devices: Part 10 Tests for irritation and delayed hypersensitivity | Rabbits treated with test samples were<br>non-irritating. Therefore, non-irritant. |
| Hemolysis Assay<br>According to ASTM F756-08<br>FDA Consensus Standard Number 2-154 | Rabbit blood treated with test samples was<br>found to be non-hemolytic. |
{7}------------------------------------------------
| Biocompatibility Tests | Results |
|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Genotoxicity<br>In Vitro Mouse Lymphoma Assay<br>ISO 10993-3:2003 | Cell culture with mouse lymphoma cells in<br>the presence of trifluorothymidine<br>exhibited a mean mutant frequency<br>equivalent to the negative control<br>Therefore, non-genotoxic. |
| Genotoxicity<br>In vivo Mouse Micronucleus Assay<br>ISO 10993-3:2003 | Adult CD-1 mice treated with test sample<br>were considered non-mutagenic |
| Genotoxicity<br>Bacterial Mutagenicity Test - Ames Assay<br>ISO 10993-3:2003 | Salmonella typhimurium histidine<br>auxotrophs and E. coli were considered<br>non-mutagenic |
| Pyrogenicity<br>Materials Mediated Rabbit Pyrogen Test | Albino rabbits treated with test samples<br>exhibited a negative response.<br>Therefore, non-pyrogenic. |
| Acute Systemic Toxicity<br>ISO 10993-11 | Albino mice treated with test samples were<br>considered non-toxic. |
| Endotoxin Testing | Less than 2.15 EU/device. Non-pyrogenic. |
| Subchronic Toxicity<br>90 day animal study | Rabbits treated with test samples for 90<br>days show the device to be non-toxic. |
| Chronic Toxicity<br>180 day animal study | Rabbits treated with test samples for 180<br>days show the device to be non-toxic. |
### Non-Clinical Testing – Side-by-Side Animal Study Comparison
Side-by-side animal implantation studies were performed between the subject and predicate device. Results show equivalent safety and performance between the subject and predicate device.
#### Conclusions
The subject and predicate device underwent non-clinical evaluation that confirmed device equivalency in the indication for use, device classification, product code, biocompatibility, safety, efficacy, environment of use, and the principles of operation. Therefore, the subject device demonstrates equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.