← Product Code [NCX](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/NCX) · K014241

# COOLGARD 3000, COOLINE, START-UP, MODELS 3000, CL-2085B, CL-2295A, CG-500D (K014241)

_Alsius Corp. · NCX · Aug 1, 2003 · Cardiovascular · SESU_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94cardiovascular-therapeutic-devices/NCX/K014241

## Device Facts

- **Applicant:** Alsius Corp.
- **Product Code:** [NCX](/submissions/CV/subpart-f%E2%80%94cardiovascular-therapeutic-devices/NCX.md)
- **Decision Date:** Aug 1, 2003
- **Decision:** SESU
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5900
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Alsius CoolGard® 3000 and Cool Line ™ Catheter Thermal Regulation System is indicated for use in fever reduction, as an adjunct to other antipyretic therapy, in patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.

## Device Story

System comprises CoolGard 3000 electro-mechanical heater/cooler unit, disposable start-up kit, and Cool Line intravascular catheter. Catheter placed percutaneously in central venous circulation; features closed-loop heat exchange balloons. System pumps temperature-controlled sterile saline through catheter balloons to perform conductive heat exchange with patient blood; saline does not enter patient. Physician sets target temperature; system automatically adjusts saline bath temperature based on patient temperature probe data. Used in clinical settings by healthcare providers. Provides fever reduction as adjunct to antipyretic therapy. Benefits include precise thermal regulation for specific neurological patient populations.

## Clinical Evidence

Clinical evidence includes a randomized controlled trial of 296 patients across four cohorts (cerebral infarction, intracerebral hemorrhage, primary traumatic brain injury, subarachnoid hemorrhage). Primary endpoint was mortality. Results showed no statistically significant difference in mortality between Cool Line and control groups for cerebral infarction (p=0.74) and intracerebral hemorrhage (p=1.00). Safety not demonstrated for subarachnoid hemorrhage or primary traumatic brain injury.

## Technological Characteristics

Integrated electro-mechanical heater/cooler with temperature monitor, controller, heat exchanger, and roller pump. Catheters are multi-lumen, biocompatible polyurethane with Duraflo heparin coating. Closed-loop fluid path for saline heat exchange. Nominal 5mm distal balloon diameter. Redundant safety controls, alarms, and temperature displays.

## Regulatory Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

## Predicate Devices

- Alpha Omega Dual Cooler Heater ([K001520](/device/K001520.md))
- MTRE Allon 2001 Thermal Regulation Device ([K001546](/device/K001546.md))
- ARROWgard Blue Quad-Lumen Central Venous Catheter ([K962577](/device/K962577.md))
- Arrow-Howes Large Bore Multi Lumen central venous catheter ([K970864](/device/K970864.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for ALSIUS Corporation. Below the logo, the text "Company Confidential" is written in a simple, sans-serif font. The logo is in black and white and appears to be a stylized version of the company's name.

## Section 4: 510(k) Summary

## 510(K) Summary: Alsius Corportation's CoolGard™ | Cool Line™ Catheter Thermal Regulation System

## Submitter's Name, Address, Telephone Number, and Contact Person:

ALSIUS CORPORATION 15770 Laguna Canyon Road, Suite 150 Irvine, CA 92618

AUG 01 2003

Contact: Ken Collins Phone: 949-453-0150 Fax: 949-453-0250 Email: kcollins@alsius.com

## Name of Device:

The Alsius CoolGard And Cool Line Catheter Thermal Regulation System.

## Common or Usual Name:

Central Venous Catheter (short term) and Thermal Regulating System.

## Classification Name:

Venous heat exchange catheters and associated temperature control systems have not been specifically classified by the FDA. However, FDA has classified venous catheters and thermal regulating systems as Class II devices under 21 C.F.R. §§ 880.5200 and §§ 870.5900 respectively.

{1}------------------------------------------------

014241

ALSIVS
CORPORATION

Company Confidential

## Predicate Device(s):

Alpha Omega Dual Cooler Heater; Alpha Omega, Inc., Austin TX K001520; S.E. 11-22-2000

MTRE Allon 2001 Thermal Regulation Device; MTRE Advanced Technologies, McCordsville, IN K001546: S.E. 06-08-2000

ARROWgard Blue Quad-Lumen Central Venous Catheter Arrow International, Inc., Reading PA K962577; S.E. 08-21-1997

Arrow-Howes Large Bore Multi Lumen central venous catheter Arrow International, Inc., Reading PA K970864: S.E. 09-17-1997

## Indications for Use

The Alsius CoolGard® 3000 and Cool Line ™ Catheter Thermal Regulation System is indicated for use in fever reduction, as an adjunct to other antipyretic therapy, in patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.

