K092192 · FHC, Inc. · HAW · Feb 12, 2010 · Neurology
Device Facts
Record ID
K092192
Device Name
WAYPOINT STEREOTACTIC SYSTEM
Applicant
FHC, Inc.
Product Code
HAW · Neurology
Decision Date
Feb 12, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The WayPoint™ Stereotactic System is intended to be used with commercially available stereotactic systems for neurosurgical procedures which require the accurate positioning of microelectrodes, stimulating electrodes, implantable electrodes or other instruments in the brain or nervous system.
Device Story
WayPoint™ Stereotactic System (WPS) provides patient-customized stereotactic fixtures for neurosurgical procedures; facilitates microelectrode recording, stimulation, and device implantation. System inputs include scanned patient images; software (WayPoint™ Planner or VoXim®) processes images to design a patient-specific Stereotactic Platform Adapter. Components include anchor/locator system, platform adapter, and mounting hardware. Used in clinical settings by neurosurgeons to secure instruments to skull-based anchor sets. Provides sub-millimeter targeting accuracy; allows alignment of surgical instruments with patient anatomy. Benefits include improved precision in target reaching compared to conventional frames.
Clinical Evidence
Bench testing only. Phantom-based studies confirmed sub-millimeter accuracy of the Stereotactic Platform Adapter. Comparison of intraoperative stereotactic coordinates to postoperative CT images demonstrated accuracy comparable to conventional frames. Preoperative target selection and final placement verification based on CT images showed error measurements comparable to standard stereotactic frame procedures.
Technological Characteristics
Stereotaxic instrument (21 CFR 882.4560). System comprises patient-customized stereotactic fixtures, skull-based anchor mounting sets, platform adapters, and planning software. Connectivity involves integration with patient imaging data (CT).
Indications for Use
Indicated for patients undergoing neurosurgical procedures requiring accurate positioning of microelectrodes, stimulating electrodes, implantable electrodes, or other instruments within the brain or nervous system.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
FHC, Inc. microTargeting 100 Drive System (K003776)
Howmedica Leibinger, Inc. Ost-Reg Marker System (K961120)
Elekta Instruments AB Leksell Stereotactic System (K972324)
Submission Summary (Full Text)
{0}------------------------------------------------
K0921.92
Image /page/0/Picture/1 description: The image shows the letters FHC in a bold, sans-serif font. The letters are black, and the background is white. The letters appear to be slightly distorted, as if they have been copied or scanned. The letters are arranged horizontally, with the F on the left, the H in the middle, and the C on the right.
FEB 1 2 2010
Microelectrodes and Instrumentation for Neuroscience
Research and Clinical microTargetingTM
**FHC Corporate & Manufacturing**
1201 Main Street
Bowdoin, ME 04287-7302 USA
Tel: 207-666-8190
US & Can.: 800-326-2905
Fax: 207-666-8539
E-mail: fhcinc@fh-co.com
www.fh-co.com
# 510(k) SUMMARY
(Per 807.92 510(k) content instructions)
## A. Summarv
# (1) Submitter's name, address, telephone number, a confact person and date summary was prepared:
FHC, Inc. 1201 Main Street Bowdoin, Maine 04287 · Tel: 207-666-5651 Fax: 207-666-8539 Contact: Lee D. Margolin, MS, PhD. January 12, 2010
### . (2) Name(s) of device:
Proprietary/Trade Name: WavPoint™ Stereotactic System Common Name: Stereotaxic instrument Classification Name: Stereotaxic instrument (21 CFR 882.4560) Product Code (84 HAW)
# Legally Marketed Predicate Devices to which submitter claims (3) substantial equivalence:
The WayPoint™ Stereotactic System functionality is claimed to be substantially equivalent to the following:
