K201175 · Medtronic Navigation, Inc. · HAW · Jun 3, 2020 · Neurology
Device Facts
Record ID
K201175
Device Name
StealthStation Cranial Software v1.3.0
Applicant
Medtronic Navigation, Inc.
Product Code
HAW · Neurology
Decision Date
Jun 3, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The StealthStation™ System, with StealthStation™ Cranial Software, is intended as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy.
Device Story
StealthStation Cranial Software v1.3.0 is an image-guided surgery (IGS) system component. It processes diagnostic or intraoperative images (X-ray, MR, Nuclear Medicine) to track surgical instruments in real-time relative to patient anatomy. The system uses optical or electromagnetic tracking to update instrument positions on a display. Used in ORs by neurosurgeons, the software provides registration, planning, and navigation visualization. It aids surgeons in locating anatomical structures, guiding instrument trajectories, and performing biopsies or resections. By providing continuous visual feedback of instrument position relative to pre-operative or intra-operative imaging, the device assists in precise surgical navigation, potentially improving accuracy and patient outcomes in cranial procedures.
Clinical Evidence
Bench testing only. No clinical data. Performance validated using anatomically representative phantoms under worst-case configurations. Software verification and validation testing performed for all requirement specifications. System integration performance testing confirmed accuracy metrics (3D positional error ≤ 2.0 mm; trajectory angle error ≤ 2.0 degrees).
Technological Characteristics
Software-based IGS system. C++ programming language. Supports X-ray, MR, and Nuclear Medicine imaging. Connectivity via network, CD, DVD, USB; DICOM compliant. Optical tracking (Northern Digital Vega) and electromagnetic tracking (Medtronic AxiEM III). Integrates with various stereotactic frames and surgical instruments. Software features include registration (PointMerge, Tracer, Touch, StealthAiR, O-arm, mechanical), 3D model building, and brain atlas integration.
Indications for Use
Indicated for patients undergoing neurosurgical procedures where stereotactic surgery is appropriate, including tumor resections, ventricular catheter placement (general and pediatric), depth electrode/lead/probe placement, and cranial biopsies.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 3, 2020
Medtronic Navigation Inc. Amelia Striegel Senior Regulatory Affairs Specialist 826 Coal Creek Circle Louisville, Colorado 80027
Re: K201175
Trade/Device Name: StealthStation Cranial Software v1.3.0 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: April 30, 2020 Received: May 1, 2020
Dear Amelia Striegel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce Acting Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K201175
Device Name StealthStation™ Cranial Software v1.3.0
### Indications for Use (Describe)
The StealthStation™ System, with StealthStation™ Cranial Software, is intended as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy.
This can include, but is not limited to, the following cranial procedures (including stereotactic frame-based and stereotactic frame alternatives-based procedures):
- · Tumor resections
- General ventricular catheter placement
- · Pediatric ventricular catheter placement
- · Depth electrode, lead, and probe placement
- · Cranial biopsies
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> <path d="M2,2 L10,10 M2,10 L10,2" stroke="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
cription 65c (Part 21 CFR 601 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
30 April 2020
- l. Company: Medtronic Navigation. Inc. 826 Coal Creek Circle Louisville, Colorado 80027 USA Telephone Number: 720-890-3200
- Amelia Striegel (Primary) Contact: Senior Regulatory Affairs Specialist Telephone Number: 720.890.2528 Fax Number: 720.890.3500
K. Elizabeth Waite Regulatory Affairs Manager Telephone number: 720.890.2182 Fax: 720.890.3500 Email: elizabeth.waite@medtronic.com
- Proprietary Trade Name: StealthStation™ Cranial Software v1.3.0 II.
- lll. Classification Name: Stereotaxic Instrument (21 CFR 882.4560)
- IV. Classification: Class II, Stereotaxic Instrument
- V. Product Codes: HAW
- VI. Predicate Device
510(k): K162309 Device name: StealthStation S8 Cranial Software v1.0.0 Manufacturer: Medtronic Navigation, Inc.
#### VII. Product Description
The StealthStation™ Cranial Software v1.3.0 works in conjunction with an Image Guided System (IGS) which consists of clinical software, surgical instruments, a referencing system and platform/computer hardware. Image guidance, also called navigation, tracks the position of instruments in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of the patient. During surgery, positions of specialized surgical instruments are continuously updated on these images either by optical tracking or electromagnetic tracking.
Cranial software functionality is described in terms of its feature sets which are categorized as imaging modalities, registration, planning, interfaces with medical devices, and views. Feature sets include functionality that contributes to clinical decision making and are necessary to achieve system performance.
The changes to the currently cleared StealthStation S8 Cranial Software are as follows:
- . Addition of an optional image display that allows the user to see through outer layers to increase the visibility of other models.
{4}------------------------------------------------
- . Update the imaging protocol to support overlapping slices.
- . Minor changes to the software were made to address user preferences and to fix minor anomalies.
#### VIII. Indications for Use
The StealthStation™ System, with StealthStation™ Cranial software, is intended as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy.