## Warning - Fever Reduction

The safety of this device has not been demonstrated for fever reduction in patients presenting with subarrchnoid hemorrhage or primary traumatic brain injury. The safety and effectiveness of this device was examined in a randomized controlled trial of 296 patients. The mortality results reported in this trial, for the four patient cohorts enrolled, are presented in the table below (CI - cerebral infarction, ICH - intracerebral hemorrhage, PTBI - primary traumatic brain injury, SAH - subarrachnoid hemorrhage).

|      | Cool Line |    |      | Control |    |      |      |
|------|-----------|----|------|---------|----|------|------|
|      | n         | N  | %    | n       | N  | %    | p*   |
| CI   | 3         | 16 | 18.8 | 3       | 14 | 21.4 | 0.74 |
| ICH  | 8         | 33 | 24.2 | 7       | 27 | 25.9 | 1.00 |
| PTBI | 10        | 44 | 22.7 | 4       | 38 | 10.5 | 0.24 |
| SAH  | 13        | 61 | 21.3 | 7       | 63 | 11.1 | 0.15 |

## Mortality by Diagnosis (ITT analysis)

*Fischer's exact test

For more details on the clinical trial results please refer to "Physician's Manual - Fever Reduction for Patients with Cerebral Infarction and Intracerebral Hemorrhage" #101416-001.

K014241 S3

{2}------------------------------------------------

014241 3/4

ALSIUS
CORPORATION

K014241 S3

#### Company Confidential

#### Technical Characteristics:

The CoolGard and Catheter Thermal Regulation System consists of the CoolGard™ 3000, a disposable Start Up Kit used in the CoolGard™ for interface with the cooling bath and patient catheter and the Intravascular Catheter. The Alsius CoolGard™ 3000 is an integrated electro-mechanical heater/cooler that consists of a temperature monitor, a temperature controller unit, a heat exchanger unit, and roller pump. It supplies temperature controlled sterile saline to the indwelling Catheter that is placed percutaneously in the patient.

The technical characteristics of the Catheter are essentially identical to those of widely used multi-lumen central venous catheters except for the dedicated closed loop fluid path through the heat exchange balloons. The Alsius Catheter materials are all biocompatible polyurethanes.

Likewise, the CoolGard™ 3000 heater/cooler has the same technical features as the medical heater/cooler unit Identified as the predicate device. These common technical features include connections for recirculating coolant to and intravascular catheter and all or combinations of the following: redundant safety controls and alarms, patient monitoring and control and temperature displays for the clinician users.

Two Models of Intravascular catheters are available for use with the CoolGard and Catheter Thermal Regulation System:

- 1. Cool Line™ Catheter Kit Model CL 2085B (2 lumens)
- 2. Cool Line™ Catheter Kit Model CL 2295A (3 lumens)

The Cool Line™ catheters are multi lumen intravascular catheters in two sizes. Two of the catheter's lumens are used to circulate sterile saline to exchange heat with the central venous blood supply. When the heat exchange feature of the catheter is in use, heated/chilled saline is pumped through the heat exchange lumen, expanding the diameter of the distal portion of the catheter to a nominal 5mm where the heating/cooling membranes interface with the patient's circulating blood. The inflow lumen/outflow lumen forms a closed-loop system through which the heated/chilled saline circulates. The chilled saline is not infused into the patient.

Additional lumens of the Alsius Catheters consist of a standard quide wire lumen that can be used as a primary infusion lumen, and a second or third infusion lumen within the shaft, depending on the catheter model selected by the clinician.

The Catheter blood contact surfaces are coated with Duraflo® Treatment, a heparin coating manufactured by Edwards Lifesciences Corporation.

The Alsius Catheters are supplied sterile for single-use only.

#### Principles of Operation:

The CoolGard™ 3000 system automatically adjusts the temperature of the heater/chiller saline bath to achieve the patient target temperature that has previously been set by the attending physician. This is done via data from a temperature probe in the patient that

Page 3 of 4

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for ALSIUS CORPORATION. The logo is in black and white and features a stylized font. The word "ALSIUS" is in a larger font than the word "CORPORATION", which is located underneath.

