K023651 · Brainlab AG · HAW · Feb 17, 2004 · Neurology
Device Facts
Record ID
K023651
Device Name
VECTORVISION CRANIAL/ENT
Applicant
Brainlab AG
Product Code
HAW · Neurology
Decision Date
Feb 17, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
VectorVision Cranial/ENT is intended to be an intraoperative image guided localization system to enable minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on patient image data being processed by the VectorVision workstation. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, CTA, X-ray, MR, MRA and ultrasound based model of the anatomy. Example procedures include but are not limited to: Cranial Procedures: Cranial biopsies Tumor resections Craniotomies/ Craniectomies Skull base procedures Thalamotomies/Pallidotomies ENT Procedures: Transphenoidal procedures Intranasal procedures Sinus procedures, such as Maximillary antrostomies, Ethmoidectomies, Sphenoid explorations, Turbinate resections and Frontal sinusotomies
Device Story
Intraoperative image-guided surgery system; enables minimally invasive navigation. Inputs: patient diagnostic images (CT, MR, etc.) and real-time 3D position of surgical tools. Operation: infrared cameras track retro-reflective marker arrays on tools; software registers patient anatomy to virtual image space via pointer probe, laser surface scanning, or MR marker localizers. Output: 2D/3D visualization of tool position relative to anatomy on workstation display. Used in OR by surgeons; integrates with microscopes, endoscopes, and ultrasound. Benefits: provides real-time spatial orientation for precise instrument placement and navigation during complex cranial/ENT procedures.
Clinical Evidence
Bench testing only. No clinical data presented. Safety and effectiveness validated through product design and development procedures.
Technological Characteristics
Passive infrared tracking system; infrared LED arrays and cameras. Hardware platforms: VectorVision, VectorVision Compact, VectorVision Sky (ceiling-mounted). Connectivity: network interface for external devices (microscopes, endoscopes). Registration methods: pointer probe, laser surface scanning, MR marker localizers. Software: navigation workstation for 2D/3D image processing and visualization.
Indications for Use
Indicated for patients requiring stereotactic surgery where rigid anatomical structures (skull, long bone, vertebra) can be referenced to CT, CTA, X-ray, MR, MRA, or ultrasound imaging. Applicable for cranial (biopsies, tumor resections, craniotomies, skull base, thalamotomies/pallidotomies) and ENT (transphenoidal, intranasal, sinus) procedures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
K023651
# 510 (k) Summary of Safety and Effectiveness for VectorVision Cranial/ENT
## Manufacturer:
| Address: | BrainLAB AG<br>Ammerthalstrasse 8<br>85551 Heimstetten<br>Germany<br>Phone: +49 89 99 15 68 0<br>Fax: +49 89 99 15 68 33 | |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Contact Person: | Mr. Rainer Birkenbach | |
| Summary Date: | Nov 17, 2003 | |
| Device Name: | | |
| Trade name: | | VectorVision Cranial / ENT |
| Common/Classification Name: | | Image Guided Surgery System, CAS / Stereotaxy |
Instrument
## Predicate Device: Vector Vision® 2 (K 983831) VectorVision Cranial / ENT / Spine (K003589)
Device Classification Name: Stereotaxy Instrument Regulatory Class: Class II
#### Indications For Use:
VectorVision Cranial/ENT is intended to be an intraoperative image guided localization system to enable minimally invasive surgery. It links a freehand probe, tracked by a passive market sensor system to virtual computer image space on patient image data being processed hy the VectorVision workstation.
The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skill, a long bone, or vertebra, can be identified relative to a CT, CTA, X-ray, MR, MRA and ultrasound based model of the anatomy.
Example procedures include but are not limited to:
{1}------------------------------------------------
#### Cranial Procedures:
Cranial biopsies Tumor resections Craniotomies/ Craniectomies Skull base procedures Thalamotomies/Pallidotomies
#### ENT Procedures:
Transphenoidal procedures Intranasal procedures Sinus procedures, such as Maximillary antrostomies, Ethmoidectomies, Spheno-idotomies / Sphenoid explorations, Turbinate resections and Frontal sinusotomies
## Device Description
The tracking system consists of two infrared light sources and infrared cameras. Around each camera lens is an array of infrared LED's, which flood the field of view with infrared light. The infrared light is reflected by retro-reflective material (reflects light back to the light source) to the two cameras. This tracking principle does not require electrical cords on the reflecting targets and is therefore called passive tracking system.
#### Software
The tracking system sends the targets 2D and 3D position to the navigation workstation, The navigation software uses this information considering known geometric information for tracking the 3D position of probes and other surgical tools with attached retro-reflective marker arrays.
To link the virtual diagnostic image space within the navigation system to the surgical environment, the surgeon selects points on the patient using a tracked pointer probe. The selected points are stored and interpreted by the computer and related to corresponding points extracted from the diagnostic image data sets.
A different option to register the patient anatomy to the virtual image space is provided by means of a laser scanning device. With this laser device the surface of the patients head is scanned meanwhile the IR camera system picks up the laser reflections on the patients skin. By creating a virtual surface model out of the 3D coordinates of these laser points the software matches patient data set and the patient position inside the operation poor.
