SYNTHECEL DURA REPAIR, SYNTHECEL DURA ONLAY

K131792 · Synthes, Inc. · GXQ · Dec 16, 2013 · Neurology

Device Facts

Record IDK131792
Device NameSYNTHECEL DURA REPAIR, SYNTHECEL DURA ONLAY
ApplicantSynthes, Inc.
Product CodeGXQ · Neurology
Decision DateDec 16, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5910
Device ClassClass 2
AttributesTherapeutic

Indications for Use

SyntheCel® Dura Onlay is intended for use as a dura replacement for the repair of dura mater in adults.

Device Story

SyntheCel® Dura Onlay is a non-resorbable, biosynthesized cellulose-based implant designed for neurosurgical applications. It consists of a non-woven, interconnected fiber matrix that functions as a mechanical barrier to protect dural defects and prevent cerebrospinal fluid (CSF) leakage. The device is intended for use by surgeons in clinical settings to repair dura mater in adult patients. It is immunologically inert, conforms to brain contours, and is applied directly to the defect site. By providing a physical seal, the device aids in the management of dural integrity, potentially reducing complications associated with CSF leaks.

Clinical Evidence

No new clinical performance data were collected. Safety and effectiveness were established via clinical data previously submitted in K113071, which demonstrated substantial equivalence to other legally marketed dura replacement products.

Technological Characteristics

Material: Biosynthesized cellulose. Form factor: Non-woven, interconnected fiber matrix. Dimensions: Various sizes ranging from 1in x 1in to 4in x 5in. Sterilization: Irradiation. Biocompatibility: Tested per ISO 10993 (non-irritating, non-sensitizing, non-mutagenic, non-cytotoxic, non-hemolytic, non-pyrogenic). Endotoxin levels: <0.06 EU/ml per LAL test.

Indications for Use

Indicated for use in adults as a dura replacement for the repair of dura mater. Contraindicated in patients with known allergy or sensitivity to cellulose.

