← Product Code [GXP](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXP) · K043280

# MIMIX MP BONE VOID FILLER (K043280)

_Biomet, Inc. · GXP · Dec 21, 2004 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXP/K043280

## Device Facts

- **Applicant:** Biomet, Inc.
- **Product Code:** [GXP](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXP.md)
- **Decision Date:** Dec 21, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.5300
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

Mimix™ MP Bone Void Filler is a self-setting calcium phosphate cement indicated for the following craniofacial procedures: - Repair of neurosurgical burr holes 1. - Craniotomy cuts and other cranial defects 2. - Augmentation or restoration of bony contour in the craniofacial skeleton area no larger than 25cm².

## Device Story

Mimix™ MP Bone Void Filler is a self-setting calcium phosphate cement used for craniofacial reconstruction. The device consists of two components: a powder (ceramic calcium phosphate) and a liquid (anhydrous citric acid and water). These are mixed intraoperatively to form a homogenous paste. The material is biocompatible and apatitic. It is intended for use by surgeons in clinical settings to fill neurosurgical burr holes, craniotomy cuts, and other cranial defects, or to restore bony contours in the craniofacial skeleton. The paste sets in situ to provide structural support for bone voids. It is a prescription-only device.

## Clinical Evidence

No clinical data. Substantial equivalence established via non-clinical testing.

## Technological Characteristics

Self-setting calcium phosphate cement. Components: ceramic calcium phosphate powder and liquid solution of anhydrous citric acid and water. Apatitic material. Manual mixing process.

## Regulatory Identification

Methyl methacrylate for cranioplasty (skull repair) is a self-curing acrylic that a surgeon uses to repair a skull defect in a patient. At the time of surgery, the surgeon initiates polymerization of the material and forms it into a plate or other appropriate shape to repair the defect.

## Predicate Devices

- Mimix™ Bone Void Filler ([K990290](/device/K990290.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEC 21 2004

K043280

Image /page/0/Figure/2 description: The image shows the logo for Biomet Inc. The logo is in black and white and features the company name in a stylized font. The letters are bold and blocky, with a thick outline around each letter. The "INC" is smaller and stacked vertically on the right side of the logo.

# 510(k) Summary

Walter Lorenz Surgical, Inc. Applicant/Sponsor: (A wholly owned subsidiary of Biomet, Inc.) 1520 Tradeport Drive P.O. Box 18009 Jacksonville, FL 32229-8009 Establishment Registration Number: 1032347

- Kacy Arnold, RN, MBA Contact Person: Biomet Regulatory Specialist Telephone: 574-267-6639 Fax: 574-372-1683
Mimix™ MP Bone Void Filler Proprietary Name:

bone void filler Common Name:

Classification Name: Methyl Methacrylate for Cranioplasty (882.5300)

### Legally Marketed Devices To Which Substantial

Equivalence Is Claimed: Mimix™ Bone Void Filler K990290

Device Description: The Mimix™ MP Bone Void Filler is packaged as separate, pre-measured powder and liquid Device Desemptions The Films - Internatively to be nixed intracperatively to produce a homogenous paste, evinponents - The are contracts. Ber ause of its apatitic nature, the material is highly biocompatible.

The powder component is a mixture of a ceramic calcium phosphate powder and sodium (it at to be in The powaci component is a mixtare of a colution comprised of anhydrous citric acid (C,H,O,) and water (H,O).

### Indications for Use:

Mimix™ MP Bone Void Filler is a self-setting calcium phosphate cement indicated for the following craniofacial procedures

- Repair of neurosurgical burr holes 1.
- 2. Craniotomy cuts and other cranial defects
- Augmentation or restoration of bony contour in the craniofacial skeleton area no ကဲ larger than 25cm².

Summary of Technologies: Mimix™ MP is substantially equivalent in materials and processing to the predicate device.

Non-Clinical Testing: Non-clinical testing demonstrated substantial equivalence between this device and the predicate device.

Clinical Testing: Clinical testing was not used to establish substantial equivalence.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 2 1 2004

Ms. Kacy Arnold, RN, MBA Biomet Manufacturing Corp. 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46582

Re: K043280

Trade/Device Name: Mimix™ MP Bone Void Filler Regulation Number: 21 CFR 882.5300 Regulation Name: Methyl methacrylate for cranioplasty Regulatory Class: II Product Code: GXP Dated: November 24, 2004 Received: November 26, 2004

Dear Ms. Arnold:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

| 10(k) Number (if known): _____________________________________________________________________________________________________________________________________________________ |  |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|

#### Mimix™ MP Bone Void Filler Device Name:

## Indications For Use:

Mimix™ MP Bone Void Filler is a self-setting calcium phosphate cement indicated for the following craniofacial procedures:

- Repair of neurosurgical burr holes 1.
- Craniotomy cuts and other cranial defects 2.
- Craniotomy cuts and other cranial delects
Augmentation or restoration of bony contour in the craniofacial skeleton area no larger than 3. 25cm².

Miriam C. Provost

(Division Sign-Off) Division of General, Restorative. and Neurological Devices

510(k) Number_________________________________________________________________________________________________________________________________________________________________

Prescription Use __XX (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use __NO_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, O fice of Device Evaluation (ODE)

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXP/K043280](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GXP/K043280)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com