Peer-reviewed clinical literature; Retrospective clinical cohorts; Institutional medical records
The sponsor provided a literature review of clinical studies to demonstrate the safety and effectiveness of alloplastic cranioplasty implants in pediatric patients (3.5 years and older), specifically addressing concerns regarding growth and long-term outcomes.
Fu et al. 2016; Retrospective outcomes comparison; Follow-up/Duration: Up to 9 years
Pediatric patients receiving alloplastic vs. autologous cranioplasty
Autologous reconstructions
Safety of alloplastic cranioplasty
Indications for Use
The PEEK Customized Cranial Implant Kit is indicated for the augmentation and/or restoration of bony and/or soft tissue deformities in the cranial and craniofacial skeleton (orbital rim, zygoma, and adjacent bone); including but not limited to, the correction and prevention of persistent temporal hollowing (PTH) in patients 3.5 years of age and older.
Device Story
Device consists of patient-specific PEEK cranial/craniofacial implants; manufactured based on patient CT data and surgeon input. Used to fill bony voids, defects, and contour irregularities in non-load bearing regions of cranial/craniofacial skeleton. Intended for pediatric patients (3.5+ years) and adults. Surgeon uses device to restore anatomy; benefits include improved symmetry and protection of brain. Risks include potential growth-related issues (e.g., brain compression, implant loosening) and radiation exposure from required CT scans. Mitigation strategies include specific labeling regarding growth disturbance and use of low-dose/child-size CT protocols.
Clinical Evidence
No new clinical trials conducted. Evidence provided via literature review of alloplastic cranioplasty in pediatric patients (3.5+ years). Studies (Nguyen 2018, Rocque 2018, Ma 2018, Sainsbury 2017, Fu 2016) demonstrate safety and efficacy of PEEK and other alloplastic implants for cranial reconstruction, showing good symmetry, effective brain protection, and no adverse impact on growing skeletons.
Technological Characteristics
Material: PEEK (Polyetheretherketone). Principle: Patient-specific, customized implant for non-load bearing cranial/craniofacial defects. Form factor: Customized to patient anatomy via CT data. Connectivity: None. Sterilization: Not specified (identical to predicate).
Indications for Use
Indicated for augmentation/restoration of bony/soft tissue deformities in cranial/craniofacial skeleton (orbital rim, zygoma, adjacent bone), including correction/prevention of persistent temporal hollowing (PTH) in patients 3.5 years and older.
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 11, 2019
Stryker Zainab Amini Regulatory Affairs Specialist 750 Trade Centre Way - Suite 200 Portage, Michigan 49002
Re: K190229
Trade/Device Name: Stryker PEEK Customized Cranial Implant Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO Dated: June 7, 2019 Received: June 10, 2019
Dear Zainab Amini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Matthew Krueger Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K190229
Device Name
Stryker PEEK Customized Cranial Implant Kit
#### Indications for Use (Describe)
The PEEK Customized Cranial Implant Kit is indicated for the augmentation of bony and/or soft tissue deformities in the cranial and craniofacial skeleton (orbital rim, zygoma, and adjacent bone); including but not limited to, the correction and prevention of persistent temporal hollowing (PTH) in patients 3.5 years of age and older.
Type of Use (Select one or both, as applicable)
| <span> <svg height="1em" width="1em"><path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm3.354 4.354a.5.5 0 0 1 .146.354v3.792l1.146-1.146a.5.5 0 0 1 .708 0l3 3a.5.5 0 0 1 .708 0l1.5-1.5a.5.5 0 0 1 .708 0l.146.146V4.354a.5.5 0 0 1 .146-.354H5.354z" fill="black"></path></svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K190229 stryker
# Section 5. 510(k) Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
#### SUBMITTER I.
| 510(k) Owner: | Stryker Leibinger GmbH& Co. KG<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Germany |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter/ Contact<br>Person: | Zainab Amini<br>Regulatory Affairs Specialist<br>Stryker Craniomaxillofacial<br>750 Trade Centre Way<br>Portage, MI 49002<br>Phone: 269-389-8349<br>Fax: 877-648-7114 |
| Date prepared: | July 10, 2019 |
| II.<br>DEVICE | |
| Trade Name: | Stryker PEEK Customized Cranial Implant |
| Common or Usual<br>name: | Customized Cranial Implant |
| Classification<br>name: | Preformed alterable cranioplasty plate 21 CFR §882.5320 |
| Regulatory Class: | Class II |
| Product Code: | GWO |
#### PREDICATE DEVICE III.
Predicate: K153248, Stryker PEEK Customized Cranial Implant Kit
This predicate has not been subject to a design-related recall.
