BIODESIGN ONLAY DURAL GRAFT OR BIODESIGN DURAPLASTY GRAFT
K131015 · Cook Biotech, Inc. · GXQ · Oct 8, 2013 · Neurology
Device Facts
Record ID
K131015
Device Name
BIODESIGN ONLAY DURAL GRAFT OR BIODESIGN DURAPLASTY GRAFT
Applicant
Cook Biotech, Inc.
Product Code
GXQ · Neurology
Decision Date
Oct 8, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5910
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Biodesign Onlay Dural Graft is intended for use as a dura substitute for the repair of dura mater. The device is provided sterile and intended for one-time use.
Device Story
Biodesign Onlay Dural Graft is a bioabsorbable, extracellular collagen matrix derived from porcine small intestinal submucosa (SIS). Implanted as a scaffold for dural repair, it facilitates infiltration by patient cells and remodeling into native tissue. Supplied in a lyophilized, sterile state, it can be placed without sutures during skull base dural repair. Used by neurosurgeons in clinical settings, the device serves as a physical barrier to replace or augment the dura mater, promoting healing and reducing potential for cerebrospinal fluid leaks.
Clinical Evidence
No clinical data provided in the document; evidence consists of bench testing, biocompatibility testing, and an animal study. Bench testing confirmed mechanical performance (suture retention, tensile, and burst strength) similar to predicates. Biocompatibility testing (genotoxicity, cytotoxicity, sensitization, systemic toxicity, etc.) on identical material (SurgiSIS Mesh) met ISO 10993-1 standards. Animal study in a rat model demonstrated adequate performance with no adverse physiological or histological effects.
Technological Characteristics
Material: Porcine small intestinal submucosa (SIS) containing Types I, III, IV, and VI collagen. Form factor: Lyophilized matrix, 154 µm to 680 µm thickness. Sterilization: Terminally sterilized. Biocompatibility: Tested per ISO 10993-1. Mechanical properties: Suture retention, ultimate tensile strength, and burst strength.
Indications for Use
Indicated for use as a dura substitute for the repair of dura mater in patients requiring cranial base dural repair.
Regulatory Classification
Identification
A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).
{0}------------------------------------------------
KB1015
OCT 0 8 2013
510(k) Premarket Notification: Biodesign® Dural Graft
#### 510(k) Summary
September 30, 2013
#### Cook Biotech Incorporated
#### Biodesign® Onlay Dural Graft
Manufacturer Name:
Cook Biotech Incorporated 1425 Innovation Place West Lafayette, Indiana 47906 Telephone: +1 (765) 497-3355 FAX: +1 (765) 807-7709
Official Contact:
Perry W. Guinn
DEVICE NAME AND CLASSIFICATION
Trade/Proprietary Name: Common Name: Classification Regulations: Biodesign Onlay Dural Graft Dura Substitute Class II, 21 CFR §882.5910 (84GXQ)
INDICATIONS FOR USE:
The Biodesign Onlay Dural Graft is intended for use as a dura substitute for the repair of dura mater. The device is provided sterile and intended for one-time use.
#### DEVICE DESCRIPTION:
The Biodesign Onlay Dural Graft is composed of a bioabsorbable, extracellular collagen matrix (Small Intestinal Submucosa, SIS). The SIS material that comprises the Biodesign Onlay Dural Graft is identical in composition and technology to that of its predicate Durasis Dural Substitute (K031850), also manufactured by Cook Biotech Incorporated and similar to its other predicate, DuraGen II Dural Regeneration Matrix (K043427) manufactured by Integra LifeSciences. The Biodesign Onlay Dural Graft is implanted as a dura substitute providing a scaffold which becomes infiltrated by the patient's cells and is remodeled into native tissue. The device is packaged in a lyophilized (dried) state and supplied sterile in a sealed double pouch system. The Biodesign Onlay Dural Graft can be placed without sutures during skull base dural repair.
#### EQUIVALENCE TO MARKETED DEVICES
The Biodesign Onlay Dural Graft is similar with respect to intended use, and identical with respect to materials and technological characteristics to the predicate devices in terms of section 510(k) substantial equivalence, as shown through bench (ultimate
{1}------------------------------------------------
510(k) Premarket Notification: Biodesign® Dural Graft
tensile strength, thickness, burst strength), animal (where the device was implanted as an onlay in a rat study), and biocompatibility testing (genotoxicity, direct contact in vitro hemolysis, cytotoxicity, muscle implantation, acute intracutaneous reactivity, sensitization, acute systemic toxicity, pyrogenicity, LAL endotoxins, subchronic systemic toxicity), and clinical studies using the device in cranial base dural repair.