This can include, but is not limited to, the following cranial procedures (including stereotactic frame-based and stereotactic frame alternatives-based procedures):
- · Tumor resections
- General ventricular catheter placement
- Pediatric ventricular catheter placement
- · Depth electrode, lead, and probe placement
- · Cranial biopsies
#### IX. Comparison of the Technological Characteristics
| Feature | StealthStation Cranial<br>Software v1.3.0 (Subject<br>Device) | StealthStation S8 Cranial<br>software v1.0.0 (Predicate<br>Device; K162309) |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The StealthStationTM System, with<br>StealthStationTM Cranial Software is<br>designed as an aid for locating<br>anatomical structures in either open<br>or percutaneous neurosurgical<br>procedures. | The StealthStation® System, with<br>StealthStation® Cranial software is<br>designed as an aid for locating<br>anatomical structures in either open<br>or percutaneous neurosurgical<br>procedures. |
| Feature | StealthStation Cranial<br>Software v1.3.0 (Subject<br>Device) | StealthStation S8 Cranial<br>software v1.0.0 (Predicate<br>Device; K162309) |
| Indications for<br>Use | The StealthStation™ System, with<br>StealthStation™ Cranial Software,<br>is intended as an aid for locating<br>anatomical structures in either open<br>or percutaneous neurosurgical<br>procedures. Their use is indicated<br>for any medical condition in which<br>the use of stereotactic surgery may<br>be appropriate, and where<br>reference to a rigid anatomical<br>structure, such as the skull, can be<br>identified relative to images of the<br>anatomy. | The StealthStation™ System, with<br>StealthStation™ Cranial software, is<br>intended as an aid for locating<br>anatomical structures in either open<br>or percutaneous neurosurgical<br>procedures. Their use is indicated<br>for any medical condition in which<br>the use of stereotactic surgery may<br>be appropriate, and where<br>reference to a rigid anatomical<br>structure, such as the skull, can be<br>identified relative to images of the<br>anatomy. |
| | This can include, but is not limited<br>to, the following cranial procedures<br>(including stereotactic frame-based<br>and stereotactic frame alternatives-<br>based procedures):<br>• Tumor resections<br>• General ventricular catheter<br>placement<br>• Pediatric ventricular catheter<br>placement<br>• Depth electrode, lead, and probe<br>placement<br>• Cranial biopsies | This can include, but is not limited<br>to, the following cranial procedures<br>(including stereotactic frame-based<br>and stereotactic frame alternatives-<br>based procedures):<br>• Tumor resections<br>• General ventricular catheter<br>placement<br>• Pediatric ventricular catheter<br>placement<br>• Depth electrode, lead, and probe<br>placement<br>• Cranial biopsies |
| System<br>Accuracy<br>Requirements | Under representative worst-case<br>configuration, the StealthStation S8<br>System with StealthStation Cranial<br>Software v1.3.0, has demonstrated<br>performance in 3D positional<br>accuracy with a mean error ≤ 2.0<br>mm and in trajectory angle<br>accuracy with a mean error ≤ 2.0<br>degrees. | Under representative worst-case<br>configuration, the StealthStation S8<br>System with StealthStation Cranial<br>v1.0.0 Software, has demonstrated<br>performance in 3D positional<br>accuracy with a mean error ≤ 2.0<br>mm and in trajectory angle<br>accuracy with a mean error ≤ 2.0<br>degrees. |
| Imaging<br>Modalities | X-Ray based, MR based,<br>Nuclear Medicine based | X-Ray based, MR based,<br>Nuclear Medicine based |
| View (Display)<br>Features | Ultrasound Video In, Ultrasound<br>Overlay, 3D, 2D Anatomic<br>Orthogonal, Trajectory 1 and 2,<br>Target Guidance, Trajectory<br>Guidance, Probes Eye, Look<br>Ahead, Microscope Injection, Video<br>Input, Endoscopic | Ultrasound Video In, Ultrasound<br>Overlay, 3D, 2D Anatomic<br>Orthogonal, Trajectory 1 and 2,<br>Target Guidance, Trajectory<br>Guidance, Probes Eye, Look<br>Ahead, Microscope Injection, Video<br>Input, Endoscopic |
| Feature | StealthStation Cranial<br>Software v1.3.0 (Subject<br>Device) | StealthStation S8 Cranial<br>software v1.0.0 (Predicate<br>Device; K162309) |
| Exam-to-Exam<br>Registration<br>Features | Identity Merge Registration, Manual<br>Merge Registration and Automatic<br>Merge Registration | Identity Merge Registration, Manual<br>Merge Registration and Automatic<br>Merge Registration |
| Patient<br>Registration<br>Features | PointMergeTM registration (referred<br>to as Landmark registrations),<br>TracerTM registration, Touch<br>registration (previously Touch-N-<br>GoTM), StealthAiRTM registration, O-<br>armTM registration, Mechanical<br>based registrations (Stereotactic<br>Localizer Registration and StarFixTM<br>Bone Anchor Registration) | PointMerge® registration (referred<br>to as Landmark registrations),<br>TracerTM registration, Touch<br>registration (previously Touch-N-<br>GoTM), StealthAiR® registration, O-<br>arm® registration, Mechanical<br>based registrations (Stereotactic<br>Localizer Registration and StarFixTM<br>Bone Anchor Registration) |
