The TMS Navigator TS helps users to plan, implement and document treatment involving TMS of the brain. The system provides planning and navigation functions using anatomical MR data. Regions of the brain to be stimulated can be determined on the basis of anatomy, functional areas or by entering previously calculated coordinates from brain atlases. The product is intended for use with the MagVenture therapy systems supplied by Tonica Elektronik A/S (Farum, Denmark), more precisely the R20 and R30 stimulator with MagOption, the X100 stimulator and the X100 stimulator with MagOption only with following magnetic coils: C-100, C-860, Cool-B65, Cool-B70, Cool D-B80, MC-125, MC-870, MCF-75, MCF-125, MCF-B65 and MMC-140-II.
Device Story
Navigation system for transcranial magnetic stimulation (TMS); uses optical tracking to monitor patient head and TMS coil positions; inputs include anatomical MR data and real-time spatial coordinates; software enables planning of stimulation sites, real-time visualization of coil position relative to brain anatomy, and documentation of stimulated locations; used in clinical settings by healthcare providers; facilitates accurate coil placement to target specific brain regions; benefits include improved precision in TMS delivery compared to non-navigated methods; does not include a stimulator; functions as an accessory to MagVenture therapy systems.
Clinical Evidence
Bench testing only. Study measured system accuracy across 36 navigated TMS sessions using a measurement phantom and test probe. 288 coordinate comparisons were performed. Results showed a mean system accuracy of 3.47 mm (95% CI: 3.40 mm, 3.53 mm). Excluding phantom imperfections, accuracy was 2.19 mm (95% CI: 2.11 mm, 2.26 mm). Results were compared to the predicate device (K191422), which reported mean accuracies of 3 mm and 5 mm.
Technological Characteristics
Optical tracking system for spatial localization; software-based navigation using 3D MRI data; compatible with specific MagVenture TMS coils (C-100, C-860, Cool-B65, Cool-B70, Cool D-B80, MC-125, MC-870, MCF-75, MCF-125, MCF-B65, MMC-140-II); standalone navigation workstation; no integrated energy source for stimulation.
Indications for Use
Indicated for planning, implementing, and documenting transcranial magnetic stimulation (TMS) of the brain in patients requiring navigated stimulation, using anatomical MR data to target specific brain regions.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Nexstim Navigated Brain Therapy (NBT) System 2 (K171902)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 17, 2023
Localite GmbH Arno Schmitgen Managing director Auguststr. 1 Bonn. 53229 Germany
Re: K223577
Trade/Device Name: Localite TMS Navigator TS Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: March 16, 2023 Received: March 16, 2023
## Dear Arno Schmitgen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Pamela D. Scott -S
Pamela D. Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223577
Device Name Localite TMS Navigator TS
#### Indications for Use (Describe)
The TMS Navigator TS helps users to plan, implement treatment involving TMS of the brain.
The system provides planning and navigation functions using anatomical MR data.
Regions of the brain to be stimulated can be determined on the basis of anatomy, functional areas or by entering previously calculated coordinates from brain atlases.
The product is intended for use with the MagVenture therapy systems supplied by Tonica Elektronik A/S (Farum, Denmark), more precisely the R20 and R30 stimulator with MagOption, the X100 stimulator and the X100 stimulator with MagOption only with following magnetic coils: C-100, C-B60, Cool-B70, Cool D-B80, MC-125, MC-B70, MCF-75, MCF-125, MCF-B65 and MMC-140-II.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px;"> ☒ </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px;"> ☐ </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) #: K223577
510(k) Summary
Prepared on: 2023-04-18
21 CFR 807.92(a)(1)
## Contact Details
| Applicant Name | Localite GmbH |
|-----------------------------|---------------------------------|
| Applicant Address | Auguststr. 1 Bonn 53229 Germany |
| Applicant Contact Telephone | 004922841012220 |
| Applicant Contact | Mr. Arno Schmitgen |
| Applicant Contact Email | quality@localite.de |
21 CFR 807.92(a)(2)
| Device Trade Name | Localite TMS Navigator TS |
|---------------------|-------------------------------------|
| Common Name | Stereotaxic instrument |
| Classification Name | Neurological Stereotaxic Instrument |
| Regulation Number | 882.4560 |
| Product Code | HAW |
21 CFR 807.92(a)(3)
Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K191422 | The Neural Navigator | HAW |
| K171902 | Nexstim Navigated Brain Therapy (NBT) System 2 | HAW |
21 CFR 807.92(a)(4)
Device Description SummaryThe TMS Navigator TS is a navigation system specifically designed to support transcranial magnetic stimulation (TMS).
It uses position data of the patient's head and the TMS coil acquired by an optical tracking system for display and navigation. The realtime display of the position of the TMS coil in the displayed MR data set provides visual support for navigation. With the help of the software, one or more locations for the TMS treatment can be planned as well as stimulated locations can be recorded in the used data set and thus documented.
These functionalities support the following use case:
Planning, execution and documentation of a TMS stimulation in the human brain.
Transcranial magnetic stimulation (TMS) uses so-called TMS coils to generate alternating electromagnetic fields to target brain areas. The depth of penetration, extent and nature of this effect (inhibitory or exhibitory) is influenced by the strength and shape of the magnetic field and the repetition frequency used.
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The TMS Navigator TS helps users to plan, implement and document treatment involving TMS of the brain.
The system provides planning and navigation functions using anatomical MR data.
{4}------------------------------------------------
Regions of the brain to be stimulated can be determined on the basis of anatomy, functional areas or by entering previously calculated coordinates from brain atlases.
