OsteoVation Impact is a calcium phosphate bone void filler indicated for the repair or filling of neurosurgical burr holes, other cranialbone defects and craniotomy cuts with a surface area no larger than 25cm2. OsteoVation Impact may be used in the restoration or augmentation of bony contours of the cranialbone skeleton.
Device Story
OsteoVation Impact is a self-setting, biocompatible calcium phosphate bone void filler. It consists of two components: a powder (calcium phosphate, carboxymethylcellulose) and a liquid (sodium phosphate, sodium silicate), which are mixed by the user using an included bowl, pestle, and spatula. The resulting paste is moldable and impactable, intended for application by a surgeon to fill cranial defects or augment bone contours. The device is supplied in sterile convenience kits and sterilized via gamma radiation. It is a passive implantable material; it does not involve electronic processing, software, or clinical decision-support algorithms. The device provides a structural scaffold for bone repair.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including working time, setting strength, tensile strength, temperature profile, pH profile, FTIR analysis, and crystallographic analysis (XRD).
Technological Characteristics
Calcium phosphate bone void filler; self-setting paste. Components: calcium phosphate, sodium phosphate (SPMA), carboxymethylcellulose, sodium silicate. Sterilization: Gamma radiation (25-50 kGy; SAL 10^-9). Form factor: 3cc, 5cc, 10cc convenience kits. No energy source, no connectivity, no software.
Indications for Use
Indicated for repair or filling of neurosurgical burr holes, cranial bone defects, and craniotomy cuts (surface area ≤ 25cm2) and restoration/augmentation of cranial bone contours.
Regulatory Classification
Identification
Methyl methacrylate for cranioplasty (skull repair) is a self-curing acrylic that a surgeon uses to repair a skull defect in a patient. At the time of surgery, the surgeon initiates polymerization of the material and forms it into a plate or other appropriate shape to repair the defect.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2017
Skeletal Kinetics, LLC. Alicia Hemphill Director, Regulatory Affairs 10201 Bubb Road Cupertino, California 95014
Re: K162864
Trade/Device Name: OsteoVation Impact Regulation Number: 21 CFR 882.5300 Regulation Name: Methyl Methacrylate For Cranioplasty Regulatory Class: Class II Product Code: GXP Dated: October 12, 2016 Received: October 13, 2016
Dear Ms. Hemphill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
## Michael J. Hoffmann -S
Carlos L. Peña. PhD. MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162864
Device Name OsteoVation® Impact
Indications for Use (Describe)
Osteo Vation Impact is a calcium phosphate bone void filler indicated for the repair or filling of neurosurgical burr holes, other cranialbone defects and craniotomy cuts with a surface area no larger than 25cm2. OsteoVation Impact may be used in the restoration or augmentation of bony contours of the cranialbone skeleton.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Michael J. Hoffmanr
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Skeletal Kinetics. The logo consists of a blue globe-like icon on the left, followed by the text "Skeletal Kinetics" in a blue sans-serif font. Below the company name, in a smaller, lighter font, is the text "A COLSON ASSOCIATE".
## 510(k) SUMMARY
#### I. SUBMITTER
Skeletal Kinetics 10201 Bubb Road Cupertino, CA 95014 USA
Phone: 972-677-4783 Fax: 972-677-4601 Email: ahemphill@osteomed.com
Contact Person: Alicia Hemphill Date Prepared: January 12, 2017
Establishment Registration: 3003890476
#### II. DEVICE
Name of the Device: OsteoVation® Impact Common or Usual Name: Hydroxyapatite Cement Classification Name: Methyl Methacrylate For Cranioplasty Regulation: 882.3500 Regulatory Class: Class II Product Code: GXP
#### III. PREDICATE DEVICE
| 510(k) Number | Product<br>Code | Trade Name | Manufacturer |
|---------------|-----------------|------------------|--------------------|
| K051784 | GXP | OsteoVation® CMF | Skeletal Kinetics, |
| Primary | | Bone Void Filler | LLC. |
#### IV. DEVICE DESCRIPTION
OsteoVation Impact is a self-setting, calcium phosphate bone void filler designed for single use and biocompatible. Osteo Vation Impact is an impactable (moldable) calcium phosphate.
