PEEK Patient Specific Cranial/Craniofacial Implant(PSCI)
K214109 · Kontour(Xi’An) Medical Technology Co., Ltd. · GXN · Oct 28, 2022 · Neurology
Device Facts
Record ID
K214109
Device Name
PEEK Patient Specific Cranial/Craniofacial Implant(PSCI)
Applicant
Kontour(Xi’An) Medical Technology Co., Ltd.
Product Code
GXN · Neurology
Decision Date
Oct 28, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5330
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PEEK Patient Specific Cranial/Craniofacial Implant(PSCI) is intended to replace bony voids in the cranial and /or craniofacial skeleton, excluding maxillofacial and /or oral regions.
Device Story
Patient-specific implant fabricated from medical-grade PEEK; designed using patient CT imaging data to match individual cranial/craniofacial bone defects. Device consists of a solid PEEK plate with holes; fixated to native bone using previously cleared titanium screws and plates. Provided non-sterile to healthcare facilities; requires sterilization before surgical implantation. Used by surgeons in hospital settings to restore skeletal structure and protect underlying brain/tissue. Benefits include anatomical fit and biocompatibility.
Clinical Evidence
No clinical data. Bench testing included biocompatibility (ISO 10993-1: cytotoxicity, sensitization, intracutaneous reactivity, acute systemic, material-mediated pyrogens, bone implantation), endotoxin testing (≤20 EU/device), particle testing (USP <788>), and performance evaluation (compression, fitting, and fixation).
Technological Characteristics
Material: Medical-grade PEEK. Design: Patient-specific, solid with holes. Fixation: Titanium screws and plates. Sterilization: Non-sterile (validated for facility sterilization). Connectivity: None. Software: None.
Indications for Use
Indicated for patients requiring replacement of bony voids in the cranial and/or craniofacial skeleton, excluding maxillofacial and/or oral regions.
Regulatory Classification
Identification
A preformed nonalterable cranioplasty plate is a device that is implanted in a patient to repair a skull defect and is constructed of a material, e.g., stainless steel or vitallium, that cannot be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 28, 2022
Kontour(Xi'an) Medical Technology Co., Ltd. % Joyce Yang Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square Nanshan District Shenzhen, Guangdong 518100, China
Re: K214109
Trade/Device Name: PEEK Patient Specific Craniofacial Implant(PSCI) Regulation Number: 21 CFR 882.5330 Regulation Name: Preformed Nonalterable Cranioplasty Plate Regulatory Class: Class II Product Code: GXN Dated: September 28, 2022 Received: September 29, 2022
Dear Joyce Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K214109
Device Name
PEEK Patient Specific Cranial/Craniofacial Implant (PSCI)
Indications for Use (Describe)
The PEEK Patient Specific Cranial Craniofacial Implant (PSCI) is intended to replace bony voids in the cranial and/or craniofacial skeleton, excluding maxillofacial and /or oral regions.
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Date of Summary prepared: October 28, 2022
# 1. Submission Sponsor
| Applicant Name | Kontour(Xi'an) Medical Technology Co., Ltd. |
|----------------|-----------------------------------------------------------------------------------------------------------|
| Address | No.6, Qinling Avenue, Caotang Technology<br>Industries Base, Xi'an High-tech Zone, 710304<br>Xi'an, China |
| Contact person | Rongrong Cong |
| Phone | 086-29-63364685 |
| E-mail | cong rongrong@kontmed.com |
# 2. Submission correspondent
| Name | Shenzhen Joyantech Consulting Co., Ltd |
|----------------|-----------------------------------------------------------------------------------|
| Address | 1713A, 17th Floor, Block A, Zhongguan Times<br>Square, Nanshan District, Shenzhen |
| Post Code | 518000 |
| Phone No. | +86-755-86069197 |
| Contact Person | Joyce Yang |
| Email | joyce@cefda.com |
# 3. Device Identification
| Type of 510(k) submission: | Traditional |
|----------------------------|----------------------------------------------------------|
| Trade Name: | PEEK Patient Specific Cranial/Craniofacial Implant(PSCI) |
| Model: | No specific model (Custom device) |
| Classification name: | Plate, Cranioplasty, Perfoemed, Non-alterable |
| Review Panel: | Neurology |
| Product Code: | GXN |
| Device Class: | II |
| Regulation Number: | 21 CFR § 882.5330 |
## 4. Legally Marketed Predicate Device
| Trade Name | Patient Contoured Implant - PEEK (PCI-PEEK) |
|-------------------|---------------------------------------------|
| Regulation number | 21 CFR § 882.5330 |
| Regulation class | II |
| Regulation name | Preformed nonalterable cranioplasty plate |
| 510(k) Number | K151382 |
| Product Code | GXN |
| Manufacturer | KLS-MARTIN L.P. |
{4}------------------------------------------------
# 5. Device Description
The PEEK Patient Specific Cranial/Craniofacial Implants are designed individually for each patient to correct defects in cranial/facial bone, and fabricated using the patient' s CT imaging data. The implants are made of PEEK material which is medical grade material. The PSCI consists of PEEK bone plate, and it is fixated to native bone using previously cleared Bioplate Fixation System manufactured by Bioplate, Inc.
