K020298 · Plus Orthopedics · HAW · May 20, 2002 · Neurology
Device Facts
Record ID
K020298
Device Name
GALILEO CAS/NAV TKR SYSTEM
Applicant
Plus Orthopedics
Product Code
HAW · Neurology
Decision Date
May 20, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Intended Use
The GALILEO CAS/NAV TKR System is intended for computer-assisted navigation of the GALILEO-CAS instruments with the aim to position the implants of the TC-PLUS™ Solution Knee (K000666 - S/E 10/13/00) optimally for the patient.
Device Story
System provides computer-assisted navigation for total knee replacement (TKR) surgery; used intraoperatively by surgeons. Inputs include anatomical bone landmarks captured via manual navigation key button and spatial tracking of instruments via infrared camera. System utilizes two active infrared transmitters attached to distal femur and proximal tibia, plus two passive locators for instrument arrangement. Computer processes spatial data to control small electric motors on femur-cutting device, ensuring precise positioning for bone resections. No preoperative CT scan required. Output provides real-time guidance for cutting device alignment; enables accurate, flexible resection. Benefits include improved implant positioning accuracy for TC-PLUS Solution Knee prostheses.
Clinical Evidence
Bench testing only. Biomechanical tests performed; results equivalent to similar implants and sufficient for in vivo loading.
Technological Characteristics
System consists of electronic components and instruments. Uses infrared camera for tracking active/passive locators. Instruments feature low-voltage electric motors for positioning. Operates via intraoperative landmark scanning without preoperative CT. Class II device (Stereotaxic instrument).
Indications for Use
Indicated for computer-assisted navigation of GALILEO-CAS instruments to position TC-PLUS Solution Knee implants during total knee replacement surgery.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
K213873 — SkyWalker Total Knee System · Microport Navibot International, LLC · Jul 7, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 2 0 2002
Kozo298
510/k) Summary of Safety and Effectiveness [in accordance with SMDA of 1990, 21 CFR 807.92(c)]
Contact: PLUS ORTHOPEDICS 6055 Lusk Blvd. San Diego, CA 92121 Tel: 858-550-3800 x 2506 - Fax: 858-550-3813 Attn: Mr. Hartmut Loch, RAC Director, Regulatory Affairs
Trade name: GALILEO CAS/NAV TKR System
Common name: Navigation System
Classification name: Instrument, Stereotaxic & 882.4560, Class II, Neurologic Device Panel 82
Product Code: HAW
Description The GALILEO CAS/NAV TKR System is a system for computer-Device and Characteristics: assisted navigation of the GALILEO-CAS instruments with the aim to position TC-PLUS™ Solution Knee prostheses optimally for the patient. The collection of patient data required for this occurs exclusively operatively. A preoperative CT scan is not necessary. The connection between patient and computer is made via two infrared transmitters (active locators), which are attached to the distal femur and to the Two passive locators are used for the spatial proximal tibia. arrangement of the instruments. A manual navigation key button is used for scanning the anatomical bone features (landmarks). An infrared camera locates the locators as well as the manual key button and is connected to the computer. The computer-assisted Galileo CAS Total Knee Replacement System supports the operating surgeon performing the total knee replacement procedure. The system takes the femur-cutting device to the required position and enables resections with high accuracy and flexibility.
> Due to the precise computer-assisted references and the correct positioning of the cutting device the bone resections can be performed with the utmost accuracy.
> The Galileo CAS system essentially consists of two components, the electronic components and the instruments, which are attached to the distal femur. The instruments, which are driven by small electric (lowvoltage) motors, are controlled by the computer, to ensure correct positioning of the femur-cutting device.
{1}------------------------------------------------
The GALILEO CAS/NAV TKR System is substantially equivalent to the Equivalence: OrthoPilot® (Aesculap, Inc. – K003347 – S/E 2/23/2001). Both systems are intended for the same medical indications and are technically similar.
The GALILEO CAS/NAV TKR System is intended for computer-Indications: assisted navigation of the GALILEO-CAS instruments with the aim to position the implants of the TC-PLUS™ Solution Knee (K000666 - S/E 10/13/00) optimally for the patient.
- Biomechanical tests have been performed. The test results were Performance data: equivalent to other similar implants and are sufficient for in vivo loading.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing right, stacked on top of each other, with flowing lines beneath them.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 0 2002
Mr. Hartmut Loch, RAC Director, Regulatory Affairs Plus Orthopedics 6055 Lusk Boulevard San Diego, CA 92121-2700
Re: K020298
Trade/Device Name: GALILEO CAS/NAV TKR System Regulation Number: 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: April 12, 2002 Received: April 15, 2002
Dear Mr. Loch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Mr. Hartmut Loch, RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. ·
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Muriam C. Provost
Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Kodod 510(k) Number:
Device Name(s):
# GALILEO CAS/NAV TKR SYSTEM
### Indications for Use:
The GALILEO CAS/NAV TKR System is intended for computer-assisted The GALILES Of the GALILEO-CAS instruments with the aim to position the navigation of the ORE/EB ONE "","Solution Knee (K000666 - S/E 10/13/00) optimally for the patient.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K020298
Prescription Use
OR
Over-The-Counter-Use _________________________________________________________________________________________________________________________________________________________
(Per 21 CFR 801.109)
(Optional format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.