Company Confidential

interfaces with the temperature controller. This principle of operation is identical to currently marketed devices.

## Summary of the Basis for Finding of Substantial Equivalence:

The Cool Line™ indication statements and intended use are identical to those for central venous catheters and heater/cooler units. The Cool Line™ catheter provides access to the central circulation for fluid administration. Additionally, the closed loop connections on the catheter can be attached to the heater/cooler unit for patient heating or cooling via conductive heat transfer from the temperature controlled saline. The Cool Line™ integrates venous access and temperature control which, separately, are intended uses for numerous devices.

The integration of temperature control and venous access could potentially affect safety and effectiveness in new ways that would raise new types of safety or effectiveness questions. Accepted scientific methods were used to generate performance data demonstrating the substantial equivalence of the Cool Line™ Catheters and CoolGard™ System to the predicate. In addition, the clinical study under IDE G990263 was conducted to further define safety characteristics.

## Conclusion

In summary, descriptive information and performance data demonstrate that the Alsius Cool Line™ Catheter and CoolGard™ System's different technological characteristics either do not raise new questions of safety and effectiveness or, where appropriate, performance data demonstrate equivalence.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG - 1 2003

Kenneth A. Collins, M.D. Vice President Clinical/Quality/Regulatory Alsius Corporation 15770 Laguna Canyon, Road, Suite 150 Irvine, California 92618

Re: K014241

Trade/Device Name: Alsius CoolGard and Cool Line Catheter Thermal Regulation System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal regulating system Regulatory Class: Class II Product Code: NCX Dated: June 20, 2003 Received: June 20, 2003

Dear Dr. Collins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device. subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling as a box warning immediately following the indications for use statement:

{5}------------------------------------------------

## Page 2 - Kenneth A. Collins, M.D.

The safety of this device has not been demonstrated for fever reduction in patients presenting with subarrachnoid hemorrhage or primary traumatic brain injury. The safety and effectiveness of this device was examined in a randomized controlled trial of 296 patients. The mortality results reported in this trial, for the four patient cohorts enrolled, are presented in the table below (CI - cerebral infarction. ICH - intracerebral hemorrhage, PTBI - primary traumatic brain injury, SAH - subarrachnoid hemorrhage).

|      | Cool Line |    |      | Control |                                                                                                                                                                               |      |        |  |
|------|-----------|----|------|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|--------|--|
|      | n         | Z  | 0/0  | n       | N                                                                                                                                                                             | %    | p-     |  |
|      |           |    |      |         |                                                                                                                                                                               |      | value* |  |
| CI   |           | 16 | 18.8 |         | 14                                                                                                                                                                            | 21.4 | 0.74   |  |
| ICH  | 8         | 33 | 24.2 |         | 27                                                                                                                                                                            | 25.9 | 1.00   |  |
| PTBI | 10        | 44 | 22.7 | ব       | 38                                                                                                                                                                            | 10.5 | 0.24   |  |
| SAH  | 13        | 61 | 21.3 |         | ર્ભ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં મુખ્યત્વે ખેત-ઉપયોગ વિદ્યારત તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, આંગણવાડી તેમ જ દૂધ | 11.1 | 0.15   |  |

Mortality by Diagnosis (ITT)

*Fischer's exact test

For more details on the results of this study please refer to "Physician's Manual - Fever Reduction for Patients with Cerebral Infarction and Intracerebral Hemorrhage" #101416-001.

Please note that the above labeling limitation is required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before this limitation is modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

Page 3 - Kenneth A. Collins, M.D.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4595. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers,

International, and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Daniel G. Schultz, M.D.

Daniel G. Schultz, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

# K014241

## Alsius CoolGard 3000/Cool Line Catheter Heat Exchange System

# Indications for Use

The Alsius CoolGard® 3000 and Cool Line ™ Catheter Thermal Regulation System is indicated for use in fever reduction, as an adjunct to other antipyretic therapy, in patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.

| Prescription Use | <div style="display:inline-block; width:15px; height:15px; border:1px solid black;"> <span style="position:relative; top:-2px; left:2px;">X</span> </div> |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Over the Counter | <div style="display:inline-block; width:15px; height:15px; border:1px solid black;"> </div>                                                               |

(Division Sign-Off)

Division of General, Restorative

and Neurological Devices

510(k) Number -

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94cardiovascular-therapeutic-devices/NCX/K014241](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94cardiovascular-therapeutic-devices/NCX/K014241)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
