K141608 · Collagen Matrix, Inc. · GXQ · Jan 16, 2015 · Neurology
Device Facts
Record ID
K141608
Device Name
COLLAGEN DURA MEMBRANE
Applicant
Collagen Matrix, Inc.
Product Code
GXQ · Neurology
Decision Date
Jan 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5910
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Collagen Dura Membrane is intended for use as a dura substitute for the repair of dura mater.
Device Story
Collagen Dura Membrane is a sterile, non-pyrogenic, bovine-derived collagen matrix; used as a dura substitute for dural repair. Device is flexible, non-friable, and conformable to defect site contours; possesses mechanical strength allowing for suturing. Supplied in various sizes for single-use. Implanted by surgeons during neurosurgical procedures to replace or repair dura mater. Gradual resorption profile allows for tissue integration. Benefits include effective dural closure and biocompatibility, reducing risk of cerebrospinal fluid leakage.
Clinical Evidence
No clinical data provided. Evidence based on bench testing and animal studies. Bench testing included dimensions, suture pullout strength, tensile strength, conformability, hydrothermal transition temperature, and porosity. Biocompatibility testing followed ISO 10993 standards (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, genotoxicity, pyrogenicity, implantation). Animal studies included rabbit dural defect repair, rat subcutaneous resorption, and canine craniectomy models, demonstrating performance equivalent to the predicate device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three overlapping human profiles facing to the right, with flowing lines extending from the bottom of the profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 16, 2015
Collagen Matrix, Inc. Peggy Hansen Senior VP, Marketing and Regulatory Affairs 15 Thornton Road Oakland, New Jersey 07436
Re: K141608 Trade/Device Name: Collagen Dura Membrane Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXQ Dated: December 15, 2014 Received: December 17, 2014
Dear Ms. Hansen,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Carlos L. Pena --S/^
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141608
Device Name Collagen Dura Membrane
Indications for Use (Describe)
Collagen Dura Membrane is intended for use as a dura substitute for the repair of dura mater.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
#### 1. Applicant Information
| Applicant Name: | Collagen Matrix, Inc. |
|-----------------|------------------------------------------------------------------|
| Address: | 15 Thornton Road<br>Oakland, New Jersey 07436 |
| Telephone: | (201) 405-1477 |
| Fax: | (201) 405-1355 |
| Contact Person: | Peggy Hansen, RAC<br>VP, Clinical, Regulatory, QA, and Marketing |
| Date Prepared: | January 15, 2015 |
#### 2. Name of the Device
Device Common Name: Device Trade Name: Device Classification Name: Dura Substitute Collagen Dura Membrane Dura Substitute 21 CFR 882.5910 Product Code GXQ Device Class II
#### 3. Legally Marketed Devices to Which Substantial Equivalence is Claimed
Predicate Device(s):
Durepair® Dura Regeneration Matrix K052211
#### 4. Description of the Device
Collagen Dura Membrane is a white, nonfriable, conformable, membrane matrix consisting of highly purified collagen derived from bovine dermis. It is flexible and conforms to the contours of the defect site. The product's mechanical strength allows the membrane matrix to be sutured in place. Collagen Dura Membrane is supplied sterile, non-pyrogenic, in various sizes, and for single use only.
#### 5. Intended Use
Collagen Dura Membrane is intended for use as a dura substitute for the repair of dura mater.
{4}------------------------------------------------
### 6. Summary/Comparison of Technical Characteristics
Collagen Dura Membrane has been determined to be substantially equivalent to the predicate devices having similar technological characteristics as follows:
| Parameter | Collagen Dura Membrane<br>(This submission) | Durepair® Dura Regeneration<br>Matrix |
|---------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Indications for Use | Intended for use as a dura<br>substitute for the repair of dura<br>mater. | Intended for use as a dura<br>substitute for the repair of dura<br>mater. |
| Collagen Source | Bovine dermis | Bovine dermis |
| Form | Membrane | Membrane |
| Color | White to off-white | White to off-white |
| Physical Integrity | Non-friable | Non-friable |
| Sizes | Variety of sizes | Variety of sizes |
| Suture Strength | Can be sutured | Can be sutured |
| Biocompatibility | Biocompatible | Biocompatible |
| In Vivo Stability | Gradual resorption | Gradual resorption |
| Sterility | Sterile, SAL 10-6 | Sterile, SAL 10-6 |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic |
| Single Use/Reuse | Single use only | Single use only |
| Packaging | Double peel package | Double peel package |
### 7. Discussion of Non-clinical Testing
The substantial equivalence of Collagen Dura Membrane and its predicate was demonstrated based on in vitro characterization studies, biocompatibility studies, and an animal efficacy study.
