K965182 · Radionics, Inc. · GXD · Mar 12, 1997 · Neurology
Device Facts
Record ID
K965182
Device Name
RFG-3CF
Applicant
Radionics, Inc.
Product Code
GXD · Neurology
Decision Date
Mar 12, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4400
Device Class
Class 2
Attributes
Therapeutic
Intended Use
To create lesions during neurosurgical lesion procedures.
Device Story
Radiofrequency lesion generator; used in neurosurgery to coagulate/lesion nervous tissue and stimulate nerve cells for electrode placement. Input: user-selected stimulation parameters (pulse amplitude, duration, frequency). Operation: delivers radiofrequency energy; provides constant voltage or constant current stimulation modes. Output: RF energy for lesioning; audio/visual indicators for system status. Operated by neurosurgeons/clinicians in clinical settings. Benefits: precise tissue lesioning and guided electrode positioning. Safety features: power-up self-tests for software/LEDs; watchdog timer for software monitoring/rebooting.
Clinical Evidence
Bench testing only. System and unit testing results verify accuracy and reliability of the RFG-3CF, including validation of new self-test routines and watchdog timer functionality.
Technological Characteristics
Radiofrequency lesion generator; constant voltage and constant current stimulation modes; stimulation frequency up to 200 Hz; audio/visual indicators; power-up self-test routines; watchdog timer for software monitoring.
Indications for Use
Indicated for neurosurgical patients requiring tissue lesioning or coagulation; used by clinicians to create lesions in nervous tissue and stimulate nerve cells to assist in electrode placement.
Regulatory Classification
Identification
A radiofrequency lesion generator is a device used to produce lesions in the nervous system or other tissue by the direct application of radiofrequency currents to selected sites.
Predicate Devices
Radionics RFG-3C Lesion Generator
Fischer Neuro N50 System
Leksell Neuro Generator, LNG 30-T
Grass Instruments' Stimulator Model S5 with Constant Current Unit, Accessory Model CCU1
Related Devices
K201610 — IonicRF Generator · ABBOTT MEDICAL · Oct 21, 2020
K111576 — NT 2000 LESIONING GENERATOR · Neuro Therm, Inc. · Sep 20, 2011
K231675 — OneRF Ablation System · Neuroone Medical Technologies Corp. · Dec 6, 2023
K220122 — APEX 6 · Rf Innovations, Inc. · Mar 15, 2023
K050084 — COSMAN RF LESION GENERATOR, MODEL RFG-1A; COSMAN TEW TRIGEMINAL ELECTRODE KIT; COSMAN TIC TRIGEMINAL ELECTRODE KIT; · Cosman Medical, Inc. · Mar 17, 2005
Submission Summary (Full Text)
{0}
Radionics RFG-3CF Lesion Generator
K965182
# Appendix 1
MAR 12 1997
## Summary of Safety and Effectiveness
### General Information
| Classification: | Class II |
| --- | --- |
| Common Name | Radiofrequency Lesion Generator |
| Device Trade Name: | RFG-3CF |
| Intended Uses | To create lesions during neurosurgical lesion procedures. |
| Predicate Device: | Radionics RFG-3C Lesion Generator; Fischer Neuro N50 System; Leksell Neuro Generator, LNG 30-T; Grass Instruments' Stimulator Model S5 with Constant Current Unit, Accessory Model CCU1 |
| Establishment Name and Address: | Radionics, Inc.
22 Terry Avenue
Burlington, MA 01803 |
| Contact Name and Phone: | Amy J. LaForte, Ph.D. |
| | (617) 272-1233 |
| Establishment Registration Number: | 1219140 |
| Performance Standard | None established under Section 514. |
### Substantial Equivalence Determination
A summary of the information contained in this premarket notification that addresses safety and effectiveness follows.
#### Safety Summary
The Radionics RFG-3CF lesion generator is a modification of the RFG-3C lesion generator. The stimulator feature has been modified to include a constant current mode and the stimulation parameter specification for frequency has been increased to 200 Hz. The system and unit testing results provided in this premarket notification verify that the RFG-3CF is accurate and reliable. Audio and visual indicators and displays inform the user regarding the proper function of the RFG-3CF. In addition, new self-tests routines have been included to check the software and front panel LEDs upon power-up. In addition, a watchdog timer has been added which will shut down the RFG-3CF and re-boot the software, if the software was not functioning properly.
#### General Safety and Effectiveness Concerns
The device labeling contains instructions for use. It includes indications for use, cautions, warnings and error conditions as well as general instructions for the proper operation of the device. This information promotes safe and effective use of the device.
#### Description of the Device and Basis for Substantial Equivalence
The RFG-3CF, addressed in this premarket notification, has the same intended use and technological characteristics as the commercially available Radionics RFG-3C, Fischer Neuro 50, and Leksell LNG 30-T generators. Like these devices, the RFG-3CF is a radiofrequency lesion generator used to coagulate or lesion nervous tissue. In addition, the stimulator feature of the RFG-3CF has the same intended use and technological characteristics as the that of the predicate devices mentioned as well as the Grass Stimulator with Constant Current Unit. The stimulator feature is used to stimulate nerve cells to aid in the placement of electrodes. The stimulator is the major feature modification of the RFG-3CF from the RFG-3C. In the RFG-3CF, the user may choose
Page 43
{1}
Radionics RFG-3CF Lesion Generator
from constant voltage or constant current stimulation. The user may also select the stimulator parameters such as pulse amplitude, duration and frequency. Radionics believes that the information and testing provided in this premarket notification clearly describe the modifications to the RFG-3CF (from the RFG-3C) and demonstrate that it is equivalent to the mentioned commercially marketed predicate devices.
Page 44
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.