Another contactless registration method for MR image data is provided by a localizer geometry of MR markers, which is - by fixed relation to the patient and the reference array and by automatic registration of the markers - a direct link to the virtual inage space. In combination with the integration of data transfer functionalities this registration method is preferred for intra-operative MR data use.
The device visualizes patient data including the option to overlay multiple data sets, outlined structures and trajectories. The area of interest is displayed in the virtual computer image space in 2D- and 3D-representations.
{2}------------------------------------------------
For the localization of any area or structure in the patient's body a pointer tool is used or any other surgical instrument with attached retro-reflective marker arrays. A special device is available to precisely register any surgical instrument for navigation.
Surgical microscopes are integrated similarly as a virtual pointer tool by a mounted marker array and a software interface. With image injection modules and video display a close cooperation of navigation system and microscope is achieved. Other intraoperative image sources like ultrasound transducers and endoscopes are tracked and integrated similarly and can be controlled within the software to be displayed on an external display device for the purpose of non-diagnostic image information overview.
For sending requests and commands to external software and computers via network, e.q. for controlling the content of the external display device, a general communication interface is integrated into the navigation software.
#### Hardware
The VectorVision hardware platforms consist of electronic hardware combined in different housings which allow to adapt to the respective needs in certain OR settings. There are infrared cameras and a touch-screen, signal transmission cables, the navigation workstation itself and cable interfaces to external devices (e.g. microscopes). Currently there are 3 platforms available: VectorVision?, VectorVision Compact and VectorVision Sky. With the VectorVision Sky ceiling mounted articulated arms carry camera and touchscreen. The VectorVision Sky can be prepared for use in magnet-resonance environments with shielded RF cage e.g. by providing optical signal transmission. Interfacing external devices such as a microscope and/or a surgical endoscope is realized by interface elements in a wall panel box.
VectorVision does not contain implants but provides a system for accurate implant placement.
## Substantial equivalence
VectorVision Cranial / ENT has been verified and validated according to BrainLAB's procedures for product design and development. The validation proves the safety and effectiveness of the system. The information provided by BrainLAB in this 510/ld application was found to be substantially equivalent with the predicate device VectorVision2 (K983831) and VectorVision Cranial / ENT / Spine (K003589).
{3}------------------------------------------------
## Part 3 Combination Product Algorithm (Revised March 12, 2003)
| Category Number | Type of Part 3 Combination Product |
|-----------------|---------------------------------------------------------------------------------------|
| N | Not a Part 3 Combination Product |
| 1 | Convenience Kit or Co-Package |
| 2 | Prefilled Drug Delivery Device/System (syringe, patch, etc.) |
| 3 | Prefilled Biologic Delivery Device/System (syringe, patch, etc.) |
| 4 | Device Coated/Impregnated/Otherwise Combined with Drug |
| 5 | Device Coated or Otherwise Combined with Biologic |
| 6 | Drug/Biologic Combination |
| 7 | Separate Products Requiring Cross Labeling |
| 8 | Possible Combination Based on Cross Labeling of Separate<br>Products (Temporary Code) |
| 9 | Other Type of Part 3 Combination Product (e.g.,<br>Drug/Device/Biologic Product) |
## Quick Reference Table1 -- Part 3 Combination Product Categories
Note: The above categories are intended to be mutually exclusive. If a product meets the definition of more than one category, select the category that you believe best describes the regulatory issues associated with the combination product.
Overview: A 21 CFR Part 3 combination product is a product comprised of components usually regulated under different authorities. It may be a drug combined with a device, a biologic combined with a device, or a drug combined with a biologic. Cells or tissues that are not eligible for regulation solely as cells or tissues because they are combined with a drug or a device (except for a sterilizing, preserving, or storage agent) may also be combination products. Discuss with your Center's product jurisdiction officer.
Combination products include those where the components are":
- Physically or chemically combined OR .
- . Separate but provided/packaged as a unit OR
<sup>1</sup> Category descriptions and examples follow.
<sup>2</sup> See 21 CFR § 3.2(e) for complete definition. [will link to definition on OCP website]
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird with three wing-like shapes above three wavy lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 7 2004
Mr. Ranier Birkenbach BrainLAB AG Ammerthalstrasse 8 85551 Heimstetten Germany
Rc: K023651
Trade/Device Name: VectorVision Cranial/ENT Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: November 17, 2004 Received: November 19, 2004
Dear Mr. Birkenbach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CHR 1000-1050.
{5}------------------------------------------------
#### Page 2 - Mr. Ranier Birkenbach
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
for Mark N. Mellema
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K023651
Device Name:
#### VectorVision Cranial/ENT
Indications For Use:
VectorVision Cranial/ENT is intended to be an intraoperative image guided localization system to enable minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on patient image data being processed by the VectorVision workstation.
The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, CTA, X-ray, MR, MRA and ultrasound based model of the anatomy,
Example procedures include but are not limited to:
Cranial Procedures: Cranial biopsies Tumor resections Craniotomies/ Craniectomies Skull base procedures Thalamotomies/Pallidotomies
ENT Procedures: Transphenoidal procedures Intranasal procedures Sinus procedures, such as Maximillary antrostomies, Ethmoidectomies / Sphenoid explorations, Turbinate resections and Frontal sinusotomies
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRA, Office of Device Evaluation (ODE)
al. Restorative, and Neurological Devices
Page 1 of 1
**510(k) Number** K02365
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.