The Stryker QuikFlap Sterile Procedure Pack is intended for reconstruction, stabilization and/or rigid fixation of non load-bearing bony areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 and higher).
Device Story
Stryker QuikFlap Sterile Procedure Pack provides pre-packaged, sterile sets of bone plates, burr hole covers, and screws for cranial fixation. Used in OR by surgeons for reconstruction/stabilization of non-load-bearing cranial bone post-craniotomy, craniectomy, or fracture. Implants are identical to Universal Neuro 3 system; new aspect is the sterile implant carrier/packaging. Surgeon selects appropriate pack configuration based on surgical need; plates are manually or instrument-bent to fit bony contours; screws (self-drilling or self-tapping) provide form/force fit fixation. System facilitates efficient OR workflow by providing necessary hardware in a single sterile unit, reducing hospital processing risks. Benefits include stable cranial closure, decreased palpability, and reduced tissue irritation due to rounded implant edges.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included biocompatibility (cytotoxicity, GC/MS fingerprinting), transport safety, seal strength, and sterilization validation. Mechanical stability of plates and screws verified per ASTM F382-99 and ASTM F543. Corrosion resistance demonstrated for anodized implants.
Technological Characteristics
Materials: Commercially pure titanium or Ti6Al4V alloy (ASTM F67, ASTM F136/ISO 5832-3). Implant carrier: Makrolon Rx2530. Blister: PETG ST 3020. Pouch: STS PA/PE 90 (ISO 11607). Tyvek: 1073B SBP 2000. Fixation: Form and force fit via plates and screws (self-drilling/self-tapping). Features: Countersinks, rounded edges, anodized finish. Sterile, single-use procedure pack.
Indications for Use
Indicated for reconstruction, stabilization, and rigid fixation of non-load-bearing cranial bony areas following craniotomy, craniectomy, or cranial fractures in adults and adolescents (age 12+). Contraindicated in non-reducible/unstable fractures, active local infections, metal allergies, foreign body sensitivity, non-compliant patients, poor bone quality, or where fixation contacts dura mater or developing paranasal sinuses.
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{0}------------------------------------------------
K120352
JUN 1 1 2012
750 Trade Centre Way Suite 200 Portage, MI 49002 t: 269 324 5346 f: 877 648 7114 www.stryker.com
# stryker®
Craniomaxillofacial
# 510(k) Summary of Safety and Effectiveness:
#### Stryker QuikFlap Sterile Procedure Pack
Proprietary Name:
Stryker QuikFlap Sterile Procedure Pack
Common Name:
QuikFlap Sterile Procedure Pack
Classification Name and Reference:
Sec. 882.5320 - Preformed alterable cranioplasty plate
Sec. 882.5250 - Burr hole cover
Sec. 882.5360 - Cranioplasty plate fastener
ﺍﻟﺮﺍﺑﻊ
Proposed Regulatory Class:
Class II
Product Codes:
GWO - Preformed alterable cranioplasty plate GXR – Burr hole cover HBW - Cranioplasty plate fastener
Predicate Device:
Stryker Universal Neuro 3 system - K112557
{1}------------------------------------------------
#### For Information contact:
Jamshed Badarpura Regulatory Compliance Analyst Stryker Craniomaxillofacial 750 Trade Centre Way, Suite 200 Portage, MI 49002 Phone: (269) 389-4260 jamshed.badarpura@stryker.com
#### Date Prepared:
4/5/2012
## Intended Use / Indication for Use
The Stryker QuikFlap Sterile Procedure Pack is intended for reconstruction, stabilization and/or rigid fixation of non load-bearing bony areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 and higher).
#### Device Description
The Stryker QuikFlap Sterile Procedure Pack is intended for reconstruction, stabilization and/or rigid fixation of non-load-bearing bony areas subsequent to craniectomy, craniectomy and cranial fractures. The device can be used in both adults and adolescents (age 12 and higher). Six different procedure packs of Stryker QuikFlap Sterile Procedure Pack will be offered. They consist of different combinations and number of plates and screws. The overview of these combinations is provided in Table 1. All implants which are part of the Stryker QuikFlap Sterile Procedure Pack are also part of the Universal Neuro 3 system. The instruments of the Universal Neuro 3 system are therefore used with the Stryker QuikFlap Sterile Procedure Packs. Universal Neuro 2 products are not contained in the Stryker QuikFlap Sterile Procedure Pack sold in the US.