OsteoVation Impact is comprised of two working components: a calcium phosphate, a sodium phosphate (SPMA), carboxymethylcellulose powder, a sodium silicate, sodium phosphate (SPMA) solution, and a mixing system (mixing bowl, pestle and spatula). The two working components are packaged separately and are to be mixed together by the end user prior to implantation.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Skeletal Kinetics. The logo features a blue globe-like icon on the left, followed by the text "Skeletal Kinetics" in a blue, sans-serif font. Below the company name, there is a smaller line of text that reads "A COLSON ASSOCIATE".
OsteoVation Impact is offered in 3cc, 5cc, and 10cc sterile convenience kits. Each kit includes the two working components in combination with a mixing bowl, pestle, and spatula. It is sterilized using gamma radiation with a minimum dose of 25 kGy and a maximum dose of 50 kGy with a Sterility Assurance Level of 10-9.
#### INDICATIONS FOR USE V.
OsteoVation Impact is a calcium phosphate bone void filler indicated for the repair or filling of neurosurgical burr holes, other cranialbone defects and craniotomy cuts with a surface area no larger than 25cm2. OsteoVation Impact may be used in the restoration or augmentation of bony contours of the cranialbone skeleton.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH VI. THE PREDICATE DEVICE
The subject and predicate devices are classified as Methyl Methacrylate for Cranioplasty, and are intended for use in cranialbone applications. In establishing substantial equivalence to the predicate device, Skeletal Kinetics, LLC evaluated the indications for use, materials, technology, and product specifications of those products. Additionally, performance testing has been completed to demonstrate that the Osteo Vation Impact as a bone void filler in cranialbone applications is as safe, and effective, and performs as well as the legally marketed device. The performance testing and device comparison did not raise issues of safety or efficacy.
A formulation change is being made to the powder and liquid working components of the subject device in order optimize the utility of the device during implantation. Sodium phosphate (SPMA) is being added to the liquid component along with a ratio change. A ratio change is also being made to the powder component. This formulation change increases the room temperature range for the utility of the paste from 19℃ - 21℃ to 19℃ - 25℃. Verification testing of the working and setting time shows the formulation change does not affect the performance characteristics of the bone void filler.
The following technological differences exist between the subject and predicate device:
- Increase in room temperature range in Instructions for Use from 19°C -● 21°C to 19°C = 25°C
- Addition of sodium phosphate (SPMA) to liquid component ●
- Ratio change to powder component ●
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Skeletal Kinetics. The logo features a blue circular graphic with a network pattern on the left. To the right of the graphic is the text "Skeletal Kinetics" in a blue sans-serif font. Below the text is the phrase "A COLSON ASSOCIATE" in a smaller, lighter font.
#### PERFORMANCE DATA VII.
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility Testing
No biocompatibility studies were needed to demonstrate safety and efficacy.
## Bench Testing
The following bench testing was conducted:
- Working time and Setting Strength ●
- Tensile Strength ●
- Temperature Profile ●
- pH Profile ●
- FTIR analysis ●
- Crystallographic Analysis
## Test equipment
Osteovation Impact was used as indicated in Instructions for Use, Constant Temperature Water Bath (VWR, S. San Francisco, CA), Phosphate buffered saline (Sigma Chemical Co, St. Louis), Flat Plastic Spatula (Nunc, Rochester, NY), Vortexer (VWR, S. San Francisco, CA) , Mechanical Testing Machine with high load indentor (Instron, Canton, MA), Setting Test containers: Acetal (or other rigid plastic) block molds, pH meter and probe (Corning model 530 meter, model 576156 probe Corning N.Y.), Rigaku Ultima IV XRD, FTIR Spectrometer (Thermo Nicolet, Madison, WI).