# IV. DEVICE DESCRIPTION
The Stryker PEEK Customized Cranial Implant (CCI) product offerings provide customized cranial or craniofacial patient specific implants based on CT data and surgeon input.
{4}------------------------------------------------
#### V. INDICATIONS FOR USE
The PEEK Customized Cranial Implant Kit is indicated for the augmentation and/or restoration of bony and/or soft tissue deformities in the cranial and craniofacial skeleton (orbital rim, zygoma, and adjacent bone); including but not limited to, the correction and prevention of persistent temporal hollowing (PTH) in patients 3.5 years of age and older.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The PEEK CCI for Children design is compared to its Predicate Device for substantial equivalence based on the following criteria:
- A. Principle of Operation
- B. Technological Characteristics
# A. Principle of Operation
The basic operational principle of the PEEK CCI for Children is that the PEEK Customized Cranial Implant is intended to fill bony voids, defects, and contour irregularities in non-load bearing regions of the cranial and craniofacial skeleton.
### B. Technological and Operational Characteristics
Both the Subject and Predicate Devices have similar technological characteristics, and the operating principle, mode of fixation, and design and manufacturing processes are all identical.
# VII. PERFORMANCE DATA
# Biocompatibility Testing
Biocompatibility and sterility testing of the device is not required as a basis for substantial equivalence. There is no change in the material, duration of contact, or reprocessing methods for the PEEK CCI. As the design and manufacturing processes, materials, and packaging processes are identical for both the Predicate and the Subject Devices.
### Performance Bench Testing
Both the Subject and Predicate Devices are designed and manufactured identically. As the Performance Bench testing of the Predicate Device are valid for the Subject Device, and because the performance bench testing of the both devices are identical, an additional performance bench testing was not required for the Subject Device as a basis for substantial equivalence.
### Animal Testing
Animal testing was not required as a basis for substantial equivalence.
{5}------------------------------------------------
### Clinical Testing
Clinical studies and findings show the potential benefits associated with the use of the device outweigh the potential risks associated with the Subject Device. Clinical studies provided within this submission indicates that the potential benefits for the patient will be greater from the use of the Subject Device when compared to the potential risks. The following table provides the scientific peer-reviewed publications describing the safe and effective application of alloplastic cranioplasties in patient 3.5 years of age and above.
| Source | Summary |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nguyen et al.<br>2018 | Report on clinical outcomes following cranioplasty in<br>children with patient-specific PEEK, titanium, and PMMA<br>implants. Follow-up of an average of 30 months. Patient-<br>specific implants provide significant degrees of freedom to<br>the surgeon which bone grafting would not be able to<br>accommodate because of lack of donor availability or sheer<br>complexity of contouring. The authors did not find any<br>indications that the implants pose a problem with the<br>growing skeleton and suggested safety at ages greater than<br>3 years. |
| Rocque et al.<br>2018 | Multicenter study at 13 institutions. Cranioplasty bone<br>resorption is predicted by patient age. Reconstruction with<br>PEEK implants is described in the study. |
| Ma et al.<br>2018 | Cranioplasty with autologous bone in 84 cases, 36 split<br>calvarial graft reconstructions, six PEEK implants, 33<br>titanium meshes. Average follow-up: 4 years. Good<br>symmetry and effective brain protection were achieved. |
| Sainsbury et<br>al. 2017 | Bilateral malar reconstruction with customized PEEK<br>implants based on CT data. The authors highlight the<br>improved implant-bone surface interface with<br>the<br>customized medical devices. |
| Fu et al. 2016 | Comparison of clinical outcomes achieved with autologous<br>and alloplastic cranioplasty in pediatric patients. 73% of the<br>included patients received alloplastic implants (PEEK,<br>PMMA, MEDPOR, titanium) while the remaining 28%<br>were treated with autologous reconstructions. Patients were<br>followed-up for up to 9 years. The authors conclude that<br>alloplastic cranioplasty in the pediatric population is a safe<br>alternative, when autologous cranial bone is not available. |
#### TABLE 1 RELEVANT CLINICAL DATA
### Potential Benefits
Potential benefits associated with the Subject Device outweigh the potential risks. Overall, the Subject Device implant offers similar benefits as the Predicate Device because the physical principle of replacing missing bone is identical. However, through investigation it was identified that the Subject Device's targeted age group would have many potential benefits that are of greater importance/magnitude for young patients.
