K190887 · Elekta Instrument AB · HAW · Aug 1, 2019 · Neurology
Device Facts
Record ID
K190887
Device Name
Leksell Vantage Stereotactic System
Applicant
Elekta Instrument AB
Product Code
HAW · Neurology
Decision Date
Aug 1, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intended Purpose of Leksell® Vantage™ Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.
Device Story
Leksell Vantage Stereotactic System provides coordinate referencing and skull fixation for neurosurgical procedures. System includes head frame, fixation pins, stereotactic arc, and accessories. Operates on established center-of-arc principle (190 mm radius); uses Cartesian (x,y,z) coordinates to triangulate brain targets. Input includes patient anatomical images (CT/MR) with fiducial markers. Used in operating rooms by neurosurgeons to guide interventional instruments (biopsy needles, cannulas, electrodes) to specific intracranial targets. Output is physical guidance of instruments at precise trajectory angles. Benefits include accurate, minimally invasive access to brain targets for stimulation, recording, lesioning, or drug delivery.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated through bench testing, including system/subsystem verification, accuracy testing (total system, mechanical, and handling), MR safety compliance (ASTM F2052, F2119, F2182, F2213, F2503), and design/usability validation by qualified personnel.
Technological Characteristics
Materials: PEEK with Aluminum tip (invasive components), Glass fiber reinforced Epoxy (head frame), Aluminum (arc). Sensing/Actuation: Mechanical coordinate referencing and instrument guidance. Energy: None (mechanical). Connectivity: None. Sterilization: EtO (ISO 11135:2014, SAL 10^-6). Form factor: Head frame with semi-circular arc system.
Indications for Use
Indicated for patients requiring stereotactic neurosurgical procedures, including deep brain stimulation, lesioning, biopsies, targeted injections, aspirations, and minimally invasive tumor treatments, where target localization and head fixation are necessary.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 1, 2019
Elekta Instrument AB Alf Laurell Regulatory Affairs Engineer Kungstensgatan 18 Stockholm, 10393 SE
Re: K190887
Trade/Device Name: Leksell Vantage Stereotactic System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: July 1, 2019 Received: July 3, 2019
Dear Alf Laurell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Matthew Krueger Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190887
### Device Name
Leksell® Vantage™ Stereotactic System
### Indications for Use (Describe)
The Intended Purpose of Leksell® Vantage™ Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.
Type of Use (Select one or both, as applicable)
| <span></span> <svg height="15" width="15"> <rect height="15" style="fill:none;stroke:#000000;stroke-width:1" width="15"></rect> <path d="M2 2 L13 13 M13 2 L2 13" stroke="#000000" stroke-width="1"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span></span> <svg height="15" width="15"> <rect height="15" style="fill:none;stroke:#000000;stroke-width:1" width="15"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Dokumentnamn/Name of document Traditional 510(k)
| Issuer | Ref nr/Dok nr/Ref no/Doc no | Edition |
|--------------------------------------------------------|-----------------------------|---------|
| Alf Laurell | -- | -- |
| Regarding | Directory | |
| CT accessories for Leksell Vantage Stereotactic System | -- | |
# 510(k) Summary
As Required by 21 CFR 807.92(c) 510 (k) Summary
#### 1. Subscribers Name & Address
Elekta Instrument AB Kungstensgatan 18, P.O. Box 7593 SE-103 93 Stockholm, Sweden Tel: (011) 46 8 587 254 00 Fax: (011) 46 8 587 255 00 Official Correspondent: Mats Premfors, QA Manager
Date summary prepared: 2019-07-27
#### 2. Trade Name
Leksell® Vantage™ Stereotactic System
#### 3. Common Name
Stereotaxic instrument
#### Device Classification 4.
| Trade Name | Product Code | Class | Regulation Number |
|---------------------------------------|--------------|-------|-------------------|