Regulatory Classification

Identification

A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the word "SYNTHES" in a bold, sans-serif font. To the left of the word is a circular logo with a stylized image inside. A small registered trademark symbol is located to the right of the word. A horizontal line is located underneath the word. DEC 16 2013 . . ### 510(k) Summary -- SyntheCcl® Dura Onlay | Date Prepared | November 13, 2013 | | | |----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter | Synthes | | | | | 1301 GOSHEN PARKWAY | | | | | West Chester, PA 19380 | | | | | United States of America | | | | Contact | Heather L. Guerin, Ph.D., P.E. | | | | | guerin.heather@synthes.com | | | | | phone: (610) 719-5432 | | | | Device Name | SyntheCel® Dura Onlay | | | | Classification | 21 CFR 882.5910 (Dura Substitute) | | | | Device Class | 2 | | | | Product Code(s) | GXQ | | | | Predicate Devices | SyntheCel Dura Replacement Devices (K113071) | | | | Device Description | SyntheCel® Dura Onlay is composed of biosynthesized cellulose and | | | | | water with a unique construction of non-woven, interconnected | | | | | cellulose fibers. SyntheCel® Dura Onlay functions as a mechanical layer | | | | | which protects and repairs the dural defect while preventing further<br>CSF leakage. SyntheCel® Dura Onlay is immunologically inert and has | | | | | demonstrated minimal foreign body response. It is non-resorbable. | | | | Indications for Use | SyntheCel® Dura Onlay is intended for use as a dura replacement for | | | | | the repair of dura mater in adults. | | | | Technological<br>Characteristics | | SyntheCel® Dura Onlay<br>(K131792) | SyntheCel Dura<br>Replacement Devices<br>(K113071) | | | | | | | | | | | | | Indications<br>for Use | SyntheCel® Dura Onlay is<br>intended for use as a<br>dura replacement for the<br>repair of dura mater in<br>adults. | SyntheCel® is intended<br>for use as a dura<br>replacement for the<br>repair of dura mater.<br><br>SyntheCel® Dura Onlay is<br>indicated for use in<br>adults for the repair of<br>dural defects and it can<br>be placed without<br>sutures.<br><br>SyntheCel® Dura<br>Substitute is indicated<br>for use in adults for the<br>repair of dural defects<br>and it can be sutured | | | | | | | | Contra-<br>indications | Must not be implanted in<br>patients who have known<br>allergy or sensitivity to<br>the implant materials<br>(cellulose). | into place.<br>Same | | | Material | Biosynthesized cellulose | Same | | | Dimensions | 1in x 1in (2.5cm x 2.5cm)<br>1in x 3in (2.5cm x 7.5cm)<br>2in x 2in (5.0cm x 5.0cm)<br>3in x 3in (7.5cm x 7.5cm)<br>4in x 5in (10.0cm x<br>12.0cm) | Same. | | | Function | Conforms to contours of<br>brain. | Same | | | Resorbability | Non-resorbable | Same | | | Sterilization | Sterilized by irradiation | Same | | Clinical<br>Performance Data | No new clinical performance data were collected in support of this<br>submission. Clinical data were previously collected to evaluate the<br>safety and effectiveness of the SyntheCel® Dura Replacement Devices<br>(including SyntheCel® Dura Repair and SyntheCel® Dura Onlay) as<br>compared to the Control. This data was included in K113071. The<br>SyntheCel® Dura Replacement Devices (including SyntheCel® Dura<br>Repair and SyntheCel® Dura Onlay) were demonstrated to be<br>substantially equivalent to other legally marketed dura replacement<br>products in terms of safety and efficacy. | | | | Non-Clinical<br>Performance Data | No new non-clinical performance data were collected in support of this<br>submission.<br>Mechanical testing data were previously collected to support<br>substantial equivalence of SyntheCel® Dura Onlay to predicate devices.<br>This data was included in K113071. Tensile strength, and burst strength<br>were tested and SyntheCel® Dura Onlay was demonstrated to be<br>substantially equivalent to predicate devices.<br>Biocompatibility testing according to standards set forth in ISO 10993<br>previously demonstrated that the material is non-irritating, non-<br>sensitizing, non-mutagenic, non-cytotoxic, non-hemolytic, and non-<br>pyrogenic. This data was included in K113071.<br>Pyrogenicity was evaluated using the Limulus Amebocyte Lysate (LAL)<br>test on the final sterilized SyntheCel® device and found to be less than<br>0.06 EU/ml (Endotoxin Units / ml) per FDA's Guidance Document for<br>Dura Substitute Devices. | | | {1}------------------------------------------------ # @ SYNTHES® . : . {2}------------------------------------------------ Image /page/2/Picture/0 description: The image contains the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo. There is a horizontal line below the word. The logo has a registered trademark symbol. . . . and the comments of the comments of the comments of the comments of 1999 - 1999 - 1999 | Substantial<br>Equivalence to<br>Predicate Devices | Based on the information presented in this submission, the proposed<br>changes to SyntheCel® Dura Onlay do not raise new questions of safety<br>and effectiveness. Therefore, it can be concluded that the SyntheCel®<br>Dura Onlay is substantially equivalent to the predicate device. | |----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1. and the control control control control controllers and . 1 . and the same of the same of the same {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three curved lines resembling a stylized human figure. The symbol is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the upper portion of the circle. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G605 Silver Spring, MD 20993-0002 #### December 16, 2013 Synthes® (USA) % Heather Guerin, Ph.D., P.E. Senior Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, PA 19380 Re: K131792 > Trade/Device Name: SyntheCel® Dura Repair and SyntheCel® Dura Onlay Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXO Dated: November 13, 2013 Received: November 14, 2013 Dear Dr. Guerin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Heather Guerin, Ph.D., P.E. forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Joyce M. Whang -S for Carlos Peña, Ph.D. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K131792 Device Name: SyntheCel® Dura Repair and SyntheCel® Dura Onlay Indications For Use: ంగా విశాఖ ? SyntheCel® Dura Repair is intended for use as a dura replacement for the repair of dura mater in adults. SyntheCel® Dura Onlay is intended for use as a dura replacement for the repair of dura mater in adults. Prescription Use > (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health (CDRH) # Joyce M. Whang -S Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...