The Stryker QuikFlap Sterile Procedure Pack is an easy to use implant carrier providing sterile plates, burr hole covers and screws in different groups to allow the use in OR. For a list of the
{2}------------------------------------------------
diverse implants groups provided in Stryker QuikFlap Sterile Procedure Pack please refer to (Table 1).
| Description | Stryker ®<br>QuikFlap<br>Article<br>number | Consisting | | | |
|-------------------------------------------------|--------------------------------------------|----------------------------------------------------------|--------|----------------------------------|--------|
| | | Plates<br>Article #<br>Neuro III | Amount | Screws<br>Article #<br>Neuro III | Amount |
| 2-HOLE PLATE<br>SET, SELF-<br>DRILLING<br>SCREW | 12-01530S | 53-36212<br>Image: [plate] | 3 | 56-15904<br>Image: [screw] | 6 |
| 2-HOLE PLATE<br>SET, SELF-<br>TAPPING<br>SCREW | 12-01531S | 53-36212<br>Image: [plate] | 3 | 56-15004<br>Image: [screw] | 6 |
| 2-HOLE PLATE<br>SET, LOW<br>PROFIL WITH<br>TAB | 12-01532S | 53-34212<br>Image: [plate] | 3 | 56-15904<br>Image: [screw] | 6 |
| 2-HOLE<br>PLATE/BURR<br>HOLE COVER<br>14MM SET | 12-01534S | 53-34212<br>Image: [plate]<br>53-34514<br>Image: [plate] | 2<br>1 | 56-15904<br>Image: [screw] | 10 |
| 2-HOLE<br>PLATE/BURR<br>HOLE COVER<br>20MM SET | 12-01536S | 53-34212<br>Image: [plate]<br>53-34520<br>Image: [plate] | 2<br>1 | 56-15904<br>Image: [screw] | 10 |
| BURR HOLE<br>COVER 14MM<br>SET | 12-01538S | 53-34514<br>Image: [plate] | 1 | 56-15904<br>Image: [screw] | 6 |
Table 1: Stryker QuikFlap Sterile Procedure Packs
・
{3}------------------------------------------------
The implants sold in the Stryker QuikFlap Sterile Procedure Pack are identical in design. material, physical properties and manufacturing process, with the implants of the Universal Neuro 3 system (K112557). The items are based on the same design drawings and are manufactured by the same processes as the implants of Universal Neuro 3 system.
After manufacturing the only difference between the implants sold in the Stryker QuikFlap Sterile Procedure Packs and the Universal Neuro 3 system is the packaging. Here, the Universal Neuro 3 system implants are single packed (plates, burr hole covers) or packed in groups of five (screws in screw rack) into blisters. To provide the surgeon a variety of sterile implants different groups of plates, burr hole covers and screws are packed together into an implant carrier, which is then placed into a blister similar (same material PETG, different size) to the ones used for Universal Neuro 3 system. Pouch is put around the blister of the Stryker QuikFlap Sterile Procedure Pack, and for safety reason this assembly is packed into an additional box, where as the Universal Neuro 3 implants have only a pouch as outer packaging. The Stryker QuikFlap Sterile Procedure Packs are additionally sterilized.
As the implants sold in the Stryker QuikFlap Sterile Procedure Pack are identical with that of FDA cleared Universal Neuro 3 System (K112557).the implants are stated to be safe for implantation. The implants are mechanically safe for implantation, biocompatible, possible to clean and sterilize within the Universal Neuro 3 module and container.
The new aspect of the Stryker QuikFlap Sterile Procedure Pack is the implant carrier. The biocompatibility of the implant carrier was proven including cytotoxicity and GC/MS Fingerprint tests. The handling performance and the transport safety of the whole Stryker QuikFlap Sterile Procedure Pack was determined, so that the new implant carrier of the Stryker QuikFlap Sterile Procedure Pack can be stated as safe and effective for its intended use.
The implant carrier, including the implants, is stored in a blister equal to the ones cleared in the 510/k) for the Universal Neuro 3 System. The blister is made out of the same material (PETG Kodar 6763) by the same supplier in the same process as for the Universal Neuro 3 blister, the only difference is its size. The blister itself is enclosed with the identical Tyvek material. The
{4}------------------------------------------------
safety of the blister -Tyvek - sterile barrier was successfully tested after conditioning of the samples in a transport test. As the blister type is already known and its sealed seam safety was proven by testing the blister, it can be stated as safe for its intended use.
As second sterile barrier, the blister of the Stryker QuikFlap Sterile Procedure Pack is placed in a peel pouch as used for the inlays of the Universal Neuro 3 system. The pouch is therewith a known packaging material which was tested as well for transport safety. A standard box is used as outer packaging. The box passed the transport safety test as well.
Overall following conclusion can be drawn:
The implants sold in the Stryker QuikFlap Sterile Procedure Pack are known and FDA cleared for the Universal Neuro 3 system and are hence proven to be safe and effective in use. biocompatibility, cleaning and sterilization. The new implant carriers were proven to be biocompatible, safe for transportation and efficient for use. Blister, pouch and outer box are known materials that are already in use for Stryker Leibinger products and that are proven to be safe for transportation. So, the Stryker QuikFlap Sterile Procedure Packs can be stated as safe and effective.
#### Contraindications
- . Use of plates in non-reducible and unstable fractures
- . Patients with active local infections
- . Patients with metal allergies and foreign body sensitivity
- . Potentially non-compliant patients with mental or neurological conditions who are unwilling or incapable of following postoperative care instructions
- . Patients with limited blood supply to or insufficient quality of bone
- . Use of plates where the fixation of these products could result in their peripheral edge coming into contact with the dura mater.
- Screws coming in contact with the dura mater .
{5}------------------------------------------------
- . Use of implants adjacent to developing parnasal sinuses
#### Possible System Adverse Effects
- . Metal sensitivities or allergic reaction
- Early or late infection, both deep and/or superficial .