| Leksell® Vantage™ Stereotactic System | HAW | II | 21 CFR 882.4560 |
#### Classification Panel 5.
Neurology
#### Predicate Device Identification 6.
| Legally marketed devices to which equivalence is being claimed | 510(k) # |
|----------------------------------------------------------------|----------|
| Leksell Stereotactic System | K080355 |
| Leksell® Vantage™ Stereotactic System (Primary) | K171123 |
#### 7. Other relevant submissions
N/A, no other relevant submissions
#### 8. Device Description
The Leksell® Vantage™ Stereotactic System is a device used for minimally invasive neurosurgical procedures. It enables coordinate referencing and fixation of the patient's skull
{4}------------------------------------------------
| Dokumentnamn/Name of document | Traditional 510(k) |
|-------------------------------|--------------------|
|-------------------------------|--------------------|
Traditional 510(k)
| Issuer | Alf Laurell |
|-----------------------------|--------------------------------------------------------|
| Ref nr/Dok nr/Ref no/Doc no | -- |
| Edition | -- |
| Regarding | CT accessories for Leksell Vantage Stereotactic System |
| Directory | -- |
and brain during image acquisition and treatment. The coordinate referencing enables target localization and accurate stereotactic treatment of brain targets; for example for aspirations and biopsies, electrode placements, injections of cells & drugs, heamatoma evacuations as well as lesioning using a number of methodologies.
The system consists of a head frame that is fixated to the patient skull by minimally invasive disposable fixation pins and a number of accessories for frame application and imaging as well as a stereotactic arc and operating room accessories for the sterile surgical procedure.
The system is based on the established Leksell center-of-arc principle, which allows guidance of compatible interventional instruments at any trajectory angle to the brain target. Compatible interventional instruments are based on the Leksell principle of an active working length of 190 mm (as the arc radius) and use the correct mechanical interfacing of the Arc instrument stop and guide holders. Various instruments, such as biopsy needles, cannulas, and electrodes can be used with the system to perform biopsies, hematoma evacuation and injections of radioactive nuclides and cytostatic agents, puncture of cysts, positioning electrodes for stimulation/ recording or lesioning and stereotactic endoscope guidance.
#### 9. Summary of clinical testinq
Clinical testing was deemed not required to demonstrate substantial equivalence with the predicate devices.
#### 10. Summary of Non Clinical and Performance testing
The verification activities have covered requirements at system and subsystem level to show that the design output meet the design input.
Design and usability validation of the system has been performed by competent and professionally qualified personnel to ensure that the product fulfils the intended use and user needs. The design and usability validation was also made to ensure that the risk control measures associated with functions related to safety (FRS) was effective.
Results from verification and validation testing demonstrates that conformance to applicable technical requirement specifications and user needs have been met showing that the system is safe to use and is substantially equivalent to the predicate devices.
Testing is summarized in the table below.
{5}------------------------------------------------
Dokumentnamn/Name of document Traditional 510(k)
| Issuer | Alf Laurell |
|-----------------------------|--------------------------------------------------------|
| Ref nr/Dok nr/Ref no/Doc no | -- |
| Edition | -- |
| Regarding | CT accessories for Leksell Vantage Stereotactic System |
| Directory | -- |
| Test | Test Method Summary | Results |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization | Sterility validation (EtO) according to<br>ISO 11135:2014 was performed to<br>demonstrate that the device have a<br>sterility assumed level (SAL) of 10-6.<br><br>The tests were performed on the<br>predicate device Leksell® Vantage™<br>Stereotactic System (K171123). | Results met pre-<br>established acceptance<br>criteria, demonstrating<br>equivalence to the<br>predicate device in terms<br>of sterility. |
| MR testing | The device was subjected to<br>compliance testing to applicable<br>consensus safety standards, such as:<br>ASTM F2052; ASTM F2119; ASTM<br>F2182, ASTM F2213 and ASTM F2503,<br>and FDA guidance regarding safety and<br>compatibility in the MR environment. | The result demonstrates<br>that the device is<br>compliant to the FDA<br>guidance and applicable<br>consensus standards. This<br>compliance demonstrates<br>substantial equivalence. |
| Accuracy | Accuracy was tested on different<br>levels of the subject device:<br>- Total system accuracy from<br>imaging to treatment.<br>- Total mechanical accuracy of<br>the system<br>- Mechanical accuracy of each<br>component<br>- Handling accuracy of various<br>steps like repeated mounting<br>of parts to each other and<br>scale setting accuracy.<br>- Displacement from load of<br>head when fixed in the frame. | All test cases were passed,<br>demonstrating that the<br>overall system accuracy of<br>the Vantage system when<br>used with the CT<br>accessories is substantial<br>equivalent to the predicate<br>device. |
| Design and Usability<br>validation | Design and usability validation of the<br>system has been performed by trained<br>competent and professionally qualified<br>personnel to ensure that the product<br>fulfils the intended use and user<br>needs. | No unacceptable human<br>factor risks were identified.<br>Residual risks have<br>been weighted against the<br>benefit of the product and<br>are deemed acceptable<br>demonstrating that the<br>Leksell® Vantage™<br>Stereotactic System used |
| | | with the CT accessories is<br>substantial equivalent to<br>the predicate device. |
{6}------------------------------------------------
Dokumentnamn/Name of document Traditional 510/lz)
Traditional 510(k)
| Issuer | Alf Laurell |
|-----------------------------|--------------------------------------------------------|
| Ref nr/Dok nr/Ref no/Doc no | -- |
| Edition | -- |
| Regarding | CT accessories for Leksell Vantage Stereotactic System |
| Directory | -- |
#### 11. Intended Use
The Intended Purpose of Leksell® Vantage™ Stereotactic System is target localization and fixation of the patient head in a coordinate system in order to perform stereotactic neurosurgical procedures, for example; deep brain stimulation, lesioning, biopsies, targeted injections, aspirations and minimal invasive tumor treatments.