| Planning<br>Features | Plan Entry and Target Selection<br>3D Model Building<br>Advanced Visualization<br>Create Patient Based Anatomical<br>Coordinate Space<br>Stereotactic Frame Settings<br>Brain Atlas: Schaltenbrand-Wahren<br>Atlas with Talairach Grid<br>STarFixTM Designer<br>Annotations | Plan Entry and Target Selection<br>3D Model Building<br>Advanced Visualization<br>Create Patient Based Anatomical<br>Coordinate Space<br>Stereotactic Frame Settings<br>Brain Atlas: Schaltenbrand-Wahren<br>Atlas with Talairach Grid<br>STarFixTM Designer<br>Annotations |
| Medical Device<br>Interfaces | Microscope Navigation: Zeiss, Leica<br>Ultrasound Navigation: Aloka and<br>Sonosite<br>Medtronic O-armTM<br>Stereotactic Frame Systems:<br>Fischer ZD, Fischer RM, Integra<br>CRW and Elekta Leksell<br>NexframeTM Stereotactic System<br>STarFixTMPlatform System<br>Stealth Midas Rex MR8<br>StealthStation Autoguide | Microscope Navigation: Zeiss, Leica<br>Ultrasound Navigation: Aloka and<br>Sonosite<br>Medtronic O-arm®<br>Stereotactic Frame Systems:<br>Fischer ZD, Fischer RM, Integra<br>CRW and Elekta Leksell<br>Nexframe® Stereotactic System<br>STarFixTMPlatform System |
| Compatible<br>Medtronic<br>Optical<br>Instrumentation | Medtronic instruments tracked via<br>optical markers or LEDs located on<br>instrument and patient trackers via<br>the optical localizing system. | Medtronic instruments tracked via<br>optical markers or LEDs located on<br>instrument and patient trackers via<br>the optical localizing system. |
| Compatible<br>Medtronic EM<br>Instrumentation | Medtronic instruments tracked via<br>electromagnetic localization<br>technology located within the<br>instrument and patient trackers | Medtronic instruments tracked via<br>electromagnetic localization<br>technology located within the<br>instrument and patient trackers |
| Feature | StealthStation Cranial<br>Software v1.3.0 (Subject<br>Device) | StealthStation S8 Cranial<br>software v1.0.0 (Predicate<br>Device; K162309) |
| Software<br>Interface<br>(GUI) | Black and gray style with procedure<br>task overview in left menu option<br>and next/back task flow at bottom of<br>the screen. Software controls for<br>images, planning and instrument<br>management are contained in a<br>right side bar. | Black and gray style with procedure<br>task overview in left menu option<br>and next/back task flow at bottom of<br>the screen. Software controls for<br>images, planning and instrument<br>management are contained in a<br>right side bar. |
| Programming<br>Language | C++ | C++ |
| Scanner<br>Interface<br>Technology (to<br>imaging<br>devices) | Network Connectivity<br>CD, DVD, USB<br>DICOM Import<br>DICOM Export | Network Connectivity<br>CD, DVD, USB<br>DICOM Import<br>DICOM Export |
| Optical<br>Technology | Manufacturer: Northern Digital Inc.<br>Localizer: Vega | Manufacturer: Northern Digital Inc.<br>Localizer: Vega |
| Electromagnetic<br>Technology | Manufacturer: Medtronic<br>Navigation, Inc.<br>Localizer: AxiEM III<br>Emitter Types: Side, Flat | Manufacturer: Medtronic<br>Navigation, Inc.<br>Localizer: AxiEM III<br>Emitter Types: Side, Flat |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
#### Identification of Legally Marketed Devices X.
StealthStation S8 Cranial Software v1.0.0 (K162309)
#### XI. Discussion of the Performance Testing
The following table summarizes the testing conducted on the StealthStation Cranial Software v1.3.0
### Description
Under representative worst-case configuration, the StealthStation S8 System with StealthStation Cranial Software v1.3.0 has demonstrated performance in 3D positional accuracy with a mean error ≤ 2.0 mm and in trajectory angle accuracy with a mean error of ≤ 2.0 degrees. This performance was determined using anatomically representative phantoms and utilizing a subset of system components and features that represent the worst-case combinations of all potential system components. Software verification and validation testing for each requirement specification. System integration performance testing for cranial surgical procedures using anatomical phantoms.
The following table summarizes the quality assurance measures that were applied during development of the software component of the system:
| Description |
|-------------------------------------------------------|
| Software Development Life Cycle |
| Software Risk Assessment |
| Software Configuration Management and Version Control |
{8}------------------------------------------------
Design verification and validation was performed using the StealthStation S8 System with StealthStation Cranial Software v1.3.0 in laboratory and simulated use settings.
The results support the safety of the device and demonstrate that the software should perform as intended in the specified use conditions.
Clinical testing was not considered necessary prior to release as this is not new technology.
#### XII. Conclusions
The StealthStation Cranial Software v1.3.0 has been shown through testing and comparison to be substantially equivalent to the identified predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.