← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K180690

# Navigation Instruments (K180690)

_Globus Medical, Inc. · OLO · Jun 25, 2018 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO/K180690

## Device Facts

- **Applicant:** Globus Medical, Inc.
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Jun 25, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

Globus Navigation Instruments are intended to be used during the preparation and placement of Globus screws (QUARTEX®, CREO®, REVERE®, REVOLVE®, ELLIPSE®, PROTEX® CT, and SI-LOK®) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

## Device Story

Navigation Instruments are nonsterile, reusable surgical tools including awls, probes, drill bits, drill guides, taps, and drivers. Used in spinal surgery, these instruments interface with the Medtronic StealthStation® System to provide real-time tracking of instrument position relative to patient anatomy. Surgeons operate these tools manually or with power drills (e.g., POWEREASE™) to prepare sites and place Globus spinal screws. The system uses CT/MR models, fluoroscopy, or digitized landmarks to register anatomy. By providing visual guidance on the StealthStation® display, the instruments assist the surgeon in precise anatomical localization, potentially improving screw placement accuracy and reducing surgical invasiveness.

## Clinical Evidence

Bench testing only. Design validation testing performed to confirm registration and accuracy. Testing demonstrated functionality and compatibility with the Medtronic StealthStation® System using the NavLock tracker and verified performance against predicate instruments.

## Technological Characteristics

Reusable, nonsterile surgical instruments (awls, probes, drill bits, drill guides, taps, drivers). Materials meet applicable biocompatibility standards. Designed for integration with Medtronic StealthStation® System (v 2.1.0) for stereotactic navigation. Mechanical operation, manual or power-assisted.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Navigation Instruments ([K153203](/device/K153203.md))
- Medtronic instruments ([K143628](/device/K143628.md))
- QUARTEX™ Occipito-Cervico-Thoracic Spinal System ([K161591](/device/K161591.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is a symbol of a stylized human figure. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Globus Medical Inc. Lori Burns Director, Regulatory Affairs 2560 General Armistead Avenue Audubon, Pennsylvania 19403

Re: K180690

Trade/Device Name: Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: May 22, 2018 Received: May 23, 2018

Dear Lori Burns:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

June 25, 2018

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K180690

Device Name Navigation Instruments

#### Indications for Use (Describe)

Globus Navigation Instruments are intended to be used during the preparation and placement of Globus screws (QUARTEX®, CREO®, REVERE®, REVOLVE®, ELLIPSE®, PROTEX® CT, and SI-LOK®) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

Type of Use (Select one or both, as applicable)

| <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">☑</span> </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </div> | <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">☐</span> </span>     Over-The-Counter Use    </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

# 510(k) Summary: Navigation Instruments

| Company: | Globus Medical Inc.         |
|----------|-----------------------------|
|          | 2560 General Armistead Ave. |
|          | Audubon, PA 19403           |
|          | 610-930-1800                |

| Primary Contact: | Lori Burns<br>Director, Regulatory Affairs |
|------------------|--------------------------------------------|
|------------------|--------------------------------------------|

April 3, 2018 Date Prepared:

Device Name: Navigation Instruments

- Classification: Per 21 CFR as follows: §882.4560 Stereotaxic Instrument Product Code: OLO Regulatory Class: II, Panel Code: 84
Primary Predicate: Navigation Instruments (K153203)

Predicate(s): Medtronic instruments (K143628) QUARTEX™ Occipito-Cervico-Thoracic Spinal System (K161591)

### Purpose:

The purpose of this submission is to request clearance for additional instruments for use with Medtronic StealthStation® System.

## Device Description:

Navigation Instruments are nonsterile, reusable instruments including awls, probes, drill bits, drill guides, taps, and drivers that can be operated manually or with a power drill such as POWEREASE™. These instruments are intended to be used with the Medtronic Synergy Spine and Trauma StealthStation® System (v 2.1.0).

### Indications for Use:

Globus Navigation Instruments are intended to be used during the preparation and placement of Globus screws (QUARTEX®, CREO®, REVERE®, REVOLVE®, ELLIPSE® PROTEX® CT, and SI-LOK®) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified

{4}------------------------------------------------

relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

#### Performance Data:

Design validation testing, including registration, and accuracy, was conducted to ensure the Navigation Instruments are acceptable for their intended use, to ensure functionality and compatibility with the Medtronic StealthStation® using the NavLock tracker, and to demonstrate substantial equivalence to the predicate instruments. Registration testing was performed to ensure that the instruments can be registered to the StealthStation®. Accuracy testing was completed for comparison to the predicate instruments.

### Technological Characteristics:

Navigation Instruments have the same technological characteristics as the predicate devices including design, intended use, material composition, function, and range of sizes. Biocompatibility of patient-contacting materials was demonstrated by using materials that meet applicable standards or are used in 510(k) cleared devices.

### Basis of Substantial Equivalence:

Navigation Instruments have been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject instruments to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO/K180690](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO/K180690)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