| Test | Test Method Summary | Results |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Working time | Tests were performed to determine<br>the workability time of the cement.<br>Measurements included mixing<br>and molding of the paste. | All samples passed the targeted<br>working time and setting<br>strength.<br>The test results demonstrated<br>working time comparable to the<br>predicate device. |
| Setting Time | Testing was conducted to<br>determine the Setting strength<br>(Mean $\ge$ 450N and $\ge$ 700N) at<br>various time points in a target<br>temperature solution, post-<br>sterilization after a target working<br>time. | All samples passed the targeted<br>setting time.<br>The test results demonstrated<br>setting time comparable to the<br>predicate device. |
| Tensile Strength | Testing evaluated the subject | PASS |
| Test | Test Method Summary | Results |
| | device minimum tensile strength at<br>24 hours. | The test results demonstrated<br>tensile strength comparable to |
| | Samples were prepared through<br>the mortal/pestle combination into<br>molds in varying diameters and<br>thickness. Samples were placed in<br>a prepared solution at 37°C for 24<br>hours. Tensile strength testing is<br>conducted using a mechanical<br>testing system. | the predicate device. |
| Temperature<br>Profile | The subject device is mixed and<br>placed into a centrifuge tube and<br>into a 37°C water bath. The<br>temperature profile is recorded at<br>specified intervals using a<br>computer controlled thermocouple<br>for 24 hours. | PASS<br>The test results demonstrated<br>temperature profile comparable<br>to the predicate device. |
| pH Profile | The subject device is placed<br>mixed and placed into a centrifuge<br>tube and into a 37°C water bath.<br>The pH is recorded at specified<br>intervals using a calibrated probe<br>with a TC temperature<br>compensator. | PASS<br>The test results demonstrated<br>pH profile comparable to the<br>predicate device |
| | The test evaluates physiological<br>characteristics of the device. | |
| FTIR Analysis | Mix and incubate subject device<br>for 24 hours. Dry cured sample at<br>37°C for 72 hours. Perform 64<br>scans at typical resolution of 4cm-<br>1.<br>This test determines the formation | PASS<br>FTIR analysis show nearly<br>identical<br>hydroxyapatite<br>structure as the predicate device |
| | of hydroxyapatite. | |
| Crystallographic<br>Analysis | Analyze prepared sample using x-<br>ray diffractometer (XRD). Perform<br>Rietveld analysis determine ratios<br>of each compound in cured<br>cement. | PASS<br>XRD analysis show<br>comparable results as the<br>predicate device. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Skeletal Kinetics. The logo consists of a blue circular graphic to the left of the company name, "Skeletal Kinetics," which is also in blue. Below the company name, in smaller letters, it says "A COLSON ASSOCIATE."
## Skeletal Kinetics OsteoVation® Impact Submission January 12, 2017
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Skeletal Kinetics. The logo consists of a circular graphic to the left of the text "Skeletal Kinetics". Below the text is the phrase "A COLSON ASSOCIATE" in a smaller font size.
## Animal Study
No animal studies were needed to demonstrate safety and efficacy.
## Clinical Studies
No clinical studies were needed to demonstrate safety and efficacy.
#### CONCLUSIONS VIII.
The indication for use, technological characteristics, materials, technology, and product specifications are equivalent. The performance testing conducted, demonstrate that the OsteoVation Impact as a bone void filler in cranialbone applications is as safe, and effective, and performs as well as the legally marketed device and did not raise new issues of safety or efficacy.
Differences in the subject and predicate device formulation change increases the room temperature range for the utility of the paste. Verification testing of the working and setting time shows the formulation change does not affect the performance characteristics of the bone void filler
Additional non-clinical evaluations (tensile strength, temperature profile, pH profile, FTIR analysis, crystallographic analysis) demonstrate comparable results as the test performed for the predicate device.
The conclusions drawn from the nonclinical demonstrate that the device is as safe, and effective, and performs as well as the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.