SMITH & NEPHEW IMAGE GUIDED SURGICAL INSTRUMENTS FOR HIP APPLICATIONS
K021798 · Smith & Nephew, Inc. · HAW · Dec 9, 2002 · Neurology
Device Facts
Record ID
K021798
Device Name
SMITH & NEPHEW IMAGE GUIDED SURGICAL INSTRUMENTS FOR HIP APPLICATIONS
Applicant
Smith & Nephew, Inc.
Product Code
HAW · Neurology
Decision Date
Dec 9, 2002
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.4560
Device Class
Class 2
Intended Use
Image Guided Surgical Instruments for Hip Applications are intended to assist the surgeon in precisely locating anatomical structures in either open or percutaneous procedures. Image Guided Surgical Instruments for Hip Applications are indicated for use in surgical hip procedures, in which the use of stereotactic surgery may be appropriate, and where a reference to a rigid anatomical structure such as a long bone can be identified relative to a CT or MR based model or fluoroscopy images of the anatomy. These procedures include but are not limited to, acetabular cup placement as part of a partial or total hip arthroplasty (primary or revision).
Device Story
Device consists of surgical instruments modified to attach image-guided arrays (Fighters) containing infrared LEDs or passive spheres. Used in OR by surgeons to assist in anatomical localization during hip procedures. Arrays interact with an external IGS Platform System to track instrument position in real-time. Reference frames with similar markers are rigidly attached to patient anatomy to allow the system to track bone position and compensate for movement. Output is visual navigation data displayed to the surgeon, facilitating precise placement of implants like acetabular cups. System provides real-time spatial feedback, potentially improving surgical accuracy and patient outcomes.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and design similarities to predicate devices.
Technological Characteristics
Stereotaxic instrument system utilizing infrared LED or passive sphere tracking arrays. Components include modified surgical instruments and rigid anatomical reference frames. System integrates with external IGS Platform for real-time spatial tracking. Connectivity is via infrared signal transmission. No specific material standards or software architecture details provided.
Indications for Use
Indicated for patients undergoing surgical hip procedures, including primary or revision partial or total hip arthroplasty, where stereotactic navigation relative to CT, MR, or fluoroscopic images is appropriate.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
K021980 — HIP MODULE FOR THE STEALTHSTATION SYSTEM · Medtronic Surgical Navigation Technologies · Nov 19, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Koz179f
# 510ck) Summarv Image Guided Surgical Instruments For Hip Applications
DEC 0 9 2002
| Submitter's name: | Smith & Nephew, Inc., Orthopaedic Division |
|-----------------------------------|--------------------------------------------------------------------------|
| Submitter's address: | 1450 Brooks Road, Memphis, TN 38116 |
| Submitter's telephone number: | 901-399-6707 |
| Contact person: | Gino J. Rouss |
| Date summary prepared: | May 29, 2002 |
| Trade or proprietary device name: | Smith & Nephew Image Guided Surgical<br>Instruments for Hip Applications |
| Common or usual name: | Stereotaxic Instrument |
| Classification name: | Stereotaxic Instrument |
| Device Class: | Class II |
| Device Product Code: | HAW |
| Panel Code: | Neurology/84 |
### Subject device description:
The Smith & Nephew Image Guided Instruments for Hip Applications are instruments that have been modified to allow image-guided arrays (Fighters) to be fixed onto the instruments. The image-guided arrays can use either infrared LEDs (light emitting diodes) or universal passive spheres to transmit or reflect the infrared LEDs (light emitting diodes) that are emitted by an IGS Platform System. Along with commercially available software, this will allow the instruments to be recognized and tracked in real time in the surgical field. The infrared LEDs (light emitting diodes) or universal passive spheres do not come in contact with the open wound during surgical procedures. The image-guided arrays that are affixed to the instruments work in conjunction with reference frames that are rigidly attached to the anatomy. Each reference frame is also fitted with either infrared LEDs (light emitting diodes) or universal passive spheres to transmit or reflect the infrared LEDs (light emitting diodes) that are emitted by the IGS Platform System. The reference frames will allow the IGS Platform System to continuously track the position of the anatomy during navigation. If any movement of the IGS Platform System or anatomy is detected, the system can compensate for it, thereby maintaining accurate navigation.
