Streamline Navigated Instruments

K200095 · Pioneer Surgical Technology, Inc., Dba Rti Surgical, Inc. · OLO · Jun 26, 2020 · Neurology

Device Facts

Record IDK200095
Device NameStreamline Navigated Instruments
ApplicantPioneer Surgical Technology, Inc., Dba Rti Surgical, Inc.
Product CodeOLO · Neurology
Decision DateJun 26, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

Streamline Navigated Instruments are intended to be used during the preparation and placement of screws from the Streamline TL and MIS Systems and preparation of screws from the Streamline OCT System during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT- or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

Device Story

Streamline Navigated Instruments are non-sterile, reusable orthopedic stereotaxic instruments; include pedicle finders, modular handles, taps, drills, and screw inserters. Used with Medtronic StealthStation Navigation System; incorporate NavLock collar for tracker attachment. Surgeon operates instruments during spinal surgery to prepare and place screws. Navigation system provides real-time tracking of instrument tip relative to patient anatomy (CT/MR/fluoroscopy/digitized landmarks). Output allows surgeon to precisely locate anatomical structures; facilitates accurate screw placement in open or MIS procedures. Benefits include improved surgical precision and guidance during spinal fixation.

Clinical Evidence

Bench testing only. Performance evaluated via dimensional comparison, tolerance analysis, and positional accuracy testing per ASTM F2554-18 to ensure compatibility with the Medtronic StealthStation System. No clinical data provided.

Technological Characteristics

Materials: Stainless steel (ASTM F899). Components: Pedicle finders, modular handles, taps, drills, screw inserters. Sensing: Optical navigation via NavLock tracker interface. Connectivity: Compatible with Medtronic StealthStation. Sterilization: Reusable, provided non-sterile. Accuracy: Tested per ASTM F2554-18.

Indications for Use

Indicated for use in spinal surgery during preparation and placement of Streamline TL, MIS, and OCT system screws to assist surgeons in locating anatomical structures in open or minimally invasive procedures. Designed for use with Medtronic StealthStation System for conditions where stereotactic surgery is appropriate and rigid anatomical structures (e.g., vertebrae) can be referenced via CT, MR, fluoroscopy, or digitized landmarks.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ June 26, 2020 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square. Pioneer Surgical Technology, Inc. DBA RTI Surgical, Inc. Kurtis Hunsberger Principal Regulatory Affairs Specialist 375 River Park Circle Marquette, Michigan 49855 Re: K200095 Trade/Device Name: Streamline Navigated Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: June 19, 2020 Received: June 22, 2020 Dear Kurtis Hunsberger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K200095 Device Name Streamline Navigated Instruments #### Indications for Use (Describe) Streamline Navigated Instruments are intended to be used during the preparation and placement of screws from the Streamline TL and MIS Systems and preparation of screws from the Streamline OCT System during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT- or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K200095 Streamline Navigated Instruments As required by 21 CFR 807.92 | Date Prepared: | June 25, 2020 | |---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Pioneer Surgical Technology, Inc.<br>DBA RTI Surgical, Inc. | | | 375 River Park Circle<br>Marquette, MI 49855 USA<br>Phone: (906) 226-9909<br>Fax: (906) 225-5868 | | Contact Information: | Kurtis Hunsberger<br>Principal Regulatory Affairs Specialist<br>RTI Surgical, Inc.<br>khunsberger@rtix.com<br>952-698-9931 | | Name of Device: | Streamline® Navigated Instruments | | Common Name: | Orthopedic Stereotaxic Instrument | | Classification Name: | Stereotaxic Instrument | | Regulation Number: | 21 CFR 882.4560 | | Regulatory Class: | Class II | | Product Code: | OLO | | Panel: | Orthopedic Device Panel, Panel Code 84 | | Predicate Device: | Primary Predicate:<br>K140454 Medtronic Navigated Instruments | | | Reference Predicate:<br>K173338 Medtronic Navigated Instruments (Navigated INFINITY™<br>Instruments) | | Device Description: | The Streamline Navigated Instruments are non-sterile, reusable<br>instruments for use with the Medtronic StealthStation Navigation<br>System to assist surgeons in precisely locating anatomical structures<br>in either open or minimally invasive (MIS) procedures for preparation<br>and placement of screws from the Streamline TL Spinal Fixation<br>System and Streamline MIS Spinal Fixation System and preparation<br>of screws from the Streamline OCT Occipito-Cervico-Thoracic<br>System. | | | The Streamline Navigated Instruments include pedicle finders,<br>modular handle, taps, drills, and screw inserters designed for use with<br>Streamline TL, Streamline MIS, and Streamline OCT Systems. The<br>instruments are manufactured from stainless steel (ASTM F899). | | Indications for Use: | Streamline Navigated Instruments are intended to be used during the<br>preparation and placement of screws from the Streamline TL and MIS<br>Systems and preparation of screws from the Streamline OCT System<br>during spinal surgery to assist the surgeon in precisely locating<br>anatomical structures in either open or minimally invasive procedures.<br>These instruments are designed for use with the Medtronic<br>StealthStation System, which is indicated for any medical condition in<br>which the use of stereotactic surgery may be appropriate, and where<br>reference to a rigid anatomical structure, such as vertebra, can be<br>identified relative to a CT- or MR-based model, fluoroscopy images,<br>or digitized landmarks of the anatomy. | | Comparison of<br>Technological<br>Characteristics with<br>the Predicate Device: | Both the subject and predicate devices are stereotactic instruments<br>designed for use with optical navigation. At a high level, the subject<br>and predicate devices are based on the following same technological<br>elements. The subject and predicate devices: | | | Are used with the Medtronic StealthStation Navigation System Are used during preparation and placement of screws during spinal surgery Include the same instrument types (probes, drills, taps, drivers, modular handles) Incorporate a NavLock collar feature designed for connection to the NavLock tracker for use with navigation Have the same/similar critical dimensions of critical length from NavLock tracker to distal tip of instruments Are used with pedicle screws of same diameters and lengths Have accuracy testing performed per ASTM F2554 Are constructed from ASTM F899 stainless steel Are reusable and provided non-sterile Are designed for use with their respective spinal fixation systems | | | There are no technological differences between the subject and<br>predicate devices. However, there are minor dimensional differences<br>inherent to the instruments in the respective spinal fixation systems<br>such as instrument/implant interface features and tap threading.<br>These minor dimensional differences do not affect the critical<br>dimensions of the instrument and do not affect performance with<br>navigation. | | Performance Data: | Streamline Navigated Instruments have been tested per ASTM<br>F2554-18 "Standard Practice for Measurement of Positional Accuracy<br>of Computer Assisted Surgical Systems." Dimensional comparison<br>and tolerance analysis of the Streamline Navigated Instruments to the<br>predicate Medtronic Navigated Instruments was conducted to ensure<br>the Streamline Navigated Instruments are acceptable for their | | | intended use, ensure functionality and compatibility with the<br>Medtronic StealthStation® System. The results of this non-clinical<br>testing show that performance of the Streamline Navigated<br>Instruments is sufficient for their intended use and demonstrate<br>substantial equivalence to legally marketed predicate devices. | | Conclusion: | The supporting evidence in this submission concludes the subject<br>Streamline Navigated Instruments are substantially equivalent to the<br>predicate devices. | {4}------------------------------------------------ {5}------------------------------------------------
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...