← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K162341

# Stryker OrthoMap Precision Knee system (K162341)

_Stryker Leibinger GmbH & Co KG · OLO · Oct 12, 2016 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO/K162341

## Device Facts

- **Applicant:** Stryker Leibinger GmbH & Co KG
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Oct 12, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Stryker OrthoMap Precision Knee system, which is comprised of the OrthoMap Precision Knee 5.0 Software and a platform of the NAV3i platform family, is intended as a planning and intraoperative guidance system to enable open or percutaneous image guided surgery. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. The system is indicated for conditions of the knee joint in which the use of computer assisted surgery may be appropriate.

## Device Story

System provides planning and intraoperative guidance for knee surgery; utilizes wireless optical tracking to display real-time location of surgical instruments relative to patient anatomy. Input: infrared light emitted by active diodes on smart instruments (trackers/pointers) sensed by navigation camera array. Processing: software computes spatial information based on intra-operative anatomy survey of femur and tibia; maps instruments to computed anatomical model. Output: visual display of instrument position/movement relative to target anatomy on monitor. Used in OR by surgeons; supports open or percutaneous procedures. Benefits: enables precise alignment of mechanical axes and implant positioning; reduces reliance on manual estimation. Healthcare providers use visual feedback to guide bone resection and implant placement.

## Clinical Evidence

No clinical studies performed. Evidence consists of bench testing, including ASTM F2554-10 accuracy testing, system accuracy verification (± 2 mm, ± 1°), clinical workflow simulation using Sawbones, and functional/safety testing.

## Technological Characteristics

Infrared optical active sensing technology; wireless tracking of smart instruments via camera array. Components: computer, navigation camera, IO-Tablet, monitor. Software: OrthoMap Precision Knee 5.0. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2. Accuracy: ± 2 mm translational, ± 1° rotational.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Stryker OrthoMap Express Knee system ([K153240](/device/K153240.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 12, 2016

Stryker Leibinger Gmbh & Co. Kg % Calley Herzog Senior Consultant Biologics Consulting Group, Inc. 400 North Washington Street, Suite 100 Alexandria, Virginia 22314

Re: K162341

Trade/Device Name: Stryker OrthoMap Precision Knee System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: August 18, 2016 Received: August 22, 2016

Dear Ms. Herzog:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/8 description: The image shows the name "Mark N. Melkerson -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or label. The letters are uniformly sized and spaced, contributing to the legibility of the name.

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K162341

Device Name Stryker OrthoMap Precision Knee system

#### Indications for Use (Describe)

The Stryker OrthoMap Precision Knee system, which is comprised of the OrthoMap Precision Knee 5.0 Software and a platform of the NAV3i platform family, is intended as a planning and intraoperative guidance system to enable open or percutaneous image guided surgery.

The system can be used for intraoperative guidance where a rigid anatomical structure can be identified. The system is indicated for conditions of the knee joint in which the use of computer assisted surgery may be appropriate.

| Type of Use (Select one or both, as applicable)                                                                       |                                                                                                                      |
|-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✖</span>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span> <span style="font-size:16px">❏</span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Stryker OrthoMap Precision Knee system is provided below.

| Device Common Name:        | Orthopedic Stereotaxic Instrument                                                                                                                                                                            |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name:         | Stryker OrthoMap Precision Knee system                                                                                                                                                                       |
| Applicant:                 | Stryker Leibinger GmbH & Co. KG<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Germany<br>Phone number: +49-761-45120<br>Fax number: +49-761-45120                                                            |
| Contact:                   | Bjoern Lampart<br>Stryker Leibinger GmbH & Co. KG<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Germany<br>P +49 761 4512 3991<br>C +49 160 97870607<br>F +49 761 4512 49 3991<br>bjoern.lampart@stryker.com |
| Prepared by:               | Calley Herzog<br>Senior Consultant<br>Biologics Consulting Group, Inc.<br>cherzog@biologicsconsulting.com<br>Phone: 720-883-3633                                                                             |
| Date Prepared:             | August 18, 2016                                                                                                                                                                                              |
| Classification Regulation: | §882.4560 - Stereotaxic Instrument                                                                                                                                                                           |
| Panel:                     | Orthopedic                                                                                                                                                                                                   |
| Product Code:              | OLO - Orthopedic Stereotaxic Instrument                                                                                                                                                                      |
| Predicate Device:          | K153240 - Stryker OrthoMap Express Knee system                                                                                                                                                               |