- . Deformation of affected bone
- Poor implant fixation resulting in mal or nonunion .
- . Implant breakage resulting in mal or nonunion
- . Device migration
- . Fibrous encapsulation of implant
#### Technological Characteristics
Stryker QuikFlap Sterile Procedure Pack is a sterile procedure pack containing bone screws, plates and burr hole covers for the respective anatomical and indicated areas. Different combinations of screws and plates are available in this pack. The surgeon can take out only the implants out of this pack. The offered combinations of screws and plates are adequate to fix common defects. For more complicated cases more than one Stryker QuikFlap Sterile Procedure Pack can be used.
Stryker QuikFlap Sterile Procedure Pack offers an easy, cost-effective solution for cranial flap fixation. Six procedure packs provide surgeons a variety of options for cranial closure. Since the Stryker QuikFlap Sterile Procedure Pack is provided sterile, it can limit processing risks for the hospitals.
The Stryker QuikFlap Sterile Procedure Pack contains bone plates, burr hole covers and screws. The implants are designed for a decreased palpability. The screws are used to fix the plates on bony structures and Stryker custom implants. The contained plates can be bent by hand or by using instruments to have a good fit to the shape of the above mentioned bony structures.
{6}------------------------------------------------
K120352
#### Performance data
This procedure pack is designed in such a way that the surgeon can take out of this pack all the implants he needs for the above mentioned surgeries. For some set configurations (12-01530S. 12-01531S) he can take out the implants needed for every step of the above mentioned surgeries. The packaging and the implant carrier are designed for easy handling of the pack and an easy taking out of the implants.
There are self-drilling and self-tapping screws contained in the Stryker QuikFlap Sterile Procedure Pack. Self-drilling screws can be used without pre-drilling, for self-tapping screws pre-drilling is required. A cross-pin mechanism on the screw head is used for all screws.
In the Stryker QuikFlap Sterile Procedure Pack there are plates with and without tabs. The tab is used to allow an easier handling of the plates. 2-hole plates and burr hole covers are offered in the Stryker QuikFlap Sterile Procedure Pack. The plates have countersinks to allow a secure fit of the screw in the plate. The plates are fixed by form and force fit.
All implants have rounded edges to avoid tissue irritation. All implants are anodized and the corrosion resistance was demonstrated.
Stryker bone plates and bone screws are made of either commercially pure titanium or Ti6AI4V alloy (acc. to ASTM F67, ASTM F136/ISO 5832-3). Both materials are biocompatible according to ISO 10993-1, corrosion-resistant and non-toxic in the biological environment, and produce negligible artefacts on X-ray and CT scans. For the implant carrier Makrolon Rx2530 is used. The packaging material for the blisters is made of PETG ST 3020, the Tyvek out of Tvvek 1073B SBP 2000. The peel pouches are made of STS PA/PE 90 according to ISO 11607.
The Stryker QuikFlap Sterile Procedure Pack along with all the contained implants is offered in double-sealed sterile packaging. All necessary tests were performed to show that these products are clean, there are no cytotoxic effects, and the products and packaging are biocompatible, there is sufficient stability of plates and screws according to ASTM F 382-99 and ASTM F 543, that
{7}------------------------------------------------
this product can be re-sterilized. There are no damages during transport, the implant carrier is sufficiently stable and there is sufficient strength of the seal.
### Substantial Equivalence Discussion
The Stryker QuikFlap Sterile Procedure Pack has been verified and validated according to Stryker procedures for product design and development. The validation proves the safety and effectiveness of the system. The subject Stryker QuikFlap Sterile Procedure Pack is substantially equivalent to legally marketed device:
- · Stryker Universal Neuro 3 System K112557
The Stryker QuikFlap Procedure Pack, as stated above, consists of devices with three product codes: GWO (plates), GXR (burr hole covers), and HBW (screws). It has the same material composition and operating principles as its predicate mentioned above. The intended use is identical to the predicate system. Further, there is no difference in dimensions and shapes between the Stryker QuikFlap Sterile Procedure Pack and the predicate device.
Specifically speaking; our Stryker QuikFlap Sterile Procedure Pack plates, burr hole covers and screws are substantially equivalent to Stryker Universal Neuro 3 System (K112557).
{8}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Stryker Mr. Jamshed Badarpura Regulatory Compliance Analyst 750 Trade Center Way, Suite 200 Portage, MI 49002
Re: K120352
Trade/Device Name: Stryker QuikFlap Sterile Procedure Pack Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed alterable cranioplasty plate Regulatory Class: Class II Product Code: GWO Dated: May 14, 2012 Received: May 15, 2012
Dear Mr. Badarpura:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
JUN 1 1 2012
{9}------------------------------------------------
#### Page 2 - Mr. Jamshed Badarpura
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
Special 510(k)
Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Stryker QuikFlap Sterile Procedure Pack
Indications for Use:
The Stryker QuikFlap Sterile Procedure Pack is intended for reconstruction, stabilization and/or rigid fixation of non load-bearing bony areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 and higher).
Prescription Use______________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over the counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K120352
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.