The PSCI is provided in non-sterile and must be sterilized before use in Health Care Facility.
# 6. Intended Use/ Indications for Use
The PEEK Patient Specific Cranial/Craniofacial Implant(PSCI) is intended to replace bony voids in the cranial and /or craniofacial skeleton, excluding maxillofacial and /or oral regions.
# 7. Technological characteristics Comparison
| .Compariso<br>n item | Subject Device:<br>PEEK Patient Specific<br>Cranial/Craniofacial Implant | Predicate Device:<br>Patient Contoured Implant-PEEK<br>(K151382) | Comments |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|------------------------|
| Indication for<br>Use | The PEEK Patient Specific<br>Cranial/Craniofacial<br>Implant(PSCI) is intended to<br>replace bony voids in the<br>cranial and /or craniofacial<br>skeleton, excluding<br>maxillofacial and /or oral<br>regions. | The Patient Contoured Implant -<br>PEEK (PCI-PEEK) is intended to<br>replace bony voids in the cranial<br>skeleton. | Same |
| Anatomical<br>Sites | Cranial or Craniofacial | Cranial | Similar<br>(Issue 1) |
| Material | PEEK | PEEK | Same |
| Style | Solid with holes | Solid or solid with holes | Same |
| Sterilization | Provided non-sterile (The PSCI<br>is provided non-sterile and must<br>be sterilized before use) | Provided Sterile (Gamma) | Different<br>(Issue 2) |
| Design and<br>manufacture | Customized device, designed<br>and manufactured in accordance | Customized device, designed<br>and manufactured in accordance | Same |
## Table 1: General Comparison
{5}------------------------------------------------
| .Compariso<br>n item | Subject Device:<br>PEEK Patient Specific<br>Cranial/Craniofacial Implant | Predicate Device:<br>Patient Contoured Implant-PEEK<br>(K151382) | Comment<br>s |
|----------------------------------------------------------|--------------------------------------------------------------------------|------------------------------------------------------------------|--------------|
| | with patient's CT-scan data | with patient's CT-scan data | |
| Method of<br>Fixation | Titanium Screws and Titanium<br>Plates | Titanium Screws and Titanium<br>Plates | Same |
| Where used<br>(hospital,<br>home,<br>ambulance,e<br>tc.) | Health Care Facilities / Hospitals | Health Care Facilities / Hospitals | Same |
lssue 1: The difference between a implant used to replace Cranial bone and a implant used to replace craniofacial bone is that the shape of the implant. The implant and its contact with the human body are the same. So this difference would not raise new issue of safety or effectiveness.
lssue 2: The subject device is provided in non-sterilized before use in Health Care Facility. The sterilization has been validated. And many similar devices are also provided in this way, such as Patient Contoured Mesh - PEEK (K072707).
### 8. Summary of Non-clinical Testing
The biocompatibility evaluation for the PEEK Patient Specific Cranial/Craniofacial Implant (PSCI) was conducted in accordance with the 2020 FDA quidance document Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process." and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" as recognized by FDA.
The evaluation tests of the PSCI contain:
- In Vitro Cytotoxicity Test
- Skin Sensitization Test
- Intracutaneous Reactivity Test
- Acute Systemic Test
- Material-mediated Pyrogens
- Bone Implantation Test
{6}------------------------------------------------
● Extractable Assessment
The LAL endotoxin testing demonstrates that the sterile PEEK implant conforms to the required ≤20 EU/device.
The Particle testing has been conducted in accordance with USP<788>.
Based on the intended use of the subject device, the performance evaluation for PEEK Patient Specific Cranial/Craniofacial Implant (PSCI) was conducted in compression, fitting, and fixation.
#### 9. Brief discussion of clinical tests
No clinical tests were performed.
#### 10.Conclusions
The subject device, PEEK Patient Specific Cranial/Craniofacial Implant has the same intended use, design, function, and is composed of the same material as the predicate device, Patient Contoured Implant-PEEK. The subject device differs from the predicate device in that it will be provided in non-sterile and must be sterilized before use in Health Care Facility. This difference does not raise new issues of safety or effectiveness. Based on the comparison and analysis above, the subject device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.