Indication for Use:

{4}------------------------------------------------

The Stryker OrthoMap Precision Knee system, which is comprised of the OrthoMap Precision Knee 5.0 Software and a platform of the NAV3i platform family, is intended as a planning and intraoperative guidance system to enable open or percutaneous image guided surgery.

The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.

The system is indicated for conditions of the knee joint in which the use of computer assisted surgery may be appropriate.

### Device Description:

The Stryker OrthoMap Precision Knee system is intended to be used as a planning and intraoperative guidance system to enable open or percutaneous image guided knee surgery. The system uses wireless optical tracking technology to display to the surgeon the intraoperative location of navigated surgical instruments relative to a computed anatomical model of the patient's leg (femur and tibia). The computed model is based on an intra-operative anatomy survey of the leg. The system consists of a Stryker surgical software application (software), which runs on a platform, consisting of a Stryker computer, a navigation camera, an IO-Tablet and a monitor. The Stryker surgical software application interfaces with smart instruments (e.g. patient trackers, instrument trackers or pointers) and several accessories enabling the tracking of surgical instruments.

The Stryker OrthoMap Precision Knee 5.0 software is compatible with the NAV31 Platform Family. The NAV3i platform family is a family of platforms that, when used with a surgical software application, displays patient specific images and/or patient specific anatomical landmark information and tracks the position and movement of surgical instruments in relation to a target anatomical site on a patient. The NAV3i platform family consists of the following three platforms that have been previously cleared independently or with other Stryker surgical software applications:

- Stryker NAV3 Platform
- Stryker NAV3i Platform ●
- Stryker NAVSuite3 Kit

The platforms consist of the following components:

- Stryker computer
- Navigation camera ●
- IO-Tablet ●
- Monitor
- Mobile cart (if applicable) ●

### Performance Data:

The following performance data were provided in support of the substantial equivalence decision:

### Electrical Safety and Electromagnetic Compatibility (EMC) Testing

{5}------------------------------------------------

Electrical safety and EMC testing were conducted on the subject device in accordance with the following standards:

- ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and ● A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) (FDA Rec# 19-4)
- . IEC 60601-1-2:2007: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (FDA Rec#19-1)

### Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "MAJOR" level of concern.

### Performance Testing - Bench

The following design verification activities have been performed to ensure the correct functionality of the system as it has been specified:

- ASTM accuracy testing verifying the accuracy performance of the localization and . tracking technology using the standardized test procedure according to ASTM Standard F2554-10.
- . System accuracy testing verifying the specified accuracy of ± 2 mm and ± 1° using a mechanical leg mimicking the patient's anatomy.
- . Clinical workflow testing verifying that all system components (application, computer platform and accessories) are compatible. Complete total knee arthroplasty procedures are simulated using Sawbones mimicking the patient's anatomy.
- Functional testing to ensure that all functional requirements are fulfilled.
- . Safety testing verifying the effectiveness of all risk controls determined in the device risk analysis and in the risk analyses of the platforms.

This strategy ensures the verification of the accuracy, system integration, software algorithms, system functionality, and correct implementation of the risk control measures. All tests have been successfully completed.

### Animal Study

No animal studies were performed to support substantial equivalence.

### Clinical Studies

No clinical studies were performed to support substantial equivalence.

