Biopor Porous Polyethylene Implants in sheet configurations are intended for the augmentation or reconstruction of the craniofacial skeleton, including the cranial skeleton, orbit, nasal bones and the zygoma.
Device Story
Biopor Porous Polyethylene Implants are porous high-density polyethylene (PPE) surgical implants; available in sheet configurations; optional water-soluble alkylene oxide copolymer (AOC) coating. Implants are contoured or carved by surgeons to meet patient-specific anatomical requirements for craniofacial augmentation or reconstruction. Provided sterile; intended for single use. Device functions as a structural scaffold for tissue ingrowth. Clinical benefit includes restoration of skeletal form and function in the craniofacial region.
Clinical Evidence
Bench testing only. Biocompatibility testing included cytotoxicity, LLNA sensitization, intracutaneous irritation, systemic toxicity, muscle implantation, and genotoxicity (Ames, chromosomal aberration, micronucleus). Performance testing included suture pull-out strength, flexibility, and porosity characterization (>40 µm). All tests met acceptance criteria.
Technological Characteristics
Material: Porous high-density polyethylene (PPE) with optional water-soluble alkylene oxide copolymer (AOC) coating. Form factor: Sheets. Sterilization: Electron beam irradiation. Porosity: >40 µm pore size. Mechanical properties: Suture pull-out strength and flexibility validated against predicate criteria.
Indications for Use
Indicated for augmentation or reconstruction of the craniofacial skeleton, including cranial skeleton, orbit, nasal bones, and zygoma in patients requiring surgical implantation.
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 27, 2016
Ceremed, Inc. Ms. Chelsea Mitchell Vice President of Regulatory Affairs 3643 Lenawee Avenue Los Angeles, California 90016
Re: K161446
Trade/Device Name: Biopor Porous Polyethylene Implants Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO Dated: August 2, 2016 Received: August 26, 2016
Dear Ms. Mitchell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena SS
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161446
#### Device Name Biopor Porous Polyethylene Implants
Indications for Use (Describe)
Biopor Porous Polyethylene Implants in sheet configurations are intended for the augmentation or reconstruction of the craniofacial skeleton, including the cranial skeleton, orbit, nasal bones and the zygoma.
Type of Use (Select one or both, as applicable)
| <input type="checkbox"/> Research and/or Educational Use Only |
|---------------------------------------------------------------|
| <input type="checkbox"/> Commercial Use |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510 (K) SUMMARY:
### Submitted by:
Chelsea Mitchell Ceremed. Inc. 3643 Lenawee Ave. Los Angeles, California 90016 Tel: (424) 258-1888 Fax: (310) 815-2130
| Contact Person: | Chelsea Mitchell |
|----------------------|----------------------------------------------------------------|
| Date Prepared | September 27, 2016 |
| Common/Usual Name: | Porous High Density Polyethylene<br>(HDPE) Surgical Implant |
| Proprietary Name: | Biopor® Porous Polyethylene Implants |
| Classification Name: | Plate, Cranioplasty, Preformed, Alterable<br>(21 CFR 882.5320) |
| Product Code: | GWO |
Predicate Device:
- 1. Ceremed, Inc. Biopor® Porous Polyethylene Surgical Implants (K141880)
### Description of the device:
Biopor® Porous Polyethylene Implants are manufactured of porous high-density polyethylene (PPE), a biomaterial that is contoured or carved to suit the anatomical and functional requirements of the patient. The implants are manufactured with the option of a coating with a water-soluble alkylene oxide copolymer blend (AOC). Biopor® Porous Polyethylene Implants are provided STERILE and must not be resterilized.
### Intended use (Indications For Use):
Biopor® Porous Polyethylene Implants in sheet configurations are intended for the augmentation or reconstruction of the craniofacial skeleton, including the cranial skeleton, orbit, nasal bones and the zygoma.