lufi
Image Guided Surgical Instruments for Hip Applications
CONFIDENTIAL
{1}------------------------------------------------
# A. Applicable 510(k)'s
| Manufacturer | Submission Name | FDA Clearance<br>Date | Exhibit<br>No. |
|----------------------------------|---------------------------------------------------------|-----------------------|----------------|
| Surgical Navigation Technologies | StealthStation | 1-24-96 | 9 |
| Surgical Navigation Technologies | StealthStation™ System Passive Instrument Option | 9-16-97 | 10 |
| Surgical Navigation Technologies | StealthStation™ System -ENT Application Addendum | 1-21-98 | 11 |
| Surgical Navigation Technologies | StealthStation® FluoroNav™ Module | 4-22-99 | 12 |
| Surgical Navigation Technologies | Indications Modifications for the StealthStation System | 2-22-00 | 13 |
| Surgical Navigation Technologies | StealthStation Generation 3 | 5-3-00 | 14 |
| BrainLAB AG | VectorVision® Hip | 9-12-01 | 15 |
| BrainLAB AG | VectorVision2 | 5-19-99 | 16 |
| Smith & Nephew, Inc. | Image Guided Surgical Instruments for Knees | 2-8-02 | 17 |
| Surgical Navigation Technologies | Knee Module For The StealthStation™ System | 1-25-02 | 18 |
## Subject device intended use:
Image Guided Surgical Instruments for Hip Applications are intended to assist the surgeon in precisely locating anatomical structures in either open or percutaneous procedures. Image Guided Surgical Instruments for Hip Applications are indicated for use in surgical hip procedures, in which the use of stereotactic surgery may be appropriate, and where a reference to a rigid anatomical structure such as a long bone can be identified relative to a CT or MR based model or fluoroscopy images of the anatomy. These procedures include but are not limited to, acetabular cup placement as part of a partial or total hip arthroplasty (primary or revision). A brief surgical technique using these instruments is provided in Exhibit 7.
# Technological Characteristics:
Image Guided Surgical Instruments for Hip Applications are similar to currently legally marketed Class II stereotactic instruments in that they incorporate infrared LED (light emitting diodes) or passive spheres onto the instruments that allow the instruments to be recognized and tracked in real time in the surgical field.
2 u t 2
Image Guided Surgical Instruments for Hip Applications
CONFIDENTIAL
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing right, with flowing lines beneath them, representing the department's mission to protect the health of all Americans and provide essential human services.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Smith & Nephew, Inc. Gino J. Rouss Orthopaedic Division 1450 East Brooks Road Memphis, Tennessee 38116
Re: K021798
Trade/Device Name: Smith & Nephew Image Guided Surgical Instruments for Hip
Applications Regulation Number: 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: HAW Dated: October 25, 2002 Received: October 28, 2002
Dear Mr. Rouss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the
{3}------------------------------------------------
Page 2 – Mr. Gino J. Rouss
quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) 1 premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Parrott
fc Celia M. Witten, Ph.D., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of
# 510(k) Number (if known): KO2
Device Name: Image Guided Surgical Instruments for Hip Applications
Indications For Use:
Image Guided Surgical Instruments for Hip Applications are intended to assist the surgeon in precisely locating anatomical structures in either open or percutaneous procedures. Image Guided Surgical Instruments for Hip Applications are indicated for use in surgical hip procedures, in which the use of stereotactic surgery may be appropriate, and where a reference to a rigid anatomical structure such as a long bone can be identified relative to a CT or MR based model or fluoroscopy you images of the anatomy. These procedures include but are not limited to, acetabular cup placement as part of a partial or total hip arthroplasty (primary or revision).
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
| Prescription Use<br>(Per 21 CFR 801.109) | OR Over-The-Counter Use |
|------------------------------------------|-------------------------|
|------------------------------------------|-------------------------|
(Optional Format 1-2-96)
Miriam C Provost
eral, Restorative
Ko21798
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.