{6}------------------------------------------------

### Substantial Equivalence Rationale:

The intended uses of the subject and predicate devices are similar. The systems are intended as planning and intraoperative guidance systems to enable open or percutaneous computer assisted surgery. Minor modifications were made to the indication for use statement to provide clarification, but do not change the meaning of the indication for use statement.

Both the subject device and the primary predicate device use the same main system components, use similar modes of operation, and use the same localization and tracking technology. Also, both systems use similar accessories (Smart Instruments, Patient Tracker Fixation, Navigated Manual Instruments and other accessories).

The technological characteristics of the subject and predicate device are equivalent. None of the changes alter the operating principle, the control mechanism, the localization and tracking technology, the main system components or the system accuracy performance. The workflow, user interaction, the software architecture and the software features are similar. As demonstrated by the performance testing described above, the subject device continues to meet the same accuracy specifications as the predicate device. Therefore, the differences in technological characteristics to do not raise new questions of safety and effectiveness.

In conclusion, the Stryker OrthoMap Precision Knee system can be found substantially equivalent to the OrthoMap Express Knee system (K153240).

The device comparison table below provides a comparison of the technological characteristics of the subject device to the primary predicate device.

|                                                     | Subject Device<br>Stryker OrthoMap Precision Knee System                                                                                                                                                                                                                                                 | Predicate Device - K153240<br>Stryker OrthoMap Express Knee system                                                                                                                                                                                                                             |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                                       | Under review                                                                                                                                                                                                                                                                                             | K153240                                                                                                                                                                                                                                                                                        |
| Clearance Date                                      | Under review                                                                                                                                                                                                                                                                                             | July 7, 2016                                                                                                                                                                                                                                                                                   |
| Device Name                                         | Stryker Navigation OrthoMap Precision Knee<br>5.0 Navigation System                                                                                                                                                                                                                                      | Stryker Navigation OrthoMap Express<br>Knee 2.0 Navigation System                                                                                                                                                                                                                              |
| Regulation<br>Number                                | §882.4560 - Stereotaxic Instrument                                                                                                                                                                                                                                                                       | §882.4560 - Stereotaxic Instrument                                                                                                                                                                                                                                                             |
| Product Code                                        | OLO - Orthopedic Stereotaxic Instrument                                                                                                                                                                                                                                                                  | OLO - Orthopedic Stereotaxic Instrument                                                                                                                                                                                                                                                        |
| Product Class                                       | II                                                                                                                                                                                                                                                                                                       | II                                                                                                                                                                                                                                                                                             |
| Localization<br>and Tracking<br>technology          | Infrared optical active sensing technology:<br>Infrared light emitted by diodes placed in a<br>known fashion on navigated surgical<br>instruments is sensed by a camera array<br>(navigation camera) on the computer platform,<br>thus allowing for computation of the spatial<br>information.           | Infrared optical active sensing technology:<br>Infrared light emitted by diodes placed in a<br>known fashion on navigated surgical<br>instruments is sensed by a camera array<br>(navigation camera) on the computer<br>platform, thus allowing for computation of<br>the spatial information. |
