The StealthStation® System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The StealthStation® System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT or MR based model or fluoroscopy images of the anatomy.
Device Story
StealthStation with FluoroMerge software is an image-guided surgical navigation system. It enables surgeons to merge pre-operative datasets (CT/MR) with intra-operative fluoroscopy images. The system tracks surgical instruments relative to rigid anatomical structures (skull, long bones, vertebrae) to assist in precise localization during open or percutaneous procedures. Used in OR settings by surgeons; the software provides visual guidance on a display, allowing real-time correlation between pre-operative planning data and live intra-operative fluoroscopic views. This integration aids in surgical accuracy, potentially improving outcomes for complex cranial, spinal, ENT, and orthopedic interventions.
Clinical Evidence
Bench testing only. Verification and validation activities were performed to confirm the software's ability to merge pre-operative and intra-operative datasets accurately.
Technological Characteristics
Software-based image registration and navigation system. Operates on the existing StealthStation hardware platform. Integrates pre-operative (CT/MR) and intra-operative (fluoroscopy) imaging data. Connectivity via standard surgical navigation workstation interfaces.
Indications for Use
Indicated for patients requiring stereotactic surgery for cranial, spinal, ENT, or orthopedic procedures (e.g., biopsies, tumor resections, pedicle screw placement, sinus procedures, knee arthroplasty) where anatomical structures can be referenced to CT, MR, or fluoroscopy images.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
K02033d
# FEB 2 7 2002
## Summary of Safety and Effectiveness StealthStation® with FluoroMerge™ Software
### I. Manufacture:
Medtronic Surgical Navigation Technologies 826 Coal Creek Circle Louisville, CO 80027 USA Telephone Number: (720) 890-3200 Fax Number: (720) 890-3500
#### II. Contact:
Victoria G. Rendon Clinical and Regulatory Affairs Associate Medtronic Surgical Navigation Technologies
### Product Name/ Classification Name: III.
Product Name: StealthStation® with FluoroMerge™ Software Classification Name: Stereotaxic Instrument (21 CFR 882.4560) Classification Panel: 84 HAW
#### IV. Date Summary Submitted
January 30, 2002
### V. Description of Device Modification:
This submission describes updates made to the StealthStation® System to include an optional software package that enables a user to merge a patient's pre-operative dataset with a patient's intra-operative dataset.
#### VI. Substantial Equivalence:
The StealthStation® with the FluoroMerge™ was shown to be substantially equivalent to the StealthStation System cleared in previous 510(k)'s. As required by risk analysis, all verification and validation activities were performed by designated individual(s) and the results demonstrated substantial equivalence.
{1}------------------------------------------------
# VII. Indications For Use:
The indications for use for the FluoroMerge™ Software are identical to the StealthStation® System indications for use. The indications for use are as follows:
The StealthStation® System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The StealthStation® System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT or MR based model or fluoroscopy images of the anatomy.
Example procedures include, but are not limited to:
Cranial Procedures: Cranial biopsies Craniotomies/Craniectomies Skull base procedures Thalamotomies/Pallidotomies Pituitary tumor removal CSF Leak Repair
Spinal Procedures: Spinal implant procedures, such as pedicle screw procedures
ENT Procedures: Transphenoidal procedures Intranasal procedures Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies /Sphenoid explorations, Turbinate resections, and Frontal sinusotomies Orbital Decompression Procedures Optic Nerve Decompression Procedures Polyposis Procedures Endoscopic Dacryocystorhinostomy Encephalocele Procedures
Orthopedic Procedures: Total Knee Arthroplasty (Primary and Revision) Unicompartmental Knee Arthroplasty
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a bird or eagle with its wings spread, with three human faces incorporated into the design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 7 2002
Ms. Victoria G. Rendon Clinical and Regulatory Affairs Associate Medironic Surgical Navigation Technologies 826 Coal Creek Circle Louisville, Colorado 80027
Re: K020338
Trade/Device Name: StealthStation® with FluoroMerge™ Software Regulation Number: 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: January 30, 2002 Received: February 1, 2002
Dear Ms. Rendon:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becalled is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations) is regists of the Medical Device Amendments, or to conninered prices to May 20, 1978, in accordance with the provisions of the Federal Food, Drug, de necs that have been received require approval of a premarket approval application (PMA). and Cosmetic rece (1101) that to nevice, subject to the general controls provisions of the Act. The I ou may, therefore, marres of the Act include requirements for annual registration, listing of general condon profiseturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it II your device is classinod toonal controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or unly I valuates and the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set er it rate 607); accems (21 (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Ms. Victoria G. Rendon
This letter will allow you to begin marketing your device as described in your Section 510(k) This icter will anow you to ocgin maisog of substantial equivalence of your device to a legally premarket notincation. The Pro Primating of Eastification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your avitro diagnostic devices), please contact the Office of additionally 21 CFR Furt 607.10 for his nally, for questions on the promotion and advertising of Compliance at (301) 594-1057. Ticalizense at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Tegulation cirities, "Ninorananing oresponsibilities under the Act may be obtained from the Other general information of your respational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Marla n Milleess
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K020338
Device Name: StealthStation® with FluoroMerge™ Software
Indications for Use:
Indications for OSC.
The StealthStation® System is intended as an aid for precisely locating anatomical structures in The StealthStation® Dystem is mentors. The StealthStation® System is indicated for any medical enner open of perodianoods probodated is surgery may be appropriate, and where reference to a condition in which and use of sters outs the skull, a long bone, or vertebra, can be identified relative to a CT or MR based model or fluoroscopy images of the anatomy.
Example procedures include, but are not limited to:
Cranial Procedures: Cranial biopsies Tumor resections Craniotomies/ Craniectomies Skull base procedures Thalamotomies/Pallidotomies Pituitary Tumor Removal CSF Leak Repair
Spinal Procedures: Spinal Implant, such as pedicle screw procedures
ENT Procedures: Transphenoidal procedures Intranasal procedures Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies Orbital Decompression Procedures Optic Nerve Decompression Procedures Polyposis Procedures Endoscopic Dacryocystorhinostomy Encephalocele Procedures
Orthopedic Indications: Total Knee Arthroplasty (Primary and Revision) Unicompartmental Knee Arthroplasty
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office Of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
| Prescription Use (Per 21 CFR 801.109) | ☑ |
|---------------------------------------|---|
|---------------------------------------|---|
OR
| Over-The-Counter Use | |
|--------------------------|--|
| (Optional Format 1-2-96) | |
*(Division Sign-Off)*
Division of General, Restorative
and Neurological Devices
| 510(k) Number | K020338 |
|---------------|---------|
|---------------|---------|
510(k) Number -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.