## 12. Technological Characteristics
The CT accessories of Leksell® Vantage™ Stereotactic System has equivalent technological characteristics as its predicate device (K080355), for further details see the table below.
#### Substantial Equivalence 13.
The intended use, the basic functionality and the fundamental technological characteristics of the device Leksell® Vantage™ Stereotactic System remain unchanged with the introduction of the CT accessories and the Leksell Vantage Head Frame Holder.
Leksell® Vantage™ Stereotactic as compared to the predicate devices has:
- Same mechanical accuracy
- . Same coordinate system
- Same Center of Arc Principle
- Intended for localizing the same intracranial anatomical targets
- Used for the same clinical interventions and/or treatments
- Intended for use by the same health care professionals
In addition, the risk evaluation regarding endotoxin testing in routine production for product FirmFix™ 1 -5 has been updated.
The conclusion from verification and validation is that the Leksell® Vantage™ Stereotactic System is substantial equivalent to its predicate devices.
The table below shows a comparison between the Leksell® Vantage™ Stereotactic System and the predicate devices.
{7}------------------------------------------------
Dokumentnamn/Name of document Traditional 510(k)
| Issuer | Alf Laurell |
|-----------------------------|--------------------------------------------------------|
| Ref nr/Dok nr/Ref no/Doc no | -- |
| Edition | -- |
| Regarding | CT accessories for Leksell Vantage Stereotactic System |
| Directory | -- |
| | Subject Device | Predicate Device (K080355) | Predicate Device (K171123) |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Leksell® Vantage™ Stereotactic System | Leksell Stereotactic System® (K080355) | Leksell® Vantage™ Stereotactic System (K171123) |
| Regulation Number | 21 CFR 882.4560 | 21 CFR 882.4560 | 21 CFR 882.4560 |
| Product Code | HAW | HAW | HAW |
| Product Class | II | II | II |
| Description | Head Frame and semi- circular arc system used for the coordinate referencing of targets in the brain and the guidance of neurosurgical instruments to these brain targets. | Head Frame and semi- circular arc system used for the coordinate referencing of targets in the brain and the guidance of neurosurgical instruments to these brain targets. | Head Frame and semi- circular arc system used for the coordinate referencing of targets in the brain and the guidance of neurosurgical instruments to these brain targets. |
| Technical Features | Coordinate System: Leksell Coordinate System using Cartesian coordinates | Coordinate System: Leksell Coordinate System using Cartesian coordinates | Coordinate System: Leksell Coordinate System using Cartesian coordinates |
| | Fixation: Head Frame applied to patient skull with four fixation pins. | Fixation: Head Frame applied to patient skull with four fixation pins. | Fixation: Head Frame applied to patient skull with four fixation pins. |
| | Fixation Pins: Single use Fixation Pins and Inserts that secures the Vantage head frame to the patient skull using Leksell Vantage Keys | Fixation Pins: Titanium Fixation Pins, Reusable Fixation Pins and Disposable Inserts that secures the Leksell Coordinate Frame G to the patient skull using Instrument Screw Drivers | Fixation Pins: Single use Fixation Pins and Inserts that secures the Vantage head frame to the patient skull using Leksell Vantage Keys |
| Imaging modality | CT and MR | CT and MR | MR |
| Technical characteristics | CT accessories: The Leksell Vantage CT Fiducial Box is used for referencing of the stereotactic coordinate system on the patient's anatomical CT images. | CT accessories: The Open CT Indicator Box is used for referencing of the stereotactic coordinate system on the patient's anatomical CT images. | CT accessories: N/A |