{6}------------------------------------------------
#### Potential Risks
The potential risks associated with the device are increasing radiation exposure and growth issues. These risks have been assessed and mitigation is provided within this submission which reduces these potential risks for the patient.
| | Potential Hazard | Potential Harm | Risk Mitigation in Labelling |
|----------------------|-------------------------------------------|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Growth related risks | Brain compression<br>against implant | Pain<br>Removal of implant<br>Brain/soft tissue injury<br>reversible<br>Brain/soft tissue injury<br>irreversible | IFU contains information<br>regarding growth disturbance |
| | Implant interference<br>with growing bone | Poor aesthetics<br>resulting in implant<br>removal<br>Annoyance | IFU contains information<br>regarding growth disturbance |
| | Loosening of<br>implant | Neurological symptoms<br>requiring surgery<br>Temporary neurological<br>symptoms | IFU contains information<br>regarding growth disturbance |
| | Radiation related risks | New / additional<br>scan for implant<br>design required | Radiation induced<br>neoplasia |
TABLE 2 POTENTIAL RISKS IDENTIFIED FOR THE SUBJECT DEVICE:
To mitigate the risks, the Instructions for Use (IFU) includes information associated with growth related risks. The IFU and the scan protocol include information to minimize ionizing radiation dose by using low-dose and child-size CT- protocols when appropriate.
# CONCLUSIONS
The clinical studies provided within this submission, show the safe and effective use of the patient specific alloplastic implants for the intended patient population. Table 1 provides references and summary of these clinical studies.
A Benefit and Risk Factor analysis was conducted for the Subject Device. Both the potential benefits and risks of the Subject Device are increased when compared to the Predicate Device. However, the overall benefit and risk ratio remains equivalent. The product labeling contains corresponding information regarding the potential risks of growth disturbance and radiation exposure for the intended patient population. Table 2 provides an overview of the risk mitigation measures.
{7}------------------------------------------------
The safety and effectiveness of both the Subject and Predicate Devices are equivalent. Both devices are designed and manufactured identically, and they both have equivalent Intended Use and Technological Characteristics.
In conclusion, the potential benefits associated with the use of the Subject Device outweigh its potential risks for the patient. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
### References
- Fu, K. J., Barr, R. M., Kerr, M. L., Shah, M. N., Fletcher, S. A., Sandberg, D. I., ... Greives, M. R. (2016). An outcomes comparison between autologous and alloplastic cranioplasty in the pediatric population. Journal of Craniofacial Surgerv, 27(3), 593-597. https://doi.org/10.1097/SCS.0000000000000000002491
- Ma, I. T., Symon, M. R., Bristol, R. E., Beals, S. P., Joganic, E. F., Adelson, P. D., ... Singh, D. J. (2018). Outcomes of Titanium Mesh Cranioplasty in Pediatric Patients. The Journal of Craniofacial Surgery, 29(1), 99-104. https://doi.org/10.1097/SCS.0000000000000004045
- Nguyen, P. D., Khechoyan, D. Y., Phillips, J. H., & Forrest, C. R. (2018). Custom CAD/CAM implants for complex craniofacial reconstruction in children: Our experience based on 136 cases. Journal of Plastic, Reconstructive and Aesthetic Surgery, 18(30251-1), 1-9. https://doi.org/10.1016/j.bjps.2018.07.016
- Rocque, B. G., Agee, B. S., Thompson, E. M., Piedra, M., Baird, L. C., Selden, N. R., ... Lam, S. K. (2018). Complications following pediatric cranioplasty after decompressive craniectomy: a multicenter retrospective study. Journal of Neurosurgery: Pediatrics, 22(3), 225-232. https://doi.org/10.3171/2018.3.PEDS17234.
- Sainsbury, D. C. G., George, A., Forrest, C. R., & Phillips, J. H. (2017). Bilateral Malar Reconstruction Using Patient-Specific Polyether Ether Ketone Implants in Treacher-Collins Syndrome Patients With Absent Zygomas. The Journal of Cramiofacial Surgery, 28(2), 515-517. https://doi.org/10.1097/SCS.0000000000003351
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.