|                                                     | Subject Device<br>Stryker OrthoMap Precision Knee System                                                                                                                                                                                                                                                 | Predicate Device – K153240<br>Stryker OrthoMap Express Knee system                                                                                                                                                                                                                             |
| Main System<br>Components                           | Computer Platform<br>OrthoMap Precision Knee 5.0 Software<br>Smart Instruments<br>Patient Tracker Fixation<br>Navigated Manual Instruments<br>Instrument Battery, Trays                                                                                                                                  | Computer Platform<br>OrthoMap Express Knee 2.0 Software<br>Smart Instruments<br>Patient Tracker Fixation<br>Navigated Manual Instruments<br>Instrument Battery, Trays                                                                                                                          |
| Compatible<br>Platforms                             | NAV3i Computer Platform Family:<br>Stryker NAV3 Platform<br>Stryker NAV3i Platform<br>Stryker NAVSuite3 Kit                                                                                                                                                                                              | NAV3i Computer Platform Family:<br>Stryker NAV3 Platform<br>Stryker NAV3i Platform<br>Stryker NAVSuite3 Kit                                                                                                                                                                                    |
| Registration<br>and Navigation<br>Work Flow         | Patient Preparation<br>System Setup<br>Patient Registration (femur and tibia)<br>Analyze Initial Alignment<br>Size and Position Implant<br>Resect Bones<br>Analyze Trial Implant<br>Analyze Final Implant                                                                                                | Patient Preparation<br>System Setup<br>Patient Registration (femur)<br>Navigation (femur)<br>Patient Registration (tibia)<br>Navigation (tibia)                                                                                                                                                |
| System<br>Accuracy                                  | The system enables the determination of the<br>mechanical axes of the leg as well as cut and<br>component alignment with a mean<br>translational error of $<$ 2 mm and a mean<br>rotational error of $<$ 1°.                                                                                             | The system enables the determination of the<br>mechanical axes of the leg as well as cut<br>and component alignment with a mean<br>translational error of $<$ 2 mm and a mean<br>rotational error of $<$ 1°.                                                                                   |
| Accessories:<br>Patient /<br>Instrument<br>Trackers | Patient/Instrument Trackers<br>Universal Tracker Tibia/Pelvic Tracker Femoral Tracker nGenius Universal Tracker nGenius Tibial Tracker nGenius Femur Tracker Hip Femur Tracker Hip Tibia/Pelvis TrackerPointers Pointer, Knee Navigation Ortho Grip Knee Pointer Pointer, Straight                       | Femur Trackers / Tibia Trackers<br>Universal Tracker Tibia/Pelvic Tracker Femoral Tracker nGenius Tibial Tracker nGenius Femur Tracker nGenius Femur Tracker Hip Femur Tracker Hip Tibia/Pelvis TrackerPointers Pointer, Knee Navigation Ortho Grip Knee Pointer Pointer, Straight             |
| Accessories:<br>Patient Tracker<br>Fixation         | OrthoLock with OrthoLock Navigation Pin or<br>OrthoLock EX-Pins<br>Anchoring Pins and Insertion Tool                                                                                                                                                                                                     | ASM Fixation Plate                                                                                                                                                                                                                                                                             |
|                                                     | Subject Device<br>Stryker OrthoMap Precision Knee System                                                                                                                                                                                                                                                 | Predicate Device – K153240<br>Stryker OrthoMap Express Knee system                                                                                                                                                                                                                             |
| Accessories:<br>Navigated<br>Manual<br>Instruments  | Dedicated Mini Jig (which consists of):  Adjustment Component Tracker Adapter Mini Cutting Guide Mini Fixation Plate              MIS Jig (which consists of):  Navigated MIS Jig-A Navigated MIS Jig-B Tracker Adapter              Plane Probes (Resection Plan Probes and a<br>Posterior Plane Probe) | Dedicated Mini Jig (which consists of):  Adjustment Component Tracker Adapter Mini Cutting Guide              ASM Resection Plane Probe                                                                                                                                                        |
| Screwdrivers                                        | Screwdriver<br>Universal Joint Screwdriver                                                                                                                                                                                                                                                               | Screwdriver<br>Universal Joint Screwdriver                                                                                                                                                                                                                                                     |

#### Device Comparison Table:

{7}------------------------------------------------

{8}------------------------------------------------

### Substantial Equivalence Conclusion:

Based on the comparison of intended use and technological characteristics, the device is similar to the predicate device. The hardware and software verification and validation testing demonstrate that the subject device meets its performance specifications and will perform as intended in the specified use conditions and that any differences between the subject device and predicate device do not raise new questions of safety and effectiveness. Therefore, the subject device can be found substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO/K162341](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO/K162341)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