#### Bench testing
The following mechanical tests were performed on finished, terminally sterilized devices:
- . Suture retention strength
- Ultimate tensile strength .
- Burst strength test .
The tests provided evidence that the Biodesign Onlay Dural Graft performed similarly to its predicate devices.
#### Biocompatibility testing
The following biocompatibility tests were performed on sterilized SurgiSIS Mesh, which is identical in composition to the Biodesign Onlay Dural Graft (according to the ISO 10993-1 standard):
- Genotoxicity �
- Direct contact in vitro hemolysis �
- Cytotoxicity .
- Muscle implantation .
- Acute intracutaneous reactivity .
- ISO Sensitization .
- Acute systemic toxicity .
- Pyrogenicity .
- LAL endotoxins .
- Subchronic systemic toxicity .
The results of these tests provided evidence that the Biodesign Onlay Dural Graft meets biocompatibility requirements of the ISO standard.
#### Animal Study
An animal study using finished devices in a rat model where the device was implanted using an onlay technique showed that the device performed adequately with no physiological, histological or functional evidence of adverse health effects were observed.
{2}------------------------------------------------
・
| Device | Biodesign® Onlay<br>Dural Graft | DuraGen II Dural<br>Regeneration Matrix | Durasis Dural<br>Substitute |
|------------------------|-------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Manufacturer | Cook Biotech<br>Incorporated | Integra LifeSciences | Cook Biotech<br>Incorporated |
| 510(k)<br>Number | K131015 | K043427 | K031850 |
| Indications for<br>Use | Intended for use as a<br>dura substitutefor<br>repair of dura mater. | DuraGen II Dural<br>Regeneration Matrix is<br>indicated as a dura<br>substitute for the repair<br>of dura mater.† | Intended for use as a dura<br>substitute for repairing<br>dura mater. |
| Material | Porcine small intestinal<br>submucosa<br>Primarily Types I, III,<br>IV and VI collagen | Bovine tendon collagen | Porcine small intestinal<br>submucosa<br>Primarily Types I, III, IV<br>and VI collagen. |
| Dimensions | 1 x 2 cm, 1 x 3 cm, 2 x<br>3 cm, 2.5 x 2.5 cm, 2.5<br>x 7.5 cm, 4 x 7 cm, 5 x<br>5 cm, 7.5 x 7.5 cm, 7 x<br>10 cm | 1 x 1 in to 5 x 7 in<br>(2.5 x 2.5 cm to 12.7 x<br>17.8 cm) | 2 x 2 cm to 10 x 20 cm |
| Thickness | 154 µm to 680 µm | 3.0 mm§ | 100 µm to 800 µm |
Table of Substantial Equivalence
† Cleared for use as an onlay.
\$ Zerris VA, James KS, et al. Repair of dura mater with processed collagen devices. J. Biomedical Materials research Part B: Applied Biomaterials. 2007; 83:580-588.
、
#### CONCLUSION:
The Biodesign® Onlay Dural Graft is substantially equivalent to its predicate devices in terms of physical characteristics and safety and effectiveness. .
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three stripes forming its body and wing. The eagle faces right and is positioned within a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
October 8, 2013
Cook Biotech Incorporated % Mr. Perry W. Guinn Vice President, Quality Assurance & Regulatory Affairs 1425 Innovation Place West Lafayette, IN 47906-1000
Re: K131015
> Trade/Device Name: Biodesign™ Onlay Dural Graft Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXQ Dated: September 5, 2013 Received: September 6. 2013
Dear Mr. Guinn:
We'have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (sec above) into cither class 11 (Special Controls) or class 111 (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Perry W. Guinn
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Division Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): K131015
Device Name: Biodesign® Onlay Dural Graft
Indications For Use:
The Biodesign Onlay Dural Graft is intended for use as a dura substitute for repair of dura mater. The device is provided sterile and is intended for one-time use.
> Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
## Joyce M. Whang -S
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.