Stryker Universal Mesh

K161821 · Stryker · GWO · Dec 8, 2016 · Neurology

Device Facts

Record IDK161821
Device NameStryker Universal Mesh
ApplicantStryker
Product CodeGWO · Neurology
Decision DateDec 8, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5320
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Universal Mesh is indicated for reconstruction, stabilization subsequent to craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12 and higher).

Device Story

Stryker Universal Mesh consists of titanium mesh implants (1.2mm, 1.5mm, 1.7mm) and associated manual instrumentation (cutters, benders, clippers). Implants are malleable for 3D anatomical adaptation and modified via trimming/bending. Used in cranial surgery to provide structural support and stabilization; fixated to bone using previously marketed Stryker CMF screws. System includes storage containers. Operated by surgeons in clinical/OR settings. Output is a physical implant providing rigid fixation; aids in bone healing and structural integrity post-cranial procedures. Non-sterile; single-use implants; reusable instruments.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing: biocompatibility (ISO 10993), cleaning/sterilization validation (ISO 17664, 14937, 17665), mechanical performance (static/dynamic load), corrosion testing, and MR compatibility (per FDA guidance).

Technological Characteristics

Material: Commercially pure titanium (ASTM F67). Form factor: 2D mesh (0.3mm-0.8mm thickness, 40x40mm to 200x200mm). Fixation: 1.2mm-1.7mm screws. Sterilization: End-user sterilized. Connectivity: None. Software: None.

Indications for Use

Indicated for reconstruction, stabilization, and rigid fixation following craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12+).