{5}------------------------------------------------
# Comparison of Technology to Predicate Device:
The differences and similarities in technological characteristics and intended use between the subject and predicate devices are listed below:
| | Biopor® PPE | Biopor® PPE |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number | K161446 | K141880 |
| Indications for<br>Use Statement | Biopor® Porous<br>Polyethylene Implants<br>in sheet configurations<br>are intended for the<br>augmentation or<br>reconstruction of the<br>craniofacial skeleton,<br>including the cranial<br>skeleton, orbit, nasal<br>bones and the zygoma. | Biopor® Porous Polyethylene Implants in block,<br>sheet, and anatomical shapes are intended for the<br>augmentation or reconstruction of the<br>"maxillofacial skeleton". |
| Materials | PPE with AOC option | PPE with Titanium and AOC options |
| Configuration | Sheets | Sheets and anatomical shapes |
| Sterility | Unchanged | Sterile via electron beam irradiation |
| Packaging | Unchanged | Inner packet, outer Tyvek pouch |
# Biocompatibility and Performance Testing:
Performance testing of the Biopor® Porous Polyethylene Implants was completed. The biocompatibility profile was leveraged from testing to support K043133 (a predicate ancestor of K141880). The endotoxin specification of the device is < 2.15 EU/device. These data are applicable to Preformed Cranioplasty Plate (21 CFR 882.5320) and support the substantial equivalence of the subject device to the predicate (K141880). The test data is summarized below:
| Test | Test Method Summary | Results |
|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Cytotoxicity<br>Study Using the<br>End-Point<br>Titration | An in vitro study of the AOC<br>coating used dilutions of an<br>extract on a confluent monolayer<br>of mouse fibroblast cells. | The extract tested negative after 24, 48<br>and 72 hours. No cytotoxicity detected. |
| Murine Local<br>Lymph Node<br>Assay (LLNA) | A study of the AOC coating for<br>delayed contract sensitization<br>using the LLNA mouse model | Under the conditions of the study, the<br>material was not considered sensitizing to<br>the mouse. |
| ISO Modified<br>Intracutaneous<br>Solution | A study of the AOC coating for<br>irritation and sensitization.<br>3 rabbits were injected into the<br>skin and observed for 72 hrs. | The primary irritation index<br>characterization for the test article was<br>negligible. |
| USP and ISO<br>Modified<br>Systemic<br>Toxicity | A study of the AOC coating for<br>systemic toxicity. 10 mice were<br>administered a dose of 50 ml/kg<br>and observed for 7 days. | There was no evidence of systemic<br>toxicity. |
| ISO Muscle<br>Implantation | The AOC coating for evaluated<br>for toxicity. Test articles were<br>implanted into the muscle of<br>rabbits. | After 2 weeks, the test article was<br>classified as a non-irritant. |
| Genotoxicity<br>Mouse Bone<br>Marrow<br>Micronucleus | The AOC coating for evaluated<br>for genotoxicity using the mouse<br>bone marrow micronucleus<br>model. | The coating showed no evidence of<br>cellular toxicity. |
| Genotoxicity:<br>Bacterial Reverse<br>Mutation | The AOC coating for evaluated<br>for genotoxicity using Bacterial<br>Reverse Mutation. | The coating showed no evidence of<br>cellular toxicity. |
| Genotoxicity: In<br>Vitro<br>Chromosomal<br>Aberration | The AOC coating for evaluated<br>for genotoxicity using In Vitro<br>Chromosomal Aberration. | The coating showed no evidence of<br>cellular toxicity. |
| Bone<br>Implantation<br>Study in the<br>Femur of the<br>Rabbit | The AOC coating was implanted<br>into the femurs of rabbits and<br>evaluated after 4 and 8 weeks. | The test article was absorbed and all sites<br>were healing normally. |
| AOC Polymer<br>Hemolysis | The AOC coating for evaluated<br>for hemolysis using In Vitro<br>rabbit red blood cells. | The test article was non-hemolytic. |
| PPE Post-<br>irradiation<br>Cytotoxicity | An <i>in vitro</i> study of the<br>irradiated PPE implant extract<br>using MEM Elution. | The test articles were non-cytotoxic. |
| ISO<br>Intramuscular<br>Implantation<br>AOC Coated &<br>Uncoated PPE | AOC Coated & Uncoated PPE<br>devices were implanted into<br>rabbit muscle and histopathology<br>was performed after 1, 2 and 4<br>weeks. | Fibrovascular ingrowth occurred into<br>coated and uncoated implants. |
| Suture Pull-out of<br>AOC-Coated<br>Biopor Porous<br>Polyethylene<br>Implant | Coated implants were evaluated<br>for strength using a suture pull-<br>out test method | The test articles met acceptance criteria. |
| Biopor Sheet<br>Performance<br>Qualification | Coated and uncoated implants<br>were evaluated for flexibility. | The test articles met acceptance criteria. |
| Porosity<br>Characterization<br>of Porous<br>Polyethylene<br>Implants | Coated and uncoated implants<br>were evaluated for porosity with<br>the criteria of pore size greater<br>than 40 µm. | The test articles met acceptance criteria. |
{6}------------------------------------------------
{7}------------------------------------------------
# Substantial Equivalence:
The Biopor® Porous Polyethylene Implants in this submission represent a line extension of additional sheet configurations intended for use in the craniofacial skeleton, and have a comparable intended use and indications for use as the predicate Biopor® Porous Polyethylene Implants indicated for use in the maxillofacial skeleton (K141880). This submission contains implants for use in the craniofacial skeleton. They are similar to one type of implants currently marketed by Ceremed.
The biocompatibility profile of the subject devices is leveraged from that of the predicate device. The mechanical properties of subject Biopor® Porous Polyethylene Implants meet the same acceptance criteria as the predicate device sheet configurations.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.