← Product Code [GWO](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GWO) · K974206

# SYNTHES CRANIAL SPRING CLIP (CSC) (K974206)

_Synthes (Usa) · GWO · Mar 16, 1998 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GWO/K974206

## Device Facts

- **Applicant:** Synthes (Usa)
- **Product Code:** [GWO](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GWO.md)
- **Decision Date:** Mar 16, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5320
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

Synthes CSC is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.

## Device Story

Synthes Cranial Spring Clip (CSC) is a mechanical fixation device for reattaching cranial bone flaps post-craniotomy. Device features T-profile with two semi-circular tabs (0.2 mm thick) and two spring-like projections designed to lodge in the diploe of surrounding cranial bone. Surgeons secure the clip using 1.5 mm Ti Alloy Bone Screws inserted laterally into the diploe. Device is single-use, MRI compatible, and low-profile (4.2 mm height). Used in clinical/surgical settings by neurosurgeons to stabilize bone flaps, facilitating healing and structural integrity of the cranium.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Material: Titanium alloy. Dimensions: 11.5 mm length, 9.0 mm width, 4.2 mm height. Fixation: 1.5 mm Ti Alloy Bone Screws. Design: T-profile with spring-loaded projections for diploe engagement. Single-use. MRI compatible.

## Regulatory Identification

A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.

## Predicate Devices

- Synthes Midfacial System ([K953806](/device/K953806.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. To the right of the word is a small circle with an R inside, indicating a registered trademark.

・

Summary of Safety and Effectiveness Information [510(k) Summary]

MAR | 6 1998

2974206

| SUBMITTER                | Synthes (USA)<br>1690 Russell Road<br>Paoli, PA 19301<br>(610) 647-9700<br>Contact: Sheri L. Musgnung                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| COMMON OR USUAL<br>NAME  | Plate, Cranioplasty, Preformed, Alterable; Fastener, Plate, Cranioplasty                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| DEVICE<br>CLASSIFICATION | Class II, 21 CFR 882.5320 and 882.5360                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| PREDICATE DEVICE         | Synthes Midfacial System (K953806)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| DESCRIPTION              | Synthes Cranial Spring Clip (CSC) is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure. The CSC has a T-profile, which consists of two semi-circular tabs, and two projections that are bent into springs to lodge in the diploe of the cranial bone surrounding the osteotomy. The two tabs are 0.2 mm thick and rest on the superior surfaces of the cranial flap and the surrounding cranial bone. The CSC has a low profile, is 11.5 mm in length, 9.0 mm in width, and is 4.2 mm in height (0.2 mm above the cranial surface). 1.5 mm Ti Alloy Bone Screws will be inserted laterally through the CSC and into the diploe. The CSC is for single use only. |
| INTENDED USE             | Synthes CSC is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

The CSC is contraindicated in the following situations: in patients
with a skull thickness less than 4 mm; or when the gap surrounding
the bone flap is greater than 2 mm.

The CSC is MRI compatible.

{1}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 6 1998

Ms. Sheri L. Musgnung Regulatory Affairs Associate Synthes USA 1690 Russell Road P.O. Box 1766 Paoli, Pennsylvania 19301

Re: K974206 Synthes (USA) Cranial Spring Clip Regulatory Class: II Product Codes: GWO and HBW Dated: February 13, 1998 Received: February 17, 1998

Dear Ms. Musgnung:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895 ... A ................................................................................ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

Image /page/1/Picture/9 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of an eagle with its wings spread.

{2}------------------------------------------------

## Page 2 - Ms. Sheri L. Musgnung

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. To the right of the word is a small circle with a capital R inside, indicating a registered trademark.

Indications for Use Statement:

K974206 510(k) Number (if known):

Synthes (USA) Cranial Spring Clip (CSC) Device Name: ...

Indications For Use:

Synthes CSC is intended to reattach a cranial bone flap to the surrounding Cranium after a craniotomy procedure.

The CSC is contraindicated in the following situations: in patients with a skull thickness less than 4 mm; or when the gap surrounding the bone flap is greater than 2 mm.

The CSC is MRI Compatible.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

signature

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number. K974206

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GWO/K974206](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/GWO/K974206)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
