21 CFR 882.5560 — External Cerebrospinal Fluid (Csf) Diversion
Neurology (NE) · Part 882 Subpart F—Neurological Therapeutic Devices · § 882.5560
Identification
A cerebrospinal fluid shunt system is a prescription device used to monitor and divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of preventing spinal cord ischemia or injury during procedures that require reduction in central nervous system pressure. A cerebrospinal fluid shunt system may include catheters, valved catheters, valves, connectors, and pressure monitors intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Classification Rationale
Class II (special controls). The special controls for this device are:
Product Codes
| Product Code | Device Name | Class | Devices | Attributes |
|---|---|---|---|---|
| PCB | External Cerebrospinal Fluid (Csf) Diversion | 2 | 2 |
Special Controls
PCB — External Cerebrospinal Fluid (Csf) Diversion
In combination with the general controls of the FD&C Act, the cerebrospinal fluid shunt system is subject to the following special controls:
De Novo Order DEN120017
PCB — External Cerebrospinal Fluid (Csf) Diversion
*Classification.* Class II (special controls). The special controls for this device are:(1) The device description must include a detailed summary of the device technical parameters, including design configuration, dimensions, engineering drawings, and a list of all components with identification of their materials of construction. (2) The patient-contacting components of the device must be demonstrated to be biocompatible. (3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Simulated use testing must be conducted to characterize fluid flow and resistance to leakage; and (ii) Mechanical integrity testing of all connections must be conducted. (4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the specified shelf life. (5) Performance data must demonstrate the sterility and pyrogenicity of patient-contacting components of the device. (6) The labeling must include: (i) Contraindications with respect to patients who should not receive a lumbar drain; (ii) A warning that the device should have 24-hour-a-day availability of trained personnel to supervise monitoring and drainage; (iii) Instructions on proper device setup, positioning, and monitoring; (iv) Warnings and precautions to inform the user of serious hazards and special care associated with the use of the device; (v) A statement that the device is not to be reused, reprocessed, or resterilized when open but unused; and (vi) Cleaning instructions for the injection sites.
eCFR
PCB — External Cerebrospinal Fluid (Csf) Diversion
(1) The device description must include a detailed summary of the device technical parameters, including design configuration, dimensions, engineering drawings, and a list of all components with identification of their materials of construction. (2) The patient-contacting components of the device must be demonstrated to be biocompatible. (3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Simulated use testing must be conducted to characterize fluid flow and resistance to leakage; and (ii) Mechanical integrity testing of all connections must be conducted. (4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the specified shelf life. (5) Performance data must demonstrate the sterility and pyrogenicity of patient-contacting components of the device. (6) The labeling must include: (i) Contraindications with respect to patients who should not receive a lumbar drain; (ii) A warning that the device should have 24-hour-a-day availability of trained personnel to supervise monitoring and drainage; (iii) Instructions on proper device setup, positioning, and monitoring; (iv) Warnings and precautions to inform the user of serious hazards and special care associated with the use of the device; (v) A statement that the device is not to be reused, reprocessed, or resterilized when open but unused; and (vi) Cleaning instructions for the injection sites.