1. FHC, Inc. microTargeting 100 Drive System (K003776; decision date: February 23, 2001; product code HAW).
2. Howmedica Leibinger, Inc. Ost-Reg Marker System (K961120; decision date: May 10. 1996; product code HAW).
3. Elekta Instruments AB Leksell SurgiPlan (K080250; decision date: May 13, 2008; product code HAW).
4. Elekta Instruments AB Leksell Stereotactic System (K972324; decision date: March 2. 1998; product code HAW).
# Description of device: (4)
The WayPoint™ Stereotactic System. (WPS System) accurately locates and rigidly supports neurosurgical or other surgical instruments relative to specific anatomical structures or targets, thus facilitating microelectrode recording, stimulation, data collection, device implantation and other surgical procedures. The WPS System provides patient-customized stereotactic fixtures designed to be secured on skull-based anchor mounting sets to hold and
{1}------------------------------------------------
align surgical instruments with patient anatomy and to allow selected targets to be reached with sub-millimeter accuracy. The WPS System consists of several components that work together to become a stereotactic system. These components are:
- 1) the WayPoint™ Stereotactic Anchor System:
- 2) the WayPoint™ Stereotactic Platform Adapter:
- 3) Device mounting hardware for the Platform; and,
- 4) WayPoint™ compatible Planning software such as:
- a. WayPoint™ Planner; or,
- b. VoXim®Planning software.
The WayPoint™ Stereotactic Anchor/Locator System provides tools and components for the temporary implantation of anchors to be used as reference points and mounting sets for the surgical procedure. The WPS System software designs a Stereotactic Platform Adapter that will mount to the WayPoint™ Stereotactic Anchor System. For each surgical instrument or device to be mounted to the Stereotactic Platform Adapter, suitable hardware is provided to act as a secure interface and allow flexibility in targeting, if required. The two existing software components of the WPS System (WayPoint™ Planner and VoXim® microTargeting™ Planning software) provide tools to create a Stereotactic Platform Adapter from scanned patient images.
## Statement of intended use: (5)
The WayPoint™ Stereotactic System is intended to be used with commercially available stereotactic systems for neurosurgical procedures which require the accurate positioning of microelectrodes, stimulating electrodes, implantable electrodes or other instruments in the brain or nervous system.
## (6) Comparison of Technological Characteristics to Predicate Devices:
The WayPoint™ Stereotactic System is substantially equivalent to the predicate devices in design, construction, materials, intended use and performance characteristics. No new issues of safety or effectiveness are introduced in the use of this device.
### B. Performance data
Both phantom- and imaging-based studies were conducted in order to evaluate the accuracy of the WPS System. Results from the phantom-based study showed that the Stereotactic Platform Adapter produced by the WPS System is accurate to a submillimetric level. Using the WPS System, a comparison of intraoperative stereotactic coordinates of an implant to its position in postoperative CT images demonstrated accuracies comparable to those of conventional frames. Finally, in a study that compared preoperative target selection and final placement verification based on CT images, error measurements were comparable to those for the same procedure as described by other groups using a standard stereotactic frame.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. The logo is black and white and appears to be a scan of a document.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
FEB 1 2 2010 -
FHC, Inc. c/o Lee D. Margolin, M.S., Ph.D. Director of Research and Compliance Quality System Officer 1201 Main Street Bowdoin, ME 04287-7302
Re: K092192
Trade/Device Name: WayPoint™ Stercotactic System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: January 15, 2010 Received: January 19, 2010
Dear Dr. Margolin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or ve and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Lee D. Margolin, M.S., Ph.D.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportuProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Erik Finholt
Malvina B. Eydelman, N Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K092192
Device Name: WayPoint™ Stercotactic System
Indications for Use:
The WayPoint™ Stercotactic System is intended to be used with commercially available stereotactic systems for neurosurgical procedures which require the accurate positioning of microclectrodes, stimulating electrodes, or other instruments in the brain or nervous system.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R.J. Hrom
Page _________________________________________________________________________________________________________________________________________________________________________
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K092192
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.