Navigated Vertex Select Instruments

K143628 · Medtronic Sofamor Danek USA, Inc. · OLO · Feb 12, 2015 · Neurology

Device Facts

Record IDK143628
Device NameNavigated Vertex Select Instruments
ApplicantMedtronic Sofamor Danek USA, Inc.
Product CodeOLO · Neurology
Decision DateFeb 12, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

Medtronic Navigated Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. Medtronic Navigated Instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Instruments are also compatible with the IPC® POWEREASE™ System.

Device Story

Navigated VERTEX SELECT® Instruments are surgical tools (screwdrivers, taps, drill bits) used for spinal screw preparation and placement. Instruments attach to NavLock™ Tracker for optical navigation via StealthStation® System; compatible with IPC® POWEREASE® System. Used in open or minimally invasive spinal surgery by surgeons. Instruments provide physical interface for bone preparation; navigation system tracks instrument position relative to patient anatomy (CT/MR/fluoroscopy/digitized landmarks). Output is visual guidance on navigation system display, assisting surgeon in precise anatomical localization. Benefits include improved accuracy in screw placement during spinal reconstruction.

Clinical Evidence

Bench testing only. Testing included navigation accuracy analysis, anatomical simulated use, navigation simulated use, CAD model evaluation, implant/instrument mating verification, and functional testing of the Spine Tools package to ensure interface compatibility with application software.

Technological Characteristics

Materials: Stainless steel. Sensing: Optical navigation via NavLock™ Tracker. Energy: Manual or power-driven (IPC® POWEREASE® System). Form factor: Surgical instruments (screwdrivers, taps, drill bits). Connectivity: Compatible with StealthStation® navigation system. Sterilization: Reusable (non-sterile) and single-use (sterile) options.

Indications for Use

Indicated for patients undergoing spinal surgery requiring preparation and placement of Medtronic screws, where stereotactic surgery is appropriate and anatomical structures can be referenced to CT, MR, fluoroscopy, or digitized landmarks.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract image of what appears to be a family. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Medtronic Sofamor Danek USA, Incorporated Mr. Tejas Patel Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132 February 12, 2015 Re: K143628 Trade/Device Name: Navigated VERTEX SELECT® Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: December 19, 2014 Received: December 22, 2014 Dear Mr. Patel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ Page 2 - Mr. Tejas Patel forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use K143628 510(k) Number (if known) Device Name Navigated VERTEX SELECT® Instruments ### Indications for Use (Describe) Medtronic Navigated Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. Medtronic Navigated Instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Instruments are also compatible with the IPC® POWEREASE™ System. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary ## December 19, 2014 - I. Company: Medtronic Sofamor Danek, USA Inc. 1800 Pyramid Place Memphis, TN 38132 Telephone Number: (901) 396-3133 - Contact: Tejas Patel Regulatory Affairs Specialist Telephone number: (901) 396-3133 Email: tejaskumar.r.patel@medtronic.com - II. Proprietary Trade Name: Navigated VERTEX SELECT® Instruments Common Name: Stereotaxic Instrument, Navigated Screwdriver, Navigated Tap, Navigated Drill Bit Classification Name: Stereotaxic Instrument (21 CFR 882.4560) Classification: Class II Product Code: OLO #### III. Predicate Device: Navigated CD HORIZON® SOLERA® Screwdrivers and Taps (K140454, S.E. 05/22/2014) This predicate has not been subject to a design-related recall. This predicate is the primary predicate for this submission. No reference devices were used in this submission. #### IV. Device Description: The Navigated VERTEX SELECT® Instruments are both non-sterile, reusable and sterile, single use instruments that can be operated manually or under power. These instruments are intended to be used when implanting components of the VERTEX® Reconstruction System. The Navigated VERTEX SELECT® Instruments are also compatible with the StealthStation® and IPC® POWEREASE® Systems. #### V. Indications for Use: Medtronic Navigated Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely {4}------------------------------------------------ locating anatomical structures in either open or minimally invasive procedures. Medtronic Navigated Instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Instruments are also compatible with the IPC® POWEREASE® System. #### VI. Comparison of the Technological Characteristics with the Predicate Device: The Navigated VERTEX SELECT® Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery and are specifically designed for use with the StealthStation® System. Identical to the predicates, the Navigated VERTEX SELECT® Instruments attach to the NavLock™ Tracker, which allows for optical navigation of the surgical instruments. These devices have similar designs as the predicate devices and incorporate the same design features to enable navigation and use with the IPC® POWEREASE® System, when desired. Like the predicate devices, the subject Navigated VERTEX SELECT® Instruments are also made from stainless steel. The following technological differences exist between the subject and predicate devices: - . The Navigated VERTEX SELECT® Instruments are intended to be used with VERTEX ® Reconstruction System screws whereas the predicate Navigated CD HORIZON® SOLERA® instruments are intended to be used with the CD HORIZON® SOLERA® screws. - The subject devices include a sterile drill bit while the predicate devices do not. - The subject Navigated Taps & Drill Bits have a single lead threadform as compared to the dual lead threadform of the predicate Navigated Taps. The instrument modifications detailed in this submission have no impact on the technological characteristics of either the existing instruments or the StealthStation® and IPC® POWEREASE® Systems. #### VII. Performance Data: Testing was completed to ensure the functionality and compatibility with the identified Medtronic products. The following table summarizes the performance testing completed: | Test | Description | |---------------------------------|---------------------------------------------------------------------| | Navigation Accuracy<br>Analysis | Confirmed navigated instrument accuracy | | Anatomical Simulated Use | Confirmed instrument functionality under expected<br>use conditions | {5}------------------------------------------------ | Navigation Simulated Use | Confirmed navigation system functionality under expected use conditions | |----------------------------------------|--------------------------------------------------------------------------------------------------------------| | CAD Model Evaluation | Verified that the CAD models are accurately reflected in the application software | | Implant/Instrument Mating Conditions | Verified that the instruments can be assembled with the appropriate devices according to their intended use | | Spine Tools Package Functional Testing | Verified that the Spine Tools package has met the required interface needs of the spine application software | #### VIII. Conclusions The Navigated VERTEX SELECT® Instruments have been shown through comparison and testing to be substantially equivalent to the identified predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...