The product is intended for use with the MagVenture therapy systems supplied by Tonica Elektronik A/S (Farum, Denmark), more precisely the R20 and R30 stimulator with MagOption, the X100 stimulator and the X100 stimulator with MagOption only with following magnetic colls: C-100, C-860, Cool-B65, Cool-B70, Cool D-B80, MC-125, MC-870, MCF-75, MCF-B65 and MMC-140-II.
## Indications for Use Comparison
The intended use of the proposed device is as follows:
The TMS Navigator TS helps users to plan, implement and document treatment involving TMS of the brain.
The system provides planning and navigation functions using anatomical MR data.
Regions of the brain to be stimulated an be determined on the basis of anatomy, functional areas or by entering previously calculated coordinates from brain atlases.
The product is intended for use with the MagVenture therapy systems supplied by Tonica Elektronik A/S (Farum, Denmark), more precisely the R20 and R30 stimulators, the R30 TMS stimulator with MagOption, the X100 stimulator with MagOption only with following magnetic colls: C-100, C-860, Cool-B65, Cool-B70, Cool D-B80, MC-125, MC-870, MCF-75, MCF-125, MCF-B65 and MMC-140-II.
The primary predicate (K191422) is "indicated for accurate positioning of the treatment coll [ ... ] with respect to target brain regions based on data obtained from MRI measurements" according to its intended use. The reference predicate (K171902) provides similar intentions as the proposed device and can thus be seen as substantially similar.
The primary predicate (K191422) mentioned reference predicate (K171902). For the reference device (K171902) the intended use is summed up as follows: Nextim Navigated Brain Therapy (NBT) System 2 is indicated for the treatment of Major Depressive Disorder in adult patients who have satisfactory improvement from prior antidepressant medication in the current episode.
Comparing the intended use statements the following rationale reasons the difference:
The reference device comes as a complete system including a Transcranial Magnetic Stimulator (product code OBP), evoked response measurement and Electromyograph (product codes GWF, IKN) besides the navigation system for transcranial magnetic stimulation (product code HAW). Refering to the device description of the reference predicate the explanation is as follows: The Nexstim NBT System 2 consists of a group of devices designed to localize the stimulation site in the brain and deliver rTMS stimulation using controlling and interpretive software. Operational control of the Nextim NBT System 2 is provided by the software. The Nexstim NBT System 2 combines magnetic resonance imaging based), three-dimensional (3-D) localization of cortical motor areas of the brain with non-invasive TMS and simultaneous electromyography (E.MG) measurement to locate areas of the brain that are capable of evoking muscle responses when stimulated, and to locate the target area for depression therapy.
The proposed device covers the application of the brain area to be stimulated with MagVenture Therapy Systems from Tonica Elektronik (Farum, Denmark), For the localization of the treatment spot 3D MRI data is used to determine and locate targets for brain stimulation. During treatment the proposed navigation system helps to apply TMS on the previously determined target region.
In conclusion the proposed device including its application, safety and effectiveness is as well comparable to the reference predicate as a product subset.
## Technological Comparison
21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
In comparison to the two legally marketed predicate and reference devices (K191422 the subject device can be seen as substancially equivalent. The subject device does not include a TMS stimulator while it is intended to be used as an accessory for such as the predicate device (K191422). The intended use of the predicate device (K191422) is comparibly similar as both devices, the subject device as well as the predicate device (K191422), support TMS coil placement by MRI guidance to identify and locate brain regions to be stimulated. A comparison of the intended use of the reference predicate (K171902) needs to be complemented with a detail look into the device descriptions. There the TMS application is combined with "magnetic resonance imagingbased (MR-based), threedimensional (3-D) localization of cortical motor areas of the brain (...) and to locate the target area for depression therapy "which is similar to the intended use of the subject device.
{5}------------------------------------------------
The table below shows and over different technological characteristics, key components and features and furthermore provide general details about the devices as an overview. In summary the proposed device can be seen as substantially equivalent to its predicate device (K191422) and reference device (K171902).
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
## Summary nonclinical tests
The TMS Navigator TS was tested for performance in accordance with the following FDA Guidance Document:
Guidance for Industry and Food and Drug Administration Staff: Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions (December 2019)
A complete test report in accordance with the referenced FDA Guidance Document is provided in the attachments:
Test Report Bench Performance Test: Study measuring the system accuracy of typical navigated TMS sessions with the TMS Navigator TS using a measurement phantom and a TMS coil test probe in a simulated use environment demonstrating substantially equivalent results compared to comparable navigated TMS systems on the market
### Conclusion
In total, 288 cordinate comparisons were deduced from eight test locations across 36 navigated TMS sessions were performed with the TMS Navigator TS using a measurement phantom and a TMS coil test probe in a simulated use environment and varying hardware setups including all steps used in a real use scenario. On average, a system accuracy of 3.47 mm with a 95% Cl of [3.40 mm. 3.53 mm] was derived across all sessions and based on eight test locations on the phantom, to which the test probe was navigated. Furthermore excluding the effects and possible imperfections of the measurement phantom on the navigation. a system accuracy of 2.19 mm with a 95% Cl of [2.11 mm, 2.26 mm] was obtained. It may be concluded that the achieved results of the TMS Navigator TS are comparable to or better than the values reported from comparion systems, especially the system chosen as predicate device "K191422" (mean system accuracy of 3 mm +/- 2.1 mm when navigating with a hand-held probe and 5 mm +/- 2.1 mm when navigating with a TMS coil).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.