Regulatory Classification

Identification

A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, with three human profiles forming the staff and a cloth draped around the base. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 8, 2016 Stryker Craniomaxillofacial Mr. Jonathan Schell Sr. Regulatory Affairs Specialist 750 Trade Centre Way, Suite 200 Portage, MI 49002 Re: K161821 Trade/Device Name: Stryker Universal Mesh Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO Dated: November 4, 2016 Received: November 7, 2016 Dear Mr. Schell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Michael J. Hoffmann -S Carlos L. Peña. PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K161821 Device Name Stryker Universal Mesh Indications for Use (Describe) The Universal Mesh is indicated for reconstruction, stabilization subsequent to craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12 and higher). | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| | | |X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92. - SUBMITTER [§807.92(a)(1)] I. | 510(k) Owner: | Stryker Leibinger GmbH & Co. KG | |---------------|---------------------------------| | | Boetzinger Strasse 41 | | | 79111 Freiburg, Germany | - Submitter/ Contact Jonathan Schell Person: Sr. Regulatory Affairs Specialist Stryker Craniomaxillofacial 750 Trade Centre Way Portage, MI 49002 Phone: 269-389-5596 Fax: 877-648-7114 - Date prepared: November 4, 2016 #### II. DEVICE [§807.92(a)(2)] | Trade Name: | Stryker Universal Mesh | |--------------------------|---------------------------------------------------------| | Common or Usual<br>name: | Bone Plate | | Classification<br>name: | Preformed Alterable Cranioplasty Plate; 21 CFR 882.5320 | | Regulatory Class: | Class II | | Product Code: | GWO | | Classification<br>Panel: | Neurology | #### PREDICATE DEVICE [§807.92(a)(3)] III. Universal CMF System – K022185 {4}------------------------------------------------ #### IV. DEVICE DESCRIPTION [§807.92(a)(4)] The Stryker Universal Mesh consists of mesh implants in different shapes, thicknesses and perforation patterns which include the 1.2mm and 1.5/ 1.7mm Dynamic Mesh implants and Hybrid Mesh implants. The meshes can be modified with trimming and bending instruments for an anatomical fit and are malleable for three-dimensional adaptations. The necessary instrumentation for mesh modification, Mesh Cutter (coarse), Mesh Step Bender, and the Mesh Clipper are included with the system. The Stryker Universal Mesh implants are provided nonsterile and are for single use only. The instrumentation and storage system is reusable. Also, the mesh can be fixated to the bone with screws available in previously marketed Stryker CMF implant systems. The previously marketed screws are not specifically packaged with the subject device, but the implant tray of the Universal Mesh Storage system contains space to store the screw disc if that option is chosen. #### INDICATIONS FOR USE [§807.92(a)(5 )] V. | | Subject Device | Predicate - K022185 | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | The Universal Mesh is indicated for<br>reconstruction, stabilization and/or<br>rigid fixation subsequent to<br>craniotomy, craniectomy,<br>cranioplasty, and cranial fractures in<br>adults and adolescents (age 12 and<br>higher). | The Stryker Leibinger Universal<br>CMF System is a Cranio-<br>maxillofacial (CMF) plate and screw<br>system intended for osteotomy,<br>stabilization and rigid fixation of<br>CMF fractures and reconstruction. | ## TABLE 1: COMPARISON OF INDICATIONS FOR USE The predicate device is indicated for use in a range of anatomic areas (cranial, craniofacial skeleton, mid-face, maxilla, mandible) for a variety of surgical applications (craniofacial surgery, reconstructive procedures). The Indications for Use statement of the Stryker Universal Mesh falls within the scope of the broader Indications statement of the predicate device. Additionally, the Subject device is intended to be used in adolescent patients 12 years of age and older. The differences in the Indications statement for the proposed device in comparison to the predicate does not constitute a new intended use. #### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE [§807.92(a)(6)] {5}------------------------------------------------ The Stryker Universal Mesh is compared to its predicate devices for substantial equivalence based on the following criteria: - A. Principle of Operation - B. Technological Characteristics ## A. Principle of Operation The Stryker Universal Mesh implants are designed for the reconstruction, stabilization and rigid fixation of the bony areas of the cranial skeleton. The necessary instrumentation and product storage are included with the system. ## B. Technological and Operational Characteristics -Substantial Equivalence of the Subject devices and Predicate devices are based on the following technological elements: | Feature | Stryker Universal Mesh | Universal CMF System<br>K022185 | Explanation of Differences | |--------------------------------------------------------|-----------------------------------|---------------------------------|----------------------------| | | Subject Device | Predicate Device | | | Sterility: End<br>User Sterilized | X | X | Same | | Sizing | | | | | Shape | Square, partly with rounded edges | Square | Similar | | Sizes | 40x40mm<br>to<br>200x200mm | 40x40mm<br>to<br>200x200m | Similar | | Implant<br>Thickness | 0.3mm -<br>0.8mm | 0.3mm -<br>0.6mm | Similar | | Design | | | | | 2-dimensional<br>(2D) | X | X | Same | | Mode of<br>Fixation | | | | | 1.2mm - 1.7mm<br>screws | X | X | Same | | Material | | | | | Titanium:<br>Commercially<br>pure titanium<br>ASTM F67 | X | X | Same | Overall these changes do not alter the Subject device significantly and it does not represent a new worst case in design. Strvk {6}------------------------------------------------ ## VII. PERFORMANCE DATA [§807.92(b)(1)] The following performance data were provided in support of the substantial equivalence determination. ## Biocompatibility Testing The biocompatibility evaluation for the Stryker Universal Mesh system was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The tests supported the biocompatibility of the device. The implant is made from titanium which conforms to ASTM F67 for chemical composition, the same as the predicate. Cytotoxicity testing was performed using DIN EN 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity. Chemical residual analysis was conducted according to validated methods with results that met the established acceptance criteria. These tests support the biocompatibility of the device. ## Performance Testing | Test | Test Method Summary | Results | |--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------| | Life time<br>reprocessing | Verify the reprocessing life of the Subject device, instruments, and storage<br>container are according to internal standard. | Internal Acceptance<br>Criteria | | Cleaning<br>Validation | Verify the cleaning process for Subject device, instruments and storage<br>container are according to recommendations in ISO 17664, AAMI TIR 30,<br>and AAMI TIR 12. | Method validated. | | Sterilization<br>Validation | Validate the sterilization instructions for the Subject device, instruments,<br>and storage container are according to internal standard. Referenced<br>standards: ISO 14937, ISO 17664, ISO 17665 - 1&2, AAMI TIR 12,<br>AAMI ST 81, AAMI ST79. | Method validated. | | Mechanical<br>performance<br>Testing | Static Load Test and Dynamic Load Test | Acceptance criterion<br>were met, tests were<br>passed. | | Corrosion<br>Testing | Verify corrosion behavior of Subject device, instruments, and storage<br>device. | Acceptance criterion<br>was met, test was<br>passed. | | MR<br>Compatibility | Magnetic Resonance (MR) testing was performed on the subject device<br>mesh implants in accordance with FDA draft guidance for industry<br>"Assessment of Radiofrequency - Induced Heating in the Magnetic<br>Resonance (MR) Environment for Multi-Configuration Passive Medical | Based on the results of<br>these evaluations, the<br>Stryker Universal mesh<br>will be labeled as MR | The following performance tests were completed: {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is black and appears to be a logo. A small circled letter 'c' is present to the upper right of the word. | Devices," June 29, 2015. MR heat testing has been performed on the<br>Subject device according to FDA final guidance "Establishing Safety and<br>Compatibility of Passive Implants in the Magnetic Resonance (MR)<br>Environment," December 11, 2014. | conditional. | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| The Subject device met all pre-defined acceptance criteria and standards and was found to not represent a new worst case. Overall, the results of the performance bench tests support the substantial equivalence of the Subject device. ## Animal Testing Animal testing was not required as a basis for substantial equivalence. ## Clinical Testing [§807.92(b)(2)] Clinical testing was not required as a basis for substantial equivalence. ## VIII. CONCLUSIONS [§807.92(b)(3)] The results of the non-clinical data demonstrate the Stryker Universal Mesh will perform as intended in the specified use conditions. According to the comparison based on the requirements of 21 CFR 807.92 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...