| | Leksell® Vantage™ Head Frame interface: | Leksell® Vantage™ Head Frame interface: | Leksell® Vantage™ Head Frame interface: |
| | Subject Device | Predicate Device<br>(K080355) | Predicate Device<br>(K171123) |
| | Leksell® Vantage™ Frame<br>Holder is a modified OR<br>adapter. In addition to<br>holding the Leksell®<br>Vantage™ Head Frame in<br>the operating room, the<br>Frame Holder is also<br>interfacing the CT Fiducial<br>box via one hole on each<br>side. | N/A | Leksell® Vantage™ OR<br>adapter |
| Principles of Use | Center of the arc<br>principle using an arc<br>radius of 190 mm.<br>Cartesian coordinates<br>applied to the brain of<br>the patient by the use of<br>a head frame fixated to<br>the skull.<br>• comprises a headframe<br>and arc;<br>• is target centered<br>• uses Cartesian (x,y,z)<br>coordinates to<br>triangulate target<br>position in frame<br>• utilizes fiducial markers<br>that can be registered<br>to a stereotactic brain<br>atlas and/or<br>intracranial image;<br>• provides support and<br>accessories for surgical<br>instruments to the<br>target; | Center of the arc<br>principle using an arc<br>radius of 190 mm.<br>Cartesian coordinates<br>applied to the brain of<br>the patient by the use<br>of a head frame fixated<br>to the skull.<br>• comprises a<br>headframe and arc;<br>• is target centered<br>• uses Cartesian (x,y,z)<br>coordinates to<br>triangulate target<br>position in frame<br>• utilizes fiducial<br>markers that can be<br>registered to a<br>stereotactic brain<br>atlas and/or<br>intracranial image;<br>• provides support and<br>accessories for<br>surgical instruments<br>to the target; | Center of the arc<br>principle using an arc<br>radius of 190 mm.<br>Cartesian coordinates<br>applied to the brain of<br>the patient by the use<br>of a head frame fixated<br>to the skull.<br>• comprises a<br>headframe and arc;<br>• is target centered<br>• uses Cartesian (x,y,z)<br>coordinates to<br>triangulate target<br>position in frame<br>• utilizes fiducial<br>markers that can be<br>registered to a<br>stereotactic brain<br>atlas and/or<br>intracranial image;<br>• provides support and<br>accessories for<br>surgical instruments<br>to the target; |
| Material | Invasive components:<br>PEEK with Aluminum tip. | Invasive components:<br>Aluminum or Titanium. | Invasive components:<br>PEEK with Aluminum tip. |
| | Head Frame:<br>Glass fiber reinforced<br>Epoxy | Head Frame:<br>Aluminum (with glued<br>joints for insulation in<br>MRI) | Head Frame:<br>Glass fiber reinforced<br>Epoxy |
| | Subject Device | Predicate Device<br>(K080355) | Predicate Device<br>(K171123) |
| | Arc:<br>Aluminum | Arc:<br>Aluminum | Arc:<br>Aluminum |
| Accuracy | Mechanical accuracy 0,9 mm | Mechanical accuracy 0,9 mm | Mechanical accuracy 0,9 mm |
{8}------------------------------------------------
Dokumentnamn/Name of document Traditional 510(k)
| Issuer | Alf Laurell |
|-----------------------------|--------------------------------------------------------|
| Ref nr/Dok nr/Ref no/Doc no | -- |
| Edition | -- |
| Regarding | CT accessories for Leksell Vantage Stereotactic System |
| Directory | -- |
{9}------------------------------------------------
Dokumentnamn/Name of document Traditional 510/k)
Traditional 510(k)
| <i>Issuer</i> | Alf Laurell |
|------------------------------------|--------------------------------------------------------|
| <i>Ref nr/Dok nr/Ref no/Doc no</i> | -- |
| <i>Edition</i> | -- |
| <i>Regarding</i> | CT accessories for Leksell Vantage Stereotactic System |
| <i>Directory</i> | -- |
### Conclusion:
The conclusion is that the subject device is substantial equivalent to its predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.