Non-clinical testing was performed in accordance with FDA recognized consensus standards and FDA quidelines as follows:
FDA Guidance Document entitled, "Guidance document for Dura Substitute Device: Guidance for Industry", issued on November 9, 2000
ISO 22442-1 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices - Part 1 Analysis and Risk Management
ISO 22442-2 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices - Part 2 Controls on Sourcing, Collection, and Handling
ISO 22442-3 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices - Part 3 Validation of the Elimination and/or Inactivation of Viruses and Transmissible Agents
{5}------------------------------------------------
ISO 10993-3:2009 Biological Evaluation of Medical Devices- Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
ISO 10993-5:2009 Biological Evaluation of Medical Devices- Part 5: Tests for in vitro cytotoxicity
ISO 10993-6:2009 Biological Evaluation of Medical Devices- Part 6: Test for local effects after implantation
ISO 10993-10:2009 Biological Evaluation of Medical Devices- Part 10 Test for local effects after implantation
ISO 10993-11:2009 Biological Evaluation of Medical devices – Part 11 Tests for systemic toxicity
#### Non-clinical Testing Conducted
In vitro product characterization testing was performed to demonstrate substantial equivalence of the subject device to its predicate device. A series of bench tests were conducted to evaluate material properties, biological properties, chemical and physical properties. The comparative bench testing is summarized in the table below.
| Test | Test Method | Results |
|-------------------------------------------|--------------------------------------------------------------------|---------------------------------------------------------------------|
| Dimensions | Measurements | Dimensions similar to predicate device |
| Suture pullout<br>strength | Internal test method using<br>mechanical test apparatus | Suture strength similar to predicate device |
| Tensile<br>strength | Internal test method using<br>mechanical test apparatus | Tensile strength similar to predicate device |
| Conformability | Internal test method to<br>measure drape angle | Conformability similar to predicate device |
| Hydrothermal<br>transition<br>temperature | Internal test method using<br>differential scanning<br>calorimeter | Hydrothermal transition temperature similar<br>to predicate device. |
| Porosity /<br>Permeability | Internal test method | Permeability similar to predicate device |
A series of in vitro and in vivo biocompatibility testing was performed to assess safety of the Collagen Dura Membrane as an implantable material. The biocompatibility testing performed is summarized in the table below.
| Test | Test Method/ Model | Results |
|-------------------------------|------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Cytotoxicity | ISO Agarose Overlay Method -<br>Extract, ISO 10993-5 | Non-cytotoxic. |
| Sensitization | Murine Local Lymph Node<br>Assay, ISO 10993-10 | Not considered to be sensitizing. |
| Intracutaneou<br>s Reactivity | Acute Intracutaneous Reactivity<br>in Rabbit, ISO 10993-10 | No erythema and no edema from the<br>test extract injected intracutaneously<br>into the rabbits. |
| Acute | ISO Systemic Toxicity in Mice, | No mortality or evidence of significant |
{6}------------------------------------------------
| Test | Test Method/ Model | Results |
|------------------------------------|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Systemic<br>Toxicity | ISO 10993-11 | systemic toxicity. |
| Genotoxicity | Bacterial Reverse Mutagenic<br>Study, ISO 10993-3 | Non-mutagenic to Salmonella<br>typhimurium tester strains TA98,<br>TA100, and TA1537, and to<br>Escherichia coli strain WP2uvrA. |
| Genotoxicity | Mouse Lymphoma Assay, ISO<br>10993-3 | None of the test article treatments<br>induced substantial increases in the<br>number of revertant colonies. Based<br>on the criteria and conditions of the<br>study protocol, the test article is<br>considered non-mutagenic. |
| Genotoxicity | In Vivo Mouse Micronucleus<br>Assay, ISO 10993-3 | None of the mice treated with the test<br>article preparations exhibited overt<br>signs of toxicity either immediately<br>post-treatment or during the induction<br>period. The levels of micronucleated<br>cells were within normal negative<br>ranges. Based on the criteria and<br>conditions of the study protocol, the<br>test article is considered non-<br>mutagenic. |
| Pyrogenicity | USP (151) Pyrogen Study -<br>Material Mediated | The test article was judged as<br>nonpyrogenic. |
| Implantation | Subcutaneous Implantation in<br>Rats, ISO 10993-6 | Minimum tissue reaction up to 4 weeks<br>of implantation and no adverse tissue<br>reaction to the host. |
| Subchronic<br>Systemic<br>Toxicity | Subcutaneous Implantation in<br>Rabbits, ISO 10993-11 | There was no evidence of systemic<br>toxicity or adverse findings attributed to<br>the test article at 13 week time point. |
| Chronic<br>Toxicity | Subcutaneous Implantation in<br>Rabbits, ISO 10993-11 | There was no evidence of systemic<br>toxicity or adverse findings attributed to<br>the test article at 26 week time point. |
An animal efficacy study and additional animal studies were conducted to evaluate the device as compared to its predicate device. No clinical tests were performed on the product; however clinical history of the predicate device was referenced in the submission. The animal studies performed are summarized in the table below.
| Test | Test Method/ Model | Results |
|----------------------------------|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Dura Repair<br>and<br>Resorption | Rabbit dural defect repair model<br>using the subject device and<br>predicate device as a control | Subject device performed as well as<br>the predicate device |
| Resorption | Rat subcutaneous model using<br>the subject device | Resorption profile verified design<br>criteria |
| Handling,<br>Implantability | Canine craniectomy model for<br>implanting the subject device<br>and predicate device | Subject device handles as well as the<br>predicate device using typical surgical<br>technique for implantation |
Viral inactivation studies were performed to ensure the viral safety of the product.
{7}------------------------------------------------
#### 8. Conclusion of Non-clinical Studies
The predicate device was cleared based on the results of non-clinical data. Subject and predicate device performance data were compared to support the safety of the subject device and demonstrate that the Collagen Dura Membrane should